Page 1: TF-DG_ops

MMIICC TTRRAACC MMTT--33000000 TTFF--DDGG DDeepptthh MMiiccrroommeetteerr FFiixxttuurree OOppeerraattiioonn MMaannuuaall

Page 2: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 2 2010 Gagemaker, LP

RCDMMT300010-10

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • 5/32" hex wrench • Depth micrometer • 50 in/lb torque wrench • Depth gage fixture (TF-DG) and cap screws • Cloth • Flat face anvil (TF-1F Block) cap screws and washers • ZEP I.D. Red cleaner • Depth gage support (TF-SS Block), for rods 6” and up 1. Inspect the depth micrometer visually as

follows: • Check for damage and excessive wear. • Check for burrs, nicks and wear on measuring

faces. • Check for damage to measuring rods, rod

seating surfaces, and rod reading surfaces. • Inspect for slack in the spindle thread by

pushing the assembly in and out. Adjust to remove excess play.

• Check for proper function of the friction or ratchet stop and spindle locking device.

• Check for smoothness of travel by rotating the thimble throughout the range.

• Ensure all rods are straight.

2. Clean both of the receiver pads and the

mounting surfaces of the flat face anvil using the cloth and ZEP I.D. Red cleaner.

3. Locate the left flat face anvil. Place the

shoulder side of the anvil against the left receiver pad shoulder.

Setup

Page 3: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 3 2010 Gagemaker, LP

RCDMMT300010-10

4. While holding the anvil against the receiver pad

shoulder, insert the two cap screws with washers into the holes on either side of the fixture.

5. While applying pressure toward the receiver

pad shoulder, use a 5/32" hex wrench to slightly tighten the screws.

6. Use a 50 in/lbs torque wrench to secure the

cap screws.

7. Locate the depth gage fixture and clean the

mounting surfaces with the cloth and ZEP I.D. Red cleaner.

8. Place the depth gage fixture against the right

receiver pad shoulder with the thumb screws facing to the right. Insert the cap screws.

Setup

Page 4: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 4 2010 Gagemaker, LP

RCDMMT300010-10

9. While applying pressure toward the receiver

pad, use a 5/32" hex wrench to slightly tighten the screws.

10. Use a 50 in/lb torque wrench to secure the cap

screws.

11. Clean the depth micrometer using the cloth and ZEP I.D. Red cleaner.

Setup

Setup

Page 5: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 5 2010 Gagemaker, LP

RCDMMT300010-10

12. Zero the depth micrometer on a flat surface

and lock the spindle.

13. Retract the thumb screws on the depth gage

fixture enough to fit the depth micrometer.

14. Mount the depth micrometer in the depth gage

fixture, aligning the depth micrometer’s rod with the center of the flat face anvil.

Page 6: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 6 2010 Gagemaker, LP

RCDMMT300010-10

15. Secure the depth micrometer by holding down

on the base and tightening the thumb knob on each side of the depth gage fixture.

16. Turn the coarse adjust knob counterclockwise

until the depth micrometer barely touches the flat face anvil.

17. Apply slight pressure until the in the

axis display lights up. The display will read 0.00000 and the gaging force will be preset to 2.0 lbs of force.

Setup

Page 7: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 7 2010 Gagemaker, LP

RCDMMT300010-10

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • 5/32" hex wrench • Seiko printer (optional) • Lightweight gage oil • Depth Micrometer • Gage Calibration Record 1. Turn the fine adjust knob clockwise to bring the

depth micrometer face in contact with the flat face anvil.

2. Verify that the readout on the CPU displays

0.00000.

3. Loosen the coarse adjust lock and turn the

coarse adjust knob clockwise to slide the movable carriage to the right.

4. Adjust the depth micrometer to the first

calibration value. Note: If you pass the calibration value, turn the

spindle back and approach the value again, from the same direction. This practice will increase the accuracy of the calibration.

If using a computer with the MT-3000, go to page 11. Calibration

Page 8: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 8 2010 Gagemaker, LP

RCDMMT300010-10

5. Turn the coarse adjust knob counterclockwise

until the depth micrometer rod barely touches the flat face anvil.

6. Secure the coarse adjust lock.

7. Turn the fine adjust knob clockwise until the

- in the axis display lights up. If you have a Seiko Printer attached to your MIC TRAC a permanent record of the measurement prints.

8. Record any deviations on the Gage Calibration

Record or in-house calibration report. Continue with remaining measurements for the first depth rod.

Note: It is recommended that depth rods 6” and longer be supported during calibration. 9. Locate and mount the TF-SS support. With the

thumbscrew facing forward, slide the support over the flat face anvil.

10. Remove the depth rod from the depth

micrometer and insert the next depth rod. Adjust the height of the depth gage support to hold the end of the depth rod, so it does not bend during calibration. Tighten the thumb screw.

Calibration

Page 9: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 9 2010 Gagemaker, LP

RCDMMT300010-10

11. Adjust the height of the depth gage support to

hold the end of the depth rod, so it does not bend during calibration. Tighten the thumb screw.

12. Turn the coarse adjust knob counterclockwise

until the depth micrometer rod barely touches the flat face anvil. Secure the coarse adjust lock.

13. Continue turning the fine adjust knob until the

. in the axis display lights up. Record any deviations on the Gage Calibration Record or in-house calibration report. Continue with the same calibration process for all remaining depth rods.

14. Remove the depth micrometer and continue

with the same calibration process for the next depth micrometer.

15. After calibrating all depth micrometers, be sure

to remove the depth gage fixture and flat face anvil from the MT-3000. Oil the fixture and anvil and return them to the storage case.

Calibration

Page 10: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 10 2010 Gagemaker, LP

RCDMMT300010-10

Materials Needed: • MIC TRAC MT-3000 base unit and CPU readout • Depth Micrometer • Computer • 5/32" hex wrench • CERTIFI software • Lightweight gage oil • MT-4-USB Digital Data Acquisition Card • Brother P-Touch Label Printer (optional) 1. Start Certifi by double clicking the Certifi icon.

CERTIFI Icon

2. Click the New Calibration Report icon.

New Report Icon

3. With the cursor in the Gage Template field,

press the Enter key. 4. Highlight the Depth Micrometer line and click

on the OK button.

Calibration If using a computer with the MT-3000, begin here.

Calibration

Page 11: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 11 2010 Gagemaker, LP

RCDMMT300010-10

5. Complete the following Client information:

• Company name and address. • OEM # • Contact name. • PO/Account - purchase order or account

number. • Calibration Date - automatically displays,

but can be changed by typing over the displayed date.

• Print #

6. Enter the following information about the gage

you are calibrating: • Serial Number • Location • Gage Manufacturer • Model • Description

7. You may change the tolerances for the caliper

if you wish. 8. You can enter additional master values by right

clicking your mouse over an existing master value. Select Insert to insert a new line. Then type in the new master value in that line.

Page 12: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 12 2010 Gagemaker, LP

RCDMMT300010-10

9. Click on the CERTIFI Monitor icon to open the

CERTIFI Monitor Window. 10. Choose the “Y” axis from the pull down box. 11. Bring the MIC TRAC jaws together and click

the Zero button twice.

12. Turn the X axis adjust knob clockwise and pull

out.

13. Turn the Y axis adjust knob counterclockwise

until the depth micrometer contacts the flat face anvil.

Calibration

Calibration

Page 13: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 13 2010 Gagemaker, LP

RCDMMT300010-10

14. Enter the first measurement in the Measured

column as follows:

• Continue turning the Y axis adjust knob until the X axis moves approximately ½ inch. Secure Locks 1 and 2.

• Click the Send to Table bar. CERTIFI records the value and displays a green box in the Pass/Fail column if the value is within tolerance. A red box displays if the value is not within tolerance. The Deviation column shows the deviation of the measured value from the master value.

15. Adjust the depth micrometer to the first

calibration value. Note: If you pass the calibration value, turn the

spindle back and approach the value again, from the same direction. This practice will increase the accuracy of the calibration.

Page 14: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 14 2010 Gagemaker, LP

RCDMMT300010-10

16. Turn the Y axis adjust knob counterclockwise

until the depth micrometer rod contacts the flat face anvil.

16. Enter the gage measurements in the Measured

column as follows:

• Continue turning the Y axis adjust knob until the X axis moves approximately ½ inch. Secure Locks 1 and 2.

• Click the Add Value button. 17. Then, using each master value, continue

measuring the depth micrometer until the values for the first depth rod are recorded in CERTIFI.

18. Loosen Lock 1 and 2 and turn the Y axis adjust knob clockwise to slide the Y axis to the right.

Note: It is recommended that depth rods 6” and longer be supported during calibration. 19. Locate the depth gage support. With the

thumbscrew facing forward, slide the support over the left flat face anvil.

20. Remove the depth rod from the depth

micrometer and insert the next depth rod.

Calibration

Page 15: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 15 2010 Gagemaker, LP

RCDMMT300010-10

21. Adjust the height of the depth gage support to

hold the end of the depth rod, so it does not bend during calibration. Tighten the thumb screw.

22. Turn the Y axis adjust knob counterclockwise

until the depth micrometer rod touches the flat face anvil.

23. Enter the gage measurement for the depth rod

in the Measured column as follows:

• Continue turning the Y axis adjust knob until the X axis moves approximately ½ inch. Secure Locks 1 and 2.

• Click the Add Value button. • Continue calibrating the remaining depth

rods as in the previous five steps. .

25. Change the Next Calibration Date, if

necessary, by typing over the displayed date.

Calibration

Page 16: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 16 2010 Gagemaker, LP

RCDMMT300010-10

Enter NIST information as necessary. With the cursor in the single box press the Enter key to bring up a list of current NIST information. For new NIST information select Edit from the menu and press the Preferences box. Select the NIST tab to enter new information.

26. Click the Save button. Note: CERTIFI automatically assigns a filename

to each calibration. DO NOT change the filename to avoid problems with the CERTIFI database.

Calibration

Page 17: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 17 2010 Gagemaker, LP

RCDMMT300010-10

27. When the confirm window displays, click the

Yes button if you want to print the calibration report.

Note: If you have a Brother P-Touch Label Printer for printing calibration stickers, the Confirm window for printing a calibration tag displays. 28. Click the Yes button to print the calibration tag.

Affix the Calibration Tag to the caliper. 29. Continue with the same calibration process for

the next caliper.

30. After calibrating all depth micrometers, be sure

to remove the depth gage fixture and flat face anvil from the MT-3000. Oil the fixture and anvil and return them to the storage case.

Calibration

Page 18: TF-DG_ops

DDeepptthh MMiiccrroommeetteerr CCaalliibbrraattiioonn RReeffeerreennccee CCaarrdd

MIC TRAC MT-3000 18 2010 Gagemaker, LP

RCDMMT300010-10

Notes: