Teledyne Analytical Instruments

TELEDYNE AAQMS Systems

Page 2: Teledyne Analytical Instruments

TELEDYNE

TELEDYNE Technologies International Corp (Appx: 2.5 Billion ) formed in1960.

Teledyne Analytical Instruments (TAI) is a company under Teledyne Corp, world leader in the design and manufacturing of high quality process gas and liquid analyzers. With over 51 years of experience, Teledyne’s product line includes electrochemical sensors, analyzers, and custom systems which combine expertise in electronics, chemistry and engineering. TAI provides scientific solutions for chemical analysis problems & water quality measurements.

Detailed attention to customer needs and dedication to the highest standards of performance and quality are company priorities. Teledyne Analytical Instruments is an ISO 9001/ 14001 certified company.

Page 3: Teledyne Analytical Instruments

3

TELEDYNE TAI has over 15 sensing technologies to continuously monitor a variety of

process gas and liquid parameters.

What We Offer

Oxygen analyzers using electrochemical, paramagnetic, and zirconium oxide sensing technologies in an array of configurations from the lightest portable to the most rugged explosion proof system

A complete line of industrial oxygen sensors, each designed for fast response, specific measurement ranges, tough background gases and other unique requirements

Medical oxygen monitors and oxygen sensors

Oxygen monitors for OEM applications in the automotive, diving, medical, safety, and food transport industries

Oxygen monitors for mixed gas diving

Page 4: Teledyne Analytical Instruments

Teledyne Analytical Instruments

TELEDYNE AAQMS Systems

Ambient Air Quality Monitoring Systems

Page 5: Teledyne Analytical Instruments

5

Scope of Supply : AQMS Project CO

NOx

SO2

VOC (BTEX)

O3

PM 10 & 2.5

WMS : Rain gauge, Wind speed, Wind direction, Solar Radiation, Relative Humidity sensors

DAS ( Data Acquisition System)

Mobile Trailer fitted with AQMS Station

Page 6: Teledyne Analytical Instruments

6

AAQMS Mobile Shelter

Page 7: Teledyne Analytical Instruments

7

AAQMS Cabinets

Page 8: Teledyne Analytical Instruments

8

Analytical Instruments for AQMS CO – GFC7001E

– Gas Filter Correlation NDIR Sensor

NO2 – 9110E– Chemiluminescence

SO2 – 6400E– UV Fluorescence

VOC – 4060– GC Speciation– Flame Ionization Detector

Ozone – 400E– UV Absorption Sensor

Particulate Matter (2.5 & 10) – Beta Ray Attenuation

Zero Air Generator – 701

Mass Flow Calibrator – 702E

DAS – Win collect

Page 9: Teledyne Analytical Instruments

9

AQMS – Sample System

Page 10: Teledyne Analytical Instruments

10

AAQMS – Operation Philosophy.

The DIVAC sample pump will draw the sample gas from ambient air and fed to the sample manifold. From the sample manifold the air is distributed to sample inlet port of each analyzer. Each Analyzer will analysis the sample air and display the actual reading.

Each analyzer data is transferred to DAS Distributed antenna system for reporting and further analysis .

The analog put from each analyzers and WMS will be connected to centralized DAS system for final data analysis and reporting.

Each Analyzers were calibrated for Zero & Span via ZAG and MFC calibrator.

Page 11: Teledyne Analytical Instruments

11

AAQMS Analyzer Panel

Page 12: Teledyne Analytical Instruments

12

Mobile AAQMS Station

Page 13: Teledyne Analytical Instruments

13

Page 14: Teledyne Analytical Instruments

14

Useful Terminology

Gas Filter Correlation comparing infrared energy absorbed by a sample to that absorbed by a reference gas according

to the Beer-Lambert law. the Beer-Lambert law The law states that there is a logarithmic dependence between the transmission T of light

through a substance and the product of the absorption coefficient of the substance, α, and the distance the light travels through the material ℓ.

where I is the light intensity after it passes through the sample and Io is the initial light intensity. the absorption coefficient α is absorption cross section, σ, time the (number) density N' of

absorbers

Page 15: Teledyne Analytical Instruments

15

Useful Terminology

– Gas Filter Correlation comparing infrared energy absorbed by a sample to that absorbed by a reference gas according

to the Beer-Lambert law. This is accomplished with a Gas Filter wheel which alternately allows a high energy light source

to pass through a CO filled chamber and a chamber with no CO present. The light path then travels through the sample cell, which has a folded path of 14 meters.

Page 16: Teledyne Analytical Instruments

16

Useful Terminology

Page 17: Teledyne Analytical Instruments

17

SO2 Analyzer – Model 6400E PARAMAGNETIC MEASUREMENT OF O2The oxygen sensor used in the 6400E utilizes the fact that oxygen is attracted into strong magnetic field while

most other gases are not, to obtain fast, accurate oxygen measurements.

The sensor’s core is made up of two nitrogen filled glass spheres, which are mounted on a rotating suspension

within a magnetic field . A mirror is mounted centrally on the suspension and light is shone onto the mirror that

reflects the light onto a pair of photocells. The signal generated by the photocells is passed to a feedback loop,

which outputs a current to a wire winding (in effect, a small DC electric motor mounted on the suspended mirror)

Oxygen from the sample stream is attracted into the magnetic field displacing the nitrogen filled spheres and

causing the suspended mirror to rotate. Therefore, the amount of light reflected onto the photocells and

therefore the output levels of the photocells. The feedback loop increases the amount of current fed into the

winding in order to move the mirror back into its original position. The more O2 present, the more the mirror

moves and the more current is fed into the winding by the feedback control loop.

A sensor measures the amount of current generated by the feedback control loop which is directly proportional

to the concentration of oxygen within the sample gas mixture.

Page 18: Teledyne Analytical Instruments

18

SO2 Analyzer – Model 6400E

Page 19: Teledyne Analytical Instruments

19

Useful Terminology

– Gas Filter Correlation comparing infrared energy absorbed by a sample to that absorbed by a reference gas according

to the Beer-Lambert law. This is accomplished with a Gas Filter wheel which alternately allows a high energy light source

to pass through a CO filled chamber and a chamber with no CO present. The light path then travels through the sample cell, which has a folded path of 14 meters.The energy loss through the sample cell is compared with the span reference signal provided by the gas filter .The signal produced is proportional to concentration, with little effect from interfering gases within the sample.

This design produces excellent zero and span stability and a high signal to noise ratio allowing extreme sensitivity.

Page 20: Teledyne Analytical Instruments

20

Useful Terminology

– NDIR Sensor A nondispersive infrared sensor is a simple spectroscopic device often used as gas detector.

It is called nondispersive because wavelength which passes through the sampling chamber is not pre-filtered instead a filter is used before the detector.

The main components are an infrared , a light tube, a wavelength sample chamber, and gas concentration is measured electro-optically , means the absorption of a specific wavelength in the infrared (IR) is mesured by electroic component.

The IR light is directed through the sample chamber towards the detector. In parallel there is another chamber with an enclosed reference gas, typically nitrogen. The detector has an optical filter in front of it that eliminates all light except the wavelength that the selected gas molecules can absorb.

– Transducer Transducer is a device that converts a signal in one form of energy to another form of energy. Flow Transducer: convert mecanical energy ( flow velocity) into electrical ( electronic signal) Pressure transducer: Convert mecanical energy ( pressure) into electrical ( electronic signal)

Page 21: Teledyne Analytical Instruments

21

Useful Terminology

– UV Fluorescence

Fluorescence is the emission of light by a substance that has absorbed light or other electromagnetic radiation. It is a form of luminescence (emission of light by a substance not resulting from heat. In most cases, the emitted light has a longer wavelength, and therefore lower energy, than the absorbed radiation. The most striking examples of fluorescence occur when the absorbed radiation is in the ultraviolet region of the spectrum, (UV Fluorescence ) and thus invisible to the human eye, and the emitted light is in the visible region

– Chemiluminescence (sometimes called "chemoluminescence") is the emission of light as the result of a chemical reaction. There may also be limited emission of heat

Page 22: Teledyne Analytical Instruments

22

Useful Terminology

– GC Speciation: Gas Chromatography Speciation Speciation is the process by which new substance and matters come to exist. The common use

for speciation is biological where new species arise. In Gas Chromatography, speciation is changing the compound structure of the element by either breaking organic compound or adding Hydrogen.

– Flame Ionization Detector FID The operation of the FID is based on the detection of ions formed during combustion of organic

compounds in a hydrogen flame. The generation of these ions is proportional to the concentration of organic species in the

sample gas stream. In order to detect these ions, two electrodes are used to provide a potential difference

Page 23: Teledyne Analytical Instruments

23

Useful TerminologyBeta Ray Attenuation

At the beginning of the sampling period, beta ray transmission is measured across a clean section of filter tape.

This section of filter tape is then mechanically advanced to the sampling inlet. Particulate matter is then drawn into the sample inlet and deposited on the filter paper. At the completion of the sampling period, the filter tape is returned to its original location and the beta ray transmission is re-measured.

The difference between the two measurements is used to determine, with exceptional accuracy, the particulate concentration.

Distributed antenna system, or DAS

Distributed antenna system is a network of spatially separated antenna nodes connected to a common source via a transport medium that provides wireless service within a geographic area or structure. DAS antenna elevations are generally at or below the clutter level and node installations are compact.

Page 24: Teledyne Analytical Instruments

24

Useful Terminology– MFC calibrator : Mass flow controller is a device used to measure and control the flow of fluids and gases. A

mass flow controller is designed and calibrated to control a specific type of fluid or gas at a particular range of flow rates. The MFC can be given a setpoint from 0 to 100% of its full scale range but is typically operated in the 10 to 90% of full scale where the best accuracy is achieved. The device will then control the rate of flow to the given setpoint. MFCs can be either analog or digital, a digital flow controller is usually able to control more than one type of fluid or gas whereas an analog controller is limited to the fluid for which it was calibrated.

– Data Acquisition System: DAS or DAQ Data acquisition is the process of sampling signals that measure real world physical

conditions and converting the resulting samples into digital numeric values that can be manipulated by a computer. Data acquisition systems typically convert analog waveforms into digital values for processing. The components of data acquisition systems include:

- Sensors that convert physical parameters to electrical signals. - Signal conditioning circuit to convert sensor signals into a form that can be converted to

digital values. - Analog-to-digital converters, which convert conditioned sensor signals to digital values.

Page 25: Teledyne Analytical Instruments

25

Useful TerminologyTrace gases– CO (currently a Criteria Pollutant)– SO2 (currently a Criteria Pollutant)– NO/NOy– NH3 (PM2.5 precursor )– HNO3 (roles in PM2.5 & O3 formation)Particulate Matter (2.5 & 10) Particles can come in almost any shape or size, and can be solid particles or liquid droplets. We

divide particles into two major groups. These groups differ in many ways. One of the differences is size, we call the bigger particles PM10 and we call the smaller particles PM2.5 PM10 are between 2.5 and 10 micrometers (from about 25 to 100 times thinner than a human hair). PM10 (which stands for Particulate Matter up to 10 micrometers in size) cause less severe health effects.PM2.5 are smaller than 2.5 micrometers (100 times thinner than a human hair).

Photo Multiplier Tube or PMT : The amount of fluoresced UV produced in the sample chamber is much less than the intensity of excitation UV source lamp. Therefore a much more sensitive device is needed to detect this light with enough resolution to be meaningful. Therefore the 6400E uses a PMT for this purpose.

Page 26: Teledyne Analytical Instruments

26

Useful Terminology

– zero and span stability : First we need to talk about Zero and Span Calibration For example, if we calibrate the level transmitter for 4 to

20 mA with the range of 0 mmH2o (The millimeter of water (mmH2O) to 5000 mmH20, for 0 mmH2o we must get 4 mA and if it not 4 mA(say 4.5mA) means, we start to adjust the zero adjustment to make 4 mA for 0 mmH2o. And for 5000 mmH2o we must get 20 mA and if it not 20 mA (say 19.69 mA) means, we start to adjust the span to 20 mA

Once Calibration is done then the stability of this calibration is very important.– high signal to noise ratio: Signal-to-noise ratio (often abbreviated SNR or S/N) is a measure used in science and

engineering that compares the level of a desired signal to the level of background noise. It is defined as the ratio of signal power to the noise power

– Scrubber Scubber systems are a diverse group of air pollution control devices that can be used to remove

some particulates and/or gases .

Page 27: Teledyne Analytical Instruments

27

Useful TerminologyInstrument Specifications

Range: The limit of measurement values that an instrument is capable of reading. The dimension being measured must fit inside this range. In our instruments the range are measured either by PPM (Parts per million or milligrams per liter) PPB (parts per billion) PPM/PPB=1000

Noise is made up of extraneous information that is unwanted because it degrades the accuracy and precision of an analysis. the zero and the span are characteristics of the noise, the span is a percentage of the signal at near full range and the zero is the offset value. Notice the zero is specified as an absolute value, while the slope is specified as a percentage. The RMS (root mean square) defines the noise measurement as an averaged value over time, rather than a peak, and also take into account slow or long term.

Span Noise is the variability of the analyzer while measuring a gas whose concentration is 80% of the analyzer’s range (span gas).

Span Gas Span gas is a gas specifically mixed to match the chemical composition of the type of gas being

measured at near full scale of the desired measurement range.

Zero Noise is the variability of the analyzer’s response when you input zero gas (gas without any of

the target gas in it.)

Lower Detectable Limit (also called LDL, Detection Limit, or Sensitivity) is defined as twice the noise. This gives a Signal-to-Noise Ratio of 2:1 at the LDL.

Page 28: Teledyne Analytical Instruments

28

Useful TerminologyInstrument Specifications

ZERO DRIFT: The change in the instrument output of a continuous monitoring instrument over a stated time period during which the instrument is not recalibrated when the input signal remains at zero. caused by factors such as temperature, humidity, pressure....

SPAN DRIFT: The change in the instrument output of a continuous monitoring instrument over a stated time period during which the instrument is not recalibrated. It is the percentage of drifts (offset of the error)compare to the signal at near full range

Span-Gas : A gas of known concentration that is used routinely to calibrate the output level of an analyzer.

Sensitivity The sensitivity of an instrument is the smallest amount it can measure, of whatever it's built to measure. Anything smaller than the sensitivity of the instrument, and the instrument doesn't even notice it. For example, a laboratory scale can measure the weight of a hair, but a truck scale can't. We say that the laboratory scale's sensitivity is much smaller (or lower) than the truck scale's sensitivity.

LINEARITY: is the amount of deviation from an instrument's ideal straight line relationship between the instrument’s output and the measured variable. Linearity in other word, is the closeness of a curve that approximates to a straight line throughout measurement range. Due to imperfections in pipe ID, coefficient of expansion of metals used in instruments, even the precise measurements have a built-in error in measurements of flow.

Page 29: Teledyne Analytical Instruments

29

ELECTRO-STATIC DISCHARGE ELECTRO-STATIC DISCHARGE

Modern electronic devices such as the types used in the various electronic assemblies of your analyzer, are very small, require very little power and operate very quickly. Unfortunately, the same characteristics that allow them to do these things also make them very susceptible to damage from the discharge of static electricity. Damage to components occurs when these static charges come into contact with an electronic device.

Temporary anti-ESD work area Always wear an Anti-ESD wrist strap when working on the electronic assemblies of your

analyzer.

Simply touching a grounded piece of metal is insufficient. While this may temporarily bleed off static

Always store sensitive components and assemblies in anti-ESD storage bags or bins

MAINTENANCE PROCEDURES please refer to P 214 of the manual

Page 30: Teledyne Analytical Instruments

TELEYNE Team Thank you All