Technical Specification FLUXUS® ADM 7x07 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 1 FLUXUS FLUXUS Measurement with transducers mounted by Variofix L ADM 7407 ADM 7907 Liquid Ultrasonic Flowmeter for Permanent Installation Designed for wall mounting or installation in 19'' rack systems Features • Precise bi-directional and highly dynamic flow measure- ment with the non-intrusive clamp-on technology • High precision at fast and slow flow rates, high temper- ature and zero point stability • Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results • User-friendly design • Transducers available for a wide range of inner pipe diameters ( ) and fluid temperatures ( ) • approved transducers for hazardous areas available • HybridTrek automatically switches between transit time and NoiseTrek mode of measurement when high partic- ulate flows are encountered Applications • Chemical industry • Petrochemical industry • Oil and gas industry • Pharmaceutical industry • Semiconductor industry • Mechanical engineering • Water and wastewater industry 6...6500 mm -40...+400 °C ATEX, IECEx, FM
36
Embed
Technical Specification FLUXUS® ADM 7x07...4 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 FLUXUS® ADM 7x07 Technical Specification Number of Sound Paths The number of sound paths is the number
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Technical Specification
FLUXUS® ADM 7x07
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 1
FLUXUS
FLUXUS
Measurement with transducers mounted by Variofix L
ADM 7407
ADM 7907
Liquid Ultrasonic Flowmeter for Permanent Installation
Designed for wall mounting or installation in 19'' rack systems
Features
• Precise bi-directional and highly dynamic flow measure-ment with the non-intrusive clamp-on technology
• High precision at fast and slow flow rates, high temper-ature and zero point stability
• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results
• User-friendly design
• Transducers available for a wide range of inner pipediameters ( ) and fluid temperatures ( )
• approved transducers for hazardous areas available
• HybridTrek automatically switches between transit time and NoiseTrek mode of measurement when high partic-ulate flows are encountered
Applications
• Chemical industry
• Petrochemical industry
• Oil and gas industry
• Pharmaceutical industry
• Semiconductor industry
• Mechanical engineering
• Water and wastewater industry
6...6500 mm-40...+400 °C
ATEX, IECEx, FM
Dan Macey
1
FLUXUS® ADM 7x07 Technical Specification
Table of Contents
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-012
Function ........................................................................................................................................................... 3Measurement Principle ..................................................................................................................................... 3Calculation of Volumetric Flow Rate ................................................................................................................. 3Number of Sound Paths.................................................................................................................................... 4Typical Measurement Setup ............................................................................................................................. 5
Clamp-on Temperature Probe (optional) ................................................................................................... 34
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 3
Technical Specification FLUXUS® ADM 7x07
Function
Measurement Principle
Transit Time Difference Principle
In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time dif-ference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a sec-ond transducer. These signals are emitted alternately in the flow direction and against it.
As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flowdirection is shorter than against the flow direction.
The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocityalong the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to ob-tain the area averaged flow velocity, which is proportional to the volumetric flow rate.
Two integrated microprocessors control the entire measuring process. This allows the flowmeter to removedisturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.
HybridTrek
If the gaseous or solid content in the medium increases occasionally during measurement, a measurementwith the transit time difference principle is no longer possible. NoiseTrek mode will then be selected by theflowmeter. This measurement method allows the flowmeter to achieve a stable measurement even with highgaseous or solid content
The transmitter can switch automatically between transit time and NoiseTrek mode without any changes tothe measurement setup.
Calculation of Volumetric Flow Rate
= kRe . A . ka . ∆t/(2 . tfl)
where
Path of the ultrasonic signal Transit time difference ∆t
- volumetric flow ratekRe - fluid mechanics calibration factorA - cross-sectional pipe areaka - acoustical calibration factor∆t - transit time differencetfl - transit time in the medium
V·
V·
4 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Number of Sound Paths
The number of sound paths is the number of transits of the ultrasonic signal through the medium in the pipe.Depending on the number of sound paths, the following methods of installation exist:
• reflection modeThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.
• diagonal modeThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe. In the case of a high signal attenuation by the medium, pipe and coatings, diagonal mode with 1 sound path will be used.
The preferred method of installation depends on the application. While increasing the number of sound pathsincreases the accuracy of the measurement, signal attenuation increases as well. The optimum number ofsound paths for the parameters of the application will be determined automatically by the transmitter.
As the transducers can be mounted with the transducer mounting fixture in reflection mode or diagonal mode,the number of sound paths can be adjusted optimally for the application..
a - transducer distance
Reflection mode, number of sound paths: 2
Diagonal mode, number of sound paths: 3
Diagonal mode, number of sound paths: 1 Diagonal mode, number of sound paths: 1,negative transducer distance
a
a
a > 0 a < 0
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 5
Technical Specification FLUXUS® ADM 7x07
Typical Measurement Setup
Example of a measurement setup in reflection mode
Example of a heat flow measurement
transducers
transmitter
RS232
analog/binary outputs analog inputs
power supplyRS485(optional)
transducers
transmitter
temperature probetemperature probe
supply line return line
volu
met
ric fl
ow r
ate
supply temperature return temperature
6 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Flow Transmitter
Technical Data
FLUXUS ADM 7407 ADM 7407 A2 ADM 7907design standard field device field device, ATEX zone 2 19 " module
measurementmeasurement principle transit time difference correlation principle,
automatic NoiseTrek selection for measurements with high gaseous or solid contentflow velocity 0.01...25 m/srepeatability 0.15 % of reading ±0.01 m/smedium all acoustically conductive liquids with < 10 % gaseous or solid content in volume
(transit time difference principle)temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5M-1985
accuracy1
with standard calibration ±1.6 % of reading ±0.01 m/swith extended calibration (optional)
±1.2 % of reading ±0.01 m/s
with field calibration2 ±0.5 % of reading ±0.01 m/sflow transmitterpower supply 100...240 V/50...60 Hz or
20...32 V DCpower consumption < 15 Wnumber of flow measuring channels
1, optional: 2
signal attenuation 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material aluminum, powder coated aluminumdegree of protectionaccording to IEC/EN 60529
IP65 IP65 IP20
dimensions see dimensional drawing 42HP x 3U(without back panel)
see dimensional drawingweight 2.8 kg 1.7 kgfixation wall mounting, optional: 2 " pipe mounting 19 " rack mountingoperating temperature -20...+60 °C display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishexplosion protection
ATEX
zone - 2 -marking -
II3G Ex nA nC ic IIC T4 GcTa -20...+60 °CII3D Ex tc IIIC T 100 °C Dc IP65
-
type of protection - gas: non sparking,dust: protection by enclosure
-
1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %
standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuescommunicationinterface - process integration:
optional: RS485 (Modbus, emitter) or HART
- diagnosis: RS232
- process integration:optional: RS485 (Modbus, emitter)
or HART- diagnosis: RS2323
- process integration:optional: RS485 (Modbus, emitter)
or HART- diagnosis: RS232
serial data kit (optional)
software (all Windows™ ver-sions)
-FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)
-FluxKoef: creating medium data setsсable RS232 RS2323 RS232
adapter RS232 - USB RS232 - USB3 RS232 - USB
outputs (optional)The outputs are galvanically isolated from the transmitter.
number on requestcurrent output
current output- range 0/4...20 mA- accuracy 0.1 % of reading ±15 μA- active output Rext < 500 Ω- passive output Uext = 4...24 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range 4...20 mA- passive output Uext = 10...24 V
voltage outputrange 0...1 V or 0...10 Vaccuracy 0...1 V: 0.1 % of reading ±1 mV
0...10 V: 0.1 % of reading ±10 mVinternal resistance Ri = 500 Ω
frequency outputrange 0...5 kHzopen collector 24 V/4 mA
binary outputReed relay - 48 V/0.25 Aopen collector - 24 V/4 mAoptorelay 26 V/100 mA -binary output as alarm output- functions limit, change of flow direction or error limit, change of flow direction or
errorbinary output as pulse output- pulse value 0.01...1000 units 0.01...1000 units- pulse width 1...1000 ms 80...1000 ms3 connection of the interface RS232 outside of explosive atmosphere (housing cover open)
FLUXUS ADM 7407 ADM 7407 A2 ADM 7907
8 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Dimensions
inputs (optional)The inputs are galvanically isolated from the transmitter.
number max. 4, on requesttemperature input
type Pt100/Pt1000connection 4-wirerange -150...+560 °C resolution 0.01 Kaccuracy ±0.01 % of reading ±0.03 K
current inputaccuracy 0.1 % of reading ±10 μA 0.1 % of reading ±10 μA 0.1 % of reading ±10 μAactive input Ui = 24 V, Ri = 50 Ω, Pi < 0.5 W,
not short-circuit proofUi = 24 V, Ri = 50 Ω, Pi < 0.5 W,
not short-circuit proofUi = 15 V, Ri = 50 Ω, Pi < 0.5 W,
not short-circuit proof- range 0...20 mA 0...20 mA 0...20 mApassive input Ri = 50 Ω, Pi < 0.3 W Ri = 50 Ω, Pi < 0.3 W Ri = 50 Ω, Pi < 0.3 W- range -20...+20 mA -20...+20 mA -20...+20 mA
voltage inputrange 0...1 Vaccuracy 0.1 % of reading ±1 mVinternal resistance Ri = 1 MΩ
FLUXUS ADM 7407
FLUXUS ADM 7407 ADM 7407 A2 ADM 7907
70*
10
Ø 4.5
163
200
287265 screw holes for wall mounting
thread: 6x M20 x 1.5cable gland: max. 6x M20
* ADM 7407 A2: +10 mm above the display
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 9
Technical Specification FLUXUS® ADM 7x07
2 " Pipe Mounting Kit (optional)
FLUXUS ADM 7907
in mm
FLUXUS ADM 7407
for vertical and horizontal pipes
208
129
222
15
123
110
M2.
544
213
129
170
198
10 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Terminal Assignment
2 The number, type and terminal assignment of the outputs and inputs will be customized.
FLUXUS ADM 7407
power supplyterminal strip KL3
terminal connection AC connection DCPE earth earthN(-) neutral - DCL(+) phase + DC
terminal strip KL2temperature probe passive current source active current source
terminal connection connection with extension cable
connection of an active input
connection of a passive input
T1a...T4a red red not connected not connectedT1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connectedT1B...T4B white white not connected -S1...S4 shield shield not connected not connected
terminal strip KL3temperature probe passive current source active current source
terminal connection connection with extension cable
connection of an active input
connection of a passive input
T1a...T4a red red not connected not connectedT1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connectedT1B...T4B white white not connected -S1...S4 shield shield not connected not connected
# # # #
# #
& &
# #
) #
&
&
'
&
&
'
#
#
#
%
%
%
$
$
$
%
$
$
%
$
$
%
%
12 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Transducers
Transducer Selectiontransducer order code
FSK 100 200 3600 6500
FSM 50 100 2000 3400
FSP 25 50 200 600
FSQ 10 25 150 400
FSS 6 10 70
5 10 50 100 500 1000 5000inner pipe diameter [mm]
recommended possible
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 13
Technical Specification FLUXUS® ADM 7x07
Transducer Order Code
1, 2 3 4 5, 6 7, 8 9...11 12, 13 no. of character
tra
nsdu
cer
tra
nsdu
cer
fre
quen
cy
- oper
atin
g te
mpe
ratu
re
expl
osi
onpr
otec
tion
conn
ectio
nsy
stem
- exte
nsio
n ca
ble
/ optio
n
description
FS set of ultrasonic flow transducers for liquids measurement, shear wave
K 0.5 MHz
M 1 MHz
P 2 MHz
Q 4 MHz
S 8 MHz
N normal temperature range
E extended temperature range (shear wave transducers with trans-ducer frequency M, P, Q)
A1 ATEX zone 1 and IECEx zone 1
A2 ATEX zone 2
F2 FM Class I Div. 2
NN not explosion proof
AS with Amphenol connector (not explosion proof transducers)
TS direct connection or connection via junction box
XXX cable length in m, for max. length of extension cable see page 29
connection system TS:
0 m: without junction box
> 0 m: with junction box JB01 (zone 1), JB02 (ATEX zone 2, FM), JB03 (not explosion proof), JBP2 (transducers IP68, ATEX zone 2), JBP3 (transducers IP68, not explosion proof)
LC long transducer cable (only FSK)
IP68 degree of protection IP68 with connection system TS
OS housing with stainless steel 316 (with connection system TS)
example
FS M - N A1 TS - 030 shear wave transducer 1 MHz, normal temperature range, zone 1, connection system TS with junction box JB01 and exten-sion cable 30 m
inner pipe diameter dmin. extended mm 100 50 25 10min. recommended mm 200 100 50 25max. recommended mm 3600 2000 200 150max. extended mm 6500 3400 600 400pipe wall thicknessmin. mm - - - -max. mm - - - -materialhousing PEEK with stainless steel
cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PEEK PEEK PEEK PEEKdegree of protection according toIEC/EN 60529
IP65 IP65 IP65 IP65
transducer cabletype 1699 1699 1699 1699length m 5 4 4 3dimensionslength l mm 126.5 62.5 62.5 39width b mm 51 32 32 22height h mm 67.5 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -40 -40 -40 -40max. °C +130 +130 +130 +130temperaturecompensation
x x x x
explosion protection
ATEX/IECEx
transducer FSK-NA1TSFSK-NA1TS/OS
FSM-NA1TSFSM-NA1TS/OS
FSP-NA1TSFSP-NA1TS/OS
FSQ-NA1TSFSQ-NA1TS/OS
categoryzone
gas: 2G dust: 2D1 21
gas: 2G dust: 2D1 21
gas: 2G dust: 2D1 21
gas: 2G dust: 2D1 21
explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C +180 +180 +180 +180marking
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,
powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
l
hb
l
hb
l
hb
hb
l
II2GII2D
0637 II2GII2D
0637 II2GII2D
0637 II2GII2D
0637
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 15
Technical Specification FLUXUS® ADM 7x07
Shear Wave Transducers (zone 1, IP68)
technical type CDK1LI1 CDM2LI1 CDP2LI1order code FSK-NA1TS/IP68 FSM-NA1TS/IP68 FSP-NA1TS/IP68transducer frequency MHz 0.5 1 2
inner pipe diameter dmin. extended mm 100 50 25min. recommended mm 200 100 50max. recommended mm 3600 2000 200max. extended mm 6500 3400 600pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PEEK with stainless steel
cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)
PEEK with stainless steel cap 316Ti (1.4571)
contact surface PEEK PEEK PEEKdegree of protection according toIEC/EN 60529
IP68 IP68 IP68
transducer cabletype 2550 2550 2550length m 12 12 12dimensionslength l mm 128.5 70 70width b mm 54 32 32height h mm 83.5 46 46dimensional drawing
operating temperaturemin. °C -40 -40 -40max. °C +100 +100 +100temperaturecompensation
explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +180 +180 +180marking
Ex q IIC T6...T3 GbEx tb IIIC TX Db
Ex q IIC T6...T3 GbEx tb IIIC TX Db
Ex q IIC T6...T3 GbEx tb IIIC TX Db
certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling
dust: protection by enclosure
gas: powder fillingdust: protection by enclosure
gas: powder fillingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
1 test conditions: 3 months/2 bar (20 m)/20 °C
1 1 1
l
hb
l
hb
l
hb
II2GII2D
0637 II2GII2D
0637 II2GII2D
0637
16 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Shear Wave Transducers (zone 1, extended temperature range)
technical type CDM2E85 CDP2E85 CDQ2E85order code FSM-EA1TS
FSM-EA1TS/OSFSP-EA1TSFSP-EA1TS/OS
FSQ-EA1TSFSQ-EA1TS/OS
transducer frequency MHz 1 2 4
inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 2000 200 150max. extended mm 3400 600 400pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PI with stainless steel
cap 304 (1.4301), option OS: 316L (1.4404)
PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PI PI PIdegree of protection according toIEC/EN 60529
IP56 IP56 IP56
transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 62.5 62.5 39width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperaturecompensation
x x x
explosion protection
ATEX/IECEx
transducer FSM-EA1TSFSM-EA1TS/OS
FSP-EA1TSFSP-EA1TS/OS
FSQ-EA1TSFSQ-EA1TS/OS
categoryzone
gas: 2G dust: 3D1 22
gas: 2G dust: 3D1 22
gas: 2G dust: 3D1 22
explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +225 +225 +225marking
Ex e q IIC T6...T2 GbEx tb IIIA TX Db
Ex e q IIC T6...T2 GbEx tb IIIA TX Db
Ex e q IIC T6...T2 GbEx tb IIIA TX Db
certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,
powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
l
hb
l
hb
hb
l
II2GII2D
0637 II2GII2D
0637 II2GII2D
0637
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 17
Technical Specification FLUXUS® ADM 7x07
Shear Wave Transducers (ATEX zone 2, FM or not explosion proof)
technical type CDK1N52 CLK1N52 CDS1N52order code FSK-NA2TS, FSK-NA2TS/OS
transducer frequency MHz 0.5 0.5 8inner pipe diameter dmin. extended mm 100 100 6min. recommended mm 200 200 10max. recommended mm 3600 3600 70max. extended mm 6500 6500 70pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PEEK with stainless steel cap 304
(1.4301), option OS: 316L (1.4404)PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
stainless steel 304 (1.4301)
contact surface PEEK PEEK PEIdegree of protection according toIEC/EN 60529
IP67 IP67 IP65
transducer cabletype 1699 1699 1699length m 5 9 2dimensionslength l mm 126.5 126.5 25width b mm 51 51 13height h mm 67.5 67.5 17dimensional drawing
operating temperaturemin. °C -40 -40 -30max. °C +130 +130 +130temperaturecompensation
x x x
explosion protection
ATEX
transducer FSK-NA2TSFSK-NA2TS/OS
FSK-NA2TS/LCFSK-NA2TS/LC/OS
-
categoryzone
gas: 3G dust: 3D2 22
gas: 3G dust: 3D2 22
-
explosion protection temperature (pipe surface)min. °C -55 -55 -max. °C +190 +190 -marking
II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc
II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc
-
certification - - -type of protection gas: non sparking
dust: protection by enclosuregas: non sparkingdust: protection by enclosure
-
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C -
FM
transducer FSK-NF2TSFSK-NF2TS/OS
FSK-NF2TS/LCFSK-NF2TS/LC/OS
FSS-NF2TS
explosion protection temperaturemin. °C -40 -40 -40max. °C +125 +125 +125marking NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860type of protection non incendive non incendive non incendive
l
hb
l
hb
hb
l
18 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Shear Wave Transducers ATEX zone 2, FM or not explosion proof)
technical type CDM2N52 CDP2N52 CDQ2N52order code FSM-NA2TS, FSM-NA2TS/OS
inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 2000 200 150max. extended mm 3400 600 400pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PEEK with stainless steel cap 304
(1.4301), option OS: 316L (1.4404)PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PEEK PEEK PEEKdegree of protection according toIEC/EN 60529
IP67 IP67 IP67
transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 62.5 62.5 39width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperaturecompensation
x x x
explosion protection
ATEX
transducer FSM-NA2TSFSM-NA2TS/OS
FSP-NA2TSFSP-NA2TS/OS
FSQ-NA2TSFSQ-NA2TS/OS
categoryzone
gas: 3G dust: 3D2 22
gas: 3G dust: 3D2 22
gas: 3G dust: 3D2 22
explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +190 +190 +190marking
II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc
II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc
II3G Ex nA IIC T6...T3 GcTa -55...+190 °CII3D Ex tc IIIC TX Dc
certification - - -type of protection gas: non sparking
dust: protection by enclosuregas: non sparkingdust: protection by enclosure
gas: non sparkingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
FM
transducer FSM-NF2TSFSM-NF2TS/OS
FSP-NF2TSFSP-NF2TS/OS
FSQ-NF2TSFSQ-NF2TS/OS
explosion protection temperaturemin. °C -55 -55 -55max. °C +190 +190 +190marking NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860type of protection non incendive non incendive non incendive
l
hb
l
hb
hb
l
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 19
Technical Specification FLUXUS® ADM 7x07
Shear Wave Transducers (ATEX zone 2 or not explosion proof, IP68)
technical type CDK1LI8 CDM2LI8 CDP2LI8order code FSK-NA2TS/IP68
FSK-NNNTS/IP68FSM-NA2TS/IP68FSM-NNNTS/IP68
FSP-NA2TS/IP68FSP-NNNTS/IP68
transducer frequency MHz 0.5 1 2
inner pipe diameter dmin. extended mm 100 50 25min. recommended mm 200 100 50max. recommended mm 3600 2000 200max. extended mm 6500 3400 600pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PEEK with stainless steel
cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)
PEEK with stainless steel cap 316Ti (1.4571)
contact surface PEEK PEEK PEEKdegree of protection according toIEC/EN 60529
IP68 IP68 IP68
transducer cabletype 2550 2550 2550length m 12 12 12dimensionslength l mm 128.5 70 70width b mm 54 32 32height h mm 83.5 46 46dimensional drawing
operating temperaturemin. °C -40 -40 -40max. °C +100 +100 +100temperaturecompensation
explosion protection temperature (pipe surface)min. °C -40 -40 -40max. °C +90 +90 +90marking
II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc
II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc
II3G Ex nA IIC T6...T5 GcTa -40...+90 °CII3D Ex tc IIIC TX Dc
certification - - -type of protection gas: non sparking
dust: protection by enclosure
gas: non sparkingdust: protection by enclosure
gas: non sparkingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
1 test conditions: 3 months/2 bar (20 m)/20 °C
1 1 1
l
hb
l
hb
l
hb
20 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Shear Wave Transducers (connection system AS, not explosion proof)
technical type CDK1NZ7 CLK1NZ7order code FSK-NNNAS FSK-NNNAS/LCtransducer frequency MHz 0.5 0.5
inner pipe diameter dmin. extended mm 100 100min. recommended mm 200 200max. recommended mm 3600 3600max. extended mm 6500 6500pipe wall thicknessmin. mm - -max. mm - -materialhousing PEEK with stainless steel
cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)
contact surface PEEK PEEKdegree of protection according toIEC/EN 60529
IP67 IP67
transducer cabletype 1699 1699length m 5 9dimensionslength l mm 126.5 126.5width b mm 51 51height h mm 67.5 67.5dimensional drawing
operating temperaturemin. °C -40 -40max. °C +130 +130temperaturecompensation
x x
l
hb
l
hb
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 21
Technical Specification FLUXUS® ADM 7x07
Shear Wave Transducers (connection system AS, not explosion proof)
technical type CDS1NZ7order code FSS-NNNAStransducer frequency MHz 8
inner pipe diameter dmin. extended mm 6min. recommended mm 10max. recommended mm 70max. extended mm 70pipe wall thicknessmin. mm -max. mm -materialhousing stainless steel 304
(1.4301)contact surface PEIdegree of protection according toIEC/EN 60529
IP65
transducer cabletype 1699length m 2dimensionslength l mm 25width b mm 13height h mm 17dimensional drawing
operating temperaturemin. °C -30max. °C +130temperaturecompensation
x
hb
l
22 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Shear Wave Transducers (connection system AS, not explosion proof)
technical type CDM2NZ7 CDP2NZ7 CDQ2NZ7order code FSM-NNNAS FSP-NNNAS FSQ-NNNAStransducer frequency MHz 1 2 4
inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 2000 200 150max. extended mm 3400 600 400pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PEEK with stainless steel
cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)
PEEK with stainless steel cap 304 (1.4301)
contact surface PEEK PEEK PEEKdegree of protection according toIEC/EN 60529
IP67 IP67 IP67
transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 62.5 62.5 39width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperaturecompensation
x x x
l
hb
l
hb
hb
l
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 23
Technical Specification FLUXUS® ADM 7x07
Shear Wave Transducers (extended temperature range, ATEX zone 2, FM or not explosion proof)technical type CDM2E52 CDP2E52 CDQ2E52order code FSM-EA2TS, FSM-EA2TS/OS
inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 2000 200 150max. extended mm 3400 600 400pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PI with stainless steel cap 304
(1.4301), option OS: 316L (1.4404)PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PI PI PIdegree of protection according toIEC/EN 60529
IP56 IP56 IP56
transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 62.5 62.5 39width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperaturecompensation
x x x
explosion protection
ATEX
transducer FSM-EA2TSFSM-EA2TS/OS
FSP-EA2TSFSP-EA2TS/OS
FSQ-EA2TSFSQ-EA2TS/OS
categoryzone
gas: 3G dust: 3D2 22
gas: 3G dust: 3D2 22
gas: 3G dust: 3D2 22
explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +235 +235 +235marking
II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc
II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc
II3G Ex nA IIC T6...T2 GcTa -45...+235 °CII3D Ex tc IIIB TX Dc
certification - - -type of protection gas: non sparking
dust: protection by enclosuregas: non sparkingdust: protection by enclosure
gas: non sparkingdust: protection by enclosure
necessary trans-ducer mountingfixture
Variofix L or Variofix C Variofix L or Variofix C Variofix L or Variofix C
FM
transducer FSM-EF2TSFSM-EF2TS/OS
FSP-EF2TSFSP-EF2TS/OS
FSQ-EF2TSFSQ-EF2TS/OS
explosion protection temperaturemin. °C -45 -45 -45max. °C +235 +235 +235marking NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860type of protection non incendive non incendive non incendive
l
hb
l
hb
hb
l
24 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Shear Wave Transducers (extended temperature range, not explosion proof,connection system AS)
technical type CDM2EZ7 CDP2EZ7 CDQ2EZ7order code FSM-ENNAS FSP-ENNAS FSQ-ENNAStransducer frequency MHz 1 2 4
inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 2000 200 150max. extended mm 3400 600 400pipe wall thicknessmin. mm - - -max. mm - - -materialhousing PI with stainless steel
cap 304 (1.4301)PI with stainless steel cap 304 (1.4301)
PI with stainless steel cap 304 (1.4301)
contact surface PI PI PIdegree of protection according toIEC/EN 60529
IP65 IP65 IP65
transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 62.5 62.5 39width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing
operating temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperaturecompensation
x x x
l
hb
l
hb
hb
l
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 25
Technical Specification FLUXUS® ADM 7x07
Transducer Mounting Fixture
Order Code
1, 2 3 4 5 6 7...9 10, 11 no. of character
tra
nsdu
cer
mou
ntin
g fix
ture
tra
nsdu
cer
- mea
surin
g m
ode
size
- fixat
ion
oute
r pi
pe
diam
eter
/ optio
n
description
VL Variofix L
VС Variofix C
WI transducer clamping fixture for WaveInjector
K transducers with transducer frequency K
M transducers with transducer frequency M, P
Q transducers with transducer frequency Q
S transducers with transducer frequency S
D reflection mode or diagonal mode
R reflection mode
S small
M medium
L large
S tension straps
W welding
N without fixation
002 10...20 mm
004 20...40 mm
T36 40...360 mm
013 10...130 mm
036 130...360 mm
092 360...920 mm
200 920...2000 mm
450 2000...4500 mm
940 4500...9400 mm
NDR any
IP68 degree of protection IP68
OS housing with stainless steel 316
Z special design
example
VL M - D S - S 200 Variofix L and tension straps for transducers with transducer frequency M, P
inner length:VCK-*L: 500 mmVCK-*S: 350 mmVCM: 400 mmVCQ: 250 mm
dimensions:VCK-*L: 560 x 122 x 102 mm,option IP68: 560 x 126 x 120 mmVCK-*S: 410 x 122 x 102 mm,option IP68: 410 x 126 x 120 mmVCM: 460 x 96 x 80 mmVCQ: 310 x 85 x 62 mm
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 27
Technical Specification FLUXUS® ADM 7x07
transducer clamping fixture for WaveInjector WI
see Technical SpecificationTSWaveInjectorVx-x
transducer clamping fixture
28 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Coupling Materials for Transducers
Technical Data
normal temperature range(4th character of transducer order code = N)
extended temperature range(4th character of transducer order code = E)
WaveInjector WI-400
< 100 °C 100...170 °C < 150 °C 150...200 °C < 280 °C 280...400 °C< 2 h coupling com-
pound type Ncoupling com-pound type E
coupling com-pound type E
coupling com-pound type E or H
coupling foiltype A
coupling foiltype B
< 24 h coupling com-pound type N
coupling com-pound type E
coupling com-pound type E
coupling foiltype VT
coupling foiltype A
coupling foiltype B
long time measure-ment
indoor coupling com-pound type N
coupling com-pound type E
coupling foiltype VT1
coupling foiltype VT2
coupling foiltype A
coupling foiltype B
outdoor coupling foiltype VT
coupling foiltype VT
coupling foiltype VT1
coupling foiltype VT2
coupling foiltype A
coupling foiltype B
1 < 5 years2 < 6 months
type order code operating temperature
material remark
°Ccoupling compound type N
990739-1 -30...+130 mineral grease paste
coupling compound type E
990739-2 -30...+200 silicone paste
coupling compound type H
990739-3 -30...+250 fluoropolymer paste
coupling foil type A 990739-7 max. 280 plomb
coupling foil type B 990739-8 > 280...400 silver
coupling foil type VT 990739-0 -10...+150,short-time peak max. 200
fluoroelastomer for transducers with transducer frequency G, H, K
990739-6 for shear wave transducers with transducer frequency M, P
990739-14 for shear wave transducers IP68 and Lambwave transducers with transducer frequency M, P
990739-15 for shear wave transducers with transducer frequency Q
990739-5 for Lambwave transducers with transducer frequency Q
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 29
Technical Specification FLUXUS® ADM 7x07
Connection Systemsconnection system TSconnection with extension cable direct connection transducers
technical type
*****8*
****LI*
*****52
connection system ASconnection with extension cable transducers
technical type*****Z7*****C3
transducer frequency(3d character of transducer
order code)
G, H, K M, P Q S
TS
x l x l x l x lcable length m 5 ≤ 300 4 ≤ 300 3 ≤ 90 2 ≤ 40
x, y - transducer cable lengthl - max. length of extension cable
(only ADM 7407, ADM 7407 A2)
tran
smitt
er
x
l
JB01
tran
smitt
er
x
tran
smitt
erx
l
JB01, JBP2, JBP3
tran
smitt
er
xtr
ansm
itte
r
x
l
JB02, JB03
tran
smitt
er
x
tra
nsm
itter
l x y
30 TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01
FLUXUS® ADM 7x07 Technical Specification
Transducer Cable
Technical Data
transducer cable extension cabletype 1699 2550 6111 2551 2615connection system AS TSstandard length m see table above
10-
max. length m - see table aboveoperating temperature °C -55...+200 -40...+100 -100...+225 -25...+80 -40...+70properties longitudinal water
tighthalogen free
fire propagation test according to IEC 60332-1
combustion test according to IEC 60754-2
sheathmaterial stainless steel 304
(1.4301)- stainless steel 304
(1.4301)- -
option OS: 316L (1.4404)
option OS: 316L (1.4404)
outer diameter mm 8 - 8 - -cable jacketmaterial PTFE PUR PFA TPE-O PURouter diameter mm 2.9 5.2 ±0.2 2.7 8 12thickness mm 0.3 0.9 0.5 2color brown gray white black blackshield x x x x x
1
TSFLUXUS_F7V1-5-1EN_Leu, 2012-07-01 31
Technical Specification FLUXUS® ADM 7x07
Junction Box
Technical Data
Dimensions
technical type JB01S4E3M JB02 JB03 JBP2 JBP3dimensions see dimensional