Top Banner
Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences Division
11

Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

Dec 23, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

Tech Talk 5/2014

NETL Raman gas composition monitor overview

Michael Buric, Steven Woodruff, Benjamin ChorpeningSensors and Controls Team, Thermal Sciences Division

Page 2: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

2

Intro to Raman spectroscopy

• Optical technique which measures hydrocarbons and homonuclear diatomics

• Linear response to concentration• Raman spectrometers are becoming common; but

still no fast, gas Raman spectrometers• Small signals

Merit Review May 10, 2011

Page 3: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

3

Why do Raman on gases in Energy Applications? (available technologies)

• Need to measure N2, H2, O2, CO, CO2, H2O, and hydrocarbons simultaneously and FAST

• Electrochemical: SLOW – many seconds for a measurement• TDLAS (tunable diode laser absorption): IR detectors and tunable

lasers (somewhat costly) needed, IR water absorption, possibly multiple lasers for multiple species

• GC (gas chromatography): elution times are LONG, multiple columns required, some species hard to discriminate

• FTIR: low SNRs in some cases (low spectral power density), water absorption problems, moving parts, calibrations

• Mass spectrometers: low pressure operation, slower• Raman: measures all species simultaneously except noble

gases, single visible laser, cheap detectors, LOW SNRs• Enhanced Raman using waveguides: Promising!

Page 4: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

4

Hollow guide (NEW!)

Free-space configuration (OLD) (1- lens imaging)

Review: Waveguide-enhanced Raman

Waveguide

Laser Spectrometer

Notch filter

L1 L2

Laser

Spectrometer

L3

L2

L1

MNotch filter

Ps/P0 is very small! (10-13 or less)

L can enhance by 103 or more

00PLKPs

Page 5: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

5

NETL Raman instrument capabilities

• Measure various gas streams of interest• Natural gas component analysis (methane, ethane,

propane, butane, etc.)• Homonuclear diatomic measurements (N2, H2, O2)

• Real-time (<1second) continuous measurement• Turbine or other dynamic system control• Replace GC analysis at similar (or lower) cost• Provide continuous monitoring with little human

input• Operate from 5-800psig (fuel pipelines, etc.)• Maintain a high-level of user safety

Page 6: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

6

Enhanced Spontaneous Raman waveguide-based system

• SPEED: One-second response time (or less)

• ACCURACY: Sub-percent for all species with little cross sensitivity

• SIMPLICITY/STABILITY: Obtains all species at once with no tunable lasers, no pump power control

2

Page 7: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

7

Industrial prototype operations timeline• 2007-2008 – Research on various types of

waveguides for Raman enhancement• 2009-2010 – Initial work with capillary waveguides

and Patent Application• 2011-2012 – Industrial testing of first prototype• 2012-Present – Redesign and industrial testing of

second prototype

Page 8: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

8

Current Detection Limits (1s Integration, 100 mW, 1 m Capillary, OO Spectrometer, computed for binary mixtures)

Co

mp

on

ent

0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

Methane, 0.08 -> .014

Ethane, 0.046 -> 0.014

Propane, 0.035 -> 0.006

iso-Butane, 0.082 -> 0.014

Ethylene, 0.10 -> 0.018

n-Butane, 0.044 -> 0.0076

CO2, 0.57 -> 0.097

N2, 0.72 -> 0.12

O2, 0.68 -> 0.12

H2, 0.11 -> 0.018 CO, 0.79 -> 0.13

Detection Limit (%)

Red: Commercial capillaryBlue: NETL Capillary

Page 9: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

9

Example: Gasifier Measurement Results(NCCC, 3/14, coal derived syngas)

• Many more data points– Track transient

events• Agreement within

GC/Raman error

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

3/25/14 12:00 PM 3/25/14 1:55 PM 3/25/14 3:50 PM 3/25/14 5:45 PM

Com

posi

tion

%

Time

GC CO

GC CO2

GC CH4

GC H2

Raman CO

Raman CO2

Raman CH4

Raman H2

Page 10: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

10

Example: Chemical Looping Reactor Measurements(NETL CLR, 4/14, Hematite and NG)

• CO2, methane and other species monitored

• Good agreement with IR absorption measurements (Horiba Via-510), but more species measured

0

1

2

3

4

5

6

4/2/14 8:24 PM 4/2/14 9:21 PM 4/2/14 10:19 PM 4/2/14 11:16 PM

Com

posi

tion

%

Time

Analyzer CO2Raman CO2

Page 11: Tech Talk 5/2014 NETL Raman gas composition monitor overview Michael Buric, Steven Woodruff, Benjamin Chorpening Sensors and Controls Team, Thermal Sciences.

11

Conclusions

• RGA measures 10 or more gases simultaneously

• Measurements made to <1% in 1 second• Continuous gasifier or other energy-

system measurements are possible• Recalibration not necessary, only

background re-measurement to maintain accuracy

• Capillary waveguides are durable• Portable analyzers suitable for haz. ops