TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2 ,M. Fernandez 3 , C. Figueiredo 1 , M. Moll 1 , H. Neugebauer 1,4 1 CERN / 2 INFN, University of Florence 3 IFCA-Santander/ 4 University of Hamburg
23

TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

Dec 23, 2015

Download

Documents

Emil Mills
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab

Christian Gallrapp1

M. Bruzzi2 ,M. Fernandez3, C. Figueiredo1, M. Moll1, H. Neugebauer1,4

1CERN / 2INFN, University of Florence3IFCA-Santander/ 4University of Hamburg

Page 2: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

2

TCT Setups in the CERN SSD Lab

• I-DLTS setup based on former TCT setup• eTCT setup • TCT+ setup combines TCT and eTCT

09/05/2014

I-DLTS setup

TCT+ setup

eTCT setup

Page 3: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

3

Current-Deep Level Transient Spectroscopy (I-DLTS)

09/05/2014

Page 4: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

4

I-DLTS setup• Equipment

– Huber CC505 chiller Temperature controlled (PT100)Minimum Temperature on the sample ca. -25C

– μs-pulsed LASER• Red (660nm)• IR (1064nm)

– Optics for red and IR illumination from top and bottom

– Temperature measurement on the DUT (PT1000 with Keithley 2410)

– Bias voltage up to 1000V– Bias Tee (Vmax = 200V)– Shielded Box (Louvain-Box)– Agilent Scope (2.5GHz Bandwidth)– Reference diode for red and IR

• LabView based software to loop parameters– temperature, bias voltage, pulse width, pulse intensity and repetition

09/05/2014

Page 5: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

5

I-DLTS

• Motivation– Improve understanding of charge carrier detrapping

for defect characterization – Investigate energy levels and cross-section of detrapping centers

• Previous work:– G. Kramberger, et. al.; 2012 - JINST 7 P04006

Determination of detrapping times in semiconductor detectors – M. Gabrysch, et.al.; 2012 - 21st RD50 Workshop

Charge carrier detrapping in irradiated silicon sensors after microsecond laser pulses

09/05/2014

Page 6: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

6

I-DLTS Layout

• Illumination with μs-pulsed red and IR LASER pulses (> 0.5μs)• Biasing up to 200V with maximum bandwidth (20kHz - 10GHz)• No amplifier to keep maximum bandwidth• Temperature controlled (> -25C)• Scope with upper bandwidth limit 2.5GHz09/05/2014

μs - pulsed LASER (red, IR)Front and back side illumination200Hz trigger frequency

Bias Tee

HV Power Supply

Scope 2.5GHz

DUT

Page 7: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

7

Samples and Parameter Space

• Parameter space– Temperature

• -20C to 20C in 5C steps

– Voltage• 50V, 100V, 150V, 200V

– Pulse width• 1us, 2us, 3us, 4us, 5us

– LASER Intensity• 1.7V, 1.8V, 1.9V, 2.0V

– Repetition: • Five repetitions for each scan point

• Micron Samples:– Thickness: 300μm– FZ and MCz n-in-p– Irradiation at CERN PS:

24GeV/c protons– Fluence:

non irrad; 5×1013 p/cm2 ; 5 ×1014 p/cm2 ; 1 ×1015 p/cm2

• Illumination: Red front

09/05/2014

• Further measurements: – Scan parameter space with IR front, Red and IR back– Determination of most suitable parameter space for analysis

Page 8: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

8

Current with Temperature

09/05/2014

-25C -10C

5C 20C

FZ n-in-p 1×1015 p/cm2 at 200VIntensity 2V

• Signal height after LASER pulse varies between 0mV at -25C and -1mV at 20C• Current drop during pulse varies with temperature (also voltage)

Page 9: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

9

Double Boxcar

09/05/2014D. Lang; 1974 - J. Appl. Phys. 45 Deep level transient spectroscopy: A new method to characterize traps in semiconductors‐

• Select rate window defined by t1 and t2

• Determine signal variation Signal(t1)-Signal(t2) for different temperatures

• Operation modes– t1 fixed, vary t2

– t2 fixed, vary t1

– t1/t2 fixed, vary t1 and t2

Page 10: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

10

Box Plot for variable pulse width

• t1 fixed, 200ns after pulse

• vary t2

– Box width: 1μs, 2μs, 3μs and 4μs

• Box plot varies with LASER pulse width

09/05/2014

1μs pulse3μs pulse

5μs pulse

FZ n-in-p 5 ×1014 p/cm2 at 200VIntensity 2V

Page 11: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

11

Next Steps• Influence of electronics on

signal– Undershoot in unirradiated

sample– Undershoot visible at low

temperatures (low de-trapping)

• Analysis following approach for TCT pulses (see: Kramberger, 2012 - JINST 7 P04006 )

• Simulation– transient after laser pulse– Current drop during laser pulse

09/05/2014

Page 12: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

12

Edge-TCT

09/05/2014

Page 13: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

13

eTCT setup• Equipment

– Computer controlled Peltier cooling (PT1000)Min. Temperature on the DUT -20C

– Annealing up to 60C directly in the setup– picosecond-pulsed IR LASER– Optics to illuminate the sample edge– Bias voltage up to 1000V– Wide bandwidth amplifier– Bias Tee for DC readout – EM shielded Box– Agilent Scope (2.5GHz Bandwidth)– XYZ stages with μm step width

• LabView based software to loop parameters– temperature, bias voltage and position

09/05/2014

Page 14: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

14

Recent eTCT measurements

• eTCT measurements on HV-CMOS– D. Muenstermann, et.al.; 2013 –

23rd RD50 Workshop Active pixel sensors in 180 nm HV CMOS technology for HL-LHC detector upgrades

• Irradiated Micron strip sensors– S. Wonsak, et.al.;

2014 – 24th RD50 Workshop Status of Silicon Strip Sensor Measurements at Liverpool

09/05/2014

Page 15: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

15

TCT+ a common setup for TCT and eTCT

09/05/2014

Page 16: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

16

TCT+ Setup• Equipment

– Computer controlled Peltier cooling (PT1000) with Huber CC505 chillerMin. Temperature on the DUT -20C

– picosecond-pulsed LASER• Red (660nm)• IR (1064nm)

– Optics for illumination• Top red and IR• Bottom red and IR• Sample edge IR

– Bias voltage up to 1000V– Wide bandwidth amplifier– Bias Tee for AC readout – EM shielded Box– Agilent Scope (2.5GHz Bandwidth)– XYZ stages with μm step width

• LabView based software to loop parameters– temperature, bias voltage, position and repetition

09/05/2014

Page 17: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

17

TCT+ Layout

• Combination of a conventional red and IR TCT setup with an edge-TCT setup

• Temperature controlled Peltier/Chiller cooling system.• Stage system provides μm steps in X, Y and Z

09/05/2014

IR (1064nm)

Red (660nm)

Pulsed laser with a trigger frequency of 200Hz

Page 18: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

18

• Irradiated MCz diodes (70MeV protons)

• R. Carney – Master Thesis; University of EdinburghInvestigation of Magnetic Czochralski diodes using novel TCT setup for future silicon detectors

Measurements with TCT+Irradiated MCz diodes

09/05/2014

Φ = 1 ×1013 neq/cm2

Φ = 5 ×1014 neq/cm2

Page 19: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

19

Measurements with TCT+Diodes with amplification

• V. Greco , et.al.; 2014 – 24th RD50 Workshop Preliminary results on proton irradiated LGAD PAD detectors

09/05/2014

Bias = 50V Bias = 150V Bias = 150V

Surface scan with red LASER front

Voltage scan with red LASER back

Page 20: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

20

Measurements with TCT+3D strip diodes

• Single sided 3D strip diodes on SOI from CNM

09/05/2014

Surface scan with IR LASER at 70V Surface scan with red LASER at 70V

columns columns

Page 21: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

21

Measurements with TCT+irradiated strip sensors from Micron

• TCT and eTCT on 1×1015 neq/cm2

irradiated strip sensors– S. Wonsak, et.al.;

2014 – 24th RD50 Workshop Status of Silicon Strip Sensor Measurements at Liverpool

09/05/2014

Q(z,V) eTCT

Q(x,V) IR top

eTCT

Page 22: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

22

Upcoming measurement

• Irradiated Micron diodes 24GeV/c protons (FZ, MCz)– non-irrad., 5.9×1013 p/cm2 , 1.0×1014 p/cm2 ,

5.3×1014 p/cm2 , 9.8×1014 p/cm2 , 2.0×1015 p/cm2 , 4.4×1016 p/cm2

• Irradiated Micron strip sensor 24GeV/c protons (FZ, MCz)– non-irrad., 6.9 ×1014 p/cm2, 9.7 ×1014 p/cm2,

1.9 ×1015 p/cm2, 3.1 ×1016 p/cm2

• Pion irradiated STMicroelectronics diodes– non-irrad., 1×1011 π/cm2, 3×1011 π/cm2,

1×1012 π/cm2, 3×1012 π/cm2,1×1013 π/cm2, 3×1013 π/cm2,1×1014 π/cm2, 3×1014 π/cm2, 7×1014 π/cm2

09/05/2014

Page 23: TCT+, eTCT and I-DLTS measurement setups at the CERN SSD Lab Christian Gallrapp 1 M. Bruzzi 2,M. Fernandez 3, C. Figueiredo 1, M. Moll 1, H. Neugebauer.

23

Thanks for your attentionQuestions?

09/05/2014