Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York
23

Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Apr 01, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Tandem Mass Spectrometry QA/QC for

Newborn Screening: Routine Operations.

Mark A. Morrissey, PhD

Wadsworth Center

Department of Health

State of New York

Page 2: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Outline

a. Routine, but not daily considerationsb. Daily analysis of samples

i.    Sample Preparation 1.      Derivatized

2.      Un-derivatized

  ii.    Documentationc. Daily Data Evaluation   i.     Data Reduction

ii.     Evaluation of QC results iii. Evaluation of abnormal samplesd. Summary

Page 3: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Routine, but not daily considerations

Instrument Maintenance • Follow the manufacturers recommendations.

• Most labs use a check-off sheet with signature and date

• Activities may include: clean cone, ballast the rough pump and check oil level, check chiller temp, etc…

• Maintained in the instrument logbook

• Ultimately up to the area supervisor and laboratory director

Page 4: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Routine, but not daily considerations

Non-routine maintenance • Documented with nature of the problem, who found the problem, resolution

and the person fixing the problem.

• We include preventative maintenance from the manufacturer. They have their own specification for function checks and mass calibration.

Tuning • Concentrated Standard Solution. May not include all the applicable analytes

(C14:1, C5OH, etc). .

• Collected material from a used plate (not documented or traceable). We do this as needed if the QC results are out of range.

Page 5: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Reagent Preparation

• HPLC grade, or Reagent grade, or better • To filter, or not to filter? • Hazards

– Acetonitrile -- toxic, flammable

– Methanol -- toxic, flammable

– Butanol -- flammable, irritant

– Acetyl Chloride -- corrosive, water reactive. Use a chilled bath and mix with butanol under argon.

Page 6: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Internal Standard Preparation

• MMWR recommends a deuterated standard to correspond to each analyte.

• Pre-prepared Internal Standards

– Available from Perkin-Elmer or Cambridge Isotope Laboratories.

– Concentration and Expiration are provided by the supplier

Page 7: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Internal Standard Preparation

• In-house Prepared Internal Standards – Available from Dr. H.J. ten Brink

(acylcarnitines) and Cambridge Isotope Labs (amino acids and acylcarnitines).

– Cheaper, but requires more labor– I estimate that for 250,000 samples it requires:

• Approximately 80 hours to prepare the standard

• 40 hours to go through a qualification

Page 8: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Internal Standard Preparation Procedure

• Weigh individual deuterated stds into a suitable container.

• Dissolve in 0.01N HCl (d-C16 in MeOH) for an exact concentration of 1.0 mg/mL or 0.1 mg/mL.

• Add the appropriate amount of each to individual 15-mL polypropylene centrifuge tubes. (Adds up to about 7.3 mL). We make about 40 tubes.

• Seal with Teflon tape and store at –70 C. Stable for at least 1-year

• Should check concentrations and adjust, if necessary

Page 9: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Internal Standard Preparation Procedure

• To prepare working internal standard dissolve the entire tube in 2-L of methanol. Store at –20C. Lasts about 10-days, (10,000 samples).

• 2-years worth of standards costs us approximately $10,000 (500,000 samples)

Page 10: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Preparation of Quality Control Samples

• Standards available from Life Science Resources, H.J. ten Brink (carnitines) and Sigma (amino acids).

• Blood obtained from the Red Cross.

• Individual stock standards of amino acids and acylcarnitines are prepared in saline.

• Added individually to 50 mL aliquots of whole blood. Stored frozen along with a desiccant. Lasts approximately six months. (24/day at each level)

Page 11: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Preparation of Quality Control Samples

• One level is approximately equal to the request for a repeat specimen level. The other level is approximately 2 to 3 times the concentration of the first.

• Determination of control limits.

• QC Samples available from CDC (bi-annually) for periodic use.

Page 12: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Sample Preparation and Analysis

• Underivatized - fewer steps, faster– Add Internal Standard (critical step)

– Cover plates and extract for 20 –30 minutes

– Transfer

– Analyze by MS/MS

• Derivatized - published, more sensitive• Both Methods require care and consistency in

sample preparation

Page 13: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Derivatized Method

1. Add Internal Standard (critical step) we use 12-channel pipetter.

2. Extract 20 minutes on a shaker (room temperature).

3. Transfer extract to a fresh plate. (We keep original plate).

4. Evaporate Solvent using warm air and gentle warming.

5. Add Butanolic HCl, cover (we use Teflon plates) and heat to approximately 60 C for 20 minutes (ovens corrode over time).

6. Evaporate Butanolic HCl using warm air.

7. Add reconstitution solvent.

8. Cover with Aluminum foil and analyze by MS/MS.

Page 14: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Documentation -- Preparation

Page 15: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Documentation -- MS/MS Analysis

Page 16: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Derivatized Method, Challenges or Hints

0

10

20

30

40

50

60

70

80

90

1stQtr

2ndQtr

3rdQtr

4thQtr

EastWestNorth

High Quality Control C4

0.50

1.00

1.50

2.00

2.50

3.00

3.50

0 50 100 150 200

Day

Con

c (µ

mol

e/L)

9413

9412

9405

9406

LCL

UCL

• We have seen high C4 when samples were extracted in polystyrene plates.

• Overheating can cause high leucine results (QC results may appear normal)

• We have observed low response for Methionine internal standard. This is under investigation.

Page 17: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Data Reduction and Evaluation

• Objectives – Determine Concentration of Analytes– Incorporate Results into a Database of Samples– Compare Values to a set of Rules– Produce Reports of Normal and Abnormal

Samples

Page 18: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Data Reduction and Evaluation

• First Step –review all the “chromatograms.” – Missed injections – Failure of the pump program

• Data Conversion – Neolynx generates a tab-delimited file that can

be read by Excel or a database program. – Ratios can be calculated by the MS software,

Excel, or the database.

Page 19: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Evaluation of Quality Control Samples.

Page 20: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Evaluation of Abnormal Results

Page 21: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Evaluation of Results

Page 22: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Analysis of Trends Sample Average Phenylalanine

0.25

0.45

0.65

0.85

1.05

1.25

1.45

1.65

1.85

0 50 100 150 200 250 300 350

Day

Conc

(mg/

dL)

Average High Quality Control Phenylalanine

2

3

4

5

6

7

8

9

10

0 50 100 150 200 250 300 350

Day

Conc

(mg/

dL)

9413

9412

9405

9406

LCL

UCL

Page 23: Tandem Mass Spectrometry QA/QC for Newborn Screening: Routine Operations. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York.

Summary

• Start with well defined rules and written Standard Operating Procedures

• Be consistent• Follow long-term trends as well as the daily

results.