1 Study of characteristics of 4-GEM detector prototype Institute of Physics, Bhubaneswar
21

Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

Mar 06, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

1

Study of characteristics of 4-GEM detector prototype

Institute of Physics, Bhubaneswar

Page 2: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 2

o  ALICE TPC upgrade

Field cage HV electrode (100 kV)

5 m

TPC ALICE

•  Continuous read-out

•  No Gating grid

•  High rate 50 kHz (Pb - Pb)

•  Increase luminosity L= 6 x 10 27 cm

•  GEM replace MWPC

•  IBF < 1%

•  Ne/CO2 /N2 – (90/10/5)

•  Large pitch(280 𝝁𝒎) use with ) use with

Standard pitch(140 𝝁𝒎) )

Page 3: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,PunjabUniversity 3

STANDARD PITCH (S) LARGE PITCH (LP)

140𝝁𝒎280𝝁𝒎

•  4 single mask GEM

•  S-LP-LP-S

•  Energy resolution : 12 %

5.9keV

•  IBF < 1%

o  GEM arrangement in TPC

Page 4: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 4

o  4GEM setup at IOP

•  Building of first 4-GEM detector prototype

•  High voltage testing with good V-I characteristics

•  Tested with Cosmic signal

Page 5: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 5

4-GEM detector prototype High voltage testing with good V-I characteristics

o  4GEM setup at IOP

Cosmic signal after Pre-amplifier Cosmic signal after Amplifier

Page 6: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,PunjabUniversity 6

Study of 4GEM at GSI

Page 7: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 7

o  Measurements done using IOP 4GEM

1.  Fe 55 Source •  ADC spectrum

•  Anode current

•  Gain and Resolution measurement

2.  X-ray generator

•  X-ray voltage scan

•  X- ray current scan

Page 8: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

•  Gas mixture: Ar/CO2: 70/30

•  7 channel HVG210 power supply

•  1 sum-up board is used for signal (1×128, 9×9 mm2 pads)

•  PXI Lab-View based DAQ is used

8

o  Set up required for experiment

Page 9: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 9

o  4GEM set up with Electronics 4 gem

mini X-ray generator

Scope and Counter

Page 10: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 10

o  Experimental set up

(Anode Current measurement)

4 gem Pre- amplifier HV

Fe 55 source

Counter

Scope Interface with Lab-VIEW MCA

Keithley Pico-ammeter

mini X-ray generator (ADC spectrum)

Page 11: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 11

o  Signals

Interface with Lab-VIEW Signal

Page 12: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 12

Tests with Fe 55 source

Page 13: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 13

GEM Voltage = -1600 V, ΔV↓𝒈𝒆𝒎 = 120 V, Ar/𝐂𝐎↓𝟐  (70/30)

o  ADC spectrum with Fe 55 source

5.9 keV

2.9 keV

Argon main peak

Page 14: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 14

•  Gain = 𝐐↓𝐨𝐮𝐭𝐩𝐮𝐭 /𝐍↓𝐩 𝐪↓𝐞  

•  Gain vs. applied voltage

where, 𝐐↓𝐨𝐮𝐭𝐩𝐮𝐭  = 𝐩𝐞𝐚𝐤 (𝐢𝐧 𝐦𝐕)/𝐆↓𝐩𝐫𝐞  

𝐍↓𝐩 =no of primary electrons 𝐪↓𝐞  = electron charge

𝐆�𝐩𝐫𝐞 =pre amplifier gain =pre amplifier gain peak =ADC main peak mean

Page 15: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 15

•  Resolution = 𝟐.𝟑𝟓 𝛔 /𝐩𝐞𝐚𝐤  ×𝟏𝟎𝟎%

•  Energy resolution vs. applied voltage

𝝈 and peak are in channel no. peak =ADC main peak mean

where, 𝛔 = Standard deviation = Standard deviation

Page 16: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 16

o  Anode current with Fe 55 source

Current is increasing exponentially with voltage

Page 17: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 17

Tests with X-ray generator

Page 18: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 18

GEM voltage = -1976 V, X-ray voltage 10kV, X-ray current 5 𝛍A A

o  Spectrum

Page 19: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 19

X-ray voltage = 10kV – 50 kV

GEM voltage = -1976 V, X-ray current 5 𝛍A A

o  X-ray voltage scan

Page 20: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •

29/11/16 ALICEIndiameet-2016,Gauha9University 20

X-ray current = 5 𝛍A - 194 𝛍A A - 194 𝛍A A

Voltage = -1976 V, X-ray voltage = 20 kV

o  X-ray current scan

Page 21: Study of characteristics of 4-GEM detector prototype · 29/11/16 ALICE India meet - 2016, Gauha University 4 o 4GEM setup at IOP • Building of first 4-GEM detector prototype •