Top Banner
Study of avalanche fluctuations and energy resolution with an InGrid-TimePix detector D. ATTIÉ 1) , M. CAMPBELL 2 ) , M. CHEFDEVILLE 3) , P. COLAS 1) , E. DELAGNES 1) , K. FUJII 4) , I.GIOMATARIS 1) , H. VAN DER GRAAF 5) , X. LLOPART 2) , M. LUPBERGER 1) , H. SCHINDLER 2) ,J. SCHMITZ 6) , M. TITOV 1) 1) CEA/Irfu Saclay, 2) CERN, 3)LAPP Annecy, 4)KEK, 5)Nikhef, 6)Twente U. Vienna Conference on Instrumentation 2010 16/02/2010, Vienna 1 Avalanche flutuations with InGrid/TimePix Avalanche fluctuations are an old problem, motivated by recent applications in MPGDs, which can be addressed with new tools.
23

Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Jan 14, 2016

Download

Documents

Scott Cooper
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 1

Study of avalanche fluctuations and energy resolution with an

InGrid-TimePix detectorD. ATTIÉ1), M. CAMPBELL2 ), M. CHEFDEVILLE3) , P. COLAS1), E. DELAGNES1) , K. FUJII4) , I.GIOMATARIS1) , H. VAN DER GRAAF5) , X. LLOPART2) , M. LUPBERGER1) , H. SCHINDLER2) ,J. SCHMITZ6), M. TITOV1)

1) CEA/Irfu Saclay, 2) CERN, 3)LAPP Annecy, 4)KEK, 5)Nikhef, 6)Twente U.

Vienna Conference on Instrumentation 2010

16/02/2010, Vienna

Avalanche fluctuations are an old problem, motivated by recent applications in MPGDs, which can be addressed with new tools.

Page 2: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 2

Avalanche statistics, an old problem…Wijsman 1949,Yule-Furry : exponentialLegler 1955, 1961: it takes a

distance UI/E to get ionization energy from the E field

Alkhazov 1970 : avalanche process is iterative, moments of the size distribution can be calculated, turnover at low Z

Approx. solution for wires: Polya

16/02/2010, Vienna

q = 0 : exponential distribution

K. Fujii

Page 3: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 3

threshold

…with new motivationsMotivated by practical consequences: - energy resolution of MPGD, contributes 1/√(1+q)N- spatial resolution of a Micromegas TPC, avalanche fluctuations

lower by a factor (1+q)/(2+q) the number of effective electrons, D. Arrogancia et al., NIM A 602 (2009) 403

- efficiency for single electron detection

16/02/2010, Vienna

Page 4: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 4

New tool : TimePix chip + InGrid

Idea : take a medical imaging chip (Medipix 2), add a clock to each pixel, replace ‘grey levels’ by ‘clock ticks’

(Michael Campbell, Xavi Lloppart, CERN)

65000 pixels, 14-bit counter, 100 MHz tunable clock frequency -> more voxels than the ALEPH TPC, but tiny!

Cover the chip with a deposited grid over SU8 pillars 50 µ high to obtain gas amplification

The chip is protected by 7 µ SiN to avoid destruction by sparks.

55 mm

55 mm

Pixel

14111 m

1612

0 m

1408

0 m

(pi

xel a

rray

)

11 22 33

44

55

55

μ m

55 μ m

Pre

am

p/

shap

er

TH

L dis

c. Con

fig

ura

tion

la

tch

es

Interface

Counter

Syn

chro

niz

ati

on L

og

ic

CERN-Nikhef-Saclay Collaboration within EUDET

16/02/2010, Vienna

Page 5: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 516/02/2010, Vienna

Micromegas + TimePix

TimePixREADOUT

MICROMESH

DRIFT

DRIFT

SPACE

ED ~ 0.7 kV/cm

EA ~ 80 kV/cm

InGrid (Nikhef-Twente)

2.5 cm

50 µm

Fe 55 source

Chamber operated with an Ar+5% isobutane mixture

Page 6: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

16/02/2010, Vienna 6

See electrons from an X-ray conversion one by one (55Fe) and count them, study their fluctuations.

Measure Time Over Threshold (linear with charge above 5 ke-) for single isolated pixels : direct access to avalanche charge distribution.

Avalanche flutuations with InGrid/TimePix

Page 7: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 7

Gain fluctuations from Time Over Threshold

Select isolated clusters with only 1 pixel. These are single electron avalanches (~10 µ rms radius).TOT is linear with number of electrons seen by the amplifier above 5 ke- : Ne = 167 TOT – 6700(red curve, corresponding to the threshold setting of our data taking)Valid up to 30 000 electrons.

This gives a direct access to avalanche size.

Number of electrons

16/02/2010, Vienna

TOT

(in 2

8 ns

tim

e bi

ns)

10000 20000 30000

U in (Volts)

Page 8: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 8

Distributions of avalanche size from

TOT at different gainsZ = G/<G> = (167*TOT-6700)/G(V)

G(V) from a measurement using a source.

16/02/2010, Vienna

Gain =2900

Gain =12600Gain =6000

Z=G/<G>

Z=G/<G> Z=G/<G>

Page 9: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

916/02/2010, Vienna Avalanche flutuations with InGrid/TimePix

Same Z from TOT distributions but log scale to see the tails

Unfortunately, the tails are dominated by TOT resolution effects.

Gain =2900

Gain =6000Gain =12600

Z=G/<G>

Z=G/<G>Z=G/<G>

Page 10: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 10

Avalanche size distribution from TOT

16/02/2010, Vienna

Polya fits above Zmin = 5000/<G> (region of linearity of TOT) are goodHowever theta values are not reliable (very correlated with the gain measurement and the TOT scale. There is a discrepancy between the average number of electrons and the gain: this is a possible effect from the protection layer and from the shaping by electronics.

HV mesh Gain fitted q

310 V 2900 5.3±1.3

330 V 6000 3.8±0.1

350 V 12600 4.7

We do not regard these fitted values as measurements of theta. They point to a value of 4.3 but with very large systematic errors (factor of 2?)

Page 11: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 11

Monte Carlo simulation

16/02/2010, Vienna

ELECTRON COUNTING

Gas : Ar+5% isobutane

Page 12: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 1216/02/2010, Vienna

In our setup, we use the Chromium K-edge to cut the Kb line (Center for X-Ray Optics)

Page 13: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 13

Monte Carlo simulation. Shows that we need enough drift distance to separate the clusters. Also shows that the escape peak is better contained than the photopeak.

16/02/2010, Vienna

Page 14: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

14

Data Ugrid = 350 V

NUMBER OF CLUSTERS

16/02/2010, Vienna Avalanche flutuations with InGrid/TimePix

Drift distance (z) cut performed using the diffusion : sqrt(rmsx 2 + rmsy 2 ) > 28 pixels (cluster separation)Cloud center within a window around the chip center (containment)

Gas : Ar+5% isobutane

Page 15: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 15

Use escape peak (only one line, better contained)

Then correct for collection efficiency (96.5 +- 1 % from MC, in this range of field ratios : 80-90)Convert U_grid into gain/threshold (threshold = 1150 e-)

16/02/2010, Vienna

Page 16: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 16

Collection efficiency from simulation : 96.5±1 %

Gain measurements (from a 80x80 mm2 copper mesh with the same gap 50 µm, gas : Ar+5% isobutane

16/02/2010, Vienna

Prediction from R. Veenhof et al., Data (in red) from D. Attié et al.(see also D. Arrogancia et al. 2009)

2µ thickness

1µ thickness

Page 17: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 17

q~1 at moderate gain (few 1000). Maybe higher at gains above 5000Exponential behaviour (q=0) strongly excluded, as well as >2q

16/02/2010, Vienna

Page 18: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 18

Determination of W and F

16/02/2010, Vienna

The background is totally negligible (time cut taking 30 time buckets around the electron cloud among 11000)The probability for merging two clusters is small, with the rms cuts. The probability for loosing electrons by the containment cuts is small. Attachment also is negligible.The main inefficiency comes from collection : 96.5+-1 % from simulation.

Using the escape peak:

W= 2897 eV / 120.4 = 24.06 +- 0.25 eV

This translates to 245+-3 electrons for the 5.9 keV line, larger than what is usually admitted for pure Ar (227). Photoelectric effect on the mesh is not excluded. This could also be a Penning effect in the conversion region.

The Fano factor could be derived from the rms of the escape line (6.8 e- ) but needs large corrections from inefficiencies.

Gas : Ar+5% isobutane

Page 19: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

1916/02/2010, Vienna Avalanche flutuations with InGrid/TimePix

Energy resolution

55Fe

Cr filter

Micromegas detector

5.9 keV Xrays giving N=245 elec. in argon with rms sqrt(F/245)

Peak width: contribution from primary (Fano) fluctuations and gain fluctuations (assuming high detection efficiency)

Width=√(F+B)/N

Page 20: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 20

5% rms resolution

Resolution measurements with InGrid

(Grid integrated on a Si wafer by post-processing technique)

16/02/2010, Vienna

Without Cr filter

With Cr filter

Page 21: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

16/02/2010, Vienna 21Avalanche flutuations with InGrid/TimePix

Page 22: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

16/02/2010, Vienna 22Avalanche flutuations with InGrid/TimePix

Width=√(F+B)/N

With a measured relative width of 0.05 and assuming F=0.2, taking the measurement of N=245, we obtain B=0.41, thus q =1.4This is roughly consistent with the results from single electron counting efficiency.

Page 23: Study of avalanche fluctuations and energy resolution with an InGrid- TimePix detector D. ATTIÉ 1), M. CAMPBELL 2 ), M. CHEFDEVILLE 3), P. COLAS 1), E.

Avalanche flutuations with InGrid/TimePix 23

CONCLUSIONS

16/02/2010, Vienna

InGrid, Microbulk, and TimePix are new detectors which allow to study the conversion and avalanche processes with unprecedented accuracy.

Time Over Threshold measurements give access to direct measurement of the fluctuations, provided absolute gain and TOT calibration can be better controlled.

The onset of single electron efficiency with Micromegas gain allows the exponential fluctuations to be excluded and favours Polya fluctuations with q close to 1 at moderate gain and reaching a few units at gains of 10 000.

To measure Fano fluctuations will require an improved setup with a longer drift and better controled field. Energy resolution measurements assuming F=0.20 favour a value of ~1. 4. q

Special thanks to R. Veenhof, J. Timmermans and Y. Bilevych