Status Report on Mk.II Pepperpot Simon Jolly Imperial College 13 th June 2007
25

Status Report on Mk.II Pepperpot

Jan 14, 2016

Download

Documents

parson

Status Report on Mk.II Pepperpot. Simon Jolly Imperial College 13 th June 2007. Pepperpot Components. Pepperpot head: Tungsten intercepting screen, 50 m m holes on 3mm pitch in 41x41 array. Tungsten sandwiched between 2mm/10mm copper support plates. Quartz scintillator images beamlets. - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Status Report on Mk.II Pepperpot

Status Report on Mk.II Pepperpot

Simon Jolly

Imperial College

13th June 2007

Page 2: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 2

Pepperpot Components• Pepperpot head:

– Tungsten intercepting screen, 50m holes on 3mm pitch in 41x41 array.

– Tungsten sandwiched between 2mm/10mm copper support plates.

– Quartz scintillator images beamlets.• Camera system:

– PCO 2000 camera with 2048 x 2048 pixel, 15.3 x 15.6 mm CCD.– Firewire connection to PC.– 105 mm Micro-Nikkor macro lens.– Bellows maintains light tight path from vacuum window to

camera.• Main support:

– Head and camera mounted at either end of 1100 mm linear shift mechanism, with 700 mm stroke.

– All mounted to single 400 mm diameter vacuum flange.

Page 3: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 3

Vacuum bellows

Camera

Moving rod

Shutter

Mounting flange

Pepperpot head

Bellows

Tungsten mesh

Beam profile head

FETS Pepperpot Design

Page 4: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 4

Pepperpot Installation

Page 5: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 5

“0 mm” position of pepperpot head is 57 mm downstream of cold box exit. 458 mm dynamic range means “300 mm” position is approximately 20 mm upstream of slit-slit scanners.

57 mm

Pepperpot Location Inside Ion Source

458 mm 100 mm

Measurements taken at 100 mm intervals.

Page 6: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 6

Mk.II Pepperpot Recent Improvements

• Completed:– Multiple calibration markings on pepperpot and profile

heads.– Sliding camera mount to improve resolution.– Larger, permanent camera-bellows mount (not

cardboard!).• To do:

– Background light replacement to improve calibration.– Support for new bellows mount.– Clamp to support sliding mount to prevent it toppling

forward.– Install new components: Tuesday 16th June?

Page 7: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 7

Pepperpot Data ImageRaw data Calibration image

Colour enhanced raw data image, 60 x 60 mm2.

Calibration image: use corners of 9 mm x 9 mm square on copper plate to give image scaling, tilt and spot spacing.

Page 8: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 8

Pepperpot Emittance Extraction

Pepperpot image spots: hole positions (blue) and beam spots (red)

Emittance profiles

Y

X

Page 9: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 9

13 kV Extract: 0 mm

Page 10: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 10

13 kV Extract: 100 mm

Page 11: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 11

13 kV Extract: 200 mm

Page 12: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 12

13 kV Extract: 300 mm

Page 13: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 13

Position Variation for 13 kV Extract

0 mm

100 mm

200 mm

300 mm

x = 1.36y = 1.47 mm mrad

x = 1.82y = 1.96 mm mrad

x = 1.65y = 1.78 mm mrad

x = 1.90y = 2.04 mm mrad

Page 14: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 14

Results of Position Variation

• Behaviour as expected: steady expansion in beam size along the beam axis.

• Emittances show no obvious nonlinearities; increases may be a result of space charge effects.

• Level of space charge compensation needs to be investigated with beam dynamics simulations.

• Flat top and bottom of the beam profile are likely due to collimation occurring upstream of the ion source exit.

Page 15: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 15

6 kV Extract: 0 mm

Page 16: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 16

9 kV Extract: 0 mm

Page 17: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 17

13 kV Extract: 0 mm

Page 18: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 18

17 kV Extract: 0 mm

Page 19: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 19

Variation of Extract Voltage at 0 mm

6 kV

9 kV

13 kV

17 kV

x = 0.25y = 0.77 mm mrad

x = 1.36y = 1.47 mm mrad

x = 0.74y = 1.14 mm mrad

x = 2.02y = 1.92 mm mrad

Page 20: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 20

Results of Extract Variation• Distinct change in beam shape, density distribution and emittance:

– 6 kV shows strong asymmetric divergence in both planes.– 9 kV shows mostly vertical expansion.– 13 kV is nearly symmetric and slightly collimated.– 17 kV dominated by collimation.

• A number of factors contribute to changes:– Upstream collimation.– Change in extract potential: changes shape of field within extract

region, and affects shape of discharge plasma.– Increase in extract potential also requires change in 90° sector

magnet current: change focussing effect of fringe fields.– Decrease in post-acceleration voltage means transverse

focussing from fringe fields within post-acceleration gap will change.

Page 21: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 21

Results Summary

Extract (kV)

Current (mA)

Z (mm) Beam size ( mm mrad)

x (mm) y (mm) x,rms y,rms

± 1 ± 1 ± 0.1 ± 3 ± 5 %

6 13 0 18 33 0.25 0.77

9 22 0 27 63 0.74 1.14

13 34 0 51 60 1.36 1.47

13 34 100 57 72 1.65 1.78

13 34 200 69 87 1.82 1.96

13 34 300 78 102 1.90 2.04

17 36 0 57 63 2.02 1.92

Page 22: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 22

Scintillator Measurements

5 kV Ext 5.5 kV Ext 6 kV Ext 6.5 kV Ext

7 kV Ext 8 kV Ext 9 kV Ext 11 kV Ext

Page 23: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 23

Scintillator/Pepperpot Comparison: 6 kV

Page 24: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 24

Scintillator/Pepperpot Comparison: 9 kV

Page 25: Status Report on Mk.II Pepperpot

13/6/07 Simon Jolly, Imperial College 25

Conclusions

• First results already promising.• Detailed research program under discussion:

start with categorising scintillator.• Clear comparison between scintillator and

pepperpot measurements: calibration markings should allow precise correlation between the two.

• Quality of results should improve with latest modifications.

• Lots of data to take…