As it is Hydramotion’s policy constantly to develop and improve its products, some information contained in this document may change without notice M04-01-035 Rev5 06/16 any pipe, any fluid no guards or shields Integral temperature sensor no moving parts — no maintenance SPECIFICATION ViscoJet VJ1-100 standalone process viscometer 0 - 1,000 cP 1 second 1% of reading or ±0.1 cP 0.3% of reading or ±0.05 cP 2 × 4-20mA configurable, representing viscosity, referenced viscosity or temperature RS485 ModBus data link 24V DC @ 80 mA Integrated temperature measurement –50 - +150 °C * 50 bar * ¾" or 1" BSPP/NPS, or 1½" cone seat as required standard instrument cable all-welded, crevice-free head: SS316 (EN 1.4401) sensor: SS316 (EN 1.4401) * (optional: S32507 duplex steel [EN 1.4410], alloy C22, alloy C276) N6 0.8 μm Ra 3 kg IP67 ViscoLink data viewing and logging package (Windows-based for PC) Viscosity range Measurement time Accuracy Repeatability Analogue outputs Digital outputs Power supply Integral thermometer Temperature range Pressure range Pipe connection Cable length Cable type Construction Materials Surface finish Weight Ingress protection Software * Custom variations to specification available to order ViscoJet 100 transducer 191mm 164mm 50mm Ø24mm M20 cable gland Heatsink Wrench flats Process connection Integral temperature sensor