EQUIPMENT TECHNICAL SPECIFICATIONS

Temperature

Range → 20°C TO 60°C

Opera�ng Accuracy → ≤ +/- 0.2°C

Uniformity→ ≤ +/- 1.0°C

Range → 40%RH to 90%RH

Opera�ng Accuracy → ≤ +/- 1%RH

Uniformity → ≤ +/- 3.0%

Humidity Opera�ng Condi�ons As Per ICH

Special Condi�ons

25°C / 60%RH

30°C / 65%RH

40°C / 75%RH

30°C / 75%RH

25°C / 40%RH

35°C / 35%RH

40°C / 25%RH

Panel Sandwich type with Tongue and Groove Interlock and Metal cam lock

Internal MOC Stainless Steel 304 grade with Ma� finish

External MOC Stainless Steel 304 grade with Ma� finish

Insula�on 80 mm thick Rigid PU Foam with density of 40kg / m3 ± 2 kgs

Tray Heavy Duty SS Perforated tray with addi�onal support and unique Highly efficient perfora�ons for be�er uniformity.

Racks Heavy duty constructed with stainless steel 304 grade square pipe

Door Full view, toughened glass with air gap - installed with imported Hinges and Lock for viewing stored samples without opening.

U�lity Externally installed at rear of the chamber constructed with Internal SS-304 & External GI powder Coated, PU Foam Insulated for be�er performance and ease of Opera�on during preven�ve maintenance/breakdown of the equipment.

Refrigera�on System Set of Energy Efficient Copeland compressor, fin and tube type evaporator & condenser.

Auto Mode Instant Start (Applicable only with Stand by Refrigera�on system) of compressors to assure opera�onal con�nuity during power outage and minimise Alarm dura�ons.

Humidifica�on system Set of reservoir tanks, Steam generator tanks, ke�le heaters, safety thermostats, float valves.

Air hea�ng system Built-in set of stainless-steel tubular heaters provided for efficient air hea�ng.

Control system

Panel MOC PCGI

Controller Advanced PLC based control system installed with HMI

PLC Make Delta with built-in Ethernet connec�vity

HMI MakeDelta 4.3” colour TFT LCD touch screen

Controlling Sensor- 1 no Imported Transmi�er type, 4 to 20 MA with Accuracy of +/- 0.3°C & +/- 3% for RH

Control Switchgear L&T Make Contactor & MCB, etc.

Auto Changeover Refrigera�on and Humidifica�on system in case of primary failure PLC Controlled Auto & manual Changeover to standby

· Refrigera�on: Dura�on & failure inclusive based for refrigera�on · Humidifica�on: Failure based Auto Changeover for humidifiers

Data Storage 1500 Lines for Controlling and Monitoring Sensor data

Illumina�on LEDs for longer life and energy efficiency

Valida�on Access Port Hole will not be provided since Wireless Data loggers are used for calibra�on & Valida�on.

Page 3: STABILITY / HUMIDITY CHAMBER

Safety features PLC based 10 levels of Safety cut offs to protect undershoot and overshoot of Temperature and Humidity.

Temperature high cut off 1st Level Temperature Posi�ve Excursion- Sum of Set Value and Process Value. (°C.THC1) Ac�on- Steam generator and Air heater Cut-Off.

Temperature higher cut off 2nd Level Temperature Posi�ve Excursion- Sum of Set Value and Process Value.(°C.THC2) Ac�on- Mains Supply Cut-Off

Temperature low cut off 1st Level Temperature Nega�ve Excursion- Difference between Set Value and Process (°C.TLC1) Value. Ac�on- Compressor Cut-Off

Temperature lower cut off 2nd Level Temperature Nega�ve Excursion- Difference between Set Value and (°C.TLC2) Process Value. Ac�on- Mains Supply Cut-Off

Humidity high cut off 1st Level Rela�ve Humidity Posi�ve Excursion- Sum of Set Value and Process Value.(%RHC1) Ac�on- Alarm Buzzer ac�va�on

Humidity higher cut off 2nd Level Rela�ve Humidity Posi�ve Excursion- Sum of Set Value and Process Value.(%RHC2) Ac�on- Steam Generator (Primary and Stand By) Cut-Off

Humidity low cut off 1st Level Rela�ve Humidity Nega�ve Excursion- Difference between Set Value and(%RHLC1) Process Value, Ac�on- Mains Supply Cut-Off

Low water steam Decreased Water Level in Primary SG Reservoir below required level.generator cut off Ac�on- Primary SG Shut Off and Switch Over to Stand By.

Chamber High Chamber Temperature exceeding CHT Thermostat range.Temperature cut off Ac�on- Mains Supply Cut off, Alarm Buzzer Ac�va�on

U�lity high Chamber Temperature exceeding UHT Thermostat range.temperature cut off Ac�on- Mains Supply Cut off, Alarm Buzzer Ac�va�on

SAFETY FEATURES

STANDARD CAPACITYModels Cap. in Internal size in mm External size in mm Rack Tray Tray size in mm Connected

litres Qty. Qty. Load

AO/WSC-400 4000 2000WX1000DX2000H 2160WX1630DX2200H 2 10 730WX625DX30H 4.0 KW

AO/WSC-600 6000 2000WX1500DX2000H 2160WX2130DX2200H 4 20 600WX625DX30H 4.25 KW

AO/WSC-800 8000 2000WX2000DX2000H 2160WX2630DX2200H 4 20 850WX625DX30H 4.25 KW

AO/WSC-1000 10000 2000WX2500DX2000H 2160WX3130DX2200H 6 30 720WX625DX30H 4.25 KW

AO/WSC-1200 12000 2000WX3000DX2000H 2160WX3630DX2200H 6 30 885WX625DX30H 4.25 KW

AO/WSC-1600 16000 2000WX4000DX2000H 2160WX4630DX2200H 8 40 905WX625DX30H 5.5 KW

AO/WSC-2000 20000 2000WX4000DX2500H 2160WX4630DX2700H 8 48 905WX625DX30H 6.0 KW

AO/WSC-2500 25000 2300WX4500DX2500H 2460WX5130DX2700H 10 60 815WX750DX30H 6.0 KW

AO/WSC-3000 30000 4000WX3000DX2500H 4160WX3630DX2700H 12 72 885WX625DX30H 9.0 KW

AO/WSC-4000 40000 4000WX4000DX2500H 4160WX4630DX2700H 16 96 905WX625DX30H 9.0 KW

AO/WSC-5000 50000 4500WX4500DX2500H 4660WX5130DX2700H 20 120 815WX725DX30H 12.75 KW

Complete so�ware module designed to provide seamless and comprehensive control in real �me, with g ra n u l a r l eve l d ata a n a l ys i s & t ra n s p a re nt representa�on with conforming to 21 CFR part 11 compliance requirements pertaining to Electronic Records and Electronic Signatures.

Auto Backup

Built-inLDAP

3 Tier Dura�on based equipment wise

alarm ageing A1, A2, A3

Automated Genera�on Of

Scheduled Reports.

Real Time Ac�vity

Log

Trend Analysis Reports

Page 4: STABILITY / HUMIDITY CHAMBER

B-239, MIDC Malegaon, Sinnar, Dist. Nasik, Maharashtra - 422113 (INDIA)

CORPORATE OFFICE

ALLYONE ENVIRONMENTAL TECHNOLOGIES INDIA PVT LTD

ELEM TECHNOLOGIES+91 [email protected]

Andhra Pradesh and Telangana

JABS PHARMATECH+880 01715572988 / [email protected]@jabspharmatech.com

Bangladesh

TRADE TIME INTERNATIONAL+977 [email protected]

Nepal

CHANNEL PARTNERSSERVICE SUPPORT

[email protected]+91 8767804736

[email protected]+91 8767804734

[email protected]+91 8767804733

SAMA SUPPLIES+962 [email protected]

Jordan

®