Spirometry & Related Testing Equipment

RET 2414LPulmonary Function TestingModule 2.0

Page 2: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Learning Objectives

Describe at least one volume displacement spirometer

List at least two principles used by flow-sensing spirometers to measure volume

Page 3: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Spirometer Definition

Any Device used to measure lung volumes and flows

Page 4: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Volume Displacement Spriometers

Water-Seal Spirometers (Collins, Stead-Wells)

Page 5: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Volume Displacement Spriometers

Water-Seal Spirometers

Bell factor Kymograph

32,160,1920 mm/min

ATPS to BTPS Potentiometer

Flow/Volume FV Loop

Page 6: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometers

ATPS to BTPS Coversion

Actual values must be corrected to body temperature pressure saturated (BTPS)

Page 7: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Volume Displacement Spriometers

Water-Seal Spirometers

Problems Cracks → Leaks Positioning

Too high Too low

Infection Control Hoses Mouth Pieces Filters

Page 8: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Volume Displacement Spriometers

Dry Rolling-Seal Spirometers

Page 9: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Volume Displacement Spriometers

Bellows – Type Spirometer

Page 10: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Flow-Sensing Spirometers (pneumotachometers)

Devices that measure gas flow

Page 11: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Turbine (impeller)

Page 12: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Turbine (impeller)

Wright Spirometer VC, VT, VE

Flows 3-300L/min Not used for FVC Compact Gas sterilized

Page 13: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Pressure Differential Flow Sensors

Page 14: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Pressure DifferentialFlow Sensors

Fleich (capillary tubes) Silverman (screens) Heating element to

reduce condensation Recalibrate after cleaning

Page 15: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Flow-Sensing Devices

Heated-Wire Flow Sensors

Page 16: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Portable Spirometers Flow-sensing spirometer

interface with a laptop computer

Page 17: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Portable Spirometers Portable spirometer using plug in

cards

Page 18: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Portable Spirometers Handheld “office” spirometers

Page 19: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometer Calibration

3 – Liter Syringe

Used for volume calibration of volume-based and flow-based spirometers

ATS Standards +/- 3% Or +/- 0.05 L

Whichever is greater

Page 20: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Peak Expiratory Flow Meters

Page 21: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Peak Expiratory Flow Meters Recommended Ranges

ATS 60-400 L/min (children) 100-850 L/min (adults)

NAEPP

100-400 L/min (children) 100 – 700 L/min (adults)

Page 22: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

Respiratory Pressure Manometers Measures MIP and MEP

Page 23: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

MIP

Page 24: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

MIP

Page 25: Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.

Spirometry & Related Testing Equipment

MIP