Top Banner
FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY Service and Parts Manual SF-1855 Part No. 004-0454-00 Rev. D (6/25/12) SpeedClave Steam Sterilizers Model Numbers: M7 M7 -011 thru -016 -020 thru -022 Serial Number Prefixes: MH, MJ, MK, ML, MM, MN, V V ®
112

SpeedClave Steam Sterilizers M7 M7

Mar 14, 2023

Download

Documents

Khang Minh
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: SpeedClave Steam Sterilizers M7 M7

FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY

Service andParts Manual

SF-1855 Part No. 004-0454-00 Rev. D (6/25/12)

SpeedClaveSteam Sterilizers

Model Numbers:

M7M7

-011 thru -016

-020 thru -022

Serial Number Prefixes:

MH, MJ, MK, ML, MM, MN, V

V

®

Page 2: SpeedClave Steam Sterilizers M7 M7

Tabl

e O

f Con

tent

sGENERAL INFORMATION

Symbols .................................... iOrdering Parts .......................... iModel / Serial Number

Location .................................. iWeights, Dimensions,

Electrical Specifications ....... iiModel Identification /

Compliance Chart .................. iiiSpecial Tools ............................ ivWarranty Information ............... iv

OPERATION &TROUBLESHOOTING

Electrical System:M7 (-011 thru -016) .......................A-2M7 (-020 thru -022) .......................A-4

Filling the Chamber ................. A-8Heat-Up / Sterilization.............. A-14Venting the Chamber ............... A-20

TESTING & REPAIRChecking for

Pressure Leaks ...................... B-2Fuse........................................... B-3Bellows ..................................... B-4Fill / Vent Valve:

Manual ..........................................B-6Electronic ......................................B-8

Temperature Regulator Assy .. B-12Heating Element ....................... B-18Overheat Thermostats ............. B-22Pressure Relief Valve ............... B-25Timer ......................................... B-26Timer Buzzer ............................ B-30Temperature Gauge ................. B-31Door Assembly ......................... B-32Reservoir Tank .......................... B-34Chamber Assembly ................. B-36

ACCESS PROCEDURES Removing & Installing:

Covers / Panels ........................ C-2Tray Plate / Rack ...................... C-3Draining / Filling Reservoir ..... C-4

WIRING DIAGRAMS & SCHEMATICS115 VAC models:

M7 (-011 / -013 / -014 / -015) ........ D-2M7 (-020 / -022) ............................D-3

230 VAC models:M7 (-012 / -016) ............................D-4M7 (-021) ...................................... D-5

EXPLODED VIEWS / PARTS LISTSM7 (-011 thru -016) .................... E-2M7 (-020 thru -022): ................... E-3

Gen

eral

In

form

atio

n

Sec

tion

ES

ectio

n E

Sec

tion

C

Sec

tion

AS

ectio

n B

Sec

tion

D

Page 3: SpeedClave Steam Sterilizers M7 M7

i© Midmark Corporation 2004 SF-1854

General Information

MA511502i

R

Symbols

CautionIndicates a potentially hazardous situationwhich could result in injury if not avoided.

Equipment AlertIndicates a potentially hazardous situationwhich could result in equipment damage if not avoided.

NoteAmplifies a procedure, practice, or condition.

Indicates that the component the check mark appearsbeside should be tested before replacing it.In Section A, test the components in the order indicated.(ex. 1st then, 2nd )

Refer to Section B for component testing procedures.

Ordering Parts

The following information is required when ordering parts:• Serial number & model number• Part number for desired part.

[Refer to Section E: Exploded Views / Parts Lists]

Non-warranty parts orders may be faxed to Midmark usingthe Fax Order Form in the back of this manual.

For warranty parts orders, call Midmark's Technical ServiceDepartment with the required information.Hours: 8:00 am until 5:00 pm EST [Monday - Friday]Phone: 1-(800)-Midmark

Model / Serial Number Location

Gen

eral

Info

rmat

ion

Serial Number

(all models)

Model & Serial Number

M7 (-011 thru -016)Location: bottom of unit

M7 (-020 thru -022)Location: back of unit

Page 4: SpeedClave Steam Sterilizers M7 M7

ii

General Information

© Midmark Corporation 2004 SF-1854

Weights, Dimensions, Electrical Specifications

Dimensions [Refer to illustration]:Height (A) ........................................... 12.8 in. (32.5 cm)Width (B) ............................................ 13.5 in. (34.3 cm)Depth (C) ........................................... 18.1 in. (46 cm)

Chamber Size: ..................................... Diameter: 7.5 in. (19 cm) Depth: 14.25 in. (36.2 cm)

Shipping Carton:(Length x Width x Height) .................. 24 in. x 16 in. x 16 in. (61 cm x 40.6 cm x 40.6 cm)

Weight:Shipping Weight ................................. 39 lbs (17.7 kg)w/reservoir empty .............................. 30 lbs (13.6 kg)w/reservoir full ................................... 41.8 lbs (19 kg)

Reservoir Capacity: ............................ Approx. 1.3 gallon (4.9 liters) at FULL markPressure Relief Valve:

opens at approximately: .................... 34 psi (234 kPa)

Electrical Requirements: .................... [See Model Identification / Compliance Chart]

Power Consumption:100 VAC models ................................ 1150 watts, 12 amps @ 100 VAC115 VAC models ................................ 1150 watts, 10 amps @ 120 VAC230 VAC models ................................ 1150 watts, 5 amps @ 240 VAC

M7 (-011 thru -016)

ATTENTIONA separate (dedicated) electrical circuit isrecommended for all models.Do not connect to a circuit with other devices,unless the circuit is rated for the additional load.

B

A

C

R

Dimensions [Refer to illustration]:Height (A) ........................................... 13 in. (33 cm)Width (B) ............................................ 14 in. (35.6 cm)Depth (C) ........................................... 19 in. (48.3 cm)

Chamber Size: ..................................... Diameter: 7.5 in. (19 cm) Depth: 14.25 in. (36.2 cm)

Shipping Carton:(Length x Width x Height) .................. 24 in. x 16 in. x 16 in. (61 cm x 40.6 cm x 40.6 cm)

Weight:Shipping Weight ................................. 39 lbs (17.7 kg)w/reservoir empty .............................. 30 lbs (13.6 kg)w/reservoir full ................................... 41.8 lbs (19 kg)

Reservoir Capacity: ............................ Approx. 1.3 gallon (4.9 liters) at FULL markPressure Relief Valve:

opens at approximately: .................... 34 psi (234 kPa)

Electrical Requirements: .................... [See Model Identification / Compliance Chart]

Power Consumption:115 VAC models ................................ 1300 watts, 10 amps @ 115 VAC230 VAC models ................................ 1300 watts, 5 amps @ 230 VAC

Fuse (back of unit):115 VAC models ................................ 12 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"230 VAC models ................................ 8 amp, 250 V, Fast-Acting, 5 x 20 mm

M7 (-020 thru -022)

Rev.12/04

Page 5: SpeedClave Steam Sterilizers M7 M7

iii© Midmark Corporation 2004 SF-1854

General InformationModel Identification / Compliance Chart

ledoM noitpircseDlaireSrebmuNsexiferP

:oTseilpmoC :sgnitaRlacirtcelE

LU445

LU1-A01016

140-2-01016

ASC/NAC,2.22C

151#

ASC/NAC,2.22C

0101#69-140-2.0101# CAV spmA

selcyC)zH(

110-7MreziliretS7MrettiR

)CAV511(HM & V x x 511 01 06

210-7MreziliretS7MkramdiM

)CAV032(JM & V x x 042/022 5 05

310-7MreziliretS7MkramdiM

)CAV001(KM & V x x 001 21 06

410-7MreziliretS7MkramdiM

)CAV511(V&LM x x 511 01 06

510-7MreziliretS7MetnalAibaD

)CAV511(MM & V x x 511 01 06

610-7MreziliretS7MetnalAibaD

)CAV032(NM & V x x 032 5 06

020-7MreziliretS7MkramdiM

)CAV511(V x x 511 01 06

120-7MreziliretS7MkramdiM

)CAV032(V x x 032 5 05

220-7MreziliretS7MrettiR

)CAV511(V x x 511 01 06

Page 6: SpeedClave Steam Sterilizers M7 M7

iv

General Information

© Midmark Corporation 2004 SF-1854

Special ToolsThis table lists all special tools needed to diagnose and repair the sterilizer.

looTlaicepS rerutcafnaM rebmuNtraP looTfoesopruP

retemitluMlatigiD elbaliavayllaicremmoC epytyna skcehcegatlov/ytiunitnocmrofrepoT

retemomrehTlatigiD elbaliavayllaicremmoC epytyna erutarepmetrebmahcyfirevoT

Warranty Information

SCOPE OF WARRANTYMidmark Corporation (“Midmark”) warrants to the original purchaser its new Alternate Careproducts and components (except for components not warranted under “Exclusions”) manufac-tured by Midmark to be free from defects in material and workmanship under normal use andservice. Midmark’s obligation under this warranty is limited to the repair or replacement, atMidmark’s option, of the parts or the products the defects of which are reported to Midmarkwithin the applicable warranty period and which, upon examination by Midmark, prove to bedefective.APPLICABLE WARRANTY PERIODThe applicable warranty period, measured from the date of delivery to the original user, shall beone (1) year for all warranted products and components.EXCLUSIONSThis warranty does not cover and Midmark shall not be liable for the following: (1) repairs andreplacements because of misuse, abuse, negligence, alteration, accident, freight damage, ortampering; (2) products which are not installed, used, and properly cleaned as required in theMidmark “Installation” and or “Installation / Operation Manual for this applicable product. (3)products considered to be of a consumable nature; (4) accessories or parts not manufacturedby Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, orother work performed upon or in connection with such products which is not expresslyauthorized in writing in advance by Midmark.EXCLUSIVE REMEDYMidmark’s only obligation under this warranty is the repair or replacement of defective parts.Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, damages for loss of profits or loss of use.NO AUTHORIZATIONNo person or firm is authorized to create for Midmark any other obligation or liability inconnection with the products.

ADDITIONAL INFORMATIONFailure to follow the guidelines listed below will void thewarranty and/or render the table unsafe for use.

• If a malfunction is detected, do not use the tableuntil necessary repairs are made.

• Do not attempt to disassemble table, replacecomponents, or perform adjustments unless you area Midmark authorized service technician.

• Do not use another manufacturer's parts to replacemalfunctioning components. Use only Midmarkreplacement parts

THIS WARRANTY IS MIDMARK’S ONLY WARRANTYAND IS IN LIEU OF ALL OTHER WARRANTIES, EX-PRESS OR IMPLIED. MIDMARK MAKES NO IMPLIEDWARRANTIES OF ANY KIND INCLUDING ANY WAR-RANTIES OF MERCHANTABILITY OR FITNESS FORANY PARTICULAR PURPOSE. THIS WARRANTY ISLIMITED TO THE REPAIR OR REPLACEMENT OFDEFECTIVE PARTS.

SF-1487 REV. A1

Page 7: SpeedClave Steam Sterilizers M7 M7

A-1

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Sec

tion

AOperation &Troubleshooting

Mode PageElectrical System:

M7 (-011 thru -016) ........................... A-2

M7 (-020 thru -022) ........................... A-4

Filling the Chamber .............................. A-8

Heat Up / Sterilization ........................... A-14

Venting the Chamber ............................ A-20

Page 8: SpeedClave Steam Sterilizers M7 M7

A-2

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

OverheatThermostat(Manual-Reset)

OverheatThermostat

(Auto-Reset)

Heating Element

TemperatureRegulator

Relay

HeaterLight

TimerBuzzer

Timer

Com.

Com.

MA670901i

Electrical System - [M7 (-011 thru -016)]The illustration shows all of the electrical components of the sterilzer.Refer to the following page for a detailed description of current flow.

Troubleshooting[Electrical System]

Problem: PageHeating element does not turn ON:

- Heater light is OFF ............................... A-6

- Heater light is ON ............................. A-7

Sterilizer shuts down before

timer setting expires ......................... A-17

Timer buzzer does not function ........... A-19

Electrical System M7 (-011 thru -016)all

Page 9: SpeedClave Steam Sterilizers M7 M7

A-3

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Electrical System - [M7 (-011 thru -016)]

With the power cord properly connected...Overheat ThermostatsCurrent (115 / 230 VAC) continuously flows thru the two (normallyclosed) overheat thermostats. This current supplies power to the timer.

If either thermostat opens (overheat or malfunction), voltage is removedfrom the timer until the thermostat is reset or replaced.

NOTEThe Manual-Reset Thermostat contacts open at approximately 285°F (140°C).To reset, allow unit to cool, then press RESET button on front of unit.

The Auto-Reset Thermostat contacts open at approximately 295°F (146°C).This thermostat automatically resets when the unit cools to approx. 265°F (129°C).

TimerCurrent is supplied to the timer thru the two overheat thermostats.

When the timer is turned ON...TimerThe (normally open) timer contacts close, and voltage is supplied to thetimer motor and the temperature regulator relay. The timer motor runs, andbegins to count down the time it was set for.(The contacts to the timer buzzer remain open).

Temperature Regulator RelayCurrent is supplied to the temperature regulator relay thru the timer. If thechamber temperature is lower than the temperature knob setting*, the relaycontacts are closed. When these contacts are closed, current flows thru therelay to the heating element and the heater light.

The diaphragm cup of the relay expands as the temperature & pressureinside the chamber increase. When the chamber temperature reaches thetemperature knob setting, the relay contacts open, and voltage is removedfrom the heating element & heater light.

[* The minimum temperature knob setting is approx. 220°F (104°C)]

When the timer is turned ON (continued)...Heater Light & Heating ElementWhen the contacts of the temperature regulator relay are closed,current is supplied to the heater light and the heating element.

As the relay contacts open and close, the heating element cycles ON / OFF.This continues until the timer setting expires.

The heater light is illuminated whenever the heating element is ON.

When the timer setting expires...Timer & Timer BuzzerThe contacts to the temperature regulator relay open, stopping the currentflow to the heater light & heating element.

The contacts to the timer buzzer close and current flows to the timer buzzer.When voltage is applied, the buzzer emits an audible signal.

The contacts to the timer motor remain closed for one minute. After oneminute the contacts to the timer motor & the timer buzzer open, stopping thecurrent flow to these two components.

Electrical SystemM7 (-011 thru -016)all

Page 10: SpeedClave Steam Sterilizers M7 M7

A-4

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Heating Element

TemperatureRegulator

Relay

HeaterLight

TimerBuzzer

Timer

Com.

Com.

OverheatThermostat(Manual-Reset)

OverheatThermostat

(Auto-Reset)

Fill / VentSwitch

Fill / VentValve

Fuse

Com. MA670900i

Electrical System - [M7 (-020 thru -022)]The illustration shows all of the electrical components of the sterilzer.Refer to the following page for a detailed description of current flow.

Electrical System M7 (-020 thru -022)all

Troubleshooting[Electrical System]

Problem: PageWhen Fill/Vent Switch is pressed:

- Chamber does not FILL ....................... A-11

- Chamber does not VENT ..................... A-23

Heating element does not turn ON:

- Heater light is OFF ............................... A-6

- Heater light is ON ............................. A-7

Sterilizer shuts down before

timer setting expires ......................... A-17

Timer buzzer does not function ........... A-19

Page 11: SpeedClave Steam Sterilizers M7 M7

A-5

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Electrical System - [M7 (-020 thru -022)]

With the power cord properly connected...FuseCurrent (115 / 230 VAC) continuously flows thru the fuse located in theback of the unit. This current supplies power to the fill / vent switch andthe overheat thermostats.

Fill / Vent SwitchCurrent is supplied to the fill / vent switch thru the fuse.

Overheat Thermostats & TimerCurrent is supplied to the two overheat thermostats thru the fuse.Current continuously flows thru the thermostats to the timer.

If either thermostat opens (overheat or malfunction), voltage is removedfrom the timer until the thermostat is reset or replaced.

NOTEThe Manual-Reset Thermostat contacts open at approximately 285°F (140°C).To reset, allow unit to cool, then press RESET button on front of unit.

The Auto-Reset Thermostat contacts open at approximately 295°F (146°C).This thermostat automatically resets when the unit cools to approx. 265°F (129°C).

When filling the chamber (pressing the fill/vent switch)...Fill / Vent SwitchThe contacts of the (normally open) switch close. When the contacts of theswitch are closed, current is supplied to the fill / vent valve.

Fill / Vent ValveWhen current is applied to the (normally closed) valve, the valve opens.When the valve is open, water flows into the chamber.

When the Timer is turned ON...TimerThe (normally open) timer contacts close, and voltage is supplied to thetimer motor and the temperature regulator relay. The timer motor runs, andbegins to count down the time it was set for.(The contacts to the timer buzzer remain open).

Electrical SystemM7 (-020 thru -022)all

When the timer is turned ON (continued)...Temperature Regulator RelayCurrent is supplied to the temperature regulator relay thru the timer. If thechamber temperature is lower than the temperature knob setting*, the relaycontacts are closed. When these contacts are closed, current flows thru therelay to the heating element and the heater light.[* The minimum temperature knob setting is approx. 220°F (104°C)]

The diaphragm cup of the relay expands as the temperature & pressureinside the chamber increase. When the chamber temperature reaches thetemperature knob setting, the relay contacts open, and voltage is removedfrom the heating element & heater light.

Heater Light & Heating ElementWhen the contacts of the temperature regulator relay are closed,current is supplied to the heater light and the heating element.As the relay contacts open and close, the heating element cyclesON / OFF. This continues until the timer setting expires.The heater light is illuminated whenever the heating element is ON.

When the timer setting expires...Timer & Timer BuzzerThe contacts to the temperature regulator relay open, stopping the currentflow to the heater light & heating element.

The contacts to the timer buzzer close and current flows to the timer buzzer.When voltage is applied, the buzzer emits an audible signal.

The contacts to the timer motor remain closed for one minute. After oneminute the contacts to the timer motor & the timer buzzer open, stopping thecurrent flow to these two components.

When pressing the Fill / Vent Switch (to VENT the chamber)...Fill / Vent SwitchThe contacts of the (normally open) switch close. When the contacts of theswitch are closed, current is supplied to the fill / vent valve.

Fill / Vent ValveWhen current is applied to the (normally closed) valve, the valve opens.When the valve is open, steam is released thru the condensing coil & thewater is returned to the reservoir .

Page 12: SpeedClave Steam Sterilizers M7 M7

A-6

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA670800i

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Check supply voltage(A dedicated circuit is recommended)

2nd

3rdM7 (-020 thru -022) onlyFuse

5thOverheat Thermostats

1stPress RESET buttonAllow unit to cool for 15-20 minutesbefore pressing RESET button.

6thTemperatureRegulator Assy.

7thTimer

4thLoose / Damaged Wire ConnectionsCheck all wiring connections.(Power cord, overheat thermostats, etc.)

Problem: Heating element does not turn ON.[Heater light is OFF]

Electrical System M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 13: SpeedClave Steam Sterilizers M7 M7

A-7

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672300i

R

Electrical System

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Problem: Heating element does not turn ON.[Heater light is ON]

HeatingElement

1st

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 14: SpeedClave Steam Sterilizers M7 M7

A-8

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA671500i

Filling the ChamberThe illustrations show the water flow when filling the chamber.Refer to the following page for a detailed description of this process.

Troubleshooting[Filling the Chamber]

Problem: PageChamber does not fill:

- M7 (-011 thru -016) .............................. A-10

- M7 (-020 thru -022) .............................. A-11

Water continuously flows into chamber:

- M7 (-011 thru -016) .............................. A-12

- M7 (-020 thru -022) ........................... A-13

Filling the Chamber

Fill / Vent Valve(open)

Reservoir

Fill / Vent Valve(open)

Reservoir

Filter Screen

M7 (-011 thru -016)

M7 (-020 thru -022)

= Water

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 15: SpeedClave Steam Sterilizers M7 M7

A-9

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Filling the Chamber

When the Fill / Vent Lever is pressed and held...Fill Vent ValveThe (normally closed - manual) valve opens. When the valve is open, waterfrom the reservoir flows thru the fill / vent valve into the chamber.

When the Fill / Vent Lever is released...Fill / Vent ValveThe valve closes, and stops the flow of water into the chamber.

Filling the Chamber

M7 (-011 thru -016)

When the Fill / Vent Switch is pressed and held...Fill / Vent Switch & ValveCurrent (line voltage) flows thru the fill/vent switch to the fill/vent valve. Whenvoltage is applied, the (normally closed) fill / vent valve opens. When the valveis open, water from the reservoir flows into the chamber thru the valve and filterscreen.

When the Fill / Vent Switch is released...Fill / Vent Switch & ValveThe fill/vent switch opens, stopping the current flow to the fill/vent valve.When voltage is removed, the valve closes. When the valve closes, waterstops flowing into the chamber.

M7 (-020 thru -022)

MA512903i

R R

Fill / VentLever

Fill / VentSwitch

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 16: SpeedClave Steam Sterilizers M7 M7

A-10

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672400i

R

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-011 thru -016)all

Filling the Chamber

Problem: Chamber does not fill.

1stLow water in reservoir?Add distilled water if necessary.

2ndFill / Vent Valve & TubingClean / adjust / replace as necessary.

Page 17: SpeedClave Steam Sterilizers M7 M7

A-11

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672500i

R

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-020 thru -022)all

Filling the Chamber

Check supply voltage(A dedicated circuit is recommended)

3rd

4thFuse

Fill / Vent Switch 6th

5thFill / Vent Valve & TubingClean / adjust / replace as necessary.

Problem: Chamber does not fill.

1stLow water in reservoir?Add distilled water if necessary.

2ndFilter ScreenClean / replace as necessary.

Page 18: SpeedClave Steam Sterilizers M7 M7

A-12

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672400i

R

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-011 thru -016)all

Filling the Chamber

Problem: Water continuously flows into chamber.

1stFill / Vent Valve & TubingClean / adjust / replace as necessary.

Page 19: SpeedClave Steam Sterilizers M7 M7

A-13

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA672600i

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-020 thru -022)all

Filling the Chamber

Fill / Vent Switch 2nd

1stFill / Vent Valve & TubingClean / adjust / replace as necessary.

Problem: Water continuously flows into chamber.

Page 20: SpeedClave Steam Sterilizers M7 M7

A-14

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672000i

Heat Up / SterilizationThe illustrations show the water / steam flow during heat up & sterilization.Refer to the following page for a detailed description of this process.

Troubleshooting[Heat-Up / Sterilization]

Problem: PageHeating element does not turn ON:

- Heater light is OFF ............................... A-6- Heater light is ON ............................. A-7

Heating element turns ON, but does not

reach required temperature .............. A-16

Sterilizer shuts down before

timer setting expires ......................... A-17

Biological test strips indicate items

are not sterile ................................... A-18

Timer buzzer does not function ........... A-19

Heat-Up / Sterilization

TemperatureRegulator Relay

(closed)

Sterilization

Heat-Up

Bellows(open)

Pressure Relief Valve(closed)

= Water

= Heated Water

= Steam

Bellows(closed)

TemperatureRegulator Relay(closed / open)

Fill / Vent Valve(closed)

Note: Electronic valve is shown.Manual valve is also closed.

Fill / Vent Valve(closed)

Note: Electronic valve is shown.Manual valve is also closed.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 21: SpeedClave Steam Sterilizers M7 M7

A-15

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672100i

R

Heat-Up / Sterilization

Heat-Up / Sterilization

When the timer is turned ON...TimerThe (normally open) timer contacts close, and voltage is supplied to thetimer motor and the temperature regulator relay. The timer motor runs, andbegins to count down the time it was set for.(The contacts to the timer buzzer remain open).

Temperature Regulator RelayCurrent is supplied to the temperature regulator relay thru the timer. If thechamber temperature is lower than the temperature knob setting*, the relaycontacts are closed. When these contacts are closed, current flows thru therelay to the heating element and the heater light.[* The minimum temperature knob setting is approx. 220°F (104°C)]

The diaphragm cup of the relay expands as the temperature & pressureinside the chamber increase. When the chamber temperature reaches thetemperature knob setting, the relay contacts open, and voltage is removedfrom the heating element & heater light.

Heater Light & Heating ElementWhen the contacts of the temperature regulator relay are closed,current is supplied to the heater light and the heating element.

As the relay contacts open and close, the heating element cycles ON / OFF.This continues until the timer setting expires.

The heater light is illuminated whenever the heating element is ON.

Bellows & Pressure Relief ValveHeat-Up:As the water in the chamber begins to boil, air is forced out of the chamber.This air passes thru the bellows into the reservoir.

Sterilization:When pure steam begins to flow thru the bellows, the bellows closesallowing pressure to build in the chamber. If the pressure in the chamberexceeds 34 psi (234 kPa), the pressure relief valve opens to prevent unsafeconditions.

When the timer setting expires...Timer & Timer BuzzerThe contacts to the temperature regulator relay open, stopping the currentflow to the heater light & heating element.

The contacts to the timer buzzer close and current flows to the timer buzzer.When voltage is applied, the buzzer emits an audible signal.

The contacts to the timer motor remain closed for one minute. After oneminute the contacts to the timer motor & the timer buzzer open, stopping thecurrent flow to these two components.

Temperature Knob

Timer Knob

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 22: SpeedClave Steam Sterilizers M7 M7

A-16

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672700i

R

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Heat-Up / Sterilization

Problem: Heating element turns ON, but doesnot reach required temperature.[Heater light is ON]

2ndTemperatureRegulator Assy.

3rdHeating Element

1stCheck for pressure leaks.Refer to Section B for checkpoints.

Page 23: SpeedClave Steam Sterilizers M7 M7

A-17

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA672800i

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Heat-Up / Sterilization

Problem: Sterilizer shuts down before timer setting expires.

3rdOverheat Thermostats

2ndPress RESET buttonAllow unit to cool for 15-20 minutesbefore pressing RESET button.

4thTemperatureRegulator Assy.

5thTimer

1stCheck for pressure leaks.Refer to Section B for checkpoints.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 24: SpeedClave Steam Sterilizers M7 M7

A-18

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA672900i

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Heat-Up / Sterilization

5thTemperatureRegulator Assy.

4thCheck for pressure leaks.Refer to Section B for checkpoints.

Problem: Biological test strips indicate items are not sterile.

Is the sterlizer overloaded?Large loads or heavy linen packs mayprevent strips from changing.

2nd

Type / condition of indicator stripsThis unit requires test strips rated for:Gravity Displacement Steam Sterilizers

Test strips must be stored in a cool, dry location.Failure to do so will result in faulty readings.

(Follow all instructions provided with test strips)

1st

3rdAre the correct trays being used?Some trays may prevent proper air flow.Be sure trays are designed for this sterilizer.

Page 25: SpeedClave Steam Sterilizers M7 M7

A-19

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA672900i

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Heat-Up / Sterilization

Problem: Timer buzzer does not function.

1stTimer Buzzer

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 26: SpeedClave Steam Sterilizers M7 M7

A-20

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672200i

Venting the ChamberThe illustrations show the steam / water flow when venting the chamber.Refer to the following page for a detailed description of this process.

Troubleshooting[Venting the Chamber]

Problem: PageChamber will not vent:

- M7 (-011 thru -016) .............................. A-22

- M7 (-020 thru -022) .............................. A-23

Venting the Chamber

Fill / Vent Valve(open)

M7 (-011 thru -016)

M7 (-020 thru -022)

= Water

Fill / Vent Valve(open)

Condensing Coil

= Steam

CondensingCoilFilter Screen

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 27: SpeedClave Steam Sterilizers M7 M7

A-21

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA512903i

R R

Venting the Chamber

AttentionThe door handle must be moved to the VENT position

before pressing the Fill / Vent Lever or Switch.

When the Fill / Vent Lever is pressed and held...Fill / Vent ValveThe (normally closed - manual) valve opens. Pressure forces waterand steam back into the reservoir thru the valve and the condensing coil.When all of the pressure has been vented, the door will "pop".

Note: Release the lever when the door "pops".If the lever is held too long, the chamber will begin to fill.

When the Fill / Vent Lever is released...Fill / Vent ValveThe valve closes.

Venting the Chamber

M7 (-011 thru -016)

When the Fill / Vent Switch is pressed and held...Fill / Vent Switch & ValveCurrent (line voltage) flows thru the fill/vent switch to the fill/vent valve. Whenvoltage is applied, the (normally closed) fill / vent valve opens. Pressure forceswater and steam back into the reservoir thru the valve and the condensing coil.When all of the pressure has been vented, the door will "pop".

Note: Release the lever when the door "pops".If the lever is held too long, the chamber will begin to fill.

When the Fill / Vent Switch is released...Fill / Vent Switch & ValveThe fill/vent switch opens, stopping the current flow to the fill/vent valve.When voltage is removed, the valve closes.

M7 (-020 thru -022)

Fill / VentLever

Fill / VentSwitch

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 28: SpeedClave Steam Sterilizers M7 M7

A-22

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA672400i

R

1stFill / Vent Valve & TubingClean / adjust / replace as necessary.

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-011 thru -016)all

Problem: Chamber will not vent.

Venting the Chamber

Page 29: SpeedClave Steam Sterilizers M7 M7

A-23

Operation & Troubleshooting

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA673000i

R

Fill / Vent Switch 3rd

2ndFill / Vent Valve & TubingClean / adjust / replace as necessary.

1stFilter ScreenClean / replace as necessary.

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Problem: Chamber will not vent.

Venting the ChamberM7 (-020 thru -022)all

Page 30: SpeedClave Steam Sterilizers M7 M7
Page 31: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-1Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Sec

tion

BComponent / Procedure PageChecking For Pressure Leaks .............. B-2

Fuse [M7 (-020 thru -022) only] ............ B-3

Bellows ............................................... B-4

Fill / Vent Valve:

Manual .............................................. B-6

Electronic .......................................... B-8

Temperature Regulator Assy. ............... B-12

Heating Element ................................... B-18

Overheat Thermostats .......................... B-22

Pressure Relief Valve ........................... B-25

Timer Assembly.................................... B-26

Timer Buzzer ........................................ B-30

Temperature Gauge ............................. B-31

Door Assembly ..................................... B-32

Reservoir Tank ..................................... B-34

Chamber Assembly .............................. B-36

Testing & Repair

Page 32: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-2Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA673100i

Components Page

Bellows ............................................... B-4

Fill / Vent Valve:

Manual .............................................. B-6

Electronic .......................................... B-8

Pressure Relief Valve ........................... B-25

Door Assembly ..................................... B-32

Checking forPressure Leaks

Checking for Pressure LeaksThis illustration shows the areas to check for pressure leaks.

Water leaking from fitting(s)?Tighten / replace fittings if necessary.

Steam from pressure relief valve?If YES, replace the pressure relief valve.

Water leaking from door?If YES, replace door gasket.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Steam from bellows tubing?If YES, replace the bellows.

Note: The bellows should close at approx. 215°F(101°C). Periodic "hissing' or "spitting" isnormal.

Bubbles from condensing coil?If YES, clean or replace the fill / vent valve.

Page 33: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-3Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA511501i

R

Fuse

Fuse Ratings:115V models ...... 12 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"230V models ...... 8 amp, 250 V, Fast-Acting, 5mm x 20mm

Fuse

LocationFuses Page

Location ............................................... B-2

Fuse Test ............................................. B-2

Wiring Diagrams ................................... D-1

Fuse Rating ...............................6.3A, 250V

Type T, 5 x 20mm

Part Number ............................ 015-0346-20

Fuses Page

Location ............................................... B-3

Fuse Test ............................................. B-3

Wiring Diagrams ................................... D-1

Part Numbers ....................................... E-14.1

Fuse Test

FuseHolder

M7 (-020 thru -022)all

FuseTestIf reading is OL...Replace fuse.

If reading is within acceptable range...Fuse is OK.

less than 5 ΩΩΩΩΩ

Fuse TestStep 1: Place meter probes on ends of fuse.

[Set meter to 200 ohms (Ω)]Acceptable Range

Page 34: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-4Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA673301i

Bellows M7 (-020 thru -022)all

M7 (-011 thru -016)all

Bellows

Location & Function

As the water in the chamber begins to boil...Air & steam are forced out of the chamber,thru the open bellows, and back into the reservoir.

When pure steam begins flowing thru bellows...The bellows closes allowing pressure to build in the chamber.[Note: The bellows will periodically "hiss" or "spit", this is normal.

Bellows PageLocation & Function .............................. B-4

Testing - refer to:

Checking for Pressure Leaks ............ B-2

Replacement ........................................ B-5

Exploded View / Part Numbers ............. E-11

The arrow indicates the directionof flow toward the reservoir.

The bellows must be oriented correctlyto ensure proper operation.

Page 35: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-5Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA673401i

Bellows - continued

Replacement

Refer to: PageCover Removal ..................................... C-2

RemovalStep 1: Loosen two compression nuts.

Remove bellows.

NOTE:The arrow on the bellows indicates the direction of flow toward the reservoir.The bellows must be oriented correctly to ensure proper operation.

Arrow must pointtoward the reservoir

as shown

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Bellows

Installation(See NOTE for proper orientation of bellows)Step 1: Install bellows.

Tighten compression nuts.

Page 36: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-6Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA673500i

R

Fill / Vent Valve (manual)

Location & Function

When the fill / vent lever is pressed (no pressure in chamber)...The (normally closed) fill / vent valve opens.Water from the reservoir flows thru the open valve into the chamber.The valve closes when the lever is released.

When the fill / vent lever is pressed (chamber is pressurized)...The (normally closed) fill / vent valve opens.Water and steam from the chamber are forced thru the open valve back into thereservoir. When all of the pressure has been released, the door will "pop".The valve closes when the lever is released.

Fill / Vent Valve PageLocation & Function .............................. B-6

Testing - refer to:

Checking for Pressure Leaks ............ B-2

Removal / Installation / Adjustment ...... B-7

Exploded View / Part Numbers ............. E-9

Fill / Vent Valve M7 (-011 thru -016)all

Fill / Vent Lever

Early Models Later Models

Page 37: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-7Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA673600i

DISTANCE "A"

Fill / Vent Valve (manual) - continued

Removal / Installation / Adjustment

Refer to: PageCover Removal ..................................... C-2

Draining / Filling the Reservoir ............. C-4

Fill / Vent ValveM7 (-011 thru -016)all

RemovalStep 1: Drain water from reservoir.

RemovalStep 2: Loosen two compression nuts.

Lever Bracket

RemovalStep 3: Disengage lever bracket from lever.

Remove valve.

Lever

InstallationStep 1: Engage lever bracket with lever.

Press valve into mtg. brackets.

Lever Bracket

InstallationStep 2: Align tubing with valve.

Tighten two compression nuts.Adjust lever bracket. (See Adjustment)

AdjustmentAdjust lever bracket so that Distance "A"is approx. 1/4" (0.63 cm). To secure position,tighten nut against lever bracket.

Page 38: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-8Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Fill / Vent Valve (electronic)

Location & Function

When the fill / vent switch is pressed (no pressure in chamber)...Current (line voltage) flows thru the fill/vent switch to the fill / vent valve.When voltage is applied, the (normally closed) fill/vent valve opens.Water from the reservoir flows thru the open valve into the chamber.The valve closes when the switch is released.

When the fill / vent lever is pressed (chamber is pressurized)...Current (line voltage) flows thru the fill/vent switch to the fill/vent valve.When voltage is applied, the (normally closed) fill / vent valve opens.Water and steam from the chamber are forced thru the open valve back intothe reservoir. When all of the pressure has been released, the door will "pop".The valve closes when the lever is released.

Fill / Vent Valve PageLocation & Function .............................. B-8

Testing - refer to:

Checking for Pressure Leaks ............ B-2

Electrical Test .................................... B-9

Replacement ........................................ B-10

Disassembly / Cleaning ........................ B-11

Wiring Diagrams ................................... D-1

Exploded View / Part Numbers ............. E-9.3

Fill / Vent Valve M7 (-020 thru -022)all

R

MA673800iFill / Vent Switch

Page 39: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-9Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA674000i

Fill / Vent Valve (electronic) - continued

Electrical Test

Refer to: PageCover Removal ..................................... C-2

Fill / Vent ValveM7 (-020 thru -022)all

Electrical TestIf reading is out of acceptable range...Replace valve.

If reading is within acceptable range...Electrical component of valve is OK.

Electrical TestStep 1: Disconnect wires from valve terminals.

Electrical TestStep 2: Place meter probes on terminals.

[Set meter to M ohms (Ω)]

Acceptable Range

115 VAC models: ........ 3.24 to 3.96230 VAC models: ....... 3.24 to 3.96

Page 40: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-10Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA677901i

Fill / Vent Valve (electronic) - continued

Replacement

Fill / Vent Valve

Refer to: PageCover Removal ..................................... C-2

RemovalStep 1: Tag and disconnect wires

from valve terminals.

RemovalStep 2: Disconnect compression fittings.

RemovalStep 4: Remove valve from bracket.

RemovalStep 3: Remove bracket / valve from base plate.

InstallationStep 2: Secure bracket / valve to base plate.

InstallationStep 3: Connect compression fittings.

InstallationStep 4: Connect wires to valve terminals.

M7 (-020 thru -022)all

Rev. 7/07

InstallationStep 1: Install valve onto bracket.Locate "IN" port and direct flow intothe reservoir as shown.

Page 41: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-11Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Fill / Vent Valve (electronic) - continued

Disassembly / Cleaning

Refer to: PageValve Removal ..................................... B-10

Fill / Vent Valve

Disassembly / CleaningStep 1: Remove nut.

Disassemble valve.

Disassembly / CleaningStep 2: Remove any debris.

Inspect components for damage.

M7 (-020 thru -022)all

Page 42: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-12Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA674200i

Temperature RegulatorAssembly

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Regulator Assembly

Location & Function

When the Temperature Control knob is adjusted...The flexible shaft rotates, increasing or decreasing the distancebetween the relay contacts. This adjusts the point (i.e. temperature)at which the relay contacts will open & close*.

As the temperature & pressure inside the chamber increase...The diaphragm cup expands, pushing the relay contacts apart.When the relay contacts are open, the heating element is de-energized.

As the temperature & pressure inside the chamber decrease...The diaphragm cup contracts, allowing the relay contacts to close.When the relay contacts are closed, the heating element is energized.

Temp. Regulator Assy. PageLocation & Function .............................. B-12

Troubleshooting .................................... B-13

Temperature Relay:

Adjustment ........................................ B-14

Removal ............................................ B-15

Installation ......................................... B-16

Diaphragm Cup Replacement .............. B-17

Wiring Diagrams ................................... D-1

Exploded View / Part Numbers ............. E-7

Diaphragm Cup

Flexible Shaft

TemperatureRelay

Page 43: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-13Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA674201i

Diaphragm Cup

TemperatureRelay

Temperature Regulator Assembly - continued

Troubleshooting

Refer to: PageRelay Adjustment ................................. B-14

Relay Removal ..................................... B-15

Diaphragm Cup Replacement .............. B-17

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature RegulatorAssembly

Sterilizer does not reach desired temperature...

Required action: Perform Relay Adjustment.Replace relay if necessary.

Sterilizer shuts down before timer setting expires...

Required action: Inspect relay.(If contacts are corroded or "fused" together - replace relay)

Inspect diaphragm cup.(If water / steam is leaking from this area, replace diaphragm cup)

Page 44: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-14Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA674400i

Temperature RegulatorAssembly

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Regulator Assembly - continued

Relay Adjustment

Refer to: PageCover Removal ..................................... C-2

Relay AdjustmentStep 1: Fill chamber with water.

Close & latch door.Set Timer knob to 30 minutes.

Relay AdjustmentStep 2: Turn Temperature knob to max.

Relay AdjustmentStep 3: Loosen setscrew 2-3 turns.

Relay AdjustmentStep 4: Allow chamber to reach its max.

temperature [>270°F (132°C)].

Relay AdjustmentStep 5: Adjust screw until temperature gauge

reads slightly above 270°F (132°C).

Relay AdjustmentStep 6: Adjust setscrew until temperature

gauge reads 270-271°F (131-132°C).

Page 45: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-15Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA674500i

Temperature RegulatorAssembly

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Regulator Assembly - continued

Relay Removal

Refer to: PageCover Removal ..................................... C-2

RemovalStep 1: Remove setscrew from flexible shaft.

RemovalStep 3: Disconnect two relay wires.

(one from heating element &one from timer - not shown)

RemovalStep 4: Remove timer buzzer.

Remove relay & spacer.

RemovalStep 2: Rotate Temperature knob until

flexible shaft unscrews from relay.

Page 46: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-16Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA674501i

Temperature RegulatorAssembly

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Regulator Assembly - continued

Relay Installation

Refer to: PageRelay Removal ..................................... B-15

Relay Adjustment ................................. B-14

Adjustment Screw

Fine AdjustmentSetscrew

InstallationStep 1: Install relay & spacer.

Install timer buzzer.

Note: Do not tighten Adjustment Screw.Fine Adjustment Setscrew should be flush with bracket.

InstallationStep 2: Connect two relay wires.

(one to heating element &one to timer - not shown)

Installation(Apply hi-temp lubricant to threads of flexible shaft)Step 3: Turn the Temperature knob until flexible

shaft screws all the way into relay.Rotate knob back (counter-clockwise)1 1/2 turns, then install Shaft Setscrew.

InstallationStep 4: Turn the Temperature knob counter-clockwise

until it hits the stop.Hold knob in place, then loosen Knob Setscrew.Rotate knob so that white mark is at 3:00 position, then tighten knob setscrew.

Shaft Setscrew

Knob Setscrew

InstallationStep 5: Perform Relay Adjustment.

Page 47: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-17Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA674700i

Temperature RegulatorAssembly

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Regulator Assembly - continued

Diaphragm Cup Replacement

Refer to: PageRelay Removal ..................................... B-15

Relay Installation .................................. B-16

RemovalStep 2: Remove nut & lockwasher.

RemovalStep 3: Remove diaphragm cup

Remove gasket from diaphragm cup.

InstallationStep 1: Install gasket onto diaphragm cup.

Install diaphragm cup.

InstallationStep 2: Install lockwasher & nut.

RemovalStep 1: Remove relay.

InstallationStep 3: Install relay.

Page 48: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-18Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Heating Element

Location & Function

Heating Element PageLocation & Function .............................. B-18

Resistance Test .................................... B-19

Replacement ........................................ B-20

Wiring Diagrams ................................... D-1

Exploded View / Part Numbers ............. E-13

Heating Element

MA674800i

R

Heater Light

When the timer is turned ON...The timer supplies current to the temperature relay. If thechamber temperature is lower than the temperature knob setting*,the relay contacts are closed. When these contacts are closed, currentflows thru the relay to energize the heating element and the heater light.

When the chamber temperature reaches the temperature knob setting,the relay contacts open, and voltage is removed from the heating element& heater light.

[* The minimum temperature knob setting is approx. 220°F (104°C)]

When the timer is OFF...Timer contacts to the temperature relay open, stopping the current flowto the heater light & heating element.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 49: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-19Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA674900i

Heating Element - continued

Resistance Test

Refer to: PageCover Removal ..................................... C-2

Heating Element

Resistance TestIf reading is out of acceptable range...Replace heating element.

If reading is within acceptable range...Heating element is OK.

Resistance TestStep 2: Place meter probes on terminals.

[Set meter to 200 ohms (Ω)]

Acceptable Range

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Resistance TestStep 1: Disconnect wires from heating element terminals.

100 VAC models ........ 8 to 10115 VAC models: ....... 11 to 13230 VAC models: ....... 45 to 51

Page 50: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-20Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA675000i

Heating Element - continued

Replacement

Heating Element M7 (-020 thru -022)all

M7 (-011 thru -016)all

Refer to: PageCover Removal ..................................... C-2

Wiring Diagrams ................................... D-1

RemovalStep 1: Disconnect power to sterilizer.

Drain all water from reservoir. RemovalStep 2: Remove inspection cover.

RemovalStep 4: Remove two nuts ,

lockwashers & flat washers.

InstallationStep 2: Install two flat washers,

lockwashers, & nuts.

Note: Hold heating element inplace when tightening nuts.

RemovalStep 3: Tag & disconnect wires from

heating element terminals.

InstallationStep 3: Connect wires to heating element terminals.

(Refer to Section D for wiring diagrams)

InstallationStep 4: Install inspection cover.

Page 51: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-21Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA675200i

(Early Models)Spacer Bracket*

Heating Element - continued

Replacement - continued

Refer to: PageCover Removal ..................................... C-2

Heating ElementM7 (-020 thru -022)all

M7 (-011 thru -016)all

RemovalStep 5: Remove heating element.

Remove gaskets from terminal posts. (Later Models)Spacers

InstallationStep 1: Install gaskets onto terminal posts.

Install spacers or spacer bracket*.Insert terminal posts thru chamber & thermostat bracket.

* Spacer Bracket must be installed above gaskets to prevent leaking.

Page 52: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-22Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Overheat Thermostats

Location & Function

Overheat Thermostats PageLocation & Function .............................. B-22

Resistance Test .................................... B-23

Replacement ........................................ B-24

Wiring Diagrams ................................... D-1

Exploded View / Part Numbers ............. E-13

M7 (-020 thru -022)all

M7 (-011 thru -016)all

OverheatThermostats

MA675300i

Manual-ResetThermostat

Auto-ResetThermostat

With the power cord properly connected...Overheat ThermostatsCurrent (115 / 230 VAC) continuously flows thru the two (normally closed)overheat thermostats. This current supplies power to the timer.

If either thermostat opens (overheat or malfunction), voltage is removedfrom the timer until the thermostat is reset or replaced.

NOTEThe Manual-Reset Thermostat contacts open at approximately 285°F (140°C).To reset, allow unit to cool, then press RESET button on front of unit.

The Auto-Reset Thermostat contacts open at approximately 295°F (146°C).This thermostat automatically resets when the unit cools to approx. 265°F (129°C).

Page 53: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-23Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Overheat Thermostats - continued

Resistance Test

Refer to: PageCover Removal ..................................... C-2

M7 (-020 thru -022)all

M7 (-011 thru -016)all

OverheatThermostats

Resistance TestStep 2: Place meter probes on terminals.

[Set meter to 200 ohms (Ω)]

Resistance TestStep 1: Disconnect wires from thermostat terminals.

ResistanceTestIf reading is (approximately) 0.00 ...Thermostat is good.

If reading is OL...Replace thermostat.

0.00 (approximately)

Acceptable Reading

Attention!Inspect thermostat for physical damage (ex. cracked plastic).If damage is apparent, replace thermostat

Page 54: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-24Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA675500i

Overheat Thermostats - continued

Replacement

M7 (-020 thru -022)all

M7 (-011 thru -016)all

OverheatThermostats

RemovalStep 1: Disconnect power to sterilizer.

Drain all water from reservoir.

RemovalStep 2: Remove inspection cover.

RemovalStep 3: Loosen two nuts on

heating element posts.

InstallationStep 1: Install thermostat.

RemovalStep 4: Disconnect wires.

Remove thermostat.

InstallationStep 2: Tighten two nuts.

InstallationStep 3: Connect wires.

InstallationStep 4: Install inspection cover.

Refer to: PageWiring Diagrams ................................... D-1

Page 55: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-25Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA675700i

R

Pressure Relief Valve

Location & Function

Pressure Relief Valve

Pressure ReliefValve

If the pressure in the chamber exceeds 34 psi (234kPa)...The pressure relief valve opens to prevent unsafe conditions.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Pressure Relief Valve PageLocation & Function .............................. B-25

Testing - refer to:

Checking for Pressure Leaks ............ B-2

Replacement ........................................ B-25

Exploded View / Part Numbers ............. E-8

Replacement

RemovalStep 1: Unscrew pressure relief valve.

InstallationStep 1: Apply hi-temp sealant to valve threads.

Install valve.

Page 56: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-26Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA675800i

Timer

Location & Function

Timer PageLocation & Function .............................. B-26

Supply Voltage Test ............................. B-27

Output Voltage Test.............................. B-28

Replacement ........................................ B-29

Wiring Diagrams ................................... D-1

Exploded View / Part Numbers ............. E-12

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Timer

NOTECurrent is supplied to the timer thru the two overheat thermostats.

When the timer is turned ON...The timer contacts to the timer motor & the temperature relayclose, and voltage is supplied to these components. Whenvoltage is applied to the timer motor, the time setting countsdown.(The contacts to the timer buzzer remain open).

When the timer setting expires...The timer contacts to the temperature relay open, stoppingthe current flow to the relay.

The timer contacts to the buzzer close for one minute.Current flows to the buzzer, resulting in a audible signal. Afterone minute, the contacts to the timer motor & the buzzeropen, stopping the current flow to these two components.

Page 57: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-27Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA676000i

Timer - continued

Supply Voltage Test

CautionThis test must be performed with the power cord connected.

Refer to: PageCover Removal ..................................... C-2

Output Voltage Test.............................. B-28

line voltage (115 / 230 VAC +10%)

Supply Voltage TestStep 1: Place meter probes on terminals as shown.

[Set meter to VAC]

Supply VoltageTestIf reading is within range...Perform Output Voltage Test.

If reading is out of range...Check voltage supply.(overheat thermostats, fuse, etc.)

Acceptable Range

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Timer

Page 58: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-28Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA675900i

Timer - continued

Output Voltage Test (perform Supply Voltage Test first)

CautionThis test must be performed with the power cord connected.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Timer

Voltage to Temp. RelayStep 1: Turn timer knob to 10 minutes.

Place meter probes on terminals as shown.[Set meter to VAC]

Voltage to Timer MotorStep 1: Turn timer knob to 10 minutes.

Place meter probes on terminals as shown.[Set meter to VAC]

Voltage to Timer BuzzerStep 1: Turn timer knob to 1 minute (or less).

Place meter probes on terminals as shown.[Set meter to VAC]

Acceptable Range

line voltage (115 / 230 VAC +10%)

Output VoltageTestIf reading is within range...Timer is functioning properly.

If reading is out of range...Replace timer.

Refer to: PageCover Removal ..................................... C-2

Supply Voltage Test ............................. B-27

Page 59: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-29Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA521501i

R

Timer - continued

ReplacementRefer to: PageCover Removal ..................................... C-2

Wiring Diagrams ................................... D-1

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Timer

RemovalStep 1: Loosen set screw.

Remove timer knob.

RemovalStep 2: Tag and disconnect wires.

RemovalStep 3: Remove two screws.

Remove timer.

InstallationStep 1: Install timer.

Install two screws.

InstallationStep 2: Connect wires.

(refer to Wiring Diagrams)

InstallationStep 3: Install timer knob.

Tighten set screw.

Page 60: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-30Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA676100i

Timer Buzzer

Location & Function

Timer Buzzer M7 (-020 thru -022)all

M7 (-011 thru -016)all

Timer Buzzer PageLocation & Function .............................. B-30

Testing - refer to:

Timer ............................................... B-28

Replacement & Volume Adjustment ..... B-30

Exploded View / Part Numbers ............. E-12

Replacement & Volume Adjustment

When the timer setting expires...Timer contacts to the buzzer close for one minute.Current flows to the buzzer, causing an audible signal.

RemovalStep 1: Disconnect buzzer wires.

RemovalStep 2: Remove buzzer.

InstallationStep 1: Install buzzer.

InstallationStep 2: Connect buzzer wires.

TimerBuzzer

Volume AdjustmentTo increase volume...Loosen adjustment screw.

To decrease volume...Tighten adjustment screw.

Adjustment Screw(present on new-style buzzers only)

Page 61: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-31Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA252301i

Temperature Gauge

Replacement

Temperature Gauge PageReplacement ........................................ B-31

Exploded View / Part Numbers ............. E-5

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Temperature Gauge

RemovalStep 1: Loosen nut.

Slide retainer bracket back.

RemovalStep 2: Unscrew compressionnut.

Remove temperature gauge.

InstallationStep 1: Install retainer bracket, nut, compression nut,

& rubber sleeve onto temperature gauge.

InstallationStep 2: Orient gauge face properly.

Tighten nut.

InstallationStep 3: Slide rubber sleeve back against fitting.

Tighten compression nut. (Do not overtighten).

Page 62: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-32Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

MA676200i

R

Door Assembly

Door Replacement

Door Assembly

Door Assembly PageTesting - refer to:

Checking for Pressure Leaks ............ B-2

Door Replacement ................................ B-32

Gasket Replacement ............................ B-32

Disassembly / Assembly ...................... B-33

Exploded View / Part Numbers ............. E-6

Gasket Replacement

threadlocking adhesive.

Door RemovalStep 1: Move door handle to

unlatched position.

Door RemovalStep 2: Remove nut,

screw, & door stop.

Door InstallationStep 1: Install door stop & screw.

Gasket InstallationStep 1: Using a brush, clean all

debris from gasket channel.

Gasket InstallationStep 2: Lubricate gasket w/soapy water.

Insert gasket into gasket channel.

Door

GasketChannel

Gasket

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Door InstallationStep 2: Install nut* (Do not overtighten)

*apply removeable threadlocking adhesive

Page 63: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-33Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA522701i

DisassemblyStep 1: Remove cap nut, washer, & door.

Remove short spring & long spring.

DisassemblyStep 3: Remove T-bolt from U-bracket.

AssemblyStep 1: Install T-bolt into U-bracket.

AssemblyStep 2: Install long spring & short spring.

Install door w/bowed edges at top & bottom.AssemblyStep 3: Apply removable threadlocking adhesive

to T-bolt threads & face of washer.Install washer & cap nut.

DisassemblyStep 2: Remove push nut, pin,

handle, & cross arm.

AssemblyStep 4: Install cross arm & handle.

Secure with pin & push nut.

Door Assembly - continued

Disassembly / Assembly

Door AssemblyM7 (-020 thru -022)all

M7 (-011 thru -016)all

Page 64: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-34Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA523001i

R

Reservoir Tank

Removal

Reservoir Tank PageRemoval ............................................... B-34

Installation ............................................ B-35

Exploded View / Part Numbers ............. E-8

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Reservoir Tank

RemovalStep 1: Drain all water from reservoir.

RemovalStep 7: Pull bellows tube out of tank.

RemovalStep 8: Remove four screws & tank.

RemovalStep 2: Loosen nut.

RemovalStep 3: Remove bulkhead fitting & washer.

RemovalStep 5: Loosen nut.

RemovalStep 6: Remove condensing coil, & washer.

RemovalStep 4: Remove clamp & drain hose.

Page 65: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-35Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA523001i

R

Reservoir Tank

Installation

InstallationStep 1: Install condensing coil & washer.

Secure with nut.

InstallationStep 3: Insert bellows tubing into tank.

InstallationStep 4: Install four screws.

InstallationStep 5: Install bulkhead fitting & washer.

Partially install nut.

InstallationStep 7: Install drain tube & clamp.

InstallationStep 8: Tighten all fittings

before filling reservoir.

NoteWhen reconnecting plumbing, apply teflon tape or sealantto threads - except where compression fittings are used.

InstallationStep 6: Reconnect plumbing.

Tighten nut.

InstallationStep 2: Reconnect plumbing.

M7 (-020 thru -022)all

M7 (-011 thru -016)all

Reservoir Tank

Page 66: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-36Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA523101i

R

Chamber Assembly

Removal

Chamber Assembly

Chamber Assembly PageRemoval ............................................... B-36

Disassembly / Assembly ...................... B-37

Installation ............................................ B-38

Exploded View / Part Numbers ............. E-10

Refer to (Removing):Covers ............................................... C-2

Tray Plate / Rack .................................. C-3

Door Assembly / Gasket ....................... B-32

Temperature Regulator Assembly ........ B-15

Heating Element ................................... B-20

Overheat Thermostats .......................... B-24

M7 (-020 thru -022)all

M7 (-011 thru -016)all

RemovalStep 1: Drain all water from reservoir.

RemovalStep 2: Remove the following components:

CoversTray Plate / RackDoor Assembly & GasketTemperature Regulator Assy.Heating ElementOverheat Thermostats

RemovalStep 3: Disconnect compression

fitting from top of chamber.

RemovalStep 4: Disconnect compression

fitting from bottom of chamber.

RemovalStep 5: Remove screw(s) securing bracket* to base.

Remove chamber.

* Bracket design may vary.

Page 67: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-37Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA253301i

Chamber Assembly - continued

Disassembly / Assembly

Chamber AssemblyM7 (-020 thru -022)all

M7 (-011 thru -016)all

DisassemblyStep 1: Remove banding & insulation.

DisassemblyStep 2: Remove bracket from chamber. Note: On later models, the bracket is welded

to the chamber - skip this step.

DisassemblyStep 3: Remove two fittings from chamber.

DisassemblyStep 3: Disconnect spring.

Remove door hinge.

AssemblyStep 1: Position door hinge & washer.

Install screw & nut*.Connect spring.

*apply removeable threadlocking adhesive

AssemblyStep 2: Install two fittings* into chamber.

*apply teflon tape / hi-temp sealant

AssemblyStep 3: (If applicable) Install bracket.

AssemblyStep 4: Install insulation & banding.

RearSpring Hole

Page 68: SpeedClave Steam Sterilizers M7 M7

Component Testing & Repair

B-38Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA523101i

R

Chamber Assembly- continued

Installation

Chamber Assembly

Refer to (Installing):Overheat Thermostats .......................... B-24

Heating Element ................................... B-20

Temperature Regulator Assembly ........ B-16

Door Assembly / Gasket ....................... B-32

Tray Plate / Rack .................................. C-3

Covers ............................................... C-2

M7 (-020 thru -022)all

M7 (-011 thru -016)all

InstallationStep 1: Secure bracket* to base.

*Bracket design may vary.

AssemblyStep 2: Connect compression fitting / plumbing

to fitting on bottom of chamber.

AssemblyStep 3: Connect compression fitting / plumbing

to fitting on top of chamber.

AssemblyStep 4: Install the following components:

Overheat ThermostatsHeating ElementTemperature Regulator Assy.Door Assembly & GasketTray Plate / RackCovers

NoteReplace compression fittings if damage is apparent.

Page 69: SpeedClave Steam Sterilizers M7 M7

C-1

Access Procedures

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Sec

tion

CRemoving & Installing: PageCovers / Panels ................................... C-2

Tray Plate / Rack .................................. C-3

Draining / Filling the Reservoir ............. C-4

Access Procedures

Page 70: SpeedClave Steam Sterilizers M7 M7

C-2

Access Procedures

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA521401i

R

Covers / Panels

Removal / Installation

Covers / Panels

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

CautionAlways unplug power cord beforeremoving any covers / panels.

Installation: M7 (-011 thru -016)Step 1: Slide cover down over

edge of front panel.

Installation: M7 (-020 thru -022)Step 1: Slide cover down over edge of front

panel & back panel (not shown).

RemovalStep 1: Remove six screws.

RemovalStep 2: Remove cover.

Installation: M7 (all)Step 2: Install six screws.

Page 71: SpeedClave Steam Sterilizers M7 M7

C-3

Access Procedures

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

Tray Plate / Rack

Tray Plate / Rack

Removal / InstallationRefer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

CautionAlways allow unit to cool beforeremoving trays or rack.

RemovalStep 1: Lift up on left side of tray plate

until it "pops" free of rack.Remove tray plate.

Offset Ends(left)

Straight Ends(right)

RemovalStep 2: Squeeze bottom of rack together.

Pull rack out of chamber.

InstallationStep 1: Position rack w/ offset ends to the left.

Squeeze bottom of rack together.Push rack into chamber.

InstallationStep 3: Push left side of plate down until it

engages w/ offset ends of rack.InstallationStep 2: Insert the right side of tray plate

under bottom wire of rack.

ALL

Bottom Wire

Page 72: SpeedClave Steam Sterilizers M7 M7

C-4

Access Procedures

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

MA514002

Draining / FillingReservoir

Draining / Filling the ReservoirRefer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

DrainingStep 1: Hold hose over a drain or suitable container.

Insert drain hose into coupling on front of unit.

Note: The max. reservoir capacity is: 1.3 gallon (4.9 liters)

Note: To remove drain hose, press release lever.

FillingStep 1: Pour distilled water into reservoir until

the water level reaches the "full mark".Do not overfill!

ALL

FullMark

Page 73: SpeedClave Steam Sterilizers M7 M7

D-1Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Wiring Diagrams

Sec

tion

DWiring Diagrams& Schematics

Model Page115 VAC models:

M7 (-011 / -013 / -014 / -015) ............ D-2

M7 (-020 / -022) ................................ D-3

230 VAC models:

M7 (-012 / -016) ................................ D-4

M7 (-021)........................................... D-5

Page 74: SpeedClave Steam Sterilizers M7 M7

D-2Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Wiring Diagrams

TIMERSWITCH

TEMPERATUREREGULATOR

RELAY

HEATING ELEMENT

HEATER LIGHT

TIMER BUZZER

TIMER MOTOR

NC

NC

"A"

BLACK WHITE

BLACK

YELLOW

WHITE YELLOW

WH

ITE

WH

ITE

YELLOW

"B" "C"

NC

MANUAL-RESETOVERHEAT

THERMOSTAT( LOW WATER )

AUTO-RESETOVERHEAT

THERMOSTAT( LOW WATER )

GR

EE

N

AUTO-RESETOVERHEAT

THERMOSTAT( LOW WATER )

BLACK

RE

D

WHITE

WHITE

YE

LLO

W

YELLOW

WHITE

WH

ITE

GR

EE

N

GROUND

HEATING ELEMENT

TEMPERATUREREGULATOR

RELAY

TIMERASSEMBLY

POWER CORD

TIMERBUZZER

YELLOWYELLOW

RE

D

MANUAL-RESETOVERHEAT

THERMOSTAT( LOW WATER )

HEATER LIGHT

YE

LLO

W

WH

ITE

3

2

1TM

B

015-1815-00WIRING HARNESS

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Wiring Diagrams

M7 (-011)all

M7 (-013)all

M7 (-014)all

M7 (-015)all

Wiring Diagrams(115 VAC Models)

Rev. 5/09

Page 75: SpeedClave Steam Sterilizers M7 M7

D-3Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Wiring Diagrams

POWERCORD

FILL / VENTVALVE

TIMER MOTOR

TIMER BUZZER

HEATER LIGHT

HEATER ELEMENT

NC

NC

NC

MANUAL- RESETOVERHEAT

THERMOSTAT(LOW WATER)

AUTO- RESETOVERHEAT

THERMOSTAT(LOW WATER)

TIMERSWITCH

TEMPERATUREREGULATOR

RELAY

F1 FUSE :115V UNIT; 12A 250V FAST ACTING

YELLOW

WH

ITE

GROUND

BLACK

GR

EE

N

WH ITE

3 B TM

Open 285 oF (+/-12

oF)

(140.5 oC)

Closes 200 oF (93

oC)

Open 295 oF (+/-12

oF)

(146oC)

Closes 265 oF (129

oC)

1

TM

FILL / VENTSWITCH

NO

115 VAC = 12 Ohms

(Lit When HeaterIs On)

(Switch Contacts OpenUntil Lever is pressed)

567

RED8

3

7

9

14

2

BLACK

BLACK

RED

BLACK

BLACK BLACK

YELLOW YELLOW

WH ITEWH ITE

YELLOW

YELLOWWH ITE

MA676900i

RED

TM

13

POWERCORD

FILL / VENTVALVE

TIMERASSEMBLY

TIMERBUZZER

PILOTLIGHT

MANUAL- RESETOVERHEAT

THERMOSTAT(LOW WATER)

AUTO- RESETOVERHEAT

THERMOSTAT(LOW WATER)

TEMPERATUREREGULATOR

RELAY

F1 FUSE: 12A, 250V FAST ACTING

WH

YL

YL

YLYL

GROUND

RD

015-1790-00WIRING

HARNESS

Open 285 oF (+/-12

oF)

(140.5 oC)

Closes 200 oF (93

oC)

Open 295 oF (+/-12

oF)

(146oC)

Closes 265 oF (129

oC)

FILL / VENTSWITCH

(Switch contacts OPENuntil lever is pressed)

21

1

2

4

4

5

3

5

6

7

6

89

8

3

7

BLACK

WH ITE

GR

EE

N

YELLOW

RED

BLACK

HEATER ELEMENT115 VAC = 12 Ohms

WH ITE

WH ITE

WH ITE

YELLOW

YELLOW

YELLOW

RE

D

REDB

LAC

K

(Lit when heater is ON)

RE

D

RED

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

Wiring Diagrams

M7 (-020)all

M7 (-022)all

Wiring Diagrams(115 VAC Models)

Page 76: SpeedClave Steam Sterilizers M7 M7

D-4Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Wiring Diagrams

TIMERSWITCH

TEMPERATUREREGULATOR

RELAY

HEATING ELEMENT

HEATER LIGHT

TIMER BUZZER

TIMER MOTOR

NC

NC

"A"

BLACK

YELLOW

WHITE YELLOW

WH

ITE

WH

ITE

YELLOW

"B" "C"

NC

MANUAL-RESETOVERHEAT

THERMOSTAT( LOW WATER )

AUTO-RESETOVERHEAT

THERMOSTAT( LOW WATER )

BROWN BLUE

GR

EE

N /

YE

LLO

W

BROWN

AUTO-RESETOVERHEAT

THERMOSTAT( LOW WATER )

RE

D

WHITE

YE

LLO

W

YELLOW

WHITE

WH

ITE

GROUND

HEATING ELEMENT

TEMPERATUREREGULATOR

RELAY

TIMERASSEMBLY

POWER CORD

TIMERBUZZER

YELLOWYELLOW

RE

D

MANUAL-RESETOVERHEAT

THERMOSTAT( LOW WATER )

HEATER LIGHT

YE

LLO

W

WH

ITE

3

2

1TM

B

BLUE

RED

RED

GR

EE

N /

YE

LLO

W

015-1815-00WIRING HARNESS

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-012)all

M7 (-016)all

Wiring Diagrams(230 VAC Models)

Rev. 5/09

Page 77: SpeedClave Steam Sterilizers M7 M7

D-5Models:

Serial Numbers:

© Midmark Corporation 2004 SF-1854

Wiring Diagrams

POWERCORD

FILL / VENTVALVE

TIMER MOTOR

TIMER BUZZER

HEATER LIGHT

HEATER ELEMENT

NC

NC

NC

MANUAL- RESETOVERHEAT

THERMOSTAT(LOW WATER)

AUTO- RESETOVERHEAT

THERMOSTAT(LOW WATER)

TIMERSWITCH

TEMPERATUREREGULATOR

RELAY

F1 FUSE :230V UNIT; 8A 250V FAST ACTING

YELLOW

WH

ITE

GROUND

BLUE

GR

EE

N /

YE

LLO

W

BROWN

3 B TM

Open 285 oF (+/-12

oF)

(140.5 oC)

Closes 200 oF (93

oC)

Open 295 oF (+/-12

oF)

(146oC)

Closes 265 oF (129

oC)

1

TM

FILL / VENTSWITCH

NO

230 VAC = 49 Ohms

(Switch contacts Openuntil leveris pressed)

567

RED8

3

7

9

14

2

BLACK

RED

BLACK

BLACK BLACK

YELLOW YELLOW

WH ITEWH ITE

YELLOW

YELLOWWH ITE

MA677100i

RED

TM

13

POWERCORD

FILL / VENTVALVE

TIMERASSEMBLY

TIMERBUZZER

PILOTLIGHT

MANUAL- RESETOVERHEAT

THERMOSTAT(LOW WATER)

AUTO- RESETOVERHEAT

THERMOSTAT(LOW WATER)

TEMPERATUREREGULATOR

RELAY

F1 FUSE: 8A, 250V FAST ACTING

WH

YL

YL

YLYL

RD

015-1790-00WIRING

HARNESS

Open 285 oF (+/-12

oF)

(140.5 oC)

Closes 200 oF (93

oC)

Open 295 oF (+/-12

oF)

(146oC)

Closes 265 oF (129

oC)

FILL / VENTSWITCH

(Switch contacts OPENuntil lever is pressed)

21

1

2

4

4

5

3

5

6

7

6

89

8

3

7

YELLOW

RED

BLACK

WH ITE

WH ITE

WH ITE

YELLOW

YELLOW

YELLOW

RE

D

REDB

LAC

K

(Lit when heater is ON)

RE

D

RED

(Lit when heater is ON)

BLACK

BLUE

GROUND

GR

EE

N /

YE

LLO

W

BROWN

HEATER ELEMENT230 VAC = 49 Ohms

Refer To: PageOperation & Troubleshooting ................ A-1

Component Testing / Repair ................. B-1

Access Procedures .............................. C-1

Wiring Diagrams ................................... D-1

Exploded Views / Part Numbers ........... E-1

M7 (-021)all

Wiring Diagrams(230 VAC Models)

Page 78: SpeedClave Steam Sterilizers M7 M7
Page 79: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Exploded Views &Parts Lists

Sec

tion

EModel Page

M7 (-011 thru -016) ...................... E-2

M7 (-020 thru -022) ...................... E-3

E-1

Page 80: SpeedClave Steam Sterilizers M7 M7

R

* Indicates multiple pages due to a serial number break for the parts illustrationE-2

M7(-011 thru -016)

MA677800i

Fuses .............. E-15

Reservoir ..................... E-8

includes:condensing coil & drain tube

Fuses .............. E-15

Covers ......................E-4

includes: main cover, base plate, leveling feet, fill cap, etc.

ChamberComponents ............... E-10

includes: fittings & mounting bracket

Door Components ...... E-6

includes: hinge, & door gasket

Front PanelComponents ............E-5

includes: heater light, thermometer, knobs, & drain coupling

Temperature RegulatorComponents ............... E-7

includes: temp. relay & diaphragm cup

Bellows & Plumbing ... E-11*

Timer / Buzzer ............. E-12Timer & Buzzer ........... E-12

Labels & Decals .......... E-17

Fill / Vent Valve& Plumbing ................. E-9*

Items not shown:Heating Element & Thermostats .......... E-13Power Cord ................. E-14Rack & Trays ............... E-15Packaging.................... E-16

Page 81: SpeedClave Steam Sterilizers M7 M7

* Indicates multiple pages due to a serial number break for the parts illustration E-3

M7(-020 thru -022)

MA677801i

R

Reservoir Tank ............ E-8.1

includes:condensing coil & drain tube

Temperature RegulatorComponents ............... E-7

includes: temp. relay & diaphragm cup

Bellows & Plumbing ... E-11.2

Timer & Buzzer ........... E-12

Labels & Decals .......... E-17.1

Fill / Vent Valve& Plumbing ................. E-9.3

includes: fill/vent switch

Items not shown:Heating Element & Thermostats .......... E-13.1Power Cord & Fuse..... E-14.1Rack & Trays ............... E-15Packaging.................... E-16

Covers ......................E-4.1

includes: main cover, base plate, leveling feet, fill cap, etc.

ChamberComponents ............... E-10.1

includes: fittings & mounting bracket

Door Components ...... E-6

includes: hinge, & door gasket

Front PanelComponents ............E-5.1

includes: heater light, thermometer, knobs, & drain coupling

Page 82: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

5 6

2

3

4

9

7

8

128

1

10

11

E-4

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Covers

050-4423-00

H98009

030-1198-00

053-1037-00

H98135

H98011

045-0001-31

040-0010-125

122357

002-0822-00

050-4286-00

M7 (-011 thru -016)all

Rev. 8/05

MA674101i

002-0643-00

Item Description Qty.

1 Cover Kit (Includes items 2 thru 4) .............. 12 • Cooling Rail ............................................... 23 • Main Cover ................................................ 14 • Push Nut .................................................... 65 Fill Cap ......................................................... 16 Rubber Seal ................................................. 17 Lockwasher .................................................. 68 Screw (#10 x 1/2", self-tapping) ................... 69 Keps Nut ....................................................... 4

10 Rubber Foot Kit (includes nut) ..................... 411 Base Plate .................................................... 112 Inspection Cover .......................................... 1

Page 83: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

11

10

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

M7 (-020 thru -022)all

Covers

Item Description Qty.

1 Fill Cap ......................................................... 12 Rubber Seal ................................................. 13 Main Cover w/ labels

(includes items 4 & 5) ................................ 14 Push Nut ....................................................... 65 Cooling Rail .................................................. 26 Lockwasher .................................................. 17 Screw (#10 x 5/8", self-drilling/tapping) ....... 78 Nut (used only on units built prior to serial

number V411489) ........................................ 49 Rubber Foot Kit (includes nut) .................... 4

10 Base Assembly ............................................ 111 Inspection Cover ......................................... 1

MA674101i

H98009

053-1037-00

H98135

H98011

002-0840-00

045-0001-31

040-0010-163

041-0010-04

002-0822-00

050-4423-00

029-3408-00

E-4.1Rev. 4/07

Page 84: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Item Description Qty.

1 Front Panel Assy. (includes items 2 & 3) ..... 12 • Front Panel (w/o label) .............................. 13 • Label .......................................................... 14 Alignment Pin ............................................... 15 Lockwasher .................................................. 16 Screw ........................................................... 17 Thermometer Bracket .................................. 18 Jam Nut ........................................................ 19 Refer to: Temp. Regulator Components E-7

10 Refer to: Fill / Vent Valve ......................... E-911 Shut-Off Coupling ........................................ 112 Lever Knob .................................................. 113 Timer Knob (includes set screw) ................ 114 Temperature Knob (includes set screw) .... 115 Refer to: Heating Element /

Overheat Thermostats ....................... E-1316 Heater Light ................................................. 117 Thermometer Kit (incl. items 18 thru 21) .... 118 • Compression Nut ...................................... 119 • Washer (#6 - 5/16" O.D.) ......................... 120 • Rubber Sleeve .......................................... 121 • Male Connector ........................................ 1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

M7 (-011 thru -016)allE-5

Front PanelComponents

MA671907i

Ritter: 002-0688-00Midmark: 002-0688-01

Ritter/Midmark: 061-0774-00Dabi Alante: 061-0774-02

050-4285-10

057-0461-00

045-0001-08 040-0010-75

H224033

041-0250-07

014-0183-01

053-0404-07

014-0192-00

053-0990-00

053-0987-00053-1008-01

053-1008-00

100 / 120 VAC: 015-1304-00230 VAC: 015-1304-01

002-0242-00

Rev. 6/09

045-0001-21

Page 85: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

MA671908i

Front PanelComponents

Item Description Qty.

1 Front Panel (w/o label) ................................ 12 Label ............................................................. 13 Alignment Pin ............................................... 14 Lockwasher ................................................. 15 Screw ........................................................... 16 Jam Nut ........................................................ 17 Refer to: Temp. Regulator Components E-78 Shut-Off Coupling ........................................ 19 Refer to: Fill / Vent Valve ......................... E-9

10 Timer Knob (includes setscrew) ................. 111 Temperature Knob (includes setscrew) ..... 112 Refer to: Heating Element /

Overheat Thermostats ....................... E-1313 Heater Light ................................................. 114 Thermometer Kit (incl. items 15 thru 18) .... 115 • Compression Nut ...................................... 116 • Washer ...................................................... 117 • Rubber Sleeve .......................................... 118 • Male Connector ........................................ 1

050-6034-10

Ritter: 061-0900-00Midmark: 061-0900-01

057-0461-00

045-0001-08 040-0010-75

041-0250-20

014-0183-01

053-0404-07

014-0192-00100 / 120 VAC: 015-1304-00230 VAC: 015-1304-01

002-0242-00

053-1008-01

053-1008-00

053-0990-00

M7 (-020 thru -022)all E-5.1

Rev. 5/09

045-0001-21

Page 86: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

19

15

16

18

17

2

4

6

8

9

3

5

12

14

7

111

10

24

23

21

20

22

13

002-0249-00

Item Description Qty.

1 Door Assembly(incl. items 2 thru 13) ........... 12 • Handle ....................................................... 13 • Push Nut .................................................... 14 • Pin ............................................................. 15 • Refer to Label Location ......................... Ref6 • Cross Arm.................................................. 17 • T-bolt ......................................................... 18 • U-bracket ................................................... 19 • Long Spring ............................................... 1

10 • Short Spring .............................................. 111 • Door ........................................................... 112 • Gasket ....................................................... 113 • Washer ................................................... 114 • Cap Nut (apply adhesive #042-0024-02) .. 115 Door Stop ..................................................... 116 Bolt ............................................................... 117 Door Gasket ................................................. 118 Bolt ............................................................... 119 Refer to Chamber Components ............... Ref20 Spring ........................................................... 121 Torque Nut ................................................... 122 Washer ......................................................... 123 Hinge ............................................................ 124 Torque Nut ................................................... 1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

E-6Door Components

MA511703i

053-0513-00

K906034

042-0005-16

050-2359-00

H96772

H96776

H96775

H96774

H106313

H96771

H96773

H97991

040-0010-103

002-0243-00

040-0250-69

H106330

041-0250-15

045-0000-04

050-4783-00

041-0010-10

ALL

053-1579-00

Rev. 5/05

Page 87: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Temperature Relay Kit(incl. items 2 & 3) ...................................... 1

2 • Spacer ....................................................... 13 • Screw (#6-32 x 7/16") ................................ 34 Flexible Shaft Assembly ............................... 15 Set Screw ..................................................... 16 Diaphragm Cup ............................................ 17 Gasket .......................................................... 18 Nut ................................................................ 19 Lockwasher .................................................. 1

Temperature RegulatorComponents

MA511902i

002-0838-00

H97951

H98064

040-0006-44

040-0008-47

H97948

H97966

H97076

H97075

ALL

E-7

040-0006-38

Rev 1/08

Page 88: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

2

3

4

7

3

11

12

149

15

8

7

6

13

10

5

1

Item Description Qty.

1 Tank with Lid (not available separately) ...... 12 Pressure Relief Valve .................................. 13 Bulkhead Fitting (includes nut) .................... 14 Neoprene Washer ........................................ 15 Condensing Coil Assy. (incl. items 6 & 7) ... 16 • Compression Fitting .................................. 17 • Bulkhead Fitting (includes nut) ................. 18 Neoprene Washer ........................................ 19 Hose Clamp ................................................. 2

10 Tank Drain Tube .......................................... 111 Refer to: Front Panel Components .......... Ref12 Drain Hose Kit (incl. items 13 thru 15) ......... 113 • Barbed Fitting ............................................ 114 • Removeable Drain Tube ........................... 115 Screw ........................................................... 2

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

M7 (-011 thru -016)all

ReservoirE-8

053-0944-00

029-2291-00

014-0183-02

014-0209-00

053-1036-01

040-0010-75

016-0748-01

002-0653-00

053-0391-03

053-0990-01

053-0391-02

053-1036-01

014-0324-00

H98089

MA675105i

Rev. (4/20/12)

Page 89: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

Item Description Qty.

1 Reservoir Assembly (incl. items 2 thru 11) 12 • Tank with Lid (not available separately) .. 13 • Pressure Relief Valve ............................... 14 • Bulkhead Fitting (includes nut) ................. 15 • Washer ...................................................... 26 • Condensing Coil ........................................ 17 • Compression Fitting .................................. 18 • Bulkhead Fitting (includes nut) ................. 19 • Neoprene Washer .................................... 1

10 • Hose Clamp .............................................. 211 • Tank Drain Tube ....................................... 112 Refer to: Front Panel Components ......... Ref13 Drain Hose Kit (incl. items 13 thru 15) ........ 114 • Barbed Fitting ............................................ 115 • Removeable Drain Tube .......................... 116 Screw ........................................................... 2

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Reservoir

MA675106i

M7 (-020 thru -022)all E-8.1

Rev. 1/09

3

7

8

9

10

1

2

5

6

4

4

12

11

14

15

16

8

13

053-0944-00

014-0432-00

014-0485-00

014-0209-00

053-1036-01

040-0010-75

016-0748-01

053-0391-03

053-0990-01

053-0391-02

H97966

014-0324-00

H98089

029-3302-00

002-0653-00

Page 90: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

15

14

13

12

16

17

1810

11

18

050-4315-00

002-0247-00

MA674600i

Fill / Vent Valve& Plumbing

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Fill/Vent Valve Kit (incl. items 2 thru 9) ....... 12 • Compression Fitting .................................. 23 • O-ring (apply hi-temp lube) ....................... 14 • No Longer Available .................................. 15 • O-ring (apply hi-temp lube) ....................... 16 • No Longer Available .................................. 17 • No Longer Available .................................. 18 • No Longer Available .................................. 19 • Bracket ...................................................... 1

10 Torque Nut ................................................... 111 Lever ............................................................ 112 Lever Bracket ............................................... 113 Screw (#10-32 x 3/4") .................................. 114 Pop Rivet ...................................................... 315 Clip ............................................................... 316 Tubing (Fill) .................................................. 117 Tubing (Vent) ............................................... 118 Elbow Fitting ................................................. 2

E-9

H98070

014-0183-02

H98069

H98136

H98138

H98068

041-0010-09

H98077

041-0010-10

050-1926-00

040-0010-96

042-0010-07

H98072

052-0332-00

052-0331-00

014-0184-01

M7 (-011)MH1000 thru MH1802

M7 (-012)MJ1000 thru MJ1091

M7 (-014)ML1000 thru ML2133

Rev. 9/11

Replaced by 002-0649-00

Replaced by 016-0843-00

Page 91: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

8

13

12

11

13

E-9.1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

MA674601i

002-0648-00014-0183-02

014-0369-00

050-4793-00

Item Description Qty.

1 Fill/Vent Valve Kit (incl. items 2 thru 6) ....... 12 • Compression Fitting .................................. 23 • Bracket ...................................................... 14 • No Longer Available .................................. 15 • Clip ............................................................ 36 • Pop Rivet ................................................... 37 Lever ............................................................ 18 Lever Bracket ............................................... 19 Torque Nut ................................................... 1

10 Screw (#10-32 x 3/4") .................................. 111 Tubing (Fill) .................................................. 112 Tubing (Vent) ............................................... 113 Elbow Fitting ................................................. 2

016-0843-00

042-0010-15

050-4315-00

050-1926-00

041-0010-10

040-0010-96052-0332-00

052-0331-00

014-0184-01

M7 (-011)MH1803 thru MH3104

M7 (-012)MJ1092 thru MJ1241

M7 (-014)ML2132 thru ML4209

Fill / Vent Valve& Plumbing

Rev. 9/11

Replaced by 002-0649-00

Replaced by H98077

Page 92: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

7

12

11

10

12

MA674602i

002-0649-00

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

E-9.2

014-0402-00

Item Description Qty.

1 Fill/Vent Valve Kit (incl. items 2 thru 5) ........ 12 • Valve Body (includes nut) ......................... 13 • Bracket ...................................................... 14 • Clip ............................................................ 35 • Pop Rivet ................................................... 36 Lever ............................................................ 17 Lever Bracket ............................................... 18 Torque Nut ................................................... 19 Screw (#10-32 x 3/4") ................................. 1

10 Tubing (Fill) .................................................. 111 Tubing (Vent) ............................................... 112 Elbow Fitting ................................................. 2

H98077

016-0843-00

042-0010-15

050-4315-00

050-1926-00

041-0010-10

040-0010-96

052-0332-00

052-0331-00

014-0184-01

M7 (-011)MH3105 thru presentV2200 thru present

M7 (-012)MJ1242 thru presentV2200 thru present

M7 (-014)ML4210 thru presentV2200 thru present

M7 (-013/-015 /-016)all

Fill / Vent Valve& Plumbing

Rev. 4/09

Page 93: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

050-6057-00

040-0010-74

040-0250-157

014-0433-00

014-0183-02

014-0184-01

115 VAC: 014-0236-00230 VAC: 014-0236-01

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

MA674605i

M7 (-020 thru -022)all E-9.3

015-1791-00

Fill / Vent Valve& Plumbing

115 VAC: 014-0236-10

014-0236-11

Item Description Qty.

1 Fill/Vent Valve .............................................. 12 Mounting Bracket ......................................... 13 Screw (#10-32 x 3/8") .................................. 24 Bolt (1/4-20 x 1/2") ....................................... 25 Elbow Fitting ................................................. 16 Male Fitting (3/8") ......................................... 17 Tubing .......................................................... 18 Fill / Vent Switch ........................................... 19 Replacement Coil ........................................ 1

10 Replacement Kit (Includes plunger, guide,and seals) .................................................... 1

Rev 2/08

Page 94: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

12

10

4

6

5

7

9

8

3

2

1

11

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Chamber Assembly - completeincludes items 2 & 3 and the following: .... 1• Heating Element & Thermostats (2)• Temperature Regulator Assembly• Door Hinge & Door Gasket

2 • Compression Fitting .................................. 13 • Chamber Shell only .................................. 24 Banding Strap (end) ..................................... 15 Chamber Bracket ......................................... 16 Screw ........................................................... 17 Screw ........................................................... 28 Elbow Fitting ................................................. 19 Nut ................................................................ 1

10 Insulation ...................................................... 111 Banding Strap .............................................. 212 Banding Clip ................................................. 2

MA676600i

115 VAC: 029-1865-00230 VAC: 029-1865-01100 VAC: 029-1865-02

014-0183-01

030-0976-00

H99388-27

051-0721-00

040-0250-47

040-0010-74

014-0184-01

041-0250-15

H97980

H99387H99388-29

The chamber bracket (item #6) on some models may be attached to the base plate w/ pop rivets.When installing a new bracket, replace rivets w/ screws (item #7).

Chamber ComponentsE-10

M7 (-011 thru -016)all

Rev. 6/25/12

Page 95: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854E-10.1

78

6

5

2

4

1

3

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Chamber Assembly - completeincludes items 2 thru 4 & the following: .... 1• Heating Element & Thermostats (2)• Temperature Regulator Assembly• Door Hinge & Door Gasket• Timer Buzzer

2 • Compression Fitting .................................. 13 • Filter Screen .............................................. 14 • Elbow Fitting .............................................. 15 Screw ........................................................... 16 Insulation ...................................................... 17 Banding Strap .............................................. 28 Banding Clip ................................................. 2

Chamber ComponentsM7 (-020 thru -022)all

MA676601i

115 VAC: 029-3278-00230 VAC: 029-3278-01

014-0183-01

040-0250-157

014-0184-01

H97980

H99387H99388-29

002-0360-00

Rev. 1/11

Page 96: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

5

3

4

3

1

2

13

7

6

10

8

9

10

11

12

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Refer to: Reservoir ................................... Ref2 Tube ............................................................. 13 Compression Nut ......................................... 24 Bellows ......................................................... 15 Tube ............................................................. 16 Chamber Manifold Tube Kit

(includes items 7 thru 9)7 • Refer to: Chamber Assembly ................ Ref8 • Tube .......................................................... 19 • Elbow Fitting .............................................. 1

10 Refer to: Front Panel Components .......... Ref11 Compression Fitting ..................................... 112 Street Tee Fitting ......................................... 213 Refer to: Chamber Assembly .................. Ref

E-11Bellows & Plumbing

014-0240-00

M7 (-011)MH1000 thru MH6044

M7 (-012)MJ1000 thru MJ1696

M7 (-014)ML1000 thru ML7297

014-0392-00

002-0375-00

014-0241-00

014-0183-01

014-0058-00

014-0184-00

H335943

Older Models w/ rubbertubing MUST use Item #6

002-0679-00

MA675600i

Page 97: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

1

12

4

11

8

9

8

6 5

3

47

1013

2

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Refer to: Reservoir ................................... Ref2 Bellows Kit (includes items 3 thru 7) ........... 13 Tube ............................................................. 14 Compression Nut ......................................... 35 Bellows ......................................................... 16 Tube ............................................................. 17 Compression Fitting ..................................... 18 Street Tee Fitting ......................................... 29 Refer to: Front Panel Components .......... Ref

10 Elbow Fitting ................................................. 111 Refer to: Chamber Assembly .................. Ref12 Tube ............................................................. 113 Compression Fitting ..................................... 1

Bellows & Plumbing

MA675601i

002-0375-01

014-0058-00

014-0184-00

H335943

M7 (-011)MH6045 thru presentV2200 thru present

M7 (-012)MJ1697 thru presentV2200 thru present

M7 (-014)ML7298 thru presentV2200 thru present

M7 (-013/-015 /-016)all E-11.1

014-0183-01

Note: Item #2 will reconfigureBellows plumbing to matchcurrent models.

(If plumbing has been reconfigured, refer to nextpage for replacement parts)

Original PlumbingConfiguration

Reconfigured PlumbingConfiguration

Page 98: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

6

4

5

4

17

7

8

910

3

2

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Refer to: Reservoir ................................... Ref2 Bellows Kit .................................................... 13 • Tube .......................................................... 14 • Compression Nut ....................................... 25 • Bellows ...................................................... 16 • Tube .......................................................... 17 • Compression Fitting .................................. 28 Refer to: Front Panel Components .......... Ref9 Tube ............................................................. 1

10 Refer to: Chamber Components .............. Ref

E-11.2Bellows & Plumbing

MA675602i014-0482-00

014-0392-00

002-0375-00

014-0481-00

014-0480-01

M7 (-020 thru -022)all

014-0484-00

002-0375-01

Rev 10/05

Page 99: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

4

2

3

5

1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Timer Assembly Kit (incl. items 2 thru 5) ..... 12 • Screw (#8-32 x 5/16") ................................ 13 • Buzzer ....................................................... 14 • Screw (#6-32 x 7/16") ................................ 15 • Extended Wiring Harness - not shown

(required for older units only) ................. 1

040-0006-44

MA671802i

Timer / Buzzer

115 VAC: 002-0373-00230 VAC: 002-0373-01230 VAC -016: Not available

040-0008-29

115 VAC: 015-1813-00230 VAC: 015-1813-01

015-1815-00[Ext. Wiring Harness - not shown]

E-12

MA671801i

Rev. 8/05

M7 (-011 thru -015)all

M7 (-020 thru -022)all

Page 100: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

9

10

11

12

1314

15

16

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Heating Element (includes items 2 & 3) ...... 12 • Nut ............................................................. 23 • Gasket ....................................................... 24 Overheat Thermostat (manual-reset) .......... 15 Overheat Thermostat (auto-reset) ............... 16 Bracket ......................................................... 17 Washer ......................................................... 28 Lockwasher .................................................. 29 Heater Spacer .............................................. 1

10 Nut ................................................................ 111 Rocker Assembly ......................................... 112 Spacer .......................................................... 113 Screw ........................................................... 114 Bracket ......................................................... 115 Reset Button ................................................ 116 Reset Rod .................................................... 1

E-13

MA673901i

115 VAC: 002-0839-00230 VAC: 002-0839-01100 VAC: 015-0894-02

H97949

H98137

H225639

002-0370-00

030-0929-00

H97971

045-0001-56

030-1484-00

057-0520-00

053-1035-00

H225619040-0010-74

H225620

041-0010-10

H225913

Old-style rocker assembly is not available.Replace with item #11.

M7 (-011 thru -016)all

Heating Element &Thermostats

Rev. 6/09

Page 101: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

9

10

11

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Heating Element Kit(includes nuts, e-clips & item 2) ............... 1

2 • Gasket ...................................................... 23 Overheat Thermostat (manual-reset) ........ 14 Overheat Thermostat (auto-reset) ............. 15 Bracket ......................................................... 16 Bushing ........................................................ 17 Spring ........................................................... 18 Reset Rod .................................................... 19 Lockwasher ................................................. 2

10 Washer ......................................................... 211 Reset Button Actuator ................................. 1

Heating Element &Thermostats

MA673903i

115 VAC: 002-0839-00230 VAC: 002-0839-01

H225639

002-0370-00

E-13.1M7 (-020 thru -022)

all

H98137

030-1521-00

H97971

045-0001-56

053-1461-00

053-1460-00

053-0068-13

025-0078-00

Rev 6/09

Page 102: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

4

2

3

1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Item Description Qty.

1 Power Cord .................................................. 12 Strain Relief .................................................. 13 Screw (#10-32 x 3/8") .................................. 14 Keps Nut ....................................................... 1

H98160

MA677200i

E-14Power Cord

115 VAC: H98159230 VAC: 015-0688-13

122357

040-0010-74

AttentionThese models do not have a fuse.

M7 (-011 thru -016)all

Page 103: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

4

6

5

1

3

2

E-14.1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

MA677201i

Power Cord / FuseM7 (-020 thru -022)all

Item Description Qty.

1 Power Cord .................................................. 12 Fuse:

a) 12A, 250V, Fast-Acting, 1/4" x 1-1/4" ... 1b) 8A, 250V, Fast-Acting, 5mm x 20mm ... 1

3 Fuse Holder .................................................. 14 Screw (#10-32 x 3/8") .................................. 15 Keps Nut ....................................................... 16 Strain Relief .................................................. 1

115 VAC: H98159230 VAC: 015-0688-13

a.) 115 VAC: 015-0346-35b.) 230 VAC: 015-0346-11

115 VAC: 015-1259-04230 VAC: 015-1259-05

122357

040-0010-74

H98160

Rev. 7/05

Page 104: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

2

3

1

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

SA102200i

Rack & Trays

Item Description Qty.

1 Tray Rack ..................................................... 12 4 inch Tray ................................................... 13 5 inch Tray ................................................... 2

002-0251-00

002-0253-00

002-0252-00

E-15ALL

Rev. 1/07

Page 105: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

2

5

3

4

6

1

R

Packaging

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

054-0326-00

MA514703i

Item Description Qty.

1 M7 Packaging Kit - Generic(includes items 2 thru 8) ............................ 1

2 • Top Front Pad ........................................... 13 • Top Rear Pad ............................................ 14 • Bottom Rear Pad ....................................... 15 • Bottom Front Pad ...................................... 16 • Carton ........................................................ 17 • Sterilizer Cleaner (not shown) ................... 18 • MSDS Sheet (not shown) ......................... 1

ALL

054-0327-00

054-0325-00054-0324-00

066-0857-00

E-16

027-0474-00

Rev. 05/11

Page 106: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

6

7

10

5

1

3

4

9

8

2

Labels & Decals

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

Ritter/Midmark: 061-0443-00Dabi Alante: 061-0594-00

MA674300i

Item Description Qty.

1 Operating Instructions Plate ........................ 12 Danger Plate (Dabi Alante only) .................. 13 Warning Label .............................................. 14 Caution Label ............................................... 15 Serial Number Label (large- n/a) ................. 16 UL Label (applicable units only) ................... 17 CSA Label (applicable units only) ................ 18 Wiring Diagram Label .................................. 19 Caution HOT Label ...................................... 1

10 Serial Number Label (small - n/a) ................ 1

E-17

Ritter/Midmark: H335669Dabi Alante: 061-0598-00

Ritter/Midmark: 061-0203-00Dabi Alante: 061-0600-00

061-0301-00

061-0509-00

M7 (-011/-013/-014): 061-0519-00M7 (-012): 061-0527-00M7 (-015): 061-0601-00M7 (-016): 061-0607-00

Ritter/Midmark: 061-0649-00Dabi Alante: 061-0599-00

M7 (-011 thru -016)all

Ritter/Midmark: H284693Dabi Alante: 061-0593-00

Rev. 5/07

Page 107: SpeedClave Steam Sterilizers M7 M7

Always Specify Model & Serial Number

Models:Serial Numbers:

© Midmark Corporation 2004 SF-1854

R

E-17.1Labels & Decals

Refer To: Page

Operation & Troubleshooting ....... A-1

Component Testing / Repair ........ B-1

Access Procedures ...................... C-1

Wiring Diagrams .......................... D-1

Exploded Views / Part Numbers .. E-1

MA674302i

061-0752-00

061-0901-00

061-0903-00H335669

MPM60455

Item Description Qty.

1 Operating Instructions Plate ........................ 12 Wiring Diagram Label .................................. 13 Warning Label .............................................. 14 Caution Label ............................................... 15 Serial Number Label (large- n/a) ................. 16 Caution HOT Label ...................................... 17 Serial Number Label (small - n/a) ................ 1

M7 (-020 thru -022)all

Page 108: SpeedClave Steam Sterilizers M7 M7
Page 109: SpeedClave Steam Sterilizers M7 M7

COMMENTS

The Technical Publications Department of Midmark Corporation takes pride in its manuals. We are sure that our manuals will fill all your needs when you are perform-ing scheduled maintenance, servicing, or repairs on a Midmark product. However, if you find any errors or feel there should be a change, addition, or deletion to themanuals, please let us know. We will correct any errors that we are made aware of and we will review requests for changes, additions, or deletions to the manuals andincorporate those requests deemed appropriate. If you see something in one of our manuals that you like or dislike, please let us know. Also, if there is something youfeel we could do to produce a better manual, please let us know.

If an error is found, please list the page and paragraph/figure in which the error was found along with a brief description of what the error is. If the correction to the erroris known, please include that information also. If a change, addition, or deletion is being requested, please list the page and paragraph/figure needing the change, alongwith a brief description of how you feel the paragraph/figure should be changed.

Please fax or mail a copy of this completed comment sheet to:

Page Number Paragraph/Figure Description

Midmark CorporationATTN: Technical Publications Dept.60 Vista DriveVersailles, Ohio 45380Fax: (937) 526-5542

Page 110: SpeedClave Steam Sterilizers M7 M7

SERVICE PARTS FAX ORDERING FORM(Do not tear out this page. Photo copy this page for use only.)

IMPORTANT NOTES:

1) Use this form for all non-warrantyorders only. Warranty orders mustbe telephoned in (1-800-643-6275).

2) FAX number to send order to:877-249-1793

3) All emergency orders must bereceived @ Midmark by 1:00 pm EST.

4) All underlined headings should befilled in prior to submittal.

AT

TE

NT

ION

: C

US

TOM

ER

SE

RV

ICE

DE

PAR

TM

EN

T

AD

DIT

ION

AL

CO

MM

EN

TS

:

NAME: SHIP TO:

ADDRESS:

CITY: STATE: ZIP:

CONTACT:

PHONE:

FAX #:

MODEL #: SERIAL #: SALES ORDER # (if applicable)

LINE # PART NUMBER QTY. DESCRIPTION COLOR(if applicable)

CREDIT CARD INFORMATION

CARD TYPE __________________________ _ CARD #_________-_________-_________-_________ EXP. DATE ______ / ______ / ________

NAME ON CARD ___________________________________________ SIGNATURE_______________________________________________________

DATE: / / TIME:

METHOD OF SHIPMENT:

PRIORITY: NON-EMERGENCY ORDER to shipwithin 72 hours if part(s) are in stock.

EMERGENCY ORDER to ship within24 hours if part(s) are in stock.[see note 3]

DEALER P.O. #:

ACCOUNT #:

ampm

Page 111: SpeedClave Steam Sterilizers M7 M7
Page 112: SpeedClave Steam Sterilizers M7 M7

Midmark Corporation60 Vista DriveP.O. Box 286Versailles, OH 45380-0286Phone: 937-526-3662Fax: 937-526-5542

www.midmark.com

Subject to change without notice.Refer to www.Documark.com for latest revision.