Special Timing Functions
• PMT
• PVC
• Crosstalk
• Auto Ventricular Blanking
for internal use only
Pacemaker - Mediated Tachycardia• Initiation• Prevention• Detection• Therapy
for internal use only
PMT – Endless Loop Tachycardia
for internal use only
PMT – Endless Loop Tachycardia
for internal use only
PMT – Endless Loop Tachycardia
for internal use only
PMT - Conditions to be fulfilled
• Initiator
•Retrograde (VA) conduction
•VDD or DDD Pacemaker
for internal use only
PMT – Cycle Length
PMT cycle time=
VA conduction time +
programmed AV delay
(or Maximum Tracking Rate interval)
for internal use only
PMT – Endless Loop Tachycardia
• Initiation• PVC• Atrial undersensing• Atrial oversensing• Atrial noncapture• Long AV delay
for internal use only
PMT – Initiation by PVC
PVCSurface ECG
Esophageal ECG
Retrograde P-waves
ASAP
VSVP
AP
VP
AS AS AS AS AS AS
VP VP VP VP VP VP VP
for internal use only
PMT – Initiation by PVC
for internal use only
PMT – Initiation
Atrial noncapture
causes ventricular stimulation
without the AV node being blocked
by anterograde event
Atrial oversense
Inhibits atrial output: causes ventricular stimulation without the AV node being
blocked by anterograde event
for internal use only
PMT – Initiation by loss of atrial capture
Surface ECG
Retrograde P-waves
Esophageal ECG
AP
VP
AP
VP
AS AS AS AS AS AS
VP VP VP VP VP VP
AP
VP
AP
VP
for internal use only
PMT – Initiation
Atrial undersense
Results in a long AV interval, allowing the AV node time to recover and conduct
retrograde
Long AVI
Allows the AV node time to recover
for internal use only
PMT – Initiation by loss of atrial sensing
Surface ECG
Esophageal ECG
Retrograde P-waves
AP AS AS AS AS AS AS
VP VP VP VP VP VP VP
AS
VP
AS
VP
AS
VP
for internal use only
PMT – Endless Loop Tachycardia
•Prevention (not SJM specific)
• PVARP• PVARP extension after PVC• Differential atrial sensing; anterograde /
retrograde p-wave morphology comparison• Rate adaptive PVARP• High max tracking rate – retrograde fatigue
for internal use only
PMT – Endless Loop Tachycardia
•Detection• P tracking at upper rate limit• P tracking at intermediate rate• AV interval modulation• Discrepancy between sensed atrial rate and
sensor indicated rate
for internal use only
PMT Options SJM
•Off (nominal for Identity and Integrity)
•10 Beats > PMT
•Auto detect (nominal for Zephyr and Victory)
• No PMT options during AMS or Magnet Application
for internal use only
PMT Options SJM - 10 Beats > PMT
Detection•10 consecutive AS-VP events at a rate > PMT Detection Rate
•PMT Detection Rate: 90 – 150 bpm, nominal 110
Response•Reset of PVARP to 480 ms for a single cycle
After PVARP extension•PMT detection is suspended for 256 cycles
•the atrial alert period is 330 ms
for internal use only
10 Beats > PMT
PVARP: 480ms
Atrial Alert Period 330ms
for internal use only
PMT Options SJM – Auto Detect
Detection• Rapid stable VP – AS interval
› Rapid: atrial rate > PMT Detection Rate› Stable: compares average of 8 VP – AS intervals to average
of all cycles. If all VP - AS intervals are with 16 ms of average stable
• On the 9th cycle the AS – VP delay is› Shortened 50 ms if AS – VP 100 ms› Lengthened 50 ms if AS – VP < 100 ms› If VP – AS cycle is within 16 ms of average VP – AS cycle
PMT. Else Sinus Tachycardia
for internal use only
PMT Options SJM – Auto Detect
(VP-AS)1 (VP-AS)2
therefore:
sinus tachycardia
(VP-AS)2(VP-AS)1
(VP-AS)1 (VP-AS)2
(VP-AS)1 = (VP-AS)2
therefore:
PMT
AS - VP AS - VP(AS-VP) -
AS - VP AS - VP(AS-VP) -
for internal use only
PMT Options SJM – Auto Detect
for internal use only
PMT – Endless Loop Tachycardia
•Termination• PVARP extension• Withhold of ventricular output• Withhold of ventricular output with early atrial
pace
for internal use only
PMT Options SJM – Auto Detect and Response
Response• After 9th interval no ventricular response
• Atrial pulse 330 ms after retrograde P-wave
for internal use only
PMT Option Auto Detect
AVI: +50ms
for internal use only
Special Timing Functions
• PMT
• PVC
• Crosstalk
• Auto Ventricular Blanking
for internal use only
PVC Options
• To distinguish between a premature ventricular
contraction (PVC) and a conducted R-wave in
DDD(R) or VDD(R) mode
• A response to a PVC to prevent the evolution into a
PMT
for internal use only
PVC Detection
A PVC is confirmed if
• An R-wave is not preceded by an atrial event
Or
• A P-wave is detected in the relative portion of the
PVARP but is not followed by an R-wave within 280ms
of the atrial event
for internal use only
PVC Options: + PVARP on PVC
After PVC Detection• PVARP extension to 480 ms
• Followed by atrial alert period of 330 ms
for internal use only
PVC Options: Atrial Pace on PVC
After PVC Detection• PVARP extension to 480 ms• If retrograde P-wave is detected PVARP is terminated• No ventricular output• Atrial Escape Interval 330 ms
for internal use only
Atrial Pace on PVC
Extension of PVARP to 480
Atrial Alert Period of 330ms
330ms
DDD 60
for internal use only
Special Timing Functions
• PMT
• PVC
• Crosstalk
• Auto Ventricular Blanking
for internal use only
What’s up Doc?
for internal use only
Crosstalk
• Sensing of an atrial output in the ventricular channel resulting in
inhibition of the ventricular pulse
• Clinical manifestation: atrial pacing with no ventricular output.
Can be overcome with Ventricular Safety Standby
• Crosstalk can occurs with
• High atrial pulse amplitude
• High ventricular sensitivity
• Short ventricular blanking
• Rapid pacing rates
• Unipolar pacing and sensing
for internal use only
Crosstalk
Decrease atrial output?
Decrease ventricular sensitivity?
Prolong ventricular blanking?
Enable Ventricular Safety Standby!
for internal use only
Ventricular Safety StandbyB
lan
kin
g
Per
iod
Cro
ssta
lk
Det
ecti
on
W
ind
ow
64ms
for internal use only
Ventricular Safety Standby
Crosstalk
Ventricular Safety Pacing
after 120 ms
V EGM
sensitivity
for internal use only
Ventricular Safety Standby
for internal use only
Is Ventricular Safety Standby Pacing Present In This Panel?
for internal use only
Ventricular Safety Standby
Sensing in
Crosstalk Detection Window
Ventricular Safety Pacing
after 120 ms
for internal use only
Ventricular Safety Standby
Blanking Period - No sensing
Ventricular Pacing after
programmed AV Interval
for internal use only
Blanking period undersensing can be dangerous at long AV intervals
for internal use only
Crosstalk – Ventricular Safety Standby
for internal use only
Special Timing Functions
• PMT
• PVC
• Crosstalk
• Auto Ventricular Blanking
for internal use only
Auto Ventricular Blanking
Maximizes the ventricular alert window.
Minimizes inappropriate inhibition due to crosstalk.
• When crosstalk is present, ventricular blanking period automatically lengthens
• In the absence of crosstalk, ventricular blanking period remains short
for internal use only
Auto Ventricular Blanking - Zephyr
Auto Ventricular Blanking - Programmability
for internal use only
Auto Ventricular Blanking - Zephyr
Programmable values
Shipped setting is Auto
for internal use only
Auto Ventricular Blanking - Zephyr
• AP - a 12 ms absolute blanking window begins.
• If ventricular activity (crosstalk) is sensed, a retriggerable 12 ms relative refractory window begins and can retrigger to a maximum of 52 ms.
• If no ventricular activity is sensed the ventricular channel enters an alert period.
time
AP
12 ms
Absolute v. Blanking
52 ms
Maximum ventricular blanking interval
64 ms
for internal use only
Auto Ventricular Blanking - Zephyr
• Crosstalk ends before 52 ms.• No need for crosstalk detection window.• Ventricular channel enters an alert period.
R R
time
AP
Ventricular Crosstalk
No Crosstalk Detection Window
170 ms VP
40 ms
64 ms
Alert12 ms
52 ms
No Sensed Ventricular Event
for internal use only
Auto Ventricular Blanking - Zephyr
• Crosstalk reaches the maximum of 52 ms
• Ventricular Safety Standby will initiate a crosstalk detection
window from 52 ms - 64 ms
• Normal Ventricular Safety Standby behavior will apply
Ventricular Crosstalk
time
AP
Crosstalk Detection Window
VP120 ms VSS
12 ms 40 ms Alert12 ms
64 ms
R R R R
52 ms
for internal use only
Auto Ventricular Blanking - Zephyr
• If there is sensing of constant ventricular events and the maximum
of 52 ms is reached
• VSS will initiate a crosstalk detection window from 52 ms - 64 ms
• Normal Ventricular Safety Standby behavior will apply
Ventricular Crosstalk
time
APCrosstalk
Detection Window
VP120 ms VSS
12 ms 40 ms Alert12 ms
64 ms52 ms
for internal use only
Ventricular Blanking
Programming EGM trigger
High Ventricular Rate V Blanking to 12ms
for internal use only
Auto Ventricular Blanking
Auto Ventricular Blanking OFF
•Ventricular Blanking will auto program to 40 ms with the following:
• Atrial output 6 Volts and higher
• Atrial pulse width 1.0 ms or longer
for internal use only