Silicon strip detectors for two- Silicon strip detectors for two- dimensional dimensional soft x-ray imaging at normal soft x-ray imaging at normal incidence incidence Pedro Rato Mendes Pedro Rato Mendes LIP Lisbon and University of Algarve, Portugal LIP Lisbon and University of Algarve, Portugal Amsterdam, 09-12 September 2002
14
Embed
Silicon strip detectors for two-dimensional soft x-ray imaging at normal incidence Pedro Rato Mendes LIP Lisbon and University of Algarve, Portugal Amsterdam,
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Silicon strip detectors for two-Silicon strip detectors for two-
dimensional dimensional
soft x-ray imaging at normal soft x-ray imaging at normal
incidenceincidence
Pedro Rato MendesPedro Rato MendesLIP Lisbon and University of Algarve, PortugalLIP Lisbon and University of Algarve, Portugal
Amsterdam, 09-12 September 2002
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Outline
• Collaboration members
• Objectives
• Experimental setup
• Energy calibration
• Imaging results
• Conclusions and future work
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Collaboration members
M. C. Abreu a, G. Baldazzi b, D. Bollini b, E. Cabal Rodriguez c, W. Dabrowski d, A. Diaz Garcia c, M. Gambaccini e, P. Giubellino f,
M. Gombia b, P. Grybos d, M. Idzik d,f, A. Marzari-Chiesa g, L. M. Montano h, F. Prino i, L. Ramello i, P. Rato Mendes a, S. Rodrigues a, M. Sitta i, P. Sousa a,
K. Swientek d, A. Taibi e, A. Tuffanelli e, R. Wheadon f, P. Wiacek d
a LIP and University of Algarve, Faro, Portugal
b Dipartimento di Fisica dell’Università di Bologna and INFN, Bologna, Italy
c CEADEN, Havana, Cuba
d Faculty of Physics and Nuclear Techniques, Academy of Mining and Metallurgy, Cracow, Poland
e Dipartimento di Fisica dell’Università di Ferrara and INFN, Ferrara, Italy
f INFN , Torino, Italy
g Dipartimento di Fisica Sperimentale dell’Università di Torino and INFN, Torino, Italy h CINVESTAV, Mexico City, Mexico
i Dipartimento di Scienze e Tecnologie Avanzate dell’Università del Piemonte Orientale and INFN , Alessandria, Italy
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Objectives
Develop a simple prototype system for 2-D
soft x-ray imaging using silicon strip detectors
• Motivations for this prototype– spin-off from high-energy physics– RX64 chip designed for this application– simple laboratory setup and data-acquisition– advanced education for young physicists
RX64 and its main applications: talk by L. Ramello, this conference
W. Dabrowski et al., RX64 – A Fully Integrated 64-channel IC for Readout of Silicon Detectors Used for X-Ray Position Sensitive Measurements, IEEE NSS/MIC 2001, San Diego
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Experimental setup
• Equipment– board with 1-sided microstrip detectors and RX64 chips
• 300 m thick Si, 132 strips at 100 m pitch (ITC-IRST, Trento, Italy)
• RX64 chip by W. Dabrowski and collaborators (Cracow, Poland)
• one or two chips per board (64 or 128 channels)
– PC with NI PCI-DIO-96 interface + LabVIEW 6i software– scanning system controlled by stepper motor– radiation sources (57Co, 109Cd, 241Am for fluorescence)– power supplies, cables
• Two-dimensional image acquisition– acquire 1-D profiles for each collimator position– form 2-D image by juxtaposition of 1-D profiles
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Experimental setup
64
silicon microstrip detector
RX64
RX64
- 64 channels (preamp, shaper, discriminator)
- 64 pseudo-random counters (20-bit)
- internal calibration circuit (1 mV to 30 mV)
- I/O on chip (interface with external bus)
PC
NationalInstruments
PCI-DIO-96LabVIEW 6i
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
First test board (one RX64 chip)
0.6 1.0 cm2
detector
RX64
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Energy calibration
• Method– perform discriminator threshold scan– differentiate integral spectrum– fit energy peaks on differential spectrum
integral spectrum differential spectrum
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Energy calibration
Calibration using Co57 and Cd109 sources (board with one RX64 chip, average on all channels)
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Energy calibration
Calibration using Am241 x-ray fluorescence source (board with two RX64 chips, average on all channels)
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Scanning system
scanning
detector board
stepper motor
radioactive source
Pb collimator
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Imaging results – one RX64 chip
Pb0.5 mm
thick
300 kBq 109Cd source – 22 keV Ag K x-rays15 mm thick Pb collimator, 300 µm wide slit
microstripdetector
scanningsystem
1.0 1.0 mm2
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Imaging results – two RX64 chips
300 kBq 109Cd source – 22 keV Ag K x-rays15 mm thick Pb collimator, 300 µm wide slit
LOW CONTRASTsnail shell
HIGH CONTRAST8-pin integrated circuit
1.0 1.0 mm2
Pedro Rato Mendes et al. IWoRID’2002 - Amsterdam
Conclusions
• Prototype main features– one-sided silicon microstrip detectors– simple readout electronics based on RX64 chip– simple 2-D scanning system
• Future developments– larger area sensors / more readout chips– new chip version with two thresholds– x-ray tube as radiation source
We have built and tested a simple prototype for 2-D soft x-ray imaging