Shielding for Diagnostic X-rays: UK Guidance Jerry Williams Royal Infirmary of Edinburgh Royal Infirmary of Edinburgh
40

Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Apr 22, 2018

Download

Documents

lambao
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Shielding for Diagnostic X-rays:UK Guidance

Jerry WilliamsRoyal Infirmary of EdinburghRoyal Infirmary of Edinburgh

Page 2: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

BIR/ IPEM Working Party ReportBIR/ IPEM Working Party Report

British Institute of Radiology

Institute of Physics and Engineering in Medicine

Working Party:

David Sutton Jerry Williams Colin Martin Don McIntosh Tony

Institute of Physics and Engineering in Medicine

David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher

Publication date: 2000

Page 3: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

BIR/ IPEM Working Party Report

Content:

BIR/ IPEM Working Party Report

• Design criteria and dose constraints

• Primary & Secondary radiations• Primary & Secondary radiations

• Building materials

• X-ray transmission factorsay t a s ss o acto s

• Assessment of shielding

• Worked examples

Working Party:

David Sutton Jerry Williams Colin Martin Don McIntosh Tony David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher

Publication date: 2000

Page 4: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

dRadiation Sources

• Radiography (film/ screen)• Radiography + Fluoroscopy• Radiography + Fluoroscopy• Angiography

CT ( i l li )• CT (single slice)• Mammography• Dental• DEXA• DEXA

Page 5: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

h lkThis talk

• Design criteria• Transmission factors• Transmission factors• Shielding materials

P i d S d R di ti• Primary and Secondary Radiations

Page 6: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Design Criteria: Design Criteria: Dosimetric Quantities

• Dose limits–Effective dose, E mSvEffective dose, E mSv

• Dose monitoringOperational quantity–Operational quantity

–Personal dose equivalent, Hp(d) mSv

Shi ldi l l ti• Shielding calculations–Air kerma, K mGy

Page 7: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Design Criteria

• Dose constraint (people)– Based on public dose limit (1 mSv/ year)

A li d ( ff bli )– Applied to everyone (staff + public)– 0.3 mGy/ year (6 µGy/ week)

• Occupancy factor • Occupancy factor – ≥5%– Dose to be < 6 mGy/ year– (Controlled area)

• Dose constraint (film or CR)/ d ( / k)– 0.4 µGy/ day (2 µGy/ week)

Page 8: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

fOccupancy factors

• Based on individual occupancy• Examples:p

Office 100%Reception area 100%Reception area 100%X-ray control room 100%Patient examination room 50%Corridor 20%Corridor 20%Toilets/ bathrooms 10%Outdoor area with seating 10%Store rooms 5%Store rooms 5%Unattended waiting rooms 5%

Page 9: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

fX-ray Transmission factors

• Adopted Archer’s empirical model1−

⎤⎡ ⎞⎛ γγα

αβ

αβ1 ⋅⋅

⎥⎦⎤

⎢⎣⎡ −⎟

⎠⎞⎜

⎝⎛ += teB

• Used Simpkin’s α, β, γ values• Added data for brick and barium Added data for brick and barium

plaster

Archer et al (1983) Health Physics, 44, 507.

Simpkin (1995) Health Physics, 68, 704

Page 10: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

h ld lShielding material

UK Standard lead thicknesses

Code t mm3 1.324 1.805 2.246 2.65

Page 11: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Primary and Secondary Primary and Secondary Radiations

• Leakage radiationg• Primary/ transmitted beams• ScatterScatter

Page 12: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

X-ray Tube LeakageX ray Tube Leakage

Assumptions:•Tube leakage factor: 150 kV/ 3.3 mA

•Field size: 1000 cm2

•Distance: 1 m from tube/ patient

WP Recommendation:WP Recommendation:

•Leakage contribution may be ignored

Page 13: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Primary & Transmitted Primary & Transmitted Radiation

• Fluoroscopy– Beam restricted to

image intensifier/ flat X-raybeam

image intensifier/ flat panel detector

– Detector has > 2mm lead equivalent

Patient

lead equivalent

• Mammography

Imagingdevice

Page 14: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Radiography: Film dose method

• Dose to film/ imaging plate– 400 speed system– Dose ≤10 µGy

Dose ≤10 µGy

• Lead equivalence (100 kV)– Cassette

• 0 19 mm• 0.19 mm– Cassette + grid

• 0.26 mm– Cassette/grid/tableCassette – Cassette/grid/table

• 0.8 mm– Cassette/grid/chest Bucky

• 0 7 mm

Bucky system

• 0.7 mm

Lead equivalence data derived from Dixon (1994) Med Phys, 21, 1785

Page 15: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Radiography: Film dose method

• Dose to film/ imaging plate

– 400 speed system– Dose ≤10 µGyDose ≤10 µGy

• Lead equivalence (100 kV)– Cassette

• 0 19 mm• 0.19 mm– Cassette + grid

• 0.26 mm– Cassette/grid/table– Cassette/grid/table

• 0.8 mm– Cassette/grid/chest Bucky

• 0 7 mm• 0.7 mm

Page 16: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Attenuated primary: limiting HVLAttenuated primary: limiting HVL

Limiting HVL: = ln(2) / α1

Limiting HVL: ln(2) / α

0.1 to eDD ⋅α−=

0 2

0.3

VL m

m

0.01

Tran

smis

sion

0.1

0.2

Lim

iting

H

0.001

T

040 60 80 100 120

kV0.0001

0 0.5 1 1.5 2 2.5 3

Lead thickness mm

Page 17: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Radiography: ESD methodRadiography: ESD method

• Unattenuated primary beam

p y– Beam outside patient

• Entrance Surface Dose (ESD)• Situations include

– Beam not collimated to cassette– Beam not directed to Bucky

• Examples– Chest radiography– Cross table radiography– Out of Bucky radiography (e.g.

extremities)

extremities)• Area of wall irradiated

– Not easy to predict– Variable– Variable

Page 18: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

lExamples

Page 19: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Chest Radiography (film dose)Chest Radiography (film dose)

• ParametersParameters– 100 films/ week– 90 kV– ESD = 0.15 mGy– Film dose = 10 µGy– FFD = 1.8 m– FSD = 1.4 m

Fil ll 0 7 – Film-wall = 0.7 m

Best Radiographic Practice

Page 20: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Calculation (Film-dose method)

• Attenuated kerma at wall per week:Film dose x Workload x ISL correction

mGy52.07.08.1

8.110010102

3 =⎥⎦⎤

⎢⎣⎡

+××× -

⎦⎣

• No of HVLs required (6 µGy/ week) = 6.5 • Limiting HVL of lead @ 90 kV = 0.23 mmLimiting HVL of lead @ 90 kV 0.23 mm• Total equivalent lead thickness = 1.5 mm• Additional lead shielding =

0 8 0.8 mm

Page 21: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Chest Radiography (ESD)Chest Radiography (ESD)

• ParametersParameters– 100 films/ week– 90 kV– ESD = 0.15 mGy– Film dose = 10 µGy– FFD = 1.8 m– FSD = 1.4 m

Fil ll 0 7 – Film-wall = 0.7 m

Page 22: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Calculation (ESD method)( )• Worst case assumption:

• Beam size greater than patient and Bucky system

• Unattenuated kerma at wall per week:ESD x Workload x ISL correction

mGy7.47.08.1

4.110015.02

=⎥⎦⎤

⎢⎣⎡

+××

• Maximum transmission permitted= 6x10-3/ 4 7= 6x10 3/ 4.7= 0.0013

• Lead shielding @ 90 kV1.4 mm

Page 23: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Summary

• Best practice –Beam collimated to patient/ image Beam collimated to patient/ image

plate/ Bucky–0.8 mm lead

• Poor practice–Beam outside patient and BuckyBeam outside patient and Bucky–1.4 mm lead

Page 24: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Scatter

Page 25: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Scatter ModelNCRP 49 method

400FKaK us ⋅⋅=

Ks – scatter kerma; Ku – primary kermaF– Field size (cm2); a – scatter factorTrout and Kelly (1972), Radiology, 104, 161

Ku . F ⇒ Area-air kerma product (AKP) orArea-air kerma product (AKP) orDose area product (DAP)

DAPSK DAPSKs ⋅=

Page 26: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

hWhy DAP?

• No assumptions on field size• Availability of DAP data• Availability of DAP data

–Requirement in UK legislation to record patient doserecord patient dose

–DAP preferred patient dose metric (fluoroscopy and radiography)(fluoroscopy and radiography)

–National surveys of patient dose

Page 27: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Scatter factor normalised to DAP (S)Scatter factor normalised to DAP (S)12

m2 )-1

125 kVp

8

10µG

y.(G

y.cm 100 kVp

85 kVp 70 kVp 50 kVp

4

6

r fac

tor,

S µ

0

2

0 30 60 90 120 150 180

Scat

ter

0 30 60 90 120 150 180

Angle of scatter

( ) ( )[ ]1f85kVdbS 234 θθθθ

JR Williams (1996) Br J Radiol, 69, 1032

( ) ( )[ ]1f85kVedcbaS 234 +−×+θ+θ+θ+θ=

Page 28: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Scatter factor normalised to DAP (S)

12

-1

Scatter factor normalised to DAP (S)

8

10

Gy.

(Gy.

cm2 )-

125 kVp100 kVp 85 kVp 70 kVp 50 kVp

4

6

fact

or, S

µG

a -1.04E-07b 3 27E 05

0

2

0 30 60 90 120 150 180

Scat

ter f b 3.27E-05

c -2.75E-03d 8.37E-02e 1 58E+000 30 60 90 120 150 180

Angle of scatter

( ) ( )[ ]1f85kVdbS 234 θθθθ

e 1.58E+00f 5.99E-03

( ) ( )[ ]1f85kVedcbaS 234 +−×+θ+θ+θ+θ=

JR Williams (1996) Br J Radiol, 69, 1032

Page 29: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Calculation vs measurement

Cli i l li tiMeasured/ Calc doseNo of DAP/ week Ave Calc Mean Min Max

Interventional Radiology (liver disease) 8 520 0.71 0.22 0.02 0.42Abdominal + lower limb angiograpy 8 1120 2.15 0.17 0.02 0.43

Clinical application positions Gy cm2 Dose mGy

Cardiac angiography 7 780 0.81 0.99 0.71 1.26Cerebral angiography 7 460 1.48 0.97 0.67 1.23Ba contrast studies (o/c tube) 9 430 0.66 0.74 0.50 1.13B t t t di ( / t b ) 10 960 1 62 0 28 0 05 0 86Ba contrast studies (u/c tube) 10 960 1.62 0.28 0.05 0.86General Radiography 10 460 0.96 0.64 0.30 1.06

Page 30: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Validation of scatter data by Validation of scatter data by Monte-Carlo

Scatter fraction (S) @ 85 kV12

Measured

6

8

10

Gy

cm2

MeasuredMC

2

4

6

S µG

y/ G

00 30 60 90 120 150 180

Angle

Page 31: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Comparison with NCRP 147

Scatter fraction (S) @ 85 kV12

UK

6

8

10

Gy

cm2 )-1

UK

NCRP 147

MonteCarlo

2

4

6

S µG

y (G

00 30 60 90 120 150 180

Angle

Page 32: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Differences in scatter factors?

• Phantom• X-ray spectrum• X ray spectrum• Measurement methods

Page 33: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

Method in practicep12

m2 )-1

125 kVp

8

10µG

y.(G

y.cm 100 kVp

85 kVp 70 kVp 50 kVp

4

6

r fac

tor,

S µ

0

2

0 30 60 90 120 150 180

Scat

ter

( )

0 30 60 90 120 150 180

Angle of scatter

( ) ( )[ ]1f85kVedcbaS 234 +−×++++= θθθθ

Page 34: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement
Page 35: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement
Page 36: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

h d ( )Method in practice (1)

0 6

0.8

1.0

doseθ d

0 2

0.4

0.6

Rel

ativ

e

r

0.0

0.2

0 30 60 90 120 150 180

S tt i l

( )[ ] ( ) 12max cmGyµGy5.2kV031.0S

−+×=

Scattering angle

( )d = 1 m; θ = 117°

Page 37: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement
Page 38: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

θ d = 1 m

∫∫ ⋅= θ

θθd

d)(SS15030

ave ∫ θd

Page 39: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

h d ( )Method in practice (2)

• Save over C-arm rotation – (30 to 150°)(30 to 150 )

• Smax for parallel beam

kV Smax Save

50 4.05 3.73

70 4.67 4.30

85 5.14 4.72

100 5 60 5 15100 5.60 5.15

125 6.38 5.86

Page 40: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement

b dTo be continued…..

• Application examples–FluoroscopyFluoroscopy–Radiography

• Comparisons with NCRP 147• Comparisons with NCRP 147