Setting Up the Orion 3 Star pH Meter with Printer 1 Setting Up the Orion 3 Star pH Meter with Printer
Jun 24, 2015
Setting Up the Orion 3 Star pH Meter with Printer 1
Setting Up the Orion 3 Star pH Meter with Printer
Setting Up the Orion 3 Star pH Meter with Printer 2
Orion 3 Star pH Meter
Front of Meter & Keypad
Setting Up the Orion 3 Star pH Meter with Printer 3
pH Meter Self-Test
Meter Connections for Self Test
Press POWER and wait until Software Version is
displayed
Initiating Self Test
Press MEASURECoveredBNC (electrode) Input
Power Cord Input
Setting Up the Orion 3 Star pH Meter with Printer 4
pH Meter Self Test 2Verify all
SEGMENTS are lit
Press MEASURE
Verify all SEGMENTS are not
lit
Press MEASURE
Setting Up the Orion 3 Star pH Meter with Printer 5
pH Meter Self Test 3
Press each button on keypad in any order
Display will read KEY
Key error if 5 second delay in pressing next button
during self-test
Press MEASURE to clear Err 033 & complete
self-test
Setting Up the Orion 3 Star pH Meter with Printer 6
Making Connections
BNC Connection for Electrode
Output to Printer
Meter Connections to Peripherals
Paper Installed
Black Cartridge Only Installed
Printer
Setting Up the Orion 3 Star pH Meter with Printer 7
Printer Self Test
While holding down FEED button, turn printer ON and keep holding FEED until printing starts
Printer Settings printed
Press FEED
Rolling ASCII Pattern printed
Remaining Rolling ASCII Pattern and Nozzle
Pattern printed
Press FEED
NV Bit Images printedDEMO MART Grocery Receipt may print here
Press FEED
Setting Up the Orion 3 Star pH Meter with Printer 8
General Configuration 19.4.ref009
Press SETUP
Current Method Displayed
Setting Up the Orion 3 Star pH Meter with Printer 9
General Configuration 19.4.ref009
Use UP/DOWN arrows to select SETUP Options
Setting Up the Orion 3 Star pH Meter with Printer 10
General Configuration 19.4.ref009
Use LINE key to move between
lines
Setting Up the Orion 3 Star pH Meter with Printer 11
General Configuration 19.4.ref009
Use POSITION key to move to the next DIGIT
Use UP/DOWN arrows to change
DIGIT
Setting Up the Orion 3 Star pH Meter with Printer 12
Global Configuration 1
GLP Methods:
ON allows storage of methods
Setting Up the Orion 3 Star pH Meter with Printer 13
Global Configuration 1
Current Date/Time: Set 2 DIGITS to
current hour in 24 hour format
Setting Up the Orion 3 Star pH Meter with Printer 14
Global Configuration 1
Current Date/Time:
Set 2 DIGITS to current minutes
Setting Up the Orion 3 Star pH Meter with Printer 15
Global Configuration 1
Current Date/Time:
Set TYPE to MDY (Month Day Year)
format
Setting Up the Orion 3 Star pH Meter with Printer 16
Global Configuration 2
Current Date/Time:
Set 4 DIGITS to current year
Setting Up the Orion 3 Star pH Meter with Printer 17
Global Configuration 2
Current Date/Time:
Set 2 DIGITS to current month
Setting Up the Orion 3 Star pH Meter with Printer 18
Global Configuration 2
Current Date/Time:
Set 2 DIGITS to current day of month
Setting Up the Orion 3 Star pH Meter with Printer 19
Global Configuration 2
RS232 Interface: Select baud rate
9600
Setting Up the Orion 3 Star pH Meter with Printer 20
Global Configuration 3
• Global Configurations only need to be set once to be the same for all methods
• Method Specific settings must be selected for each method Only the Method Number and pH Resolution differ for the two
ARCBS standard methods
RS232 Interface: Select output format
for PRINTER
Troubleshooting Tip:
If printout is not formatted like
examples, check if setting is for computer
(comma delimited) output
Setting Up the Orion 3 Star pH Meter with Printer 21
Method Specific ConfigurationMethod 1 pH:
Select resolution 0.1 (1 decimal)
Setting Up the Orion 3 Star pH Meter with Printer 22
Method Specific Configuration
All pH:Select USA Buffer
Set
Setting Up the Orion 3 Star pH Meter with Printer 23
Method Specific ConfigurationAll Calibration Alarm
Set:Set Digits to 0024
(recalibration required if > 24 hours)
Setting Up the Orion 3 Star pH Meter with Printer 24
Method Specific Configuration
All Read Measurement:
Select CONTINUOUS (user must press
MEASURE to print result)
Setting Up the Orion 3 Star pH Meter with Printer 25
Method Specific Configuration 2
All Data Archive & Retrieval:
Select NO to roll over log when memory is full
All Data Archive & Retrieval:
Select OFF to not automatically save results
to log
All Temperature Compensation (C):Set 3 DIGITS to 22.0 (average temperature
for processing platelets)
All Stirrer Speed:Select OFF (only used for special magnetic
stirrer interfaced with the pH meter)
Setting Up the Orion 3 Star pH Meter with Printer 26
Method Specific Configuration 3
• Use same password for all methods
• Default (0000) allows everyone access to settings possible inadvertent changes
to methods
All Password Protect:
Set 4 digits to lock methods and settings
All Autosampler:Select OFF (only used for automatic sampler interfaced with the pH
meter)
All Automatic Shutoff:
Select ON to have meter shut off after
20 minutes inactivity
Setting Up the Orion 3 Star pH Meter with Printer 27
Method Specific Configuration 4
•Press MEASURE to save method
•Press SETUP to enter another method, if needed
Method 1:Set 2 digits to 01
Method 2:Select resolution 0.01
(2 decimals)Method 2:
Set 2 digits to 02
Match other settings to table as for Method 1