When zero clear was performed, a charge exceeding ±5%F.S of the default zero point was present. : After approx. 1 s. error display, the sensor returns to measurement mode. This time will depend on the deviation of the product itself and ambient environment during zero clear operation. Return to a condition without charge, and retry zero clear operation. Insertion of batteries Slide the cover of the case open to insert the batteries. Use 2 x AA (LR6) alkaline dry cell batteries. When inserting the batteries, pay attention to the polarities and insert them in the correct direction. After insertion, ensure the cover is closed correctly. Assembly Error Display Error Type Troubleshooting Method The sensor is broken. Stop using immediately and contact SMC. Error in internal data. The charge has exceeded the lower limit of the measurable voltage range, or the measuring distance is not suitable. Turn off the power supply once, and turn it on again. If the error persists contact SMC. Remove the charge until the value is within the measurable voltage range. Also, check the measuring distance. The charge has exceeded the upper limit of the measurable voltage range, or the measuring distance is not suitable. LCD Charge amount Peak/Bottom hold value LIGHT button Back light ON/OFF Battery saving: "L" Peak hold: "P" Bottom hold: "b" Power button Switch on and off the power supply LR6 dry cell battery (2 pcs.) Ground terminal Mounting of the high voltage measuring handle (option) 1. Insert the sensor into the handle in the direction shown. 2. Fix the cable to the handle as shown. 3. The assembly is complete and the handle is now ready for use. Procedure Setting / Adjustment 1. Insert the batteries. 2. Connect the ground wire. Connect the ground wire to a point suitable for connection to an external protective earthing system. The position for attaching the ground wire to the Handheld meter is shown below. Stop using immediately and contact SMC. 3. Press the button. 4. Move the sensor closer to the position 50 mm away from the measured target. Gradually bring the sensor closer to the measured target from a distance, and stop measurement immediately if the displayed value indicates overflow (HHH) or underflow (LLL). (The target has a high charged potential that is highly dangerous. The measured value does not change even if the distance is shortened.) 5. Check the display. Detecting port Distance to a measured target Detecting range Reference of 50 mm Protected earth Ground wire : The detecting range of the sensor is 180 mm (at a detection distance of 50 mm). Handle Names of individual parts Display Function settings Power ON When the is pressed with the power supply off, the power supply will be turned on. All indications are displayed for 1 s. after the power supply is turned on. 1 s. after All indications Instantaneous value indicatio Detecting port Sensor label Caution label for sensor Sensor Power OFF When the is pressed for 3 s. or more with the power supply on, the power supply will be turned off. Also, no button operation for a certain period will turn off the power supply. (For details, refer to Auto power-OFF function.) Turn off the power supply after using the product to keep the life of batteries as long as possible. Auto power-OFF If no button is pressed for 5 min. or more with the power supply on, the power supply will turn off automatically. 3 s. after 6 s. All indications Instantaneous value indication Instantaneous value indication Flashing Auto power-OFF extension When the is pressed for 6 s. or more with the power supply off, the continuous operating time with no button operation will extend to 15 min. (When this auto power-off extension is activated, indications on the display will keep flashing for 3 s.) Peak / Bottom hold value When the is pressed with the power supply on, the indication will be changed to instantaneous value, peak hold value, bottom hold value and instantaneous value, in that order. : Release the after "P" or "b" is displayed so as not to turn off the power supply. Peak hold value The maximum charged potential and "P" are displayed. The maximum charged potential is continuously detected and updated from when the peak hold is started. If the value over the held maximum charged potential is detected, the display will change. •Bottom hold value The minimum charged potential and "b" are displayed. The minimum charged potential is continuously detected and updated from when the bottom hold is started. If the value under the held minimum charged potential is detected, the display will change. Zero clear A displayed value can be adjusted to zero at measured charged potential in a range of ±5%F.S. of default potential. (There will be a slight displacement, depending on the deviation of the sensor itself and ambient environment of the sensor when the zero clear is performed.) When the and are pressed for 6 s. or more with the power supply on, the displayed value is reset to zero, and then the measuring mode is recovered automatically. Once the power supply is turned off, the offset value for zero clear is cleared. Back Light When the is pressed while the charged potential is displayed, the back light will turn on. Press the button again to turn off the back light. Zero clear error Error Name Sensor failure System error Cable broken Measurement error The cable is broken and the product cannot measure correctly. The charge amount is detected, but not reflected in the displayed value. Battery low When the battery voltage becomes low "L" will be displayed. The display will vary depending on how much battery is left. Display resolution change The display resolution changes depending on the measured charged potential value. L (Flashing) The batteries are low. Prepare to replace with new batteries. Display Content L (On continuously) The batteries are very low. Replace with new batteries immediately. 3 s. kV 5 min. 6 s. Instantaneous value indication kV kV POWER LIGHT POWER LIGHT kV kV POWER LIGHT POWER LIGHT Instantaneous value indication Peak hold value Bottom hold value Handheld Electrostatic Meter Operation Manual IZH10 Thank you for purchasing an SMC IZH10 Series Handheld Electrostatic Meter. Please read this manual carefully before operating the product and make sure you understand its capabilities and limitations. Please keep this manual handy for future reference. To obtain more detailed information about operating this product, please refer to the SMC website (URL http://www.smcworld.com ) or contact SMC directly. Safety Instructions These safety instructions are intended to prevent hazardous situations and/or equipment damage. These instructions indicate the level of potential hazard with the labels of "Caution", "Warning" or "Danger". They are all important notes for safety and must be followed in addition to International standards (ISO/IEC) and other safety regulations. CAUTION indicates a hazard with a low level of risk which, if not avoided, could result in minor or moderate injury. Operator This operation manual is intended for those who have knowledge of machinery using pneumatic equipment, and have sufficient knowledge of assembly, operation and maintenance of such equipment. Only those persons are allowed to perform assembly, operation and maintenance. Read and understand this operation manual carefully before assembling, operating or providing maintenance to the product. Caution: Warning: Danger: WARNING indicates a hazard with a medium level of risk which, if not avoided, could result in death or serious injury. DANGER indicates a hazard with a high level of risk which, if not avoided, will result in death or serious injury. Safety Instructions Do not disassemble, modify (including change of printed circuit board) or repair. An injury or failure can result. Do not operate in an atmosphere containing flammable or explosive gases. Fire or an explosion can result. The product is not designed to be explosion proof. Do not handle with wet hands. This can cause electric shock. Provide grounding to ensure safety and accurate measurement. Improper grounding of the ground wire can cause electrical charge to build up at the sensor and ground terminal of the product and discharge to the user. Do not contact the sensor with the measured target. It can cause failure. Avoid strong impact on the product Do not drop, hit or apply excessive shock to the product while handling. It can cause failure. Do not operate outside of the specification. Fire, malfunction or damage can result. Please use it after confirming the specification. Warning Caution Summary of Product parts Troubleshooting Error Indication This function is to display error location and content when a problem or an error occurs. If the error cannot be reset after the above measures are taken, then please contact SMC. ±1 to ±20 kV Minimum display unit: 0.1 kV 0 to ±0.99 kV Minimum display unit: 0.01 kV <Display example> Akihabara UDX 15F, 4-14-1, Sotokanda, Chiyoda-ku, Tokyo 101-0021, JAPAN Phone: +81 3-5207-8249 Fax: +81 3-5298-5362 URL http://www.smcworld.com Note: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer. © 2011 SMC Corporation All Rights Reserved Specifications Outline with Dimensions (in mm) Refer to the product catalogue or SMC website (URL http://www.smcworld.com ) for more information about the product specifications and outline dimensions. Refer to the SMC website (URL http://www.smcworld.com ) for more information about troubleshooting.
1

Setting / Adjustment Meter Operation Manual

Oct 26, 2021

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Setting / Adjustment Meter Operation Manual

When zero clear was performed, a chargeexceeding ±5%F.S of the default zero point waspresent.∗: After approx. 1 s. error display, the sensorreturns to measurement mode.This time will depend on the deviation of theproduct itself and ambient environment duringzero clear operation.

Return to a condition withoutcharge, and retry zero clearoperation.

Insertion of batteriesSlide the cover of the case open to insert the batteries.Use 2 x AA (LR6) alkaline dry cell batteries.When inserting the batteries, pay attention to the polarities and insert them inthe correct direction.After insertion, ensure the cover is closed correctly.

Assembly

Error Display Error Type Troubleshooting Method

The sensor is broken.Stop using immediately andcontact SMC.

Error in internal data.

The charge has exceeded the lower limit of themeasurable voltage range, or the measuringdistance is not suitable.

Turn off the power supplyonce, and turn it on again. Ifthe error persists contactSMC.

Remove the charge until thevalue is within themeasurable voltage range.Also, check the measuringdistance.

The charge has exceeded the upper limit of themeasurable voltage range, or the measuringdistance is not suitable.

LCD Charge amount Peak/Bottom hold value

LIGHT button Back light ON/OFF

Battery saving: "L"

Peak hold: "P"Bottom hold: "b"

Power buttonSwitch on and offthe power supply

LR6 dry cell battery(2 pcs.)

Ground terminal

Mounting of the high voltage measuring handle (option)

1. Insert the sensor into the handle in thedirection shown.

2. Fix the cable to the handle as shown.

3. The assembly is complete and the handle isnow ready for use.

Procedure

Setting / Adjustment

1. Insert the batteries.

2. Connect the ground wire.Connect the ground wire to a point suitable for connection to an externalprotective earthing system.The position for attaching the ground wire to the Handheld meter is shownbelow.

Stop using immediately andcontact SMC.

3. Press the button.

4. Move the sensor closer to the position 50 mm away from the measured target.Gradually bring the sensor closer to the measured target from a distance, andstop measurement immediately if the displayed value indicates overflow(HHH) or underflow (LLL).(The target has a high charged potential that is highly dangerous. Themeasured value does not change even if the distance is shortened.)

5. Check the display.

Detecting port

Dista

nce

to a

mea

sure

d ta

rget

Detecting range

Reference of 50 mm

Protected earth

Ground wire

∗: The detecting range of the sensor is 180 mm (at a detection distance of 50 mm).

Handle

Names of individual parts•Display

Function settingsPower ONWhen the is pressed with the powersupply off, the power supply will beturned on.All indications are displayed for 1 s.after the power supply is turned on.

1 s. after

All indications Instantaneousvalue indicatio

Detecting port

Sensor label

Caution label for sensor

•Sensor

Power OFFWhen the is pressed for 3 s. or more with thepower supply on, the power supply will be turned off.Also, no button operation for a certain period will turnoff the power supply.(For details, refer to Auto power-OFF function.)Turn off the power supply after using the product to keep the life of batteriesas long as possible.

Auto power-OFFIf no button is pressed for 5 min. or more with thepower supply on, the power supply will turn offautomatically.

3 s. after

6 s.

All indications Instantaneousvalue indication

Instantaneousvalue indication

Flashing

Auto power-OFF extensionWhen the is pressed for 6 s. or more with the power supply off, thecontinuous operating time with no button operation will extend to 15 min.(When this auto power-off extension is activated, indications on the displaywill keep flashing for 3 s.)

Peak / Bottom hold valueWhen the is pressed with thepower supply on, the indicationwill be changed to instantaneousvalue, peak hold value, bottomhold value and instantaneousvalue, in that order.∗: Release the after "P" or "b" is displayed so as not to turn off the power supply.

•Peak hold valueThe maximum charged potential and "P" are displayed.The maximum charged potential is continuously detected and updated fromwhen the peak hold is started. If the value over the held maximum chargedpotential is detected, the display will change.•Bottom hold valueThe minimum charged potential and "b" are displayed.The minimum charged potential is continuously detected and updated fromwhen the bottom hold is started. If the value under the held minimumcharged potential is detected, the display will change.

Zero clearA displayed value can be adjusted to zero atmeasured charged potential in a range of ±5%F.S. ofdefault potential.(There will be a slight displacement, depending onthe deviation of the sensor itself and ambientenvironment of the sensor when the zero clear is performed.)When the and are pressed for 6 s. or more with the power supply on,the displayed value is reset to zero, and then the measuring mode isrecovered automatically.Once the power supply is turned off, the offset value for zero clear is cleared.

Back LightWhen the is pressed while the charged potentialis displayed, the back light will turn on.Press the button again to turn off the back light.

Zero clearerror

Error Name

Sensor failure

System error

Cable broken

Measurementerror

The cable is broken and the product cannotmeasure correctly. The charge amount isdetected, but not reflected in the displayedvalue.

Battery lowWhen the battery voltage becomes low "L" will bedisplayed. The display will vary depending on howmuch battery is left.

Display resolution changeThe display resolution changes depending on the measured charged potentialvalue.

L(Flashing)

The batteries are low. Prepare to replace with new batteries.

Display Content

L(On continuously)

The batteries are very low. Replace with new batteries immediately.

3 s. kV

5 min.

6 s.Instantaneous

value indication

kV kV

POWER LIGHT POWER LIGHT

kVkV

POWER LIGHT POWER LIGHT

Instantaneousvalue indication

Peak hold value Bottom hold value

Handheld ElectrostaticMeter

Operation ManualIZH10Thank you for purchasing an SMC IZH10 Series Handheld Electrostatic Meter.Please read this manual carefully before operating the product and make sure youunderstand its capabilities and limitations.Please keep this manual handy for future reference.

To obtain more detailed information about operating this product, pleaserefer to the SMC website (URL http://www.smcworld.com) or contact SMCdirectly.

Safety InstructionsThese safety instructions are intended to prevent hazardous situations and/orequipment damage.These instructions indicate the level of potential hazard with the labels of"Caution", "Warning" or "Danger". They are all important notes for safety and mustbe followed in addition to International standards (ISO/IEC) and other safetyregulations.

CAUTION indicates a hazard with a low level of riskwhich, if not avoided, could result in minor ormoderate injury.

OperatorThis operation manual is intended for those who have knowledge of machineryusing pneumatic equipment, and have sufficient knowledge of assembly,operation and maintenance of such equipment. Only those persons areallowed to perform assembly, operation and maintenance.Read and understand this operation manual carefully before assembling,operating or providing maintenance to the product.

Caution:

Warning:

Danger:

WARNING indicates a hazard with a medium level ofrisk which, if not avoided, could result in death orserious injury.

DANGER indicates a hazard with a high level of riskwhich, if not avoided, will result in death or seriousinjury.

Safety Instructions

Do not disassemble, modify (including change of printed circuit board) or repair.An injury or failure can result.

Do not operate in an atmosphere containing flammable or explosive gases.Fire or an explosion can result.The product is not designed to be explosion proof.

Do not handle with wet hands.This can cause electric shock.

Provide grounding to ensure safety and accurate measurement.Improper grounding of the ground wire can cause electrical charge to build up at the sensor and ground terminal ofthe product and discharge to the user.

Do not contact the sensor with the measured target.It can cause failure.

Avoid strong impact on the productDo not drop, hit or apply excessive shock to the product while handling. It can cause failure.

Do not operate outside of the specification.Fire, malfunction or damage can result. Please use it after confirming the specification.

Warning

Caution

Summary of Product parts

TroubleshootingError IndicationThis function is to display error location and content when a problem or an error occurs.

If the error cannot be reset after the above measures are taken, then pleasecontact SMC.

±1 to ±20 kV

Minimum display unit:0.1 kV

0 to ±0.99 kV

Minimum display unit:0.01 kV

<Display example>

Akihabara UDX 15F, 4-14-1, Sotokanda, Chiyoda-ku, Tokyo 101-0021, JAPANPhone: +81 3-5207-8249 Fax: +81 3-5298-5362

URL http://www.smcworld.com

Note: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer.© 2011 SMC Corporation All Rights Reserved

SpecificationsOutline with Dimensions (in mm)

Refer to the product catalogue or SMC website (URL http://www.smcworld.com) formore information about the product specifications and outline dimensions.

Refer to the SMC website (URL http://www.smcworld.com) for more informationabout troubleshooting.