Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
Rosemount 2160
www.rosemount.com
• World’s first wireless liquid level switch for reliable point
level detection
• Wireless capabilities extend the full benefits of PlantWeb® to
previously inaccessible locations
• Self-organizing network delivers information rich data with
>99% data reliability
• Designed for operation in temperature extremes of –94 to 500
°F (–70 to 260 °C)
• Virtually unaffected by flow, bubbles, turbulence, foam,
vibration, solids content, coating, properties of the liquid,
andproduct variations
• “Fast Drip” fork design gives quicker response time,
especially with viscous liquids
• Intrinsically Safe certification option
2160***E2160***S
ContentsRosemount 2160 Wireless Vibrating Fork Liquid Level
Switch . . . . . . . . . . . . . . . . . page 2Specification . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . page 5Product Certifications .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . page 7Dimensional Drawings. . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . page 9
Rosemount 2160 WirelessVibrating Fork Liquid Level Switch
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
2
Rosemount 2160 WirelessVibrating Fork Liquid Level Switch
• World's first wireless liquid level switch for reliable point
level detection• Wireless capabilities extend the full benefits of
PlantWeb to previously
inaccessible locations• Enhanced instrument health/self-checking
diagnostics of the fork and sensor• Self-checking condition
monitoring and alerts using Field Communicator or AMS• Suitable for
most liquids
Additional InformationSpares and Accessories: page
4Specifications: page 5Certifications: page 7Dimensions: page 9
2160***E2160***S
2160 Wireless Level Switches
Table 1. 2160 Ordering Information★The Standard offering
represents the most common options. The starred options (★) should
be selected for best delivery.
The Expanded offering is subject to additional delivery lead
time.
Model Product Description2160 Wireless Vibrating Fork Liquid
Level SwitchOutputStandard StandardX Wireless ★Housing
MaterialStandard StandardD Dual Compartment Housing - Aluminum
(Aluminium) ★Conduit Entry / Cable ThreadsStandard Standard8
1/2-in. NPT thread ★Operating TemperatureStandard StandardS
Standard: –40 °F (–40 °C)...302 °F (150 °C) ★E Extreme: –94 °F (–70
°C)...500 °F (260 °C) ★Material of Construction: Process Connection
/ ForkStandard StandardS(1) 316/316L Stainless Steel
(1.4401/1.4404) ★ExpandedH(2) Alloy C (UNS N10002), Alloy C-276
(UNS N10276), solidProcess Connection SizeStandard Standard9 3/4
in. ★1 1 in. / 25 mm (DN25) ★2 2 in. / 50 mm (DN50) ★5 11/2 in. /
40 mm (DN40) ★3 3 in. / 80 mm (DN80) ★4 4 in. / 100 mm (DN100) ★6 6
in. / 150 mm (DN150) ★8 8 in. / 200 mm (DN200) ★7 21/2-in. / 65 mm
(DN65) ★ExpandedX(3) Customer specific
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
3
Rosemount 2160
Process Connection RatingStandard StandardAA ASME B16.5 Class
150 flange ★AB ASME B16.5 Class 300 flange ★DB EN1092-1 PN25/40
flange ★NN For use with non-flange process connection type
★ExpandedAC ASME B16.5 Class 600 flangeDA EN1092-1 PN10/16 flangeDC
EN1092-1 PN63 flangeDD EN1092-1 PN100 flangeXX(3) Customer
specificProcess Connection TypeStandard StandardR Raised Face (RF)
flange ★B BSPT (R) thread ★G BSPP (G) thread ★N NPT thread ★P BSPP
(G) O-ring ★C Tri-Clover Clamp ★ExpandedX(3) Customer specificFork
LengthStandard StandardA Standard length 1.7-in. (44 mm) ★H(4)
Standard length flange 4.0-in. (102 mm) ★E(5) Extended, customer
specified length in tenths of Inches ★M(5) Extended, customer
specified length in millimeters ★Specific Extended Fork
LengthStandard Standard0000 Factory default length (only if fork
length A or H is selected) ★xxxx(5) Specific customer specified
length in tenths of inches, or millimeters (xxxx mm or xxx.x
inches) ★Surface FinishStandard Standard1 Standard surface finish
★2 Hand polished (Ra < 0.4 µm) ★Product CertificationsStandard
StandardNA No Hazardous Locations Certifications ★I1 ATEX Intrinsic
Safety ★I3 NEPSI Intrinsic Safety ★I5 FM Intrinsic Safety ★I6(6)
CSA Intrinsic Safety ★I7 IECEx Intrinsic Safety ★Wireless Update
Rate, Operating Frequency and ProtocolStandard StandardWA3 User
configurable update rate, 2.4 GHz DSSS, Wireless ★Omnidirectional
Wireless Antenna and SmartPowerStandard StandardWK1(7) Long range,
integral antenna, long-life power module adapter, intrinsically
safe (Power module separate) ★
Table 1. 2160 Ordering Information★The Standard offering
represents the most common options. The starred options (★) should
be selected for best delivery.
The Expanded offering is subject to additional delivery lead
time.
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
4
Spare Parts and Accessories
OPTIONSMeterStandard StandardM5 LCD meter ★Factory
ConfigurationStandard StandardC1 Factory configure Date,
Descriptor, Message Fields and Wireless Parameters ★Calibration
Data CertificationStandard StandardQ4 Certificate of functional
test ★Material Traceability CertificationStandard StandardQ8
Material traceability certification per EN 10204 3.1 ★Special
ProceduresStandard StandardP1(8) Hydrostatic testing with
certificate ★Typical Model Number: 2160 X D 8 S S 1 NN N A0000 1 I5
WA3 WK1 M5 Q8
(1) Flanges are dual certified 316 and 316L Stainless Steel
(1.4401 and 1.4404).
(2) Only available for BSPT and NPT threaded process connection
types as standard, other upon request.
(3) Other process connections available upon request.
(4) Not available for hand polished wet side.
(5) Example: Code E1181 is 118.1 inches. Code M3000 is 3000
millimeters. See Extended Lengths on page 5 for minimum and maximum
extended lengths.
(6) The requirements of CRN are met when a Rosemount 2160
CSA-approved vibrating fork level switch model
2160****S**********I6****** is configured with 316/316L stainless
steel (1.4401/1.4404) wetted parts and either NPT threaded or 2-in.
to 8-in. ASME B16.5 flanged process connections.
(7) The long-life Power Module must be shipped separately. Order
part number 00753-9220-0001.
(8) Option limited to units with extended lengths up to 59.1-in.
(1500 mm).
Table 1. 2160 Ordering Information★The Standard offering
represents the most common options. The starred options (★) should
be selected for best delivery.
The Expanded offering is subject to additional delivery lead
time.
Table 2. Spare Parts and Accessories★The Standard offering
represents the most common options. The starred options (★) should
be selected for best delivery.
The Expanded offering is subject to additional delivery lead
time.
Spares and Accessories Standard Standard02100-1000-0001 Seal for
1-in. BSPP (G1A). Material: Non-asbestos BS7531 grade X carbon
fiber with rubber binder ★
02100-1040-0001 Seal for 3/4-in. BSPP (G3/4A). Material:
Non-asbestos BS7531 grade X carbon fiber with rubber binder ★
02100-1010-0001 Hygienic adaptor boss 1-in. BSPP. Material: 316
SS fitting. FPM/FKM o-ring ★
02100-1020-0001 2-in. (51 mm) Tri-clamp kit (vessel fitting,
clamp ring, and seal). Material: 316 St. steel, NBR Nitrile ★
02120-2000-0001(1) 11/2-in. BSPP adjustable clamp gland for
1-in. extended lengths.Material: 316 St. steel, (Si) Silicone
rubber seal
★
02120-2000-0002(1) 11/2-in. NPT adjustable clamp gland for 1-in.
extended lengths.Material: 316 St. steel, (Si) Silicone rubber
seal
★
(1) The adjustable clamp gland is not explosion-proof.
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
5
Rosemount 2160
Specification
Physical
ProductThe Rosemount 2160 WirelessVibrating Fork Liquid Level
Switch
Measuring PrincipleVibrating Fork
ApplicationsMost liquids including coating liquids, aerated
liquids, and slurries.
Enclosure Housing: Low-copper aluminumPaint: PolyurethaneCover
O-ring: Nitrile butadiene
Terminal Block and Power Module PackPBT
AntennaPBT/PC integrated omnidirectional antenna
ConnectionsSee “2160 Ordering Information” on page 2.
Extended Lengths
The maximum extended length is 118.1 in. (3000 mm) except for:•
Hand-polished process (maximum length is 39.4 in./1000 mm)
Process Material316/316L Stainless Steel (1.4401/1.4404) dual
certified, orAlloy C (UNS N10002) and Alloy C-276 (UNS
N10276).Hand-polished to better than 0.4 m option available for
hygienic connections.Gasket material for 3/4-in. and 1-in. BSPP (G)
is non-asbestos BS7531 Grade X carbon fiber with rubber binder.
Dimensional DrawingsSee Dimensional Drawings on page 9.
MountingSuitable for horizontal and vertical
installations.Rotatable housing allows correct alignment of both
the forks and the omnidirectional antenna for optimal signal and
best viewing position of the LCD integral display.
Enclosure RatingsHousing is NEMA 4X and IP66 compliant.
Electrical
Wireless Power ModuleReplaceable, Intrinsically Safe
Lithium-Thionyl Chloride power module with PBT enclosure.Ten year
life at one minute transmit rate (1).
(1) Reference conditions are 70 °F (21 °C), and routing data for
three additional network devices. NOTE: Continuous exposure to
ambient temperature limits (-40 °F or 185 °F) (-40 °C or 85 °C) may
reduce specified power module life by 20 percent.
HART Communicator ConnectionsClips are permanently fixed to
terminal block.
Performance
Electromagnetic Compatibility (EMC)All models meet all relevant
requirements of EN 61326.
Hysteresis (Water)±0.039 in. (±1 mm) nominal
Switching Point (Water)0.5 in. (13 mm) from fork tip if mounted
vertically.0.5 in. (13 mm) from the fork edge if mounted
horizontally. The switch point varies with different liquid
densities.
Functional
OutputWireless 2.4 GHz DSSS
Local DisplayThe optional five-digit integral LCD display can
indicate the switch state (Dry or Wet) and diagnostic
information.
Humidity Limits0 to 100% relative humidity
Transmit RateUser-selectable: from 4 seconds up to 60
minutes.The optional integral LCD display updates at the transmit
rate up to once per minute.
Maximum Operating PressureThreaded connection: See Figure
1.Hygienic connection: 435 psig (30 bar g)Flanged connection:The
maximum operating pressure is the lower of the process pressure
(Figure 1) and flange pressure rating (Table 4).
Table 3. Minimum Extended LengthsProcess Connection Minimum
Extended Length3/4–in. Threaded 3.8 in. (95 mm)1–in. Threaded 3.7
in. (94 mm)Flanged 3.5 in. (89 mm)Tri-Clamp 4.1 in. (105 mm)
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
6
NOTE:The final maximum operating pressure rating depends on
theprocess (tank) connection.Clamp glands (order #02120-2000-0001
or 02120-2000-0002) limit the maximum operating pressure to 18.85
psig (1.3 bar g).
Figure 1. Process Pressure
TemperatureSee Figure 2 for the maximum and minimum operating
temperatures.Figure 2. Temperature
Liquid Density RangeMinimum liquid density is 31.2 lb/ft3 (500
kg/m3).
Liquid Viscosity Range0.2 to 10000 cP (centiPoise).
Solids Content and CoatingThe maximum recommended diameter of
solid particles in the liquid is 0.2 in. (5 mm).For coating
products, avoid bridging of forks (fork-to-fork).
CIP (Clean In Place) CleaningThe 2160 withstands steam
cleaning.
Table 4. Maximum Flange Pressure RatingFlange Standard SST
Flanges (1)
(1) ASTM stainless steel.
ASME B16.5 Class 150 275 psig (2)
(2) At 100 °F (38 °C), the pressure rating decreases with an
increasing process temperature.
ASME B16.5 Class 300 720 psig (2)
ASME B16.5 Class 600 1440 psig (2)
EN1092-1 PN 10/16 10/16 bar g (3)
EN1092-1 PN 25/40 25/40 bar g (3)
EN1092-1 PN 63 63 bar g (3)
(3) At 122 °F (50 °C), the pressure rating decreases with an
increasing process temperature.
EN1092-1 PN 100 100 bar g (3)
1450 (100)
1880 (80)
-14.5 (-1.0)-40
(-40)122 (50)
302 (150)
Process Temperature °F (°C)Pro
cess
Pre
ssur
e ps
ig (b
arg)
32 (0)
2160***S
1450 (100)
885 (61)
-14.5 (-1.0)-94
(-70)100 (38)
500 (260)
Process Temperature °F (°C)Pro
cess
Pre
ssur
e ps
ig (b
arg)
32 (0)
2160***E
176 (80)
122 (50)
-40 (-40)-40
(-40)302
(150)
Process Temperature °F (°C)
Am
bien
t Tem
pera
ture
°F (°
C)
122 (50)
32 (0)
32 (0)
2160***S
176 (80)
176 (80)
-40 (-40)
-94 (-70)
500 (260)
Process Temperature °F (°C)
Am
bien
t Tem
pera
ture
°F (°
C)
149 (65)
32 (0)
32 (0)
2160***E
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
7
Rosemount 2160
Product Certifications
ROSEMOUNT 2160
Approved Manufacturing LocationsRosemount Inc.– Chanhassen,
Minnesota, USAMobrey Limited– Slough, United KingdomEmerson Process
Management Asia Pacific Private Limited– Singapore
European Directive InformationThe EC declaration of conformity
for all applicable European directives for this product can be
found at www.rosemount.com.A hard copy may be obtained by
contacting an Emerson Process Management representative.ATEX
Directive (94/9/EC)• Emerson Process Management complies with
the
ATEX DirectiveEuropean Pressure Equipment Directive (PED)
(97/23/EC)• The 2160 is outside the scope of PED Directive
L.V. Directive• EN61010-1 Pollution degree 2, Category II (264 V
max),
Pollution degree 2, Category III (150 V max).Electro Magnetic
Compatibility (EMC) (2004/108/EC)• EN 61326-1:2006
Radio and Telecommunications Terminal EquipmentDirective
(R&TTE) (1999/5/EC)• Emerson Process Management complies with
the
R&TTE Directive
Telecommunication ComplianceAll wireless devices require
certification to ensure that they adhere to regulations regarding
the use of the RF spectrum. Nearly every country requires this type
of product certification. Emerson is working with governmental
agencies around the world to supply fully compliant products and
remove the risk of violating country directives or laws governing
wireless device usage. To see which countries our devices have
received certification for use in, see
www.rosemount.com/smartwireless.
FCC and ICThis device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1) This
device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may
cause undesired operation. This device must be installed to ensure
a minimum antenna separation distance of 8 in. (20 cm) from all
persons.FCC ID: LW2RM2510IC ID: 2731A-RM2510
Canadian Registration NumberCRN 0F04227.25
NOTEThe requirements of CRN are met when a Rosemount 2160
CSA-approved vibrating fork level switch model
2160****S**********I6****** is configured with 316/316L stainless
steel (1.4401/1.4404) wetted parts and either NPT threaded or 2-in.
to 8-in. ASME B16.5 flanged process connections.
Hazardous Locations Certificates
American and Canadian Approvals
Factory Mutual (FM) ApprovalsI5 Project ID: 3036541
FM Intrinsic Safety, Non-incendive, and Dust Ignition-proof
Intrinsically Safe for Class I, Division 1, Groups A, B, C, DZone
Marking: Class I, Zone 0, AEx ia llCTemperature Codes T4 (Tamb =
–50 to 70 °C)Non-incendive for Class I, Division 2, Groups A, B, C,
and DDust Ignition-proof for Class II/III, Division I, Groups E, F,
GAmbient temperature limits: –50 to 70 °CFor use with Rosemount
SmartPower® optionsP/N 753-9220-0001 only.Enclosure Type 4X /
IP66
Canadian Standards Association (CSA) ApprovalI6 Certificate
Number: 06 CSA 1786345
CSA Intrinsic SafetyIntrinsically Safe for Class I, Division 1,
Groups A, B, C,and DTemperature Code T3CEnclosure Type 4X / IP66For
use with Rosemount SmartPower optionsP/N 753-9220-0001
only.Intrinsically Safe when installed in accordance with Rosemount
drawing 71097/1271.
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
8
European Approvals
ATEX ApprovalI1 ATEX Intrinsic Safety
Certificate Number: Baseefa 09ATEX0253XII 1 G, Ex ia IIC T5-T2
(Ta = –40 to 70 °C)IP66For use with Rosemount SmartPower optionsP/N
753-9220-0001 only.Special Conditions For Safe Use:The surface
resistivity of the antenna is greater than1 gigaohm. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned
with solvents or a dry cloth.
Rest Of The World Approvals
IECEx ApprovalI7 IECEx Intrinsic Safety
Certificate Number: IECEx BAS 09.0123XEx ia IIC T5-T2 (Ta = –40
to 70 °C)IP66For use with Rosemount SmartPower optionsP/N
753-9220-0001 only.Special Conditions For Safe Use:The surface
resistivity of the antenna is greater than1 gigaohm. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned
with solvents or a dry cloth.
National Supervision and Inspection Centre(NEPSI) ApprovalsI3
NEPSI Intrinsic Safety
Certificate: GYJ101138XEx ia IIC T5-T2Special Conditions For
Safe Use:
1. Symbol “X” is used to denote specific conditions of use:a.
Model 648 WTT or Model 3051S WPT type battery
pack provided by the manufacturer should be usedb. The surface
resistivity of the antenna is greater than
1 gigaohm. To avoid electrostatic charge build-up, it must not
be rubbed or cleaned with solvents or a dry cloth
c. The Rosemount 2160 enclosure is made of aluminum alloy and
given a protective epoxy coating. However, care should be taken to
protect it from impact or abrasion if located in a Zone 0.
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
9
Rosemount 2160
Dimensional Drawings2160 Thread Mounting (Standard Length) . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
page 92160 Thread Mounting (Extended Length) . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . page 102160
Flange Mounting (Standard Length) . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . page 112160 Flange Mounting
(Extended Length) . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . page 122160 Hygienic Fitting (Standard
Length). . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . page 132160 Hygienic Fitting (Extended Length) . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. page 15
2160 Thread Mounting (Standard Length)
A A
AluminumHousing
0.5 (13) Switchpoint(When Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
3/4-in. or 1-in.Thread
6.1 (154) No LCD Fitted
1.14 (29) for 1-in. Thread0.9 (23) for 3/4-in. Thread
1.7(44)
2.7(69)
MinimumTo AllowRemovalOf Cover
View OnArrow ‘A’ShowingPossibleAntennaRotation
90°
6.9(174)
1.6 (40) A/FHexagon
0.5 (13) Switchpoint(When Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
3/4-in. or 1-in. Thread
1.14 (29) for 1-in. Thread0.9 (23) for 3/4-in. Thread1.7
(44)
2.7(69)
14.5(369)
2160***S 2160***E
3.5(90)
3.5(90)
7.9 (200)
6.8 (172) LCD Fitted
2.4(60)
3.7 (93) 1.6 (40) A/FHexagon
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
Note: Dimensions are in inches (millimeters)
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
10
2160 Thread Mounting (Extended Length)
Table 5. Thread Mounting Fork Length
A A
1.65 (40) A/F Hexagon
3/4-in. or 1-in. Thread
E(M)
1.7(44)
0.5 (13) Switchpoint(When Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
1.14 (29) for 1-in. Thread0.9 (23) for 3/4-in. Thread90°
14.4(367)
6.8(172)
1.6 (40) A/FHexagon
0.5 (13) Switchpoint(Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
3/4-in. or 1-in. Thread
1.14 (29) for 1-in. Thread0.9 (23) for 3/4-in. Thread
1.7(44)
E(M)
2160***S 2160***E
3.5(90)
3.5(90)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
Note: Dimensions are in inches (millimeters)
ProcessConnection
Standard LengthModel Code A
Minimum LengthModel Code E (M)
Maximum Length Model Code E (M)(1)
3/4-in. Thread 1.73 in. (44 mm) 3.75 in. (95 mm) 118.1 in. (3000
mm)1-in. Thread 1.73 in. (44 mm) 3.74 in. (94 mm) 118.1 in. (3000
mm)
(1) Maximum extended length of fork with hand-polished option is
39.4 in. (1000 mm).
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
11
Rosemount 2160
2160 Flange Mounting (Standard Length)
A A
1.7(44)
4.0(102)
0.5 (13) Switchpoint(When Mounted Horizontally)
1.14 (29) for 11/2-in. orLarger Flange;0.9 (23) for up to 1-in.
Flange
0.5 (13) Switchpoint(When Mounted Vertically)
90°
13.8(351)
1.7(44)
4.0(102)
0.5 (13) Switchpoint(When Mounted Horizontally)
1.14 (29) for 11/2-in. orLarger Flange;0.9 (23) for up to 1-in.
Flange
0.5 (13) Switchpoint (When Mounted Vertically)
2160***S 2160***E7.0(177)
3.5(90)
3.5(90)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
Note: Dimensions are in inches (millimeters)
Note: Dimensions are in inches (millimeters)
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
12
2160 Flange Mounting (Extended Length)
Table 6. Flange Mounting Fork Length
A A
1.7(44)
E(M)
0.5 (13) Switchpoint(When Mounted Horizontally)
1.14 (29) for 11/2-in. orLarger Flange;0.9 (23) for up to 1-in.
Flange
0.5 (13) Switchpoint(When Mounted Vertically)
7.0(177)
13.8(351)
1.7(44)
0.5 (13) Switchpoint(Mounted Horizontally)
1.14 (29) for 11/2-in. orLarger Flange;0.9 (23) for up to 1-in.
Flange
0.5 (13) Switchpoint(When Mounted Vertically)
E(M)
2160***S 2160***E
3.5(90)
3.5(90)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
90°
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
Note: Dimensions are in inches (millimeters)
ProcessConnection
Standard LengthFork Length Code H
Minimum LengthFork Length Code E(M)
Maximum Length Fork Length Code E(M)
3/4-in., 1-in. or larger flange 4.0 in. (102 mm) 3.7 in. (94 mm)
118.1 in. (3000 mm)
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
13
Rosemount 2160
2160 Hygienic Fitting (Standard Length)
A A
1.14(29)1.7
(44)
2.5(64)
1/2-in. (38) or2-in. (51) Tri-Clamp
1.6 (40) A/FHexagon
1.14(29)1.7
(44)
1.6 (40) A/FHexagon
0.5 (13) Switchpoint(When Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
0.5 (13) Switchpoint(When Mounted Vertically)
0.5 (13) Switchpoint(When Mounted Horizontally)
2.5(64)
7.1(179)
14.9(379)
1/2-in. (38) or2-in. (51) Tri-Clamp
2160***S 2160***E
3.5(90)
3.5(90)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
90°
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
Note: Dimensions are in inches (millimeters)
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
14
2160 Hygienic Fitting (Standard Length) Continued
A A
1.14(29)1.7
(44)
3.1(78)
1-in. Thread
1.14(29)1.7
(44)
1.6 (40 A/F Hexagon
0.5 (13) Switchpoint(When Mounted Horizontally)
0.5 (13) Switchpoint(When Mounted Vertically)
0.5 (13) Switchpoint(Mounted Vertically)
0.5 (13) Switchpoint(When Mounted Horizontally)
3.1(78)
7.0(178)
14.6(372)
1-in. Thread
2160***S 2160***E
3.5(90)
3.5(90)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
90°
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
1.6 (40 A/F Hexagon
Note: Dimensions are in inches (millimeters)
-
Product Data Sheet00813-0100-4160, Rev BBSeptember 2010
15
Rosemount 2160
2160 Hygienic Fitting (Extended Length)
Table 7. Hygienic Fitting Fork Length
A A
1.14(29)1.7
(44)
1/2-in. (38) or 2-in. (51) Tri-Clamp
E(M)
0.5 (13) Switchpoint(When Mounted Vertically)
0.5 (13) Switchpoint(When Mounted Horizontally)
14.9(379)
7.1(179)
E(M)
0.5 (13) Switchpoint(When Mounted Vertically)
1.7(44)
1.14(29)
2160***S 2160***E
0.5 (13) Switchpoint(Mounted Horizontally)
3.5(90)
3.5(90)
90°
AluminumHousing
View OnArrow ‘A’ShowingPossibleAntennaRotation
7.9 (200)
3.7 (93)
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
6.8 (172) LCD Fitted
2.4(60)
6.8 (172) LCD Fitted
6.1 (154) No LCD Fitted MinimumTo AllowRemovalOf Cover
2.4(60)
Note: Dimensions are in inches (millimeters)
ProcessConnection
Standard LengthFork Length Code H
Minimum LengthFork Length Code E(M)
Maximum Length Fork Length Code E(M)(1)
Tri-Clamp 1.73 in. (44 mm) 4.13 in. (105 mm) 118.1 in. (3000
mm)1-in. Threaded 1.73 in. (44 mm) 3.74 in. (94 mm) 118.1 in. (3000
mm)
(1) Maximum extended length of fork with hand-polished option is
39.4 in. (1000 mm).
-
Product Data Sheet00813-0100-4160, Rev BB
September 2010Rosemount 2160
Rosemount Level SolutionsEmerson provides a complete range of
Rosemount products for level measurement applications.
Pressure – Level or Interface MeasurementEmerson has a complete
line of Rosemount pressure transmitters and remote seals for
measuring level or interfaces in liquid applications. Optimize
performance with direct mount, Tuned Seal systems:• Rosemount
3051S_L, 3051L, and 1151LT Liquid Level
Transmitters• Rosemount 1199 Remote Diaphragm Seals with direct
mount
or capillary connectionsVibrating Fork Switches – Point Level
DetectionThe Rosemount 2100 Series is developed for reliable point
level detection of liquids and consists of:• Rosemount 2110 Compact
Vibrating Fork
Liquid Level Switch• Rosemount 2120 Full-featured Vibrating
Fork
Liquid Level Switch• Rosemount 2130 Extreme Temperature
Vibrating Fork
Liquid Level Switch• Rosemount 2160 Wireless Vibrating Fork
Liquid Level Switch
Guided Wave Radar – Level and Interface
MeasurementMultivariable, loop-powered Guided Wave Radar
transmitters with a wide range of probe styles to fit different
liquids and solids applications. The product line consists of:•
Rosemount 3300 Series – Versatile and easy-to-use
transmitter with proven reliability• Rosemount 5300 Series –
Accurate, high performance
transmitter with FOUNDATION™ fieldbus supportNon-contacting
Radar – Level MeasurementThe Rosemount non-contacting radar family
consists of:• Rosemount 5400 Series Transmitters – Loop-powered
transmitter with a wide range of antennas, for liquid level
measurement in most applications and process conditions
• Rosemount 5600 Series Transmitters – 4-wire loop-powered
transmitter giving maximum sensitivity and performance for solids,
challenging reactors, rapid level changes and excessive process
conditions.
Non-contacting Ultrasonic – Level MeasurementThe Rosemount 3100
Series ultrasonic level transmitters provide continuous
non-contacting level measurement of liquids. The range consisits
of:• Rosemount 3101 for simple continuous level measurement•
Rosemount 3102 for continuous measurement with two integral
relays for local control functionality• Rosemount 3105
Intrinsically safe certified version for
hazardous areas
The Emerson logo is a trademark and service mark of Emerson
Electric Co.Rosemount and the Rosemount logotype are registered
trademarks of Rosemount Inc.PlantWeb is a registered trademark of
one of the Emerson Process Management group of companies.HART and
WirelessHART are registered trademarks of the HART Communication
Foundation.All other marks are the property of their respective
owners.
Standard Terms and Conditions of Sale can be found at
www.rosemount.com\terms_of_sale
00813-0100-4160 Rev BB 09/10
© 2010 Rosemount, Inc. All rights reserved.
Emerson Process Management Rosemount Inc.8200 Market
BoulevardChanhassen, MN 55317 USAT (U.S.) 1-800-999-9307T
(International) (952) 906-8888F (952) 949-7001www.rosemount.com
Emerson Process ManagementBleigistrasse 23P.O. Box 1046CH 6341
BaarSwitzerlandT +41 (0) 41 768 6111F +41 (0) 41 768 6300
Emerson Process ManagementAsia Pacific Pte Ltd1 Pandan
CrescentSingapore 128461T +65 6777 8211F +65 6777 0947Service
Support Hotline: +65 6770 8711Email:
[email protected]
Emerson FZEP.O. Box 17033Jebel Ali Free ZoneDubai UAET +971 4
883 5235F +971 4 883 5312