Top Banner
Fully mechanical refurbish and upgrade of the electronic control box of Pav-Pac Laboratory Compactor.
17

RMS Refurbish PAV-PAC Laboratory Compactor

May 30, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: RMS Refurbish PAV-PAC Laboratory Compactor

Fully mechanical refurbish and upgrade of the electronic control box of Pav-Pac Laboratory

Compactor.

Page 2: RMS Refurbish PAV-PAC Laboratory Compactor

The challenge has been to supply the best service that meet all your business requirements, including performance, reliability, and maintainability with critical spare parts that are readily available locally and with technology that is readily serviceable.

Page 3: RMS Refurbish PAV-PAC Laboratory Compactor

Mechanical rebuild of your machine involve the following

steps: 1.  Fully dismantle machine, clean inspect all components 2.  Replace all pneumatic system with parts readily available 3.  Replace all limit , reed switches 4.  Service table assemble, replace Sprag clutches, seals 5.  Service latching mechanism 6.  Assemble all parts 7.  Test run

Page 4: RMS Refurbish PAV-PAC Laboratory Compactor

Control box upgrade 1.  Replace existing control box with a new control box,

this unit will be using existing plug configuration as current unit.

2.  Supply and install Panasonic PLC 8 in 6out 3.  Supply and install RedLion Kadet Operator Interface

850 4.  Supply and install 24vdc power supply 5.  Supply and install safety relay/switch 6.  Supply new enclosure 7.  Test run

Page 5: RMS Refurbish PAV-PAC Laboratory Compactor

Control panel

Page 6: RMS Refurbish PAV-PAC Laboratory Compactor

Setting counter

Page 7: RMS Refurbish PAV-PAC Laboratory Compactor

Touch screen

Page 8: RMS Refurbish PAV-PAC Laboratory Compactor

Replace main pneumatic cylinder Old cylinder New cylinder

Page 9: RMS Refurbish PAV-PAC Laboratory Compactor

Replace all solenoids- Standard Old solenoids- obsolete New solenoids

Page 10: RMS Refurbish PAV-PAC Laboratory Compactor

Supply solenoid manifold Old solenoids –obsolete

New solenoids

Page 11: RMS Refurbish PAV-PAC Laboratory Compactor

Safety Modifications Safety Relay Safety switch

Page 12: RMS Refurbish PAV-PAC Laboratory Compactor

Replace locking cylinder Old cylinder Two New cylinders

Page 13: RMS Refurbish PAV-PAC Laboratory Compactor

Replace Slew solenoid Old solenoid New Solenoid

Page 14: RMS Refurbish PAV-PAC Laboratory Compactor

Replace turn table solenoid Old solenoid New solenoid

Page 15: RMS Refurbish PAV-PAC Laboratory Compactor

Replace all reed switches Old switches New switches

Page 16: RMS Refurbish PAV-PAC Laboratory Compactor

Replace air service unit Old unit New unit

Page 17: RMS Refurbish PAV-PAC Laboratory Compactor

Service turn table

Replace clutch &seals