RFID Fall RFID Fall Detection Floor Detection Floor Mat System Mat System Andrew Heidt Andrew Heidt Brian Tippins Brian Tippins Coleman McDaniel Coleman McDaniel Zach Brannan Zach Brannan Abdirizak Mire Abdirizak Mire
Jan 27, 2016
RFID Fall Detection RFID Fall Detection Floor Mat SystemFloor Mat System
Andrew Heidt Andrew Heidt Brian TippinsBrian Tippins
Coleman McDanielColeman McDanielZach BrannanZach BrannanAbdirizak MireAbdirizak Mire
System OverviewSystem Overview
Detects patient falls using an RFID Detects patient falls using an RFID wrist tag and an antenna embedded wrist tag and an antenna embedded in a floor matin a floor mat
Is used in hospitals and elderly care Is used in hospitals and elderly care facilities, where falls are commonfacilities, where falls are common
Will cost $3,500 per unitWill cost $3,500 per unit
Design ObjectivesDesign Objectives
Detect patient falls and egressDetect patient falls and egress
< 5% detection failures< 5% detection failures
Notify staff of incidentsNotify staff of incidents
Record incidents in logfileRecord incidents in logfile
System DiagramSystem Diagram
RFID ReaderRFID Reader
Feig OBID i-Scan MR101-AFeig OBID i-Scan MR101-A Operating Frequency: 13.56 MHzOperating Frequency: 13.56 MHz Power Output: 1 WattPower Output: 1 Watt Output Impedance: 50 ohmsOutput Impedance: 50 ohms
Floor MatFloor Mat
Built with hard tapered edgesBuilt with hard tapered edges Contains antenna structureContains antenna structure Placed at patient bedsidePlaced at patient bedside
30”
48”
3”
AntennaAntenna
Copper foil antennaCopper foil antenna Resonant at 13.56MHzResonant at 13.56MHz Matched to 50 ohm readerMatched to 50 ohm reader
Antenna Design OptionsAntenna Design Options
Option 1: Single-loop antenna approximately the Option 1: Single-loop antenna approximately the area of the floor matarea of the floor mat
Option 2: Two antennas multiplexedOption 2: Two antennas multiplexed
Single-loop Multiplexed
Impedance Simulations (HFSS) Comparing Single-Impedance Simulations (HFSS) Comparing Single-loop and Multiplexed designsloop and Multiplexed designs
Single-loop Multiplexed
L= Z/(jω)=1.81 μH L= Z/(jω)=1.26 μH
Magnetic Field Simulations (HFSS) Comparing Magnetic Field Simulations (HFSS) Comparing Single-loop and Multiplexed DesignsSingle-loop and Multiplexed Designs
Single-loop Multiplexed
Summary of Antenna Design OptionsSummary of Antenna Design Options
Single-loopSingle-loop MultiplexedMultiplexed
ProsPros
Larger rangeLarger rangeLow costLow costSimpleSimple
Smaller inductance Smaller inductance (easier to match)(easier to match)
ConsCons Higher inductance Higher inductance (harder to match)(harder to match)
Multiplexing adds Multiplexing adds complexity and costcomplexity and costHigher failure rateHigher failure rateSmaller rangeSmaller range
Computer Interface: Process FlowComputer Interface: Process Flow
Interfacing with Hospital NetworkInterfacing with Hospital Network
Patient PCPatient PC• Logs incidentsLogs incidents• Displays fall alertsDisplays fall alerts• Transmits alerts to nurse stationTransmits alerts to nurse station
Nurse StationNurse Station• Displays fall alertsDisplays fall alerts• Adds/Removes patient informationAdds/Removes patient information
VB – Computer ImplementationVB – Computer Implementation
Patient PC functions:Patient PC functions:• Listener() – listens on serial port for tag Listener() – listens on serial port for tag
detectionsdetections• File_Logger() – logs detection incident File_Logger() – logs detection incident
with date/time in local filewith date/time in local file• Transmit() – forwards detection Transmit() – forwards detection
incidents and listen resets to the nurse incidents and listen resets to the nurse stationstation
• Alert() – displays alert window on PCAlert() – displays alert window on PC
VB – Computer ImplementationVB – Computer Implementation
Nurse StationNurse Station• Add(), Remove() – add and remove Add(), Remove() – add and remove
patients from the monitoring systempatients from the monitoring system• Edit() – edit patient information (name, Edit() – edit patient information (name,
room number, tag number etc.)room number, tag number etc.)• Ignore() – stops alerts for a specific Ignore() – stops alerts for a specific
patientpatient• Alert() – displays alert window with Alert() – displays alert window with
incident and patient informationincident and patient information
GUIsGUIs Nurse StationNurse Station
Patient PCPatient PC
CostsCosts
Expected Costs:Expected Costs: Updated Costs: Updated Costs:
Part Price
Feig Mid Range Reader $700
Floor Mat $100
Copper Foil $10
Cables $15
Capacitors $10
Resistors $2
Multiplexer* $50
Total $887
Part Price
Feig Mid Range Reader $532
Floor Mat** $100
Copper Foil $20
Cables $29.35
Capacitors $9.10
Resistors** $2
Monitor $120
PC $50
Total $862.45
* - We no longer anticipate the use of a multiplexer in the design
** - Estimated Cost
TimelineTimeline
Patient falls cost the health care Patient falls cost the health care industry $19.4 billion in each yearindustry $19.4 billion in each year