Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton, Kavita Gala ASCRS 2010 No Relevant Financial Disclosure
Dec 13, 2015
Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass
Measurement System
St Thomas Hospital, London
Anish Dhital, David Spalton, Kavita Gala
ASCRS 2010No Relevant Financial Disclosure
Purpose• To measure the reproducibility of optical measurements in
pseudophakic patients using a double pass measurement system.
Light scatter• Increasing interest in measurements of light scatter and impact on
visual function• Methods of measuring scatter
• Double-pass• Psychophysical• Hartmann-Shack
Variables that affect light scatter
• Tear film• Cornea• Anterior Chamber• Intra-ocular lens• Vitreous clarity
• Measurement error inherent in the system may also affect the repeatability of measurements
Montes -Mico et al IOVS 46:12:4468-44742005Nanavaty et al Ophthalmogica in press
Optical Quality Analysis System II (OQAS)
• Uses double-pass measurement system
• Images point spread function• Derives MTF, Strehl ratio, light
scatter (objective scatter index - OSI)
(Visionmetrics, Spain)
MTF - the ratio between contrast in the retinal image and contrast of the original object.OSI - objective parameter calculated by evaluating the amount of light on the outside of the double-pass image in relation to the amount of light on the centre. Higher OSI value relates to more intraocular scattering.Strehl ratio - measurement of an optical system (ratio of the MTF to that of a diffraction limited system). Thus, Strehl ratio will be a figure between 0 and 1, the higher its value, the better
Methods • Prospective reproducibility study
Population • Normal pseudophakic eyes,• At least three months post-cataract surgery
Measures• Refraction• OQAS measurements used customised OQAS II
machine
Methods• Refraction - to obtain the cylindrical correction to optimise
measurement.
• OQAS II - Pupils were dilated and measurements made for a 4mm exit pupil diameter.
• 3 sets of measurements of OSI and Strehl ratio from a single
eye at one visit were taken and the average calculated. This was repeated one to two weeks later.
• Correlation plots and Bland-Altman analyses were used to assess and compare the 2 sets of measurements.
Results
• 51 eyes from 29 patients were assessed, with a heterogenous group of intraocular lenses:• 20 eyes with an Alcon SA60AT• 5 eyes with an Akreos MI60,• 6 eyes with an Acrismart 36a IOL,• 6 eyes with an Alcon SA60WF• 6 eyes with the Morcher 89A ‘Bag in the Lens’• 2 eyes with Crystalens HD intra-ocular lens.• 6 eyes with unknown IOL type
Results
0
0.05
0.1
0.15
0.2
0.25
0 0.05 0.1 0.15 0.2 0.25
R2 = 0.78
Average of 2nd reading
Average of 1st reading
Correlation of Strehl ratio values
Results
0
2
4
6
8
10
12
14
16
18
20
0 2 4 6 8 10 12 14 16
Correlation of OSI values
R2 = 0.95
Average of 1st reading
Average of 2nd reading
Results Bland-Altman Plot of Strehl ratio
-0.05
-0.04
-0.03
-0.02
-0.01
0
0.01
0.02
0.03
0.04
0.05
0.06
-0.1 -0.05 0 0.05 0.1 0.15 0.2 0.25 0.3
Coefficient of Repeatability = 0.033
Strehl ratio
Difference
Standard Deviation = 0.017
Results Bland-Altman Plot of OSI
Difference
-2
-1.5
-1
-0.5
0
0.5
1
1.5
2
2.5
3
0 2 4 6 8 10 12 14 16 18
Ocular Scatter Index
Coefficient of Repeatability = 1.56
Standard Deviation = 0.79
Conclusion
• Optical measurements using a double-pass system are reproducible despites concerns of changes in the anterior segment between visits.