Top Banner
Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton, Kavita Gala ASCRS 2010 No Relevant Financial Disclosure
12

Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Dec 13, 2015

Download

Documents

Britton Fox
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass

Measurement System

St Thomas Hospital, London

Anish Dhital, David Spalton, Kavita Gala

ASCRS 2010No Relevant Financial Disclosure

Page 2: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Purpose• To measure the reproducibility of optical measurements in

pseudophakic patients using a double pass measurement system.

Light scatter• Increasing interest in measurements of light scatter and impact on

visual function• Methods of measuring scatter

• Double-pass• Psychophysical• Hartmann-Shack

Page 3: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Variables that affect light scatter

• Tear film• Cornea• Anterior Chamber• Intra-ocular lens• Vitreous clarity

• Measurement error inherent in the system may also affect the repeatability of measurements

Montes -Mico et al IOVS 46:12:4468-44742005Nanavaty et al Ophthalmogica in press

Page 4: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Optical Quality Analysis System II (OQAS)

• Uses double-pass measurement system

• Images point spread function• Derives MTF, Strehl ratio, light

scatter (objective scatter index - OSI)

(Visionmetrics, Spain)

MTF - the ratio between contrast in the retinal image and contrast of the original object.OSI - objective parameter calculated by evaluating the amount of light on the outside of the double-pass image in relation to the amount of light on the centre. Higher OSI value relates to more intraocular scattering.Strehl ratio - measurement of an optical system (ratio of the MTF to that of a diffraction limited system). Thus, Strehl ratio will be a figure between 0 and 1, the higher its value, the better

Page 5: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Methods • Prospective reproducibility study

Population • Normal pseudophakic eyes,• At least three months post-cataract surgery

Measures• Refraction• OQAS measurements used customised OQAS II

machine

Page 6: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Methods• Refraction - to obtain the cylindrical correction to optimise

measurement.

• OQAS II - Pupils were dilated and measurements made for a 4mm exit pupil diameter.

• 3 sets of measurements of OSI and Strehl ratio from a single

eye at one visit were taken and the average calculated. This was repeated one to two weeks later.

• Correlation plots and Bland-Altman analyses were used to assess and compare the 2 sets of measurements.

Page 7: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Results

• 51 eyes from 29 patients were assessed, with a heterogenous group of intraocular lenses:• 20 eyes with an Alcon SA60AT• 5 eyes with an Akreos MI60,• 6 eyes with an Acrismart 36a IOL,• 6 eyes with an Alcon SA60WF• 6 eyes with the Morcher 89A ‘Bag in the Lens’• 2 eyes with Crystalens HD intra-ocular lens.• 6 eyes with unknown IOL type

Page 8: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Results

0

0.05

0.1

0.15

0.2

0.25

0 0.05 0.1 0.15 0.2 0.25

R2 = 0.78

Average of 2nd reading

Average of 1st reading

Correlation of Strehl ratio values

Page 9: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Results

0

2

4

6

8

10

12

14

16

18

20

0 2 4 6 8 10 12 14 16

Correlation of OSI values

R2 = 0.95

Average of 1st reading

Average of 2nd reading

Page 10: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Results Bland-Altman Plot of Strehl ratio

-0.05

-0.04

-0.03

-0.02

-0.01

0

0.01

0.02

0.03

0.04

0.05

0.06

-0.1 -0.05 0 0.05 0.1 0.15 0.2 0.25 0.3

Coefficient of Repeatability = 0.033

Strehl ratio

Difference

Standard Deviation = 0.017

Page 11: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Results Bland-Altman Plot of OSI

Difference

-2

-1.5

-1

-0.5

0

0.5

1

1.5

2

2.5

3

0 2 4 6 8 10 12 14 16 18

Ocular Scatter Index

Coefficient of Repeatability = 1.56

Standard Deviation = 0.79

Page 12: Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,

Conclusion

• Optical measurements using a double-pass system are reproducible despites concerns of changes in the anterior segment between visits.