This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Recirculating Coolers - Basic type
Low Temp. Recirculating Coolers - General Type
Low Temp. Recirculating Coolers - Intelligent type
Providing a constant temperature control and high cooling efficiency.
Operating cost can be reduced by using the recirculator for cooling water. It is eco-friendly and reduces energy consumption.
Operating features • Control temperature range is from 3℃~40℃.
• Clear and easy-to-use LCD display. (resolution 0.1℃)
• Bright LED water level indicator can be seen from a distance.
• Low fluid level and dry-running protection when triggered shows an error symbol on the display and an audible buzzer sound.
• If the equipment operates at 40℃ or above due to the external environment, user can simply identify the error with a visual signal or an audible buzzer sound.
• Pump pressure can be controlled using a by-pass function. (only H type)
Constructional features • The refrigerating system is designed to reduce energy consumption
and to maximize in refrigerating efficiency.
• Eco-friendly R-404A refrigerant is used to prevent environmental pollution. Double service valve is used to prevent refrigerant leakage.
• Removable gridded vent for reducing condenser contamination and to facilitate cleanliness.
• A plastic solution inlet is for easy and safe pouring of solution into the chiller unit.
• Pressure gauge comes standard with chiller units to provide fluid system diagnostics at a glance and for observing the pump pressure.
• Wheels allow for easy mobility throughout the lab.
• The beautifully designed chiller is compact.
• Different types of pumps and valves come standard with the chiller to provide versatility.
• Variety of accessories are available to meet user’s different lab environment needs. (refer to accessories)
Testing in accordance with DIN EN 12876Optional Accessories •Fittings,Adapters
see page 78-79 •Tubing,Insulations •Tubeclamps •Bathfluid
Recirculating Coolers (Chillers)
LCD Display
Pressure Gauge Level Indicator
73
Pumping Capacities (Bath fluid : water)
Centrifugal Magnetic
Specifications & Ordering Informations
Model HX-20 HX-25 HX-20H HX-25H HX-35H HX-45H HX-55H
Powerful recirculating coolers, the HL series provide a wide range of cooling capacities with temperature reliability. Innovative control system, with a dynamic pumping capacity, and ideal for external temperature applications. Broad range of models to choose from based on working temperature range.
Testing in accordance with DIN EN 12876Optional Accessories •Fittings,Adapters
see page 78-79 •Tubing,Insulations •Tubeclamps •Bathfluid
Recirculating Coolers (Chillers)
Operating features • Temperature range is from -20℃~40℃.
• Bright LED water level indicator can be easily identified.
• Low fluid level and dry-running protection when triggered shows an error symbol on the display and an audible sound.
• If the equipment operates at 40℃ or above due to the external environment, user can simply identify the error with a visual or an audible sound.
• Designed to simplify operation with a touch screen type display. - Easy-to-read VFD display with interactive touch keys. - Setting and resolution indication 0.1℃/℉. - Quick keypad lock prevents accidental parameter changes. - Signal indicator for operation status.
Constructional features • The refrigerating system is designed to reduce energy consumption
and to maximize in refrigerating efficiency.
• Eco-friendly R-404A / R507 refrigerant is used to prevent enviro-nmental pollution. Double service valve is used to prevent refrigerant leakage.
• Removable gridded vent for reducing condenser contamination and to facilitate cleanliness.
• A wide (Ø120mm / 4.72″) solution inlet is for easy and safe pouring of solution into the recirculating cooler.
• Pressure gauge comes standard for observing the pump pressure
• The refrigeration system can be observed easily to diagnostic at a glance by high / low pressure gauges. (HL-25H, 35H, 45H, 55H models)
• Wheels allow for easy mobility.
• The beautifully designed and compact.
• Different types of pumps come standard with the recirculating cooler to provide versatility.
• Warning alarm/monitoring system. - Over temperature. - Lack of bath solutions.
• Variety of accessories is available to meet user’s different lab environment needs. (Refer to accessories)
Low Temp. Recirculating Coolers - Intelligent Type
Adjustable, precise PID temperature controller beneficial for various cooling tasks, for use in the science, research, and industrial laboratories.Innovative and exclusive controller can be adjusted the cooling capacity from 1% to 100%
Operating features • Temperature range is from -20℃~40℃. • As an option an external sensor can be directly installed into the
cooling circuit for monitoring and regulating of temperatures. • Precise temperature stability ± 0.2℃ (at 15℃) without heater. • Bright LED water level indicator can be easily identified. • Low fluid level and dry-running protection when triggered shows
an error symbol on the display and an audible sound. • If the equipment operates at 40℃ or above due to the external
environment, user can simply identify the error with a visual or an audible sound.
• Designed to simplify operation with a touch screen type display. - Easy-to-read VFD display with interactive touch keys - Setting and resolution indication 0.1℃/℉ - Quick keypad lock prevents accidental parameter changes - Signal indicator for operation status
Constructional features • The refrigerating system is designed to reduce energy consumption
and to maximize in refrigerating efficiency. • Available external temperature sensor (Pt 100) for measured value • Computer Interface -Software provided -Save data in excel format -Storage of program information, backup of value settings/recovery,
and temperature values -RS-232C comes standard • Eco-friendly R-404A / R507 refrigerant is used to prevent enviro-
nmental pollution. Double service valve is used to prevent refrigerant leakage.
• Removable gridded vent for reducing condenser contamination and to facilitate cleanliness.
• A wide (Ø120mm / 4.72″) solution inlet is for easy and safe pouring of solution into the recirculating cooler.
• Pressure gauge comes standard for observing the pump pressure • The refrigeration system can be observed easily to diagnostic at a
glance by high / low pressure gauges. (HL-25H, 35H, 45H, 55H models)
• The beautifully designed and compact. • Different types of pumps come standard with the recirculating
cooler to provide versatility • Warning alarm/monitoring system - Over temperature - Lack of bath solutions
• Variety of accessories is available to meet user’s different lab env-ironment needs. (Refer to accessories)
Barbed fittings Barbed fitting systems are geared for quick connecting of various size type tubes using pipe fittings, tube fittings, and hose fittings without tools.
Connectors / AdaptersCat. No. DescriptionHXE1105 3/4”Maleto3/4”FemaleHXE1106 3/4”Maleto5/8”FemaleHXE1107 3/4”Maleto1/2”FemaleHXE1108 3/4”Maleto3/8”FemaleHXE1109 3/4” Male to 3/4” MaleHXE1110 3/4” Male to 5/8” MaleHXE1111 3/4” Male to 1/2” MaleHXE1112 3/4” Male to 3/8” Male
One touch adapters set Connect and exchange rigid tubing easily.