Top Banner
Recent Updates in the CAL Kim Attig, Nina Gartman, Lee Green, and Tracy Dombek National Atmospheric Deposition Program Central Analytical Laboratory (CAL) Illinois State Water Survey Institute of Natural Resource Sustainability University of Illinois at Urbana-Champaign Champaign, IL 61820; [email protected] Abstract In the past five years, important updates have taken place in the Central Analytical Laboratory (CAL). In the fall of 2007, the bucket cleaning room was moved to a refurbished and expanded area, which allowed the CAL to purchase and start using two new washers for cleaning buckets, lids, and bottles. In the summer of 2010, a new flow injection analysis instrument (FIA) was purchased. The new instrument is capable of doing low flow analysis as well as the method currently run at the CAL. Both methods on the new instrument were tested and results compared to the current method on the original FIA. In the summer of 2011, a new system and autosampler was purchased for ion chromatography (IC). With the AS-40 autosamplers which are currently used, the new ICS-5000 system is being tested and the results compared with the ICS-2000 systems. The new AS-AP autosampler was also tested on the ICS-2000 systems and results compared to the ICS-2000 systems with the AS- 40 autosamplers. Purchases such as these new instruments are important developments and show that the CAL is striving to progress and move forward along with the changing technology, while at the same time maintaining the standards of accuracy and precision which the lab already adheres to. The NADP a is National Research Support Project-3: A Long-Term Monitoring Program in Support of Research on the Effects of Atmospheric Chemical Deposition. More than 240 sponsors support the NADP, including State Agricultural Experiment Stations; universities; private companies and other non-governmental organizations; Canadian government agencies; state, local, and tribal government organizations; and federal agencies, including the U.S. Department of Agriculture-Cooperative State Research, Education, and Extension Service (under agreement no. 2008-39134-19508). Any findings or conclusions in this poster do not reflect the views of the U.S. Department of Agriculture or other NADP sponsors. The Illinois State Water Survey is a division of the Institute of Natural Resource Sustainability at the University of Illinois. Thanks to Pam Bedient and Chris Lehmann for poster formatting and review. New Bucket Washers Two new washers started being used in fall of 2007. Four buckets can fit in each washer at one time, which doubled our total washing capacity. The washers feature heated wash and rinse cycles. New FIA A new FIA was purchased in summer of 2010 to replace the aging FIA. New IC and autosampler A new ICS-5000 system and AS-AP autosampler were purchased in the summer of 2011. Conclusions Testing of the ICS-5000 system is ongoing. The AS-AP autosampler may be replaced with a model that utilizes our current vials. We suspect that the background measured in the vials specifically designed for use with the AS-AP is not within our acceptable limits. The new FIA is working well with the current method. The bucket washers work well! R² = 0.9998 0 0.5 1 1.5 2 2.5 0 0.5 1 1.5 2 2.5 New FIA Ammonium (ppm) Old FIA Ammonium (ppm) New FIA Ammonium vs Old FIA Ammonium MDL 0.006 ppm R² = 0.9953 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 New FIA Orthophosphate (ppm) Old FIA Orthophosphate (ppm) New FIA Orthophosphate vs Old FIA Orthophosphate MDL 0.008 ppm Median 95th percentile concentration 0.045 ppm Using the low flow method, the baseline drifted drastically over the timespan of a few samples. The baseline of a normal run with the current method used for ammonium is stable and does not drift. R² = 0.9987 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 ICS-5000 chloride (ppm) ICS-2000 chloride (ppm) ICS-5000 vs ICS-2000 Chloride MDL 0.005 ppm R² = 0.9995 0 1 2 3 4 5 6 7 8 9 10 0 1 2 3 4 5 6 7 8 9 10 ICS-5000 nitrate (ppm) ICS-2000 nitrate (ppm) ICS-5000 vs ICS-2000 Nitrate MDL 0.004 ppm R² = 0.9990 0 1 2 3 4 5 6 7 8 0 1 2 3 4 5 6 7 8 ICS-5000 sulfate (ppm) ICS-2000 sulfate (ppm) ICS-5000 vs ICS-2000 Sulfate MDL 0.003 ppm The ICS-5000 is being tested and results are comparable to the ICS-2000 using the AS-40 autosampler. When analyzing DI water in the vials manufactured for the ASAP autosampler, the background of sulfate and chloride routinely exceed our current detection limits. The new FIA was tested and results were found to be comparable and reliable using the same methods as the old FIA. No significant statistical differences were observed between the instruments. Low flow analysis was investigated on the new instrument, but the results indicated a loss of sensitivity with this method. ICS-5000 (2011) 2 Models ICS-2000 (2004) QuikChem 8500 (2010) QuikChem 8000 (1996) replaced in 2010 Forma Fury Model #8887 (1982) replaced in 2007 Thermo Forma Fury Model #8891 MDL’s Cl - 0.005 ppm NO 3 - 0.004 ppm SO 4 2- 0.003 ppm 0 0.001 0.002 0.003 0.004 0.005 0.006 0.007 0.008 Cl SO4 NO3 amount of analyte detected (ppm) analyte DI Water Samples Analyzed Using the ASAP Autosampler CAL Staff (1982) replaced in 2010 CAL Staff (2011)
1

Recent Updates in the CAL - nadp.slh.wisc.edunadp.slh.wisc.edu/conf/2011/posters/attig.pdf · 2011, a new system and autosampler was purchased for ion chromatography (IC). With the

Sep 12, 2018

Download

Documents

vuongtu
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Recent Updates in the CAL - nadp.slh.wisc.edunadp.slh.wisc.edu/conf/2011/posters/attig.pdf · 2011, a new system and autosampler was purchased for ion chromatography (IC). With the

Recent Updates in the CAL Kim Attig, Nina Gartman, Lee Green, and Tracy Dombek

National Atmospheric Deposition Program

Central Analytical Laboratory (CAL)

Illinois State Water Survey

Institute of Natural Resource Sustainability

University of Illinois at Urbana-Champaign

Champaign, IL 61820; [email protected]

Abstract

In the past five years, important updates have taken place in the Central Analytical Laboratory

(CAL). In the fall of 2007, the bucket cleaning room was moved to a refurbished and

expanded area, which allowed the CAL to purchase and start using two new washers for

cleaning buckets, lids, and bottles. In the summer of 2010, a new flow injection analysis

instrument (FIA) was purchased. The new instrument is capable of doing low flow analysis as

well as the method currently run at the CAL. Both methods on the new instrument were

tested and results compared to the current method on the original FIA. In the summer of

2011, a new system and autosampler was purchased for ion chromatography (IC). With the

AS-40 autosamplers which are currently used, the new ICS-5000 system is being tested and

the results compared with the ICS-2000 systems. The new AS-AP autosampler was also

tested on the ICS-2000 systems and results compared to the ICS-2000 systems with the AS-

40 autosamplers. Purchases such as these new instruments are important developments and

show that the CAL is striving to progress and move forward along with the changing

technology, while at the same time maintaining the standards of accuracy and precision which

the lab already adheres to.

The NADP a is National Research Support Project-3: A Long-Term Monitoring Program in Support of Research on the Effects of

Atmospheric Chemical Deposition. More than 240 sponsors support the NADP, including State Agricultural Experiment Stations; universities;

private companies and other non-governmental organizations; Canadian government agencies; state, local, and tribal government organizations;

and federal agencies, including the U.S. Department of Agriculture-Cooperative State Research, Education, and Extension Service (under

agreement no. 2008-39134-19508). Any findings or conclusions in this poster do not reflect the views of the U.S. Department of Agriculture or other

NADP sponsors. The Illinois State Water Survey is a division of the Institute of Natural Resource Sustainability at the University of Illinois.

Thanks to Pam Bedient and Chris Lehmann for poster formatting and review.

New Bucket Washers

Two new washers started

being used in fall of 2007.

Four buckets can fit in each

washer at one time, which

doubled our total washing

capacity.

The washers feature

heated wash and rinse

cycles.

New FIA

A new FIA was purchased in summer of 2010 to replace the aging FIA.

New IC and autosampler

A new ICS-5000 system and AS-AP autosampler were purchased in the summer of 2011.

Conclusions

Testing of the ICS-5000 system is ongoing.

The AS-AP autosampler may be replaced with a

model that utilizes our current vials. We suspect

that the background measured in the vials

specifically designed for use with the AS-AP is not

within our acceptable limits.

The new FIA is working well with the current

method.

The bucket washers work well!

R² = 0.9998

0

0.5

1

1.5

2

2.5

0 0.5 1 1.5 2 2.5

New

FIA

A

mm

on

ium

(p

pm

)

Old FIA Ammonium (ppm)

New FIA Ammonium vs Old FIA Ammonium

MDL 0.006 ppm

R² = 0.9953

0

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16

New

FIA

Ort

ho

ph

osp

hate

(p

pm

)

Old FIA Orthophosphate (ppm)

New FIA Orthophosphate vs Old FIA Orthophosphate

MDL 0.008 ppm Median 95th percentile

concentration

0.045 ppm

Using the low flow method, the baseline drifted

drastically over the timespan of a few samples.

The baseline of a normal run with the current method

used for ammonium is stable and does not drift.

R² = 0.9987

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5

ICS

-5000 c

hlo

rid

e (

pp

m)

ICS-2000 chloride (ppm)

ICS-5000 vs ICS-2000 Chloride

MDL 0.005 ppm

R² = 0.9995

0

1

2

3

4

5

6

7

8

9

10

0 1 2 3 4 5 6 7 8 9 10

ICS

-5000 n

itra

te (

pp

m)

ICS-2000 nitrate (ppm)

ICS-5000 vs ICS-2000 Nitrate

MDL 0.004 ppm

R² = 0.9990

0

1

2

3

4

5

6

7

8

0 1 2 3 4 5 6 7 8

ICS

-5000 s

ulf

ate

(p

pm

)

ICS-2000 sulfate (ppm)

ICS-5000 vs ICS-2000 Sulfate

MDL 0.003 ppm

The ICS-5000 is being tested and results are comparable to the ICS-2000 using the AS-40 autosampler.

When analyzing DI water in the vials manufactured for

the ASAP autosampler, the background of sulfate and

chloride routinely exceed our current detection limits.

The new FIA was tested and results were found to be comparable and reliable

using the same methods as the old FIA. No significant statistical differences were

observed between the instruments.

Low flow analysis was investigated on the new instrument, but the results indicated

a loss of sensitivity with this method.

ICS-5000 (2011) 2 Models ICS-2000 (2004)

QuikChem 8500 (2010) QuikChem 8000 (1996) replaced in 2010

Forma Fury Model #8887 (1982) replaced in 2007 Thermo Forma Fury Model #8891

MDL’s

• Cl- 0.005 ppm

• NO3- 0.004 ppm

• SO42- 0.003 ppm

0

0.001

0.002

0.003

0.004

0.005

0.006

0.007

0.008

Cl SO4 NO3

am

ou

nt

of

an

aly

te d

ete

cte

d (

pp

m)

analyte

DI Water Samples Analyzed Using the ASAP Autosampler

CAL Staff (1982) replaced in 2010

CAL Staff (2011)