IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 22: Optimization of Protection in Dental Radiology IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Aug 11, 2015
IAEAInternational Atomic Energy Agency
RADIATION PROTECTION INDIAGNOSTIC AND
INTERVENTIONAL RADIOLOGY
L 22: Optimization of Protection in Dental Radiology
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 22: Optimization of Protection in Dental Radiology 2
Introduction
• Dental radiology makes use of specific types of equipment, needed for different purposes. Frequent exposures though each with low dose involve a risk for the practitioner and for the patient
• Background: general principles of x-ray diagnostic imaging
IAEA 22: Optimization of Protection in Dental Radiology 3
Dental X-ray equipment
Radiation protection in dental radiology
Quality control for dental equipment
Topics
IAEA 22: Optimization of Protection in Dental Radiology 4
• To be able to apply the principle of radiation protection to dental radiology system including design and Quality Control.
Overview
IAEAInternational Atomic Energy Agency
Part 22: Optimization of protection in dental radiology
Topic 1: Dental x-ray equipment
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 22: Optimization of Protection in Dental Radiology 6
Types of units
• “Intra-Oral” units• Standard dental tube
• uses an intra-oral image receptor
• has extra-oral x-ray tube
• Panoramic (OPG)
• Cephalometric ( Ceph)
IAEA 22: Optimization of Protection in Dental Radiology 7
Intra-Oral Dental X-Ray Equipment
IAEA 22: Optimization of Protection in Dental Radiology 8
Modern Dental X-Ray Unit
IAEA 22: Optimization of Protection in Dental Radiology 9
Panoramic X-Ray Equipment
IAEA 22: Optimization of Protection in Dental Radiology 10
Cephalometric X-Ray Equipment
IAEA 22: Optimization of Protection in Dental Radiology 11
X-Ray Tube
• stationary Anode
• avoid overheating
• tube duty cycle:• typical: 1:30 intaroral
• 1:10 OPG
• 420 mAs/hr intraoral
IAEA 22: Optimization of Protection in Dental Radiology 12
Tube Head
IAEA 22: Optimization of Protection in Dental Radiology 13
Generator Circuit
IAEA 22: Optimization of Protection in Dental Radiology 14
Generators & Pre-Heat
• Medium frequency - stable waveform
• Single phase (SP) - pulsed
• Pre-Heat: separate circuit for heating filament
• Single Phase units without a pre-heat circuit • initial pulses of variable kV
IAEA 22: Optimization of Protection in Dental Radiology 15
Collimator
1. Lead Collimator
with central hole
2. Spacer Tube
IAEA 22: Optimization of Protection in Dental Radiology 16
Applicator Cones
Good Bad Bad
IAEA 22: Optimization of Protection in Dental Radiology 17
Cephalometric Holder
IAEA 22: Optimization of Protection in Dental Radiology 18
Intra-Oral Dental X-Ray Equipment (technical data)
Exposure time from 60 ms to 2.5 s
Tube Min. 50 kV, ~7mA
Focal spot size 1 mm
Inherent filtration ~2 mm Al equivalent
Focus-skin distance 20 cm
Irradiated field 28 cm2 with round section, 6 cm
diameter collimator
IAEA 22: Optimization of Protection in Dental Radiology 19
Panoramic X-Ray Equipment (technical data)
Focal spot 0.5 mm
kV 60 - 80 kV in 2 kV steps
mA 4 - 10 mA steps 4, 5, 6, 8, 10
Exposure time 12 s (standard projections)
0.16 - 3.2 s (cephalometric projections)
Flat panoramic cassette 15x30 cm (Lanex Regular
screens))
IAEA 22: Optimization of Protection in Dental Radiology 20
Image Receptors in Dental Radiology
• Small films (2 x 3 or 3 x 4 cm) in light-tight envelopes (no screen)
• Digital intraoral sensors - compared with category E film, the radiation dose is reduced by 60%.
Intraoral Radiology
Panoramic Radiology and Cephalometry• Film-screen combination• Digital sensors - compared with film-
screen sensitivity class 200, the radiation dose is reduced by 50-70%.
IAEA 22: Optimization of Protection in Dental Radiology 21
Dental Radiology Film Types
Sensitivity class D• Very good spatial resolution• Typical delivered dose: about 0.5 mGy• Typical exposure times: 0.3 - 0.7 s
Sensitivity class E• Good spatial resolution• Typical delivered dose: about 0.25 mGy• Typical exposure times: 0.1 - 0.3 s
IAEAInternational Atomic Energy Agency
Part 22: Optimization of Protectionin Dental Radiology
Topic 2: Radiation Protection in Dental Radiology
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 22: Optimization of Protection in Dental Radiology 23
Facts Very frequent examination (about 25% of all
the radiological examinations)
Delivered doses may differ of a factor 2 or 3. (entrance doses between 0.5 and 150 mGy)
Image Quality often very low
Organs at risk: parathyroid, thyroid, larynx, parotid glands
Radiation Protection in Dental Radiology
IAEA 22: Optimization of Protection in Dental Radiology 24
Keep under control time and temperature of the developing process.
Do not use oxydized chemicals
Do not adjust development time by viewing the film
Technical hints to reduce patient dosesQuality Control of Film Processing
Radiation Protection in Dental Radiology
IAEA 22: Optimization of Protection in Dental Radiology 25
Sensitivity class D• Very good spatial resolution• Typical delivered dose: about 0.5 mGy• Typical exposure times: 0.3 - 0.7 s
Technical hints to reduce patient dosesPrivilege the use of sensitive films
Sensitivity class E• Good spatial resolution• Typical delivered dose: about 0.25 mGy• Typical exposure times: 0.1 - 0.3 s
RP in Dental Radiology
IAEA 22: Optimization of Protection in Dental Radiology 26
Technical hints to reduce patient dosesLead apron and collar
Useful when the path of primary beam intercepts the protected organs (downward bite-twin projection).
Radiation Protection in Dental Radiology
IAEA 22: Optimization of Protection in Dental Radiology 27
Panoramic examination
• Image quality not as good as in intra-oral films
• Important global information
• Relatively low dose(one panoramic examination 35 intra-oral films)
Radiation Protection in Dental Radiology
IAEAInternational Atomic Energy Agency
Part 22: Optimization of Protection in Dental Radiology
Topic 3: Quality Control for Dental Equipment
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
IAEA 22: Optimization of Protection in Dental Radiology 29
Why Dental QC ?
• Widespread use of dental units
• Lack of QC history on most units
• Dental practitioners working in the primary health care sector do not have the continuous medical physics support available in a hospital-based diagnostic imaging department
IAEA 22: Optimization of Protection in Dental Radiology 30
What Tests ?
• Collimation
• Dose Evaluation
• Exposure Time
• Half Value Layer
• Kilovoltage (kVp)
• Leakage Radiation
IAEA 22: Optimization of Protection in Dental Radiology 31
The recommended tests are consequently divided into:
those simple tests which can be performed by dental practice staff
those more complex tests which can be carried out by medical physicists.
Quality Control for Dental Equipment
IAEA 22: Optimization of Protection in Dental Radiology 32
Quality Control for Dental Equipment
Tests which can be performed bydental practice staff
Physical parameter Tolerances Frequency
Image Quality ±10% reference values
Quarterly
Developer temperature and condition of processing solutions
Specified by the film manufacturer
Every time processing solutions are used
Processing Base+Fog: >0.2 ODSpeed and Contrast >0.15 OD about baseline
Every time processing solutions are changed
IAEA 22: Optimization of Protection in Dental Radiology 33
Quality Control for Dental Equipment
Tests performed by medical physicists
Physical parameter Tolerances Frequency
Tube voltage >50 kV and error <10%
3 yearly
Beam size/collimation <60 mm diameter (intra-oral) <150 x 10 mm at cassette (panoramic)
3 yearly
Dose at cone tip 50 kV: <5.0 mGy70 kV: <2.5 mGy(E speed film)
1-3 yearly
Dose-width product for panoramic film
<75 mGy mm 1-3 yearly
IAEA 22: Optimization of Protection in Dental Radiology 34
Dental QC Methods
UnitIntra-Oral Receptors (I/O)
Cephalometric (Ceph)
Panoramic (OPG )
Test Methodas for Radiology QC
as for Radiology QC
where possible:• immobilise unit• remove slit collimator
IAEA 22: Optimization of Protection in Dental Radiology 35
Test Equipment
• kVp meter• measure kVp
average
• Programmable delay ~100 ms
• Range: 50 to 120 kV
• Aluminum filters • 4 x 1mm
• Grade 1100
• Dosimeter • small & large volume
chambers
• 2mm wide detector for OPG
• Timer • triggering at 75%
peak kV
IAEA 22: Optimization of Protection in Dental Radiology 36
• Expose film or fluorescent screen
• Measure x-ray field image
Collimation
Unit Film position Limits of X Ray field
Standard End of Cone 60 mm diameter
OPG Slit, in film cassette (2 films)
Equal slit images
Vertical image ± film height
Ceph At cassette ±1 % of FFD on all margins
IAEA 22: Optimization of Protection in Dental Radiology 37
Collimator Light Intensity
Ceph units:
• Place external detector 1m from focus
• Measure illuminance in lux ( lumen/m2)
• Read each quadrant
• Limit: >100 lux at 1m
IAEA 22: Optimization of Protection in Dental Radiology 38
“Dead man” Switch
• timer at 50 cm from focus
• set low kV, mA, long time
• start exposure
• release switch during exposure
Require exposure cut-out when switch is released. Check exposure time is less than set time
IAEA 22: Optimization of Protection in Dental Radiology 39
Kilovoltage
Accuracy:
• Set kVp meter to ~100 msec delay
• Observe kVp waveform at 70 kV if poss.
• Limit: measured kVp within 5% of set value
Reproducibility:
• Take 5 repeat exposures
• Limit: coefficient of variation ≤ 2%
IAEA 22: Optimization of Protection in Dental Radiology 40
Dose Evaluation
Skin dose from I/O units:
• place cone 10 mm from dosimeter
• set maxillary molar/ bitewing setting
• Should be (65-70 kVp):
2-3 mGy for molar view
< 5 mGy for any view
IAEA 22: Optimization of Protection in Dental Radiology 41
Output Reproducibility (1)
Standard I/O units & Ceph units:• Dosimeter position:
• I/O units ~10 mm from cone
• Ceph units: 75 cm from focus
or other recommended distance
• Five repeat exposures
Limit: coefficient of variation ≤ 5%
IAEA 22: Optimization of Protection in Dental Radiology 42
Output Reproducibility (2)
Optional Method for OPG units:
• Align detector on film cassette slit
• Measure dose rate
• Take 5 repeat exposures
Limit: coefficient of variation ≤ 5%
IAEA 22: Optimization of Protection in Dental Radiology 43
Exposure Time Accuracy
Standard I/O units & Ceph units: • Set timer to trigger at 75% peak kV
• Test times in the normal working range
Limit: ≤10 % error for I/O units
≤ 5% error for all other units
IAEA 22: Optimization of Protection in Dental Radiology 44
Timer Reproducibility
Standard I/O units & Ceph units
• Place timer in beam
• 5 repeat exposures
Limit: coefficient of variation ≤5%
IAEA 22: Optimization of Protection in Dental Radiology 45
Half Value Layer (HVL)
Standard I/O units:• Position cone facing down
• Place dosimeter at ~ 40 cm from focus
• Position Al filters near end of cone
• Measure dose• measure with no added filters
• with 2,3,4 mm Al added, then again with no filters
• Plot on semi-log paper and find HVL
IAEA 22: Optimization of Protection in Dental Radiology 46
HVL Set-up
IAEA 22: Optimization of Protection in Dental Radiology 47
HVL: Ceph & OPG Units
• Position Al filters on collimator
Cephalometric units
• Position dosimeter at 75 cm from focus
OPG Units
• Position dosimeter on film cassette slit
• Measure dose rate, dose for fixed exposure time, or dose for full scan
NB Test kVp accuracy before measuring HVL
IAEA 22: Optimization of Protection in Dental Radiology 48
OPG Quality Control (kVp/HVL measurement)
IAEA 22: Optimization of Protection in Dental Radiology 49
kVp HVL (mm Al) Intraoral Ceph/OPG
60 1.5 1.8
70 1.5 2.1
80 2.3 2.3
90 2.5 2.5
HVL- Minimum Values
IAEA 22: Optimization of Protection in Dental Radiology 50
Leakage Radiation
• Cover collimator with 2 mm lead• Set large ionization chamber (~180 cc)
or solid state detector at known distance d cm from focus
• Expose on maximum kVp for 0.5 - 1 sec (but be careful!)
• Scale up the reading to 1 hour
IAEA 22: Optimization of Protection in Dental Radiology 51
Leakage Radiation
• Choose the most likely points for leakage - • where collimator joins to housing, lateral to
the tube, joints in housing etc.
• Normalise measurement to 1 m from focus with inverse square law:
Dose1m = dosed x (distanced/100)2
IAEA 22: Optimization of Protection in Dental Radiology 52
Leakage Measurement
• In normal radiography, leakage measurements are normalised to the maximum continuous rated tube current R: Leakage = Dose (for P mAs at 100 cm) x (R x
3600/P)
BUT…...a dental unit has no continuous operation rating
IAEA 22: Optimization of Protection in Dental Radiology 53
Leakage Measurement
• Instead we normalize the dose reading to tube duty cycle:
• Leakage Limit at 1 m from focus:
≤0.25 mGy/hour - Standard I/O units
≤1.0 mGy/hour - Ceph & OPG units
IAEA 22: Optimization of Protection in Dental Radiology 54
Calculation of Leakage
• Measured leakage at 100 cm = 2mGy/hr
• Duty cycle = 1: 20 (i.e. must wait ~ 20 times the exposure time between exposures to allow tube cooling)
• Thus actual leakage = 2 / 20 mGy/hr
= 0.1 mGy/hr
i.e. acceptable
IAEA 22: Optimization of Protection in Dental Radiology 55
Radiation Protection in Dental Radiology
• Keep under control time and temperature of the developing process.
• Do not use oxidized chemicals
• Regularly check processing with phantom
Quality Control of Film Processing
IAEA 22: Optimization of Protection in Dental Radiology 56
Dental Phantom
IAEA 22: Optimization of Protection in Dental Radiology 57
Dental Phantom
IAEA 22: Optimization of Protection in Dental Radiology 58
• Code of practice for protection of persons against ionising radiations arising from medical and dental use, DHSS, HMSO, London, 1964
• Guidance notes for the protection of persons against ionising radiations arising from medical and dental use, DHSS, HMSO, London, 1988
• Radiation protection and quality assurance in dental radiology. Radiation protection 81. European Commission.(1995) CG-89-95-971-EN-C
Where to Get More Information
IAEA 22: Optimization of Protection in Dental Radiology 59
Summary
• General and specific types of x-ray equipment (and image receptors) are reviewed: (panoramic and cephalometric), with technical data about operating conditions
• Although doses are generally low, the high frequency of examinations requires radiation protection (for the practitioner) in dental radiology
• Some tests are detailed for Quality control of dental equipment.