Page 1: Radiation Protection

Rakesh C A

Page 2: Radiation Protection

Objectives of radiation protection

• The International commission of Radiation protection (ICRP) Stated that “the overall objectives of radiation protection is to provide an appropriate standard of protection for man without unduly limiting the beneficial practices giving rise to radiation exposure”.

• NCRP (1993)- “The goal of radiation protection is to prevent the occurrence of serious radiation induced conditions in exposed persons & to reduce stochastic effects in exposed persons to a degree that is acceptable in relation to the benefits to the individual & society from activities that generate such exposure”.

Page 3: Radiation Protection

Protection

• Why?• From What?• Whom to protect?• How to protect?

Page 4: Radiation Protection

Protection

• When to protect?

Page 5: Radiation Protection

MEASUREMENTS!!!

Page 6: Radiation Protection

RADIATION UNITS

• ROENTGEN– unit of radiation exposure that will liberate a charge of 2.58x10-4coulombs/kg of air.

• Independent of the area or field size

Page 7: Radiation Protection

Absorbed dose • Deposition of energy in pt by radiation exposure

• Independent of composition of irradiated material and energy of beam

• RAD: unit of absorbed dose

• GRAY: SI unit of absorbed dose

• Gray defined as the quantity of radiation that results in an energy deposition of 1 joule per kilogram.

• I GRAY = 100 RAD• 1RAD = 1 cGY

Page 8: Radiation Protection

Dose equivalent

• It is a measure of biological effectiveness of radiation• REM: unit of absorbed dose equivalent• SIEVERT : SI unit

• 1 sievert = 100 rems

• Dose equivalent=Absorbed dose x QF.• REM = RADS X QUALITY FACTOR

Page 9: Radiation Protection

Quality factor• It is the parameter used to describe the quality of beam.• Gives the amount of energy deposited per unit length travel.

Expressed in KEV per micron.

Type of radiation Q factorX rays 1 Gamma rays 1Beta particle 1Electrons 1Thermal neutrons 5Other neutrons 20Protons 20 Alpha particle 20

Page 10: Radiation Protection

EFFECTIVE DOSE EQUIVALENT

• Purpose – to relate exposure to risk• It is calculated by multiplying the dose equivalent

received by each individual organ or tissue (DT) by an appropriate tissue weighting factor (WT) and summing for all the tissues involved.

Page 11: Radiation Protection

o b ta in th eE F F EC TIVE D O SE

to th e p t in m sv

sum o f a ll th e o rg an s an d tissues irrad ia ted

m ultip ly b y th eTIS S U E W E IG H IN G F AC TO R W t

fo r th e tissue o r o rg an co n cern ed

E Q U IVAL E N T D O S Eto th e o rg an in m sv

m ultip ly b y th eR AD IATIO N W E IG H IN G F AC TO R W r

O R Q U AL ITY F AC TO Rfo r th e rad ia tio n used

fo r each o rg an an d tissue es tim ate th eAB S O R B E D D O SE

in m g y

Page 12: Radiation Protection

Protection

• Why?• From What?• Whom to protect?• How to protect?

Page 13: Radiation Protection

Is there RADIATION in this room?

Page 14: Radiation Protection
Page 15: Radiation Protection

Radiation - We live with

Natural Radiation: Cosmic rays, radiation within our body, in food we eat, water we drink, house we live in, lawn, building material etc.

Human Body: K-40, Ra-226, Ra-228e.g. a man with 70 kg wt. 140 gm of K

140 x 0.012%0.0168 gm of K-40

0.1 Ci of K-40

Page 16: Radiation Protection

Radiation - We live with

Gy/yrNew Delhi 700Bangalore 825Bombay 424Kerala 4000(in narrow coastal strip)

Page 17: Radiation Protection

Radiation – We travel with

Page 18: Radiation Protection

Radiation - We eat withFood Radioactive levels (Bq/kg)

Daily intake (g/d)

Ra-226 Th-228 Pb-210 K-40

Rice 150 0.126 0.267 0.133 62.4

Wheat 270 0.296 0.270 0.133 142.2

Pulses 60 0.233 0.093 0.115 397.0

Other Vegetables

70 0.126 0.167 -- 135.2

Leafy Vegetables

15 0.267 0.326 -- 89.1

Milk 90 -- -- -- 38.1

Composite Diet

1370 0.067 0.089 0.063 65.0

Dose equivalent=0.315 mSv/yrTotal dose from Natural sources = 1.0 to 3.0 mSv/yr

Page 19: Radiation Protection
Page 20: Radiation Protection

Natural and Manmade sources

Page 21: Radiation Protection

SOURCES OF RADIATION

• Natural radiation: 1. External: Cosmic and gamma radiation2. Internal: radionuclides with in the body

ingested or inhaled• Medical procedures:

1. Diagnostic 2. Therapeutic

• Nuclear weapons/industry/accidents

Page 22: Radiation Protection

Electromagnetic Waves

Low HighENERGY

Radio waves

Microwaves

Radar

Infrared

Visible light

Ultra-violet

X-ray

Gamma-ray

Non-ionizing radiation

Ionizing radiation

Page 23: Radiation Protection

Primary Types of Ionizing Radiation

• Alpha particles• Beta particles• Gamma rays (or

photons)• X-Rays (or photons)• Neutrons

Ionizing Radiation

alpha particle

beta particle

Radioactive Atom

X-ray

gamma ray

Page 24: Radiation Protection

Direct Ionization Caused By:

• Protons

• Alpha Particles

• Beta Particles

• Positron Particles

Page 25: Radiation Protection

Indirect Ionization Caused By:

• Neutrons

• Gamma Rays

• X-Rays

Page 26: Radiation Protection

DO WE NEED RADIATION

PROTECTION ?

Page 27: Radiation Protection
Page 28: Radiation Protection

Radiation health effects

CELL DEATH BOTH

TYPEOF

EFFECTS

CELL TRANSFORMATION

Page 29: Radiation Protection

Radiation health effects

DETERMINISTICSomaticClinically attributable in the exposed individual

CELL DEATH

STOCHASTICsomatic & hereditaryepidemiologically attributable in large populations

ANTENATALsomatic and hereditary expressed in the foetus, in the live born or descendants

BOTH

TYPEOF

EFFECTS

CELL TRANSFORMATION

Page 30: Radiation Protection

• Deterministic

(Threshold/non-stochastic)• Existence of a dose threshold

value (below this dose, the effect is not observable)

• Severity of the effect increases with dose

• A large number of cells are involved

Radiation injury from an industrial source

Deterministic effects

Page 31: Radiation Protection

• Cataracts of the lens of the eye 2-10 Gy

• Permanent sterility

• males 3.5-6 Gy

• females 2.5-6 Gy

• Temporary sterility

• males 0.15 Gy

• females 0.6 Gydose

Severity ofeffect

threshold

Threshold Doses for Deterministic Effects

Page 32: Radiation Protection

Stochastic Effects

• Stochastic(Non-Threshold)– No threshold – Probability of the effect increases with dose– Generally occurs with a single cell– e.g. Cancer, genetic effects

Page 33: Radiation Protection
Page 34: Radiation Protection
Page 35: Radiation Protection
Page 36: Radiation Protection

How much

DNA is

repaired ?

Page 37: Radiation Protection

Repair of DNA damage

• RADIOBIOLOGISTS ASSUME THAT THE REPAIR SYSTEM IS NOT 100% EFFECTIVE.

Page 38: Radiation Protection

Outcomes after cell exposure

DAMAGE REPAIRED CELL NECROSIS OR APOPTOSIS

TRANSFORMED CELL

DAMAGE TO DNA

Page 39: Radiation Protection

10-6

10-12

10-9

10-15

10-3

1 second

1 hour

1 day

1 year

100 years

1 ms

100

109

106

103

Energy deposition

Excitation/ionization

Initial particle tracks

Radical formation

PHYSICAL INTERACTIONS

PHYSICO-CHEMICAL INTERACTIONS

BIOLOGICAL RESPONSE

MEDICAL EFFECTS

Diffusion, chemical reactions

Initial DNA damage

DNA breaks / base damage

Repair processesDamage fixation

Cell killing

Promotion/completion

TeratogenesisCancer

Hereditary defects

Proliferation of "damaged" cells

Mutations/transformations/aberrations

T IM

E (s

ec)

Timing of events leading to radiation effects.

Page 40: Radiation Protection

C H A IN O F EVEN TS FO LLO W IN G EXPO SU R E TO IO N IZ IN GR A D IA TIO N

CELL DEATHDETERM INISTIC EFFECTS

CELLULAR TRANSFO RM ATIO NM AY BE SO M E REPAIRSTO CHASTIC EFFECTS

CELLULAR LEVEL

SUBCELLULAR DAM AG E(M EM BRANES, NUCLEI, CHRO M O SO M ES)

m olecular changes(DNA,RNA, ENZYM ES)

free radicals(chem ical changes)

ionisation

exposure

Page 41: Radiation Protection

Radiosensitivity [RS]

• RS = Probability of a cell, tissue or organ of suffering an effect per unit of dose.

Page 42: Radiation Protection

RS laws (Law of Bergonie & Tribondeau)

Radiosensitivity of living tissues varies with maturation & metabolism;1. Stem cells are radiosensitive. More mature

cells are more resistant2. Younger tissues are more radiosensitive3. Tissues with high metabolic activity are highly

radiosensitive4. High proliferation and growth rate, high

radiosensitivty

Page 43: Radiation Protection

Radiosensitivity

Muscle

Bones

Nervous system

Skin

Mesoderm organs (liver, heart, lungs…)

Bone Marrow

Spleen

Thymus

Lymphatic nodes

Gonads

Eye lens

Lymphocytes (exception to the RS laws)

Low RSMedium RSHigh RS

Page 44: Radiation Protection

RADIATION EFFECTS

DETERMINISTIC EFFECT• Mechanism is cell killing• Has a threshold dose• Deterministic in nature• Severity increases with dose• Occurs only at high doses• Can be completely avoided • Causal relationship between

radiation exposure and the effect

• Sure to occur at an adequate dose

STOCHASTIC EFFECTMechanism is cell modificationHas no thresholdProbabilistic in natureProbability increases with doseOccurs at even at low dosesCannot be completely avoidedCausal relationship cannot be established at low dosesOccurs only among a small percentage of those exposed

Page 45: Radiation Protection

RADIATION EFFECTS

DETERMINISTIC EFFECT• Radiation Sickness• Radiation syndromes

– Haematopoietic syndrome– GI syndrome– CNS syndrome

• Damage to individual organs• Death• Late damage

STOCHASTIC EFFECT• Chromosomal damage• Cancer Induction (Several

years after exposure to radiation)

• Genetic Effects (Hereditary in future generations only)

• Somatic Mutations

Page 46: Radiation Protection

SO WE NEED RADIATION

PROTECTION!!!

Page 47: Radiation Protection

OBJECTIVES OF RADIATION PROTECTION

• PREVENTION of deterministic effect• LIMITING the probability of stochastic effect

HOW? Up to what point?

Page 48: Radiation Protection

We live with1-3 mSv

Can kill4000 mSv

Radiation

Where to stop, where is the safe point?

Page 49: Radiation Protection

mSv

Year

Changes in Dose Limit (ICRP)(Safe levels)

0

100

200

300

400

500

1931 1947 1977 1990

Page 50: Radiation Protection

Dose Limits (ICRP 60)Occupational Public

Effective dose 20 mSv/yr averaged* 1 mSv in a yrover 5 yrs.

Annual equivalentdose to• Lens of eye 150 mSv 5 mSv• Skin 500 mSv 50 mSv• Hands & Feet 500 mSv

* with further provision that dose in any single yr > 30 mSv (AERB) and =50 mSv (ICRP)

Page 51: Radiation Protection

WHAT ISBASIS FOR

DOSE LIMITS?

Page 52: Radiation Protection
Page 53: Radiation Protection

PRINCIPLESOF

RADIATIONPROTECTION

Page 54: Radiation Protection

PRINCIPLES OF RADIATION PROTECTION

1. Justification of practices2. Optimization of protection by

keeping exposure as low as reasonably achievable

3. Dose limitation

Page 55: Radiation Protection

Justification of procedure versus the net benefit

i.e. no practice involving exposures to radiation should be adopted unless it provides sufficient benefit to offset the detrimental effects of radiation.

Page 56: Radiation Protection

Optimization of protection

Protection should be optimized in relation to

the magnitude of doses,

number of people exposed

for all social and economic strata of patients.

Page 57: Radiation Protection

• Optimization of protection can be achieved by optimizing the procedure to administer a radiation dose which is

as low as reasonably achievable,

so as to derive maximum diagnostic information with minimum discomfort to the patient

Page 58: Radiation Protection

All doses should be kept

• As

• Low

• As

• Reasonably

• AchievableThe ALARA Principle

Page 59: Radiation Protection

HOW TO APPLYTHESE PRINCIPLES IN

DIAGNOSTIC RADIOLOGY?

Page 60: Radiation Protection

How much time one works with radiation?

RADIOGRAPHY

Page 61: Radiation Protection

Radiation ON Time

Workload=100 exposures/day

CxR = 50x50 m sec = 2500 = 2.5s

LS = 50x800 m sec = 40000=40s

Total time = 45 sec/day

Not greater than 1 min/day

Page 62: Radiation Protection

Staff Doses

Dose limit ICRP = 20 mSv/yr.

Radiography work 0.1 mS/yr.

i.e. 1/200th of dose limit

Page 63: Radiation Protection

Relative Dose Received

number of chest x-rays0 50 100 150 200

Arm, head,ankle & foot (1)Head & Neck (3)

Head CT (10)Thoracic Spine (18)Mammography, Cystography (20)

Pelvis (24)Abdomen, Hip, Upper & lower femur (28)

Ba Swallow (30)Obsteric abdomen (34)

Lumbo-sacral area (43)Cholangiography (52)

Lumber Myelography (60)Lower abdomen CT male (72)Upper Abdomen CT (73)Ba Meal (76)

Angio-head, Angio-peripheral (80)Urography (87)

Angio-abdominal (120)Chest CT (136)

Lower Abd. CT fem. (142)Ba enema (154)

Lymphan. (180)

mSv.050.150.490.921.01.221.41.51.72.152.593.03.613.673.84.04.366.06.87.137.699.0

Radiation Doses in Radiological Exam. (as multiple of chest x-ray)

Page 64: Radiation Protection

IS IT POSSIBLE TO GET DETERMINISTIC EFFECTS IN

RADIOGRAPHIC WORK ?

For staff, for patient..??

Page 65: Radiation Protection

Radiography

Risk of Staff Patient Public

DeathSkin burnInfertilityCataractCancerGenetic effect

××××UU

××××UU

××××UU

U: unlikely

Page 66: Radiation Protection

FLUOROSCOPYANDCT

Page 67: Radiation Protection

Fluoroscopy

Barium study: 3-6 min/pt x 8 patients/d= 40 min/d

ANGIOGRAPHY• Diagnostic = 50 min/d• Therapeutic = 2-5 hr/d

CT = 10-45 min/d

Page 68: Radiation Protection

Fluoroscopy (excl. ther angio)

Risk of Staff Patient Public

DeathSkin burnInfertilityCataractCancerGenetic effect

××××UU

××××UU

××××UU

U: unlikely

Page 69: Radiation Protection
Page 70: Radiation Protection

X-ray tube

Primary beam

Scattered radiation

Patient

Page 71: Radiation Protection

Radiation emitted by the X Ray tube

• Primary radiation: before interacting photons• Scattered radiation: after at least one interaction;• Leakage radiation: not absorbed by the X Ray tube

housing shielding• Transmitted radiation: emerging after passage

through matter

Page 72: Radiation Protection

X-ray Tube Position

• Position the X-ray tube under the patient not above the patient.

• The largest amount of scatter radiation is produced where the x-ray beam enters the patient.

• By positioning the x-ray tube below the patient, you decrease the amount of scatter radiation that reaches your upper body.

X-ray Tube

Image Intensifier

Page 73: Radiation Protection

• TUBE CURRENT• TUBE POTENTIAL• HIGH OR LOW Z TARGET MATERIAL• FILTRATION• TYPE OF WAVEFORM

FACTORS AFFECTING X Ray BEAM

Page 74: Radiation Protection

Tube current

• Determines the quantity of the photons which also contribute to the patient dose.

• Increased exposure time also contributes to

an increased patient dose.

Page 75: Radiation Protection

X Ray spectrum: tube current

Change of QUANTITYNO change of quality

Effective kV not changed

Page 76: Radiation Protection

X Ray spectrum: tube potential

Change in QUANTITY&

Change in QUALITY - spectrum shifts to higher Energy- characteristic lines appear

Page 77: Radiation Protection

• use of high KV technique and low mAs (using the

shortest exposure time)• The high KV beam has higher energy photons,

which undergo a lesser degree of beam attenuation and greater penetration of the beam through the patient.

• Therefore the tissue deposition of photons is reduced, which reduces the radiation dose to the patient

Page 78: Radiation Protection

A. At high KVp, majority of the photons are of high energy;therefore minimum number of photons are deposited in the patient

(dark area).B. At low KVp, a large number of photons are of low energy;

therefore larger number are deposited in the patient (dark area).

Page 79: Radiation Protection

X Ray spectrum: Target Z

Higher Z

Lower Z

X Ray Energy (keV)

Number of X Rays per unit Energy

Page 80: Radiation Protection

What is beam filtration?

10 15 20 25 30

15

10

5

Energy (keV)

Nu

mb

er o

f p

ho

ton

s (a

rbit

rary

no

rmal

isat

ion

)

X Ray spectrum at 30 kV for an X Ray tube with a Mo target and a 0.03 mm Mo filter

Absorber placed betweenSource and object

Will preferably absorb the lower energy photons

Or absorb parts of spectrum(K-edge filters)

Page 81: Radiation Protection

Tube filtration • Inherent filtration (always present)

– reduced entrance (skin) dose to the patient (cut off the low energy X Rays which do not contribute to the image)

• Additional filtration (removable filter)– further reduction of patient skin and superficial tissue

dose without loss of image quality• Total filtration (inherent + added) • Total filtration must be > 2.5 mm Al for a > 110 kV

generator

Page 82: Radiation Protection

Filtration

Change in QUANTITY&

Change in QUALITY spectrum shifts to higher energy

1- Spectrum out of anode2- After window tube housing

(INHERENT filtration)3- After ADDITIONAL filtration

Page 83: Radiation Protection

Tube filtration

Page 84: Radiation Protection

CollimationCollimate tightly to the

area of interest.Reduces the patient’s

total entrance skin exposure.

Improves image contrast.

Scatter radiation to the operator will also decrease.

Page 85: Radiation Protection

• Antiscatter grids

Antiscatter grids reduce scattered radiation reaching the film thus improving the quality of the resulting the radiograph and reducing chances of repeat exposures.

Source of -rays

LeadScattered X Rays

Useful X RaysFilm and cassette

Patient

Page 86: Radiation Protection

Patient Protection

• Correct filtration– 0.5 mm Al equivalent (inherent)– Added filtration is good– Minimum total filtration (inherent + added) must

be 2.5 mm Al equivalent – Accurate collimation

• Minimum repeats

Page 87: Radiation Protection

• Good technique to avoid re-takes:– use of correct film for the view intended– use of appropriate film holder– correct film placement within film holder– correct placement (angulation) of film holder in

patient’s mouth– correct tube angulation– correct exposure time

Page 88: Radiation Protection

AMOUNT & TYPE OF RADIATION EXPOSURE

– TIME

– DISTANCE

– SHIELDING

Page 89: Radiation Protection

Time• The exposure time is related to radiation exposure

and exposure rate (exposure per unit time) as follows :

• Exposure time  =    Exposure                           Exposure rateOr

Exposure = Exposure rate x Time

The algebraic expressions simply imply that if the exposure time is kept short, then the resulting dose to the individual is small

Page 90: Radiation Protection

TIME

- Take foot off fluoro pedal if physician is not viewing the TV monitor

- Use last image hold (freeze frame)

- Five-minute timer

- Use pulsed fluoro instead of continuous fluoro

- Low-Dose mode: 40% dose of Normal fluoro

- Pulsed Low-Dose provides further reduction with respect to Normal Dose continuous mode:

- Use record mode only when a permanent record is required

- Record beam-on time for review

Page 91: Radiation Protection

• Distance

• The second radiation protection action relates to the distance between the source of radiation and the exposed individual.

• The exposure to the individual decreases inversely as the square of the distance. This is known as the inverse square law, which is stated mathematically as :              1       I ~ ———              d2

Page 92: Radiation Protection

- One step back from tableside:cuts exposure by factor of 4

- Move Image Int. close to patient:less patient skin exposureless scatter (more dose interception by tower)sharper image

- Source to Skin Distance (SSD):38 cm for stationary fluoroscopes30 cm for mobile fluoroscopes

Page 93: Radiation Protection

Equipment to Control Distance

• In case of X-ray equipment operating up to 125 kVp, the control panel can be located in the X-ray room.

AERB recommends that the distance between control panel and X-ray unit/chest stand should not be less than 3 m for general purpose fixed x-ray equipment.

Page 94: Radiation Protection

• In mobile radiography,

where there is no fixed protective control booth, the technologist should remain at least 2 m from the patient, the x-ray tube, and the primary beam during the exposure.

• In this respect, the ICRP (1982), as well as the NCRP (1989a), recommended that the length of the exposure cord on mobile radiographic units be at least 2 m long

Page 95: Radiation Protection

Shielding

• Shielding implies that certain materials (concrete, lead) will attenuate radiation (reduce its intensity) when they are placed between the source of radiation and the exposed individual.

Page 96: Radiation Protection

• Lead is used as a radiation shielding material as it has a high atomic number (i.e. 82)

• Atomic number of an element is the number of protons in the nucleus (which is equal to the number of electrons around the nucleus)

• For the photoelectric process, the mass absorption coefficient increases with the cube of the atomic number (z3)

Page 97: Radiation Protection

• It is known that

• 0.25 mm lead thickness attenuates 66% of the beam at 75kVp

• and 1mm attenuates 99% of the beam at same kVp.

• It is recommended that for general purpose radiography the minimum thickness of lead equivalent in the protective apparel should be 0.5mm.

Page 98: Radiation Protection

- Lead aprons: cut exposure by factor of 20distant scatter: 0.25 mm Pb eqdirect involvement: 0.5 mm Pb

Alpha

Beta

Gamma and X-rays

Neutron

Paper Plastic Lead Concrete

10

n

00 g

Page 99: Radiation Protection

Four aspects of shielding in diagnostic radiology

1. X-ray tube shielding

2. Room shielding       (a) X-ray equipment room shielding        (b) Patient waiting room shielding.

3. Personnel shielding

4. Patient shielding (of organs not under investigation)

Page 100: Radiation Protection

1) X-ray tube shielding (Source Shielding)

• The x-ray tube housing is lined with thin sheets of lead because x-rays produced in the tube are scattered in all directions.

• This shielding is intended to protect both patients and personnel from leakage radiation.

• Leakage radiation is that created at the X-ray tube anode but not emitted through the x-ray tube portal.

• Rather, leakage radiation is transmitted through tube housing.

Page 101: Radiation Protection

• According to AERB recommendations manufacturers of x-ray devices are required to shield

the tube housing so as to limit the leakage radiation exposure rate to

0.1 R/ hr at a distance of 1 meter from the tube anode.

Page 102: Radiation Protection

2) Room shielding (Structural Shielding)

The lead lined walls of Radiology department are referred to as protective barriers because they are designed to protect individuals located outside the X-ray rooms from unwanted radiation.

Page 103: Radiation Protection

• There are two types of protective barriers.

(a) Primary Barrier: is one which is directly struck by the primary or the

useful beam.

(b) Secondary Barrier: is one which is exposed to secondary radiation

either by leakage from X-ray tube or by scattered radiation from the patient.

Page 104: Radiation Protection

The shielding of X-ray room is influenced by the nature of occupancy of the adjoining area. In this respect two

types of areas have been identified.

Control Area:• Is defined as the area routinely

occupied by radiation workers who are exposed to an occupational dose.

• For control area, the shielding should be such that it reduces exposure in that area to

<26mSv/kg/week

Uncontrolled areas:• Are those areas which are

not occupied by occupational workers.

• For these areas, the shielding should reduce the exposure rate to

<2.6mSv/kg/week

Page 105: Radiation Protection

• AERB has laid down GUIDELINES for shielding of X-ray examination room and patient’s waiting room which are as follows.

• The room housing an X-ray unit is not less than 18m2 for general purpose radiography and conventional fluoroscopy equipment.

Page 106: Radiation Protection

• In case the installation is located in a residential complex, it is ensured that

1. Wall of the x-ray rooms on which primary x-ray beam falls is not less than 35 cm thick brick or equivalent,

2. Walls of the x-ray room on which scattered x-rays fall is not less than 23 cm thick brick or equivalent

3. There is a shielding equivalent to at least 23 cm thick brick or 1.7 mm lead in front of the doors and windows of the x-ray room to protect the adjacent areas, used by general public

Page 107: Radiation Protection

• Unshielded openings in an X-ray room for ventilation or natural light, are located above a height of 2 m.

• Rooms housing fluoroscopy equipment are so designed that adequate darkness can be achieved conveniently, when desired, in the room.

Page 108: Radiation Protection

Rooms housing diagnostic X-ray units and related equipment are located as far away as feasible from

• areas of high occupancy and general traffic, • maternity and paediatric wards • and other departments of the hospital that are not

directly related to radiation and its use.

Page 109: Radiation Protection

• Shielding of the Xray control room :

• The control room of an X-ray equipment is a secondary protective barrier which has two important aspects:

• (a) The walls and viewing window of the control booth, which should have lead equivalents of 1.5mm. (b) The location of control booth, which should not be located where the primary beam falls directly, and the radiation should be scattered twice before entering the booth

Page 110: Radiation Protection

• The AERB recommends the following shielding for the Xray control room:

• The control panel of diagnostic X-ray equipment operating at 125 kVp or above is installed in a separate room located outside but contiguous to the X-ray room and provided with appropriate shielding, direct viewing and oral communication facilities between the operator and the patient

Page 111: Radiation Protection

• Patient waiting area

• Patient waiting areas are provided outside the X-ray room.

• A suitable warning signal such as red light and a warning placard is provided at a conspicuous place outside the X-ray room and kept ‘ON’ when the unit is in use to warn persons not connected with the particular examination from entering the room

Page 112: Radiation Protection

• 3) Personnel shielding

• Shielding of occupational workers can be achieved by following methods:

• Personnel should remain in the radiation environment only when necessary (step behind the control booth, or leave the room when practical)

Page 113: Radiation Protection

• Lead aprons are shielding apparel recommended for use by radiation workers. These are classified as a secondary barrier to the effects of ionizing radiation.

Page 114: Radiation Protection

• These aprons protect an individual only from secondary (scattered) radiation, not the primary beam .

• The thickness of lead in the protective apparel determines the protection it provides.

Page 115: Radiation Protection

• It is recommended that women radiation workers should wear a customized lead apron that reaches below midthigh level and wraps completely around the pelvis.

• This would eliminate an accidental exposure to a conceptus

Page 116: Radiation Protection

Care of the lead apparel: • It is imperative that lead aprons are not abused,

such as by– dropping them on the floor,– piling them in a heap – improperly draping them over the back of a chair.

• Because all of these actions can cause internal fracturing of the lead, they may compromise the apron’s protective ability.

Page 117: Radiation Protection

• When not in use, – all protective apparel should be hung on properly

designed racks. • Protective apparel also should be radiographed for

defects such as internal cracks and tears at least once a year

Page 118: Radiation Protection

• Other protective apparel include eye glasses with side shields, thyroid shields and hand gloves.

• The minimum protective lead equivalents in hand gloves and thyroid shields should be 0.5mm.

Page 119: Radiation Protection

• 4)Patient shielding

• Most radiology departments shield the worker and the attendant, paying little attention to the radiation protection of the patient.

• It has been recommended that the thyroid, breast and gonads be shielded, to protect these organs especially in children and young adults

Page 120: Radiation Protection

Rooms

Notification of hazard presenceSigns, Posting, Warning signs

Only authorized users may have access to x-ray devices

Energized equipment must be attended at all times

Lock lab door when equipment not attended

Page 121: Radiation Protection

Posting, Warning sign

Door signWarning sign

Page 122: Radiation Protection

Exposure to X-ray radiation is reduced if:

TIME exposed to source is decreased

DISTANCE from source is increased

SHIELDING from source is increased

To sum up……

Page 123: Radiation Protection

Notable Changes: FDA regs.For equipment manufactured after 10 June 2006:

• Warning Label – “WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions and maintenance scheduled are observed.”

• Timer: audible signal every 5 min of irradiation time until resetAND

Irradiation time display at fluoroscopist’s working position:- means to reset display at zero for new exam/procedure

• Last Image Hold (LIH) after exposure termination- indicate if LIH = radiograph or ‘freeze-frame’ image

Page 124: Radiation Protection

RADIATION IN THE CT SUITE

Page 125: Radiation Protection

• It has been estimated that although CT accounts for less than 50% of all x-ray examinations it contributes upto 40% of the collective dose from diagnostic radiology .

• CT Scanners have scattered radiation levels that may prove hazardous.

• The dose unit used in CT is the computed tomography dose index “CTDI”.

Page 126: Radiation Protection

• This measurement is defined in relation to the radiation field delivered at a specific point (x, y) by the CT Scanner.

• CTDI is usually expressed in terms of absorbed dose to air and is called CTDI air.

• Absorbed dose to tissue (Dtissue) is related to absorbed dose to air (Dair) by a mathematical coefficient which has a value of about 1.06 and an error not greater than ± 1%.

Page 127: Radiation Protection

• Such measurements are made using a special pencil ionisation chamber or by a thermoluminescent dosimeter (TLD) .

• Langer et al evaluated scattered radiation in a CT suite and documented that the radiation on the floor of the C.T. suite could be as high as 0.3 Gy/day.

Page 128: Radiation Protection

RADIATION PROTECTION IN CT SUITE

Page 129: Radiation Protection

It was concluded that

• adequate shielding should be provided for the floor and roof areas of a CT suite depending on which floor the CT is located.

• It was proposed an additional thickness of 2.5mm of lead or 162mm of concrete to shield the front and rear reference points, so as to reduce the dose to 1 mGy/year

Page 130: Radiation Protection

• The highly collimated X-ray beam in CT results in markedly non uniform distribution of absorbed dose perpendicular to the tomographic plane during the CT exposure.

• Therefore the size of the CT room housing the gantry of the CT unit as recommended by AERB should not be less than 25m2

Page 131: Radiation Protection
Page 132: Radiation Protection

• The greatest risk to the fetus of chromosomal abnormalities and subsequent mental retardation is between 8 and 15 weeks of pregnancy and examinations involving radiation to the fetus should be avoided during this period.

• For examinations which may involve rather heavy doses of radiation such as Barium enemas, pelvic or abdominal CT, the examination should be carried out during the first 10 days of the menstrual cycle to avoid irradiating any possible pregnancy

Page 133: Radiation Protection

If Pelvic Area in Beam:• No possibility of pregnancy - proceed• Probably pregnant - radiologist decides

– delay X-ray until after delivery, or– use non-X-ray technique (e.g. ultrasound), or– go ahead with X-ray but keep dose low

• Possibly pregnant, low dose procedure - proceed if period is not overdue.

• High dose procedure (10s of mGy, e.g. pelvic CT)

– X-ray in first 10 days of menstrual cycle .

Page 134: Radiation Protection

Pregnancy and Mammography“There is no requirement to enquire about pregnancy prior to mammography as there is no significant dose to the fetus”

NHBSP Dec 02

For pregnant staff, • a risk assessment must be performed, • dose to fetus < 1 mSv for rest of pregnancy.

Page 135: Radiation Protection

Radiation detection and measurement

Page 136: Radiation Protection

• The instruments used to detect radiation are referred to as

radiation detection devices.

• Instruments used to measure radiation are called radiation dosimeters

Page 137: Radiation Protection

Devices monitor and record ionizing radiation doses(occupational exposure)

Must distinguish from background radiation

DOSIMETRY

Page 138: Radiation Protection

• Personnel Dosimetry

Personnel dosimetry refers to the monitoring of individuals who are exposed to radiation during the course of their work.

Personnel dosimetry policies need to be in place for all occupationally exposed individuals.

The data from the dosimeter are reliable only when the dosimeters are properly worn, receive proper care, and are returned on time.

Page 139: Radiation Protection

The radiation measurement is a time-integrated dose, i.e., the dose summed over a period of time, usually about 3 months.

The dose is subsequently stated as an estimate of the effective dose equivalent to the whole body in mSv for the reporting period.

Dosimeters used for personnel monitoring have dose measurement limit of 0.1 - 0.2 mSv

Page 140: Radiation Protection

Proper care includes • not irradiating the dosimeter except during

occupational exposure• and ensuring proper environmental conditions

Monitoring is accomplished through the use of personnel dosimeters such as

• the pocket dosimeter, • the film badge • the thermoluminescent dosimeter

Page 141: Radiation Protection

Pocket Dosimeter

• Outwardly resembles a fountain pen .

It consists of • a thimble ionization chamber

with an eyepiece and a transparent scale,

• a hollow charging rod• a fixed and a movable fiber.• electrometer----separate

-----built-in (self reading type)

Page 142: Radiation Protection

The ability of radiation to produce ionization in air is

the basis for radiation detection by the ionization chamber.

It consists of an electrode positioned in the middle of a cylinder that contains gas.

When x-rays enter the chamber, they ionize the gas to form negative ions (electrons) and positive ions (positrons).

Page 143: Radiation Protection

The electrons are collected by the positively charged rod, while the positive ions are attracted to the negatively charged wall of the cylinder.

The resulting small current from the chamber is

subsequently amplified and measured.

The strength of the current is proportional to the radiation intensity.

Page 144: Radiation Protection

• Is sensitive for exposures upto 0.2 R

• Disadvantages------

– Easily damaged

– Unreliable in inexperienced hands

– Does not provide a permanent record

Page 145: Radiation Protection

Film Badge Monitoring

• These badges use small x-ray films sandwiched between several filters to help detect radiation.

• The photographic effect, which refers to the

ability of radiation to blacken photographic films, is the basis of detectors that use film.

Page 146: Radiation Protection

Film badge

detects beta, gamma, X Ray

Page 147: Radiation Protection

Wearing the badge -wear the badge on the collar region, because the collar region

including head, neck, and lens of the eyes are unprotected.

Wearing period-• Each member of staff wears film badge for a period of 4

weeks.

• At the end of period the film inside is changed.• The exposed film is sent to BARC.• Useful for detecting radiation at or above 0.1 msv (10 mrem)

Page 148: Radiation Protection

Advantages

– inexpensive,

– easy to use,

– permanent record of exposure,

– wide range of sensitivity ( 0.2 – 2000 msv),

– identifies type and energy of exposure,

Page 149: Radiation Protection

disadvantages • they are not sensitive enough to capture very low

levels of radiation( < 0.15 msv),

• Their susceptibility to fogging caused by high temperatures , humidity and light means that they cannot and should not be worn for longer than a 4-week period at a stretch,

• Enormous task to chemically process a large number of small films and subsequently compare each to some standard test film.

Page 150: Radiation Protection

Thermo luminescent dosimetry (TLD) Monitoring

• The limitations of the film badge are overcome by the thermo luminescent dosimeter (TLD).

• Thermo luminescence is the property of certain materials to emit light when they are stimulated by heat.

• Materials such as lithium fluoride (LiF), lithium borate (Li2B4O7), calcium fluoride (CaF2), and calcium sulfate (CaSO4) have been used to make TLDs

Page 151: Radiation Protection

• When an LiF crystal is exposed to radiation, a few electrons become trapped in higher energy levels.

For these electrons to return to their normal energy levels, the LiF crystal must be heated.

As the electrons return to their stable state, light is emitted because of the energy difference between two orbital levels.

The amount of light emitted is measured (by a photomultiplier tube) and it is proportional to the radiation dose.

Page 152: Radiation Protection

• The measurement of radiation from a TLD is a two-step procedure.

• In step 1, the TLD is exposed to the radiation.

• In step 2, the LiF crystal is placed in a TLD analyzer, where it is exposed to heat.

Page 153: Radiation Protection

• As the crystal is exposed to increasing temperatures, light is emitted.

• When the intensity of light is plotted as a function of the temperature, a glow curve results.

• The glow curve can be used to find out how much radiation energy is received by the crystal because the highest peak and the area under the curve are proportional to the energy of the radiation.

Page 154: Radiation Protection
Page 155: Radiation Protection

Advantages

• The TLD can measure exposures to individuals as low as 5 mR can withstand a certain degree of heat, humidity, and pressure

• Their crystals are reusable• Is very compact ( suitable even for finger dosimetry)• And instantaneous readings are possible if the department has a TLD

analyzer.• Response to radiation is proportional upto 400 R

Disadvantages

• Very expensive• No permanent record ( other than glow curves)• Cannot distinguish radioactive contamination.

The greatest disadvantage of a TLD is its cost

Page 156: Radiation Protection

Storing TLD Badges• Badge must not be left in an area

where it could receive a radiation exposure when not worn by the individual (e.g. On a lab coat or left near a radiation source)

• Store badges in a dark area with low radiation background (in low light away from fluorescent or uv lights, heat and sunlight)

• Lost or damaged badges should be reported immediately to the radiation safety officer and a replacement badge will be issued

Page 157: Radiation Protection

The Regulatory Bodies

Page 158: Radiation Protection

• There are various Regulatory Bodies at the international and National level, which lay down norms for radiation protection.

• These are • the International Commission for Radiation

Protection ( ICRP),• the National Commission for Radiation Protection

(NCRP ) in America,• and the Atomic Energy Regulatory Board (AERB) in

India.

Page 159: Radiation Protection

• The International Commission of Radiation Protection (ICRP) was formed in 1928 on the recommendation of the first International Congress of Radiology in 1925.

• The commission consists of 12 members and a

chairman and a secretary who are chosen from across the world based on their expertise.

• The first International Congress also initiated the birth of the ICRU or the International Commission on Radiation Units and measurements

Page 160: Radiation Protection

• The Indian regulatory board is the AERB, Atomic Energy Regulatory Board.

• The Atomic Energy Regulatory Board was constituted on November 15, 1983

• by the President of India by exercising the powers conferred by Section 27 of the Atomic Energy Act, 1962

to carry out certain regulatory and safety functions under the Act.

Page 161: Radiation Protection

• Radiation safety in handling of radiation generating equipment is governed by section 17 of the Atomic Energy Act, 1962, and the Radiation Protection Rules (RPR)

• The “Radiation Surveillance Procedures of Medical Applications of Radiation,” specify general requirements for ensuring radiation protection in installation and handling of X-ray equipment. Guidance and practical aspects on implementing the requirements of this Code are provided in revised documents issued by AERB in the year 2001

Page 162: Radiation Protection

Dose Limits Recommended by ICRP (1991)

Exposure Dose Limit (mSv per year)Condition

Occupational Apprentices Public(16-18 years)

Whole body: 20 mSv per year, 6 mSv in a year 1 mSv in a year,(effective dose) averaged over defined averaged over period of 5 years with 5 years,

no more than 50 mSvin a single year

Parts of the body:(equivalent dose)Lens of the eye 150 mSv per year 50 mSv in a year 15 mSv in a yearSkin* 500 mSv per year 150mSv in a year 50 mSv in a yearHands and feet** 500 mSv per year 150 mSv in a year 50 mSv in a year

*Averaged over areas of no more than any 1 cm2 regardless of the area exposed. The nominal depth is 7.0 mg cm -2

**Averaged over areas of the skin not exceeding about 100 cm2

Note 1.Dose limit for Women upon declaration of pregnancy - 2 mSv measured on the surface of the abdomen and 1/20th of ALI for exposure to internal emitters.Note 2.Dose limits do not apply to medical exposures, to natural sources of radiation and under conditions resulting from accidents.

Page 163: Radiation Protection

Radiation protection survey and programme

Page 164: Radiation Protection

• The responsibility for establishing a radiation protection programme rests with the hospital administration / owners of the X-ray facility

• The administration is expected to appoint a Radiation Safety Committee (RSC), and a Radiation Safety Officer (RSO).

• It is recommended by NCRP that the RSC should comprise of a radiologist, a medical physicist,, a senior nurse and an internist. It is the duty of RSC to perform a regular radiation protection survey

Page 165: Radiation Protection

This survey has 5 phases which are: 1. Investigation: To obtain information regarding

layout of the department, workload, personnel monitoring and records.

2. Inspection: Each diagnostic installation in the department is examined for its protection status with respect to its operating factors, control booth and availability of protection devices.

3. Measurement: Measurements are conducted on exposure factors. In addition scattered radiation and patient dose measurements in radiography and fluoroscopy are performed.

Page 166: Radiation Protection

4. Evaluation: The radiation protection status of the department is evaluated by examination of records, equipment working, status of protective clothing and the radiation doses obtained from phase-3.

5. Recommendations: A report is prepared on the protection status of the department and the problem areas if any identified, for which recommendations are made regarding corrective measures

Page 167: Radiation Protection

Thin-window GM (Geiger-Mueller) survey meter may be used to- Check leaking radiation- Indicate x-ray production- Monitor routine operation

Ion chamber is used to determine dose rate at the x-ray field.

SurveyArea

Survey meters are calibrated annually.

Page 168: Radiation Protection

Depicts the organizational flow chart and the administrative and functional components of radiation protection program.

Page 169: Radiation Protection

CONCLUSION• Protect patient, public and staff• Remember dose is cumulative• Benefit/risk ratio• Principles of radiation protection• Dose reduction = time, distance, shielding

High speed film Lead coats toreduced exp. time steps away stop scatter radiation

Page 170: Radiation Protection

Thanks!!!