Top Banner

of 57

Quick Start Guide-MadgeTech

Jul 07, 2018

Download

Documents

Nirav Desai
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/18/2019 Quick Start Guide-MadgeTech

    1/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    2/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    3/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    4/57

     

    MadgeTech Page 7 Quick Start Guide 

    1 Quick Start

    To begin using your data logger right away, follow these six simplesteps:

    1. Install the software. (See Software Installation)2. Attach the logger to your computer using the interface cable.

    (See Hardware Installation)3. From the Communication menu, select "Auto Configure Port".

    Fig 1.1Fig 1.1

    4. From the Device menu, select “Start Device”. Fig 1.3

    Fig 1.3

    5. A message will appear (fig 1.3) Click Yes to continue. Be sureto have downloaded all previously recorded data beforecontinuing.

    MadgeTech Page 8 Quick Start Guide

    Fig 1.3

    6. The “Start Device” window will open. Note that there are twosections of this window, the “Start Method” section and the “StartParameters” section. (Fig 1.4)

    Fig 1.4

    a. The “Start Method” section allows the user to choose to“Start Now” or “Delay Start”, which allows most devicesto record data anywhere from 0 to 60 days from the timeit is started.

    b. The “Start Parameters” section allows the user to setthree options in the device: User ID, Reading Rate, andWrap Around.

    i. The “User ID” field allows the user to store acombination of six letters or numbers in thedevice to help you identify the device, or theperson who started the device, etc.

  • 8/18/2019 Quick Start Guide-MadgeTech

    5/57

     

    MadgeTech Page 9 Quick Start Guide 

    ii. The Reading Rate field allows the user to selecta specific reading rate in the device, typicallyranging from one reading every 2 seconds toone reading every 12 hours. These ranges canvary depending on the device.

    iii. The “Wrap Around” field is only available to

    devices which have the ability to performmemory wrap around. If the device is able to doso, the user chooses to overwrite the oldest datawhen the device memory becomes full. If thisbox is left unchecked, the device will stoprecording data when the memory is full.

    7. When finished, click on “Start Device”.8. After a brief pause while the software communicates with the

    device, you will see the message "Device Started". Your device

    is now started and will begin taking measurements according tothe parameters set in section 6. You may now disconnect thelogger and place the device where you need it to perform itsmeasurements. Later, when you are ready to view themeasurements, simply reconnect it to the computer, and select"Read Device Data" from the Device menu.

    For more detailed information on using our software, please see thehelp file located under the help menu in the MadgeTech software.

    MadgeTech Page 10 Quick Start Guide

    2 Software Installation

    Instal l ing from CD ROM

    Insert the CD ROM labeled “MadgeTech Data Recording Software” into

    your CD ROM drive. From the Windows Start menu, choose the Runcommand and type d:\setup into the Open field and click OK. If your CDROM drive is not the D: drive, use the correct letter in the instructionsabove.

     After installation, the software will be listed under the default MadgeTech software program group and saved under the default “C:\ProgramFiles\MadgeTech 2.00\” directory.

  • 8/18/2019 Quick Start Guide-MadgeTech

    6/57

     

    MadgeTech Page 11 Quick Start Guide 

    3 Hardware Installation

    1. Connect the DB9 end of the interface cable provided with the

    software package to the computer’s serial port. (Fig 3.1

    Fig. 3.1

    2. Connect the stereo jack end of the interface cable to the datalogger. Fig 3.2

    Fig 3.2

    MadgeTech Page 12 Quick Start Guide

    CRYOTEMP

    To start the CRYOTEMP refer to “Quick Start”. After starting, theCRYOTEMP is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    Be sure to keep the CRYOTEMP dry. If using the device in a dry iceenvironment, place the CRYOTEMP in a sealable plastic bag. If usingthe data logger in a wet ice environment, place the CRYOTEMP in a

    mason jar and close the cover.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close case and secure with screw.

    Note: Before replacing the battery, be sure to stop and download the

    data from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

  • 8/18/2019 Quick Start Guide-MadgeTech

    7/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    8/57

     

    MadgeTech Page 15 Quick Start Guide 

    OCTTEMP

    To start the OCTTEMP refer to “Quick Start”. After starting, theOCTTEMP is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The OCTTEMP has eight SMP connections which allow the user toinsert subminiature thermocouple plugs into the connectors on thedevice. Fig. A shows how to connect the individual thermocouples.

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,

    B and N. To select the correct thermocouple type in the MadgeTechsoftware:

    1. Select “Thermocouple Type” from the device menu. (Fig. B)

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish. 

     – I nput

    Individual Connector

    for thermocouples

    + Input

    Connector for IFC101

    12

    34

    56

    78

    OCTTEMP

    MadgeTech Page 16 Quick Start Guide

    Fig. B

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose per

    local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched between

    the cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

  • 8/18/2019 Quick Start Guide-MadgeTech

    9/57

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    10/57

    MadgeTech Page 19 Quick Start Guide 

    QUADTEMP

    To start the QUADTEMP refer to “Quick Start”. After starting, theQUADTEMP is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The QUADTEMP has four SMP connections which allows the user toinsert subminiature thermocouple plugs into the connectors on thedevice. The diagram below shows how to connect the individualthermocouples.

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware:

    1. Select “Thermocouple Type” from the device menu. (Fig. B)

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in the

    device.5. Click Ok, to finish.

     – I nput

    Individual Connector

    for thermocouples

    + Input

    12

    3

    4

    QUADT EMP

    Connector for IFC101

    MadgeTech Page 20 Quick Start Guide

    Fig. B

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose per

    local regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched betweenthe cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    11/57

    MadgeTech Page 21 Quick Start Guide 

    RTDTEMP101

    To start the RTDTEMP101 refer to “Quick Start”. After starting, theRTDTEMP101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The RTDTEMP101 has a four-position removable screw terminalconnection. The RTDTEMP101 accepts 2, 3 or 4-wire RTDconfigurations. Temperature probe to be used with the recorder must bea 100 ohm platinum RTD, in the standard 2-, 3- or 4-wire configuration.The RTDTEMP101 is designed to achieve exceptional accuracy with the4-wire probe, but will still yield measurements better than required for

    most applications with the 2- or 3-wire probes.

    Insure that the probe you select can be connected to the RTDTEMP101input by selecting a probe with lead wires, or by attaching an adapterthat will allow you to connect wire leads to the probe. OMEGA’s PR-10and PR-11 series of probes have the proper leads for direct connectionto the RTDTEMP101.

    For 4-wire RTD probes, connect the four lead wires to the RTDTEMP101

    as shown in Fig A.

    For 3-wire RTD probes, short inputs 3 and 4 together, then connect thelead wires to inputs 1, 2 and 3.

    For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2together, then connect the RTD lead wires to inputs 2 and 3.

    Fig A

    Connector for 100

    ohm Pt RTD

    4 – Red or Short to 3

    3 – Red

    2 – Black

    1 – Black or Short to 2

    4 – Excitation cu rrent out (+)3 – Measurement (+) input2 – Me asurement (-) Input1 – Ground (-) 

    RTDT EM P101

    IFC Connector

    RTDT EMP101

    MadgeTech Page 22 Quick Start Guide

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close case and secure with screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

  • 8/18/2019 Quick Start Guide-MadgeTech

    12/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    13/57

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    14/57

    MadgeTech Page 27 Quick Start Guide 

    TC110

    To start the TC110 refer to “Quick Start”. After starting, the TC110 isarmed; data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At this

    point the device cannot be restarted until it has been re-armed by thecomputer.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    The TC110 has a three position removable screw terminal. This enablesthe TC110 to be connected to most 2-wire thermocouples with lead

    wires, and to have the ability to shield the thermocouple with the groundconnection. If the TC110 is to be used with a 2-wire thermocouplewithout shielding, leave the ground input unconnected. Be sure toconnect the thermocouple with the right polarity as marked on the caseof the device, the ground connection is optional. Fig A.

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. B

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish. 

     – Input

    Connector for

    removable screw

    terminal

    + Input

    Wires

    GroundTC110

    MadgeTech Page 28 Quick Start Guide

    Fig. B

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the old battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw. For 10+

    year operation, be sure to set the reading rate to no more than onereading every 15 minutes.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    15/57

    MadgeTech Page 29 Quick Start Guide 

    TC4000

    To start the TC4000 refer to “Quick Start”. After starting, the TC4000 isarmed; data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At this

    point the device cannot be restarted until it has been re-armed by thecomputer.

    Depending on the physical configuration of the TC4000, the device willeither have an input for a subminiature thermocouple connector or aremovable screw terminal.

    The standard connector for the TC4000 is the SMP connection whichallows for the user to insert a subminiature thermocouple plug into the

    connector on the device. Fig. A

    Fig. A

    The non-standard connector for the TC4000 is the removable screwterminal. This enables the TC4000 to be connected to most 2-wirethermocouples with lead wires. Be sure to connect the thermocouplewith the right polarity as marked on the case of the device. Fig. B

    Fig. B

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. C

    2. Click on the change button in the Thermocouple Type window.3. Select the correct thermocouple type from the drop down list.

     – I nput

    + Input

    Thermocouple connector

    TC4000

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    Thermocouplewires

    TC4000

    MadgeTech Page 30 Quick Start Guide

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish. 

    Fig. C

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the old battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

  • 8/18/2019 Quick Start Guide-MadgeTech

    16/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    17/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    18/57

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    19/57

    MadgeTech Page 37 Quick Start Guide 

    TEMP1000P

    The serial port on the TEMP1000P is located under the cap with the keyring. To connect the logger to the interface cable, unscrew the cap andinsert the stereo jack end of the interface cable into the logger. To start

    the TEMP1000P refer to “Quick Start”. After starting, the TEMP1000P isarmed, data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At thispoint the device cannot be restarted until it has been re-armed by thecomputer.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,

    incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Contact the factory for information on replacing the battery.

    Warning

    This device is intended for water proof applications. THIS UNITMUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATEROR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETYOF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    MadgeTech Page 38 Quick Start Guide

    TEMP1000S

    The serial port on the TEMP1000S is located under the cap with the keyring. To connect the logger to the interface cable, unscrew the cap andinsert the stereo jack end of the interface cable into the logger. To start

    the TEMP1000S refer to “Quick Start”. After starting, the TEMP1000S isarmed, data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At thispoint the device cannot be restarted until it has been re-armed by thecomputer. 

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,

    incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Contact the factory for information on replacing the battery.

    Warning

    This device is intended for water proof applications. THIS UNITMUST BE PROPERLY SEALED BEFORE SUBMERSING IN WATEROR OTHER MEDIUM. FOR QUESTIONS REGARDING THE SAFETYOF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

  • 8/18/2019 Quick Start Guide-MadgeTech

    20/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    21/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    22/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    23/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    24/57

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    25/57

    MadgeTech Page 49 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched between

    the cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 50 Quick Start Guide

    QUADVOLT

    To start the QUADVOLT refer to “Quick Start”. After starting, theQUADVOLT is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The QUADVOLT allows for user defined units to be programmed into thedevice. To use the programmable engineering units, refer to the“Engineering Units” section.

    The QUADVOLT has four two-position removable screw terminalconnections. The QUADVOLT accepts 2-wire configurations. Insurethat the device that is to be measured can be connected to theQUADVOLT input by selecting an output connection with lead wires, orby attaching an adapter that will allow you to connect wire leads to thedevice to be measured.

    For a two-wire connection, connect IN+ to positive signal (typically thered wire) and IN- to the negative signal (typically the black wire). Fig. A

    Fig. A

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – I nput

    Individual Connector

    for removable screw

    terminal

    + Input

    Wires

    QUADVOLT

    1

    23

    4

    Connector for IFC101

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    26/57

    MadgeTech Page 51 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched between

    the cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 52 Quick Start Guide

    VOLT101

    To start the VOLT101 refer to “Quick Start”. After starting, the VOLT101is armed; data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At this

    point the device cannot be restarted until it has been re-armed by thecomputer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The VOLT101 allows for user defined units to be programmed into thedevice. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The VOLT101 has a two-position removable screw terminal connection.The VOLT101 accepts 2-wire configurations. Insure that the device thatis to be measured can be connected to the VOLT101 input by selectingan output connection with lead wires, or by attaching an adapter that willallow you to connect wire leads to the device to be measured.

    For a two-wire connection, connect IN+ to positive signal (typically thered wire) and IN- to the negative signal (typically the black wire). Fig. A

    Fig. A

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    Thermocouplewires

    VOLT101

     

  • 8/18/2019 Quick Start Guide-MadgeTech

    27/57

    MadgeTech Page 53 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 54 Quick Start Guide

    VOLT110 – 100mV

    To start the VOLT110 – 100 mV refer to “Quick Start”. After starting, theVOLT110 – 100 mV is armed; data will start recording. The device willstop recording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The VOLT110 – 100 mV allows for user defined units to be programmedinto the device. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The VOLT110 – 100 mV has a three-position removable screw terminalconnection. The VOLT110 – 100 mV measures a differential input andaccepts 2 or 3-wire configurations. Insure that the device that is to bemeasured can be connected to the VOLT110 – 100 mV input byselecting an output connection with lead wires, or by attaching anadapter that will allow you to connect wire leads to the device to bemeasured.

    For a two-wire connection, connect IN+ to positive signal (typically thered wire) and IN- to the negative signal (typically the black wire).

    For a three-wire connection, make the connections as described for thetwo-wire connection, also connecting Ground to signal ground. Fig. A

    Fig. A

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    Wires

    Ground

    VOLT110-100m V

  • 8/18/2019 Quick Start Guide-MadgeTech

    28/57

     

    Battery Replacemen t OCTPROCESS

  • 8/18/2019 Quick Start Guide-MadgeTech

    29/57

    MadgeTech Page 57 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw. For 10+

    year operation, be sure to set the reading rate to no more than onereading every 15 minutes.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 58 Quick Start Guide

    OCTPROCESS

    To start the OCTPROCESS refer to “Quick Start”. After starting, theOCTPROCESS is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The OCTPROCESS allows for user defined units to be programmed intothe device. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The OCTPROCESS has eight two-position removable screw terminalconnections. The OCTPROCESS accepts 2-wire configurations. Insure

    that the device that is to be measured can be connected to theOCTPROCESS input by selecting an output connection with lead wires,or by attaching an adapter that will allow you to connect wire leads to thedevice to be measured.

    For a two-wire loop, connect IN+ to positive signal (typically the red wire)and IN- to the negative signal (typically the black wire). Fig. A

    Fig. A

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – I nput

    Individual connector

    for removable screwterminal

    + Input

    w ires

    OCTPROCESS

    12

    34

    56

    78

    Connector for IFC101

     

    Battery Replacemen t PROCESS101

  • 8/18/2019 Quick Start Guide-MadgeTech

    30/57

    MadgeTech Page 59 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched between

    the cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 60 Quick Start Guide

    PROCESS101

    To start the PROCESS101 refer to “Quick Start”. After starting, thePROCESS101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The PROCESS101 allows for user defined units to be programmed intothe device. To use the programmable engineering units, refer to the“Engineering Units” section.

    The PROCESS101 has a two-position removable screw terminalconnection. The PROCESS101 accepts 2-wire configurations. Insurethat the device that is to be measured can be connected to thePROCESS101 input by selecting an output connection with lead wires,or by attaching an adapter that will allow you to connect wire leads to thedevice to be measured.

    For a two-wire loop, connect IN+ to positive signal (typically the red wire)and IN- to the negative signal (typically the black wire). Fig. A

    Fig A

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – Input

    Connector for

    removable screw

    terminal

    + Input

    Wires

    PROC ESS101

     

    Battery Replacemen t PROCESS110

  • 8/18/2019 Quick Start Guide-MadgeTech

    31/57

    MadgeTech Page 61 Quick Start Guide 

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 62 Quick Start Guide

    PROCESS110

    To start the PROCESS110 refer to “Quick Start”. After starting, thePROCESS110 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The PROCESS110 allows for user defined units to be programmed intothe device. To use the programmable engineering units, refer to the“Engineering Units” section.

    The PROCESS110 has a three-position removable screw terminalconnection. The PROCESS110 accepts 2 or 3-wire configurations.Insure that the device that is to be measured can be connected to thePROCESS110 input by selecting an output connection with lead wires,or by attaching an adapter that will allow you to connect wire leads to the

    device to be measured.

    For a two-wire loop, connect IN+ to positive signal (typically the red wire)and IN- to the negative signal (typically the black wire). You can leaveground unconnected, or connect it to signal shield. Fig. A

    Fig. A

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    Wires

    Ground

    PROC ESS110

     

    Battery Warning QUADPROCESS

  • 8/18/2019 Quick Start Guide-MadgeTech

    32/57

    MadgeTech Page 63 Quick Start Guide 

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw. For 10+year operation, be sure to set the reading rate to no more than one

    reading every 15 minutes.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 64 Quick Start Guide

    QUADPROCESS

    To start the QUADPROCESS refer to “Quick Start”. After starting, theQUADPROCESS is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The QUADPROCESS allows for user defined units to be programmedinto the device. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The QUADPROCESS has four, two-position removable screw terminalconnections. The QUADPROCESS accepts 2-wire configurations.Insure that the device that is to be measured can be connected to theQUADPROCESS input by selecting an output connection with leadwires, or by attaching an adapter that will allow you to connect wire leadsto the device to be measured

    For a two-wire loop, connect IN+ to positive signal (typically the red wire)and IN- to the negative signal (typically the black wire). Fig. A

    Fig. A

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY. 

     – I nput

    Individual Connectorfor removable screw

    terminal

    + Input

    Wires

    QUADPROCESS

    1

    23

    4

    Connector for IFC101

     

    Battery Replacemen t EVENT101

  • 8/18/2019 Quick Start Guide-MadgeTech

    33/57

    MadgeTech Page 65 Quick Start Guide 

    y p

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched between

    the cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 66 Quick Start Guide

    EVENT101

    To start the EVENT101 refer to “Quick Start”. After starting, theEVENT101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The EVENT101 has a two-position removable screw terminalconnection. The EVENT101 accepts 2-wire configurations. To use,connect IN+ to positive signal (typically the red wire) and IN- to thenegative signal (typically the black wire). Fig. A

    Fig. A

    The EVENT101 records the time at which a pulse input or contactclosure occurs on the input of the device. The device is continuouslysampling at 1 kHz, therefore, the pulse input must be at least 1 ms induration. An input voltage of greater than 30 volts on the input couldcause loss of data or damage to data logger. Only one event will berecorded per reading interval (ie. 1 second). The EVENT101 will mostoften be used with the 1 second reading rate.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    Wires

    EVENT101

     

    Battery Replacemen t EVENT110

  • 8/18/2019 Quick Start Guide-MadgeTech

    34/57

    MadgeTech Page 67 Quick Start Guide 

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 68 Quick Start Guide

    EVENT110

    To start the EVENT110 refer to “Quick Start”. After starting, theEVENT110 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    The EVENT110 has a two-position removable screw terminalconnection. The EVENT110 accepts 2-wire configurations. To use,connect IN+ to positive signal (typically the red wire) and IN- to thenegative signal (typically the black wire). Fig A

    Fig. A

    The EVENT110 records the time at which a pulse input or contactclosure occurs on the input of the device. The device is continuouslysampling at 1 kHz, therefore, the pulse input must be at least 1 ms induration. An input voltage of greater than 30 volts on the input couldcause loss of data or damage to data logger. Only one event will be

    recorded per reading interval (i.e. 1 second). The EVENT110 will mostoften be used with the 1 second reading rate.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – Input

    Connector for

    removable screw

    terminal

    + Input

    Wires

    EVENT110

     

    Battery Replacemen t PULSE101

  • 8/18/2019 Quick Start Guide-MadgeTech

    35/57

    MadgeTech Page 69 Quick Start Guide 

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 70 Quick Start Guide

    PULSE101

    To start the PULSE101 refer to “Quick Start”. After starting, thePULSE101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The PULSE101 allows for user defined units to be programmed into thedevice. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The PULSE101 has a two-position removable screw terminalconnection. The PULSE101 accepts 2-wire configurations. To use,connect IN+ to positive signal (typically the red wire) and IN- to thenegative signal (typically the black wire). Fig. A

    Fig. A

    The PULSE101 records the number of pulse inputs or contact closureson the input of the device. The input is continuously sampled at 1 kHz,therefore, the input pulse must be at least 1 ms in duration. This devicewill read up to 100 pulses per second. The counter is reset at thebeginning, and is recorded at the end of each sample interval. Thecounter is incremented when the input transitions high during the readinginterval.

    Note: An input voltage of greater than 30 volts on the input could causeloss of data or damage to data logger.

     – Input

    Connector for

    removable screw

    terminal

    + Input

    wires

    PULSE101

     

    Battery Warning PULSE110

  • 8/18/2019 Quick Start Guide-MadgeTech

    36/57

    MadgeTech Page 71 Quick Start Guide 

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 72 Quick Start Guide

    U S 0

    To start the PULSE110 refer to “Quick Start”. After starting, thePULSE110 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    The PULSE110 allows for user defined units to be programmed into thedevice. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The PULSE110 has a two-position removable screw terminalconnection. The PULSE110 accepts 2-wire configurations. To use,connect IN+ to positive signal (typically the red wire) and IN- to thenegative signal (typically the black wire). Fig. A

    Fig. A

    The PULSE110 records the number of pulse inputs or contact closureson the input of the device. The input is continuously sampled at 1 kHz,therefore, the input pulse must be at least 1 ms in duration. This devicewill read up to 100 pulses per second. The counter is reset at thebeginning, and is recorded at the end of each sample interval. Thecounter is incremented when the input transitions high during the readinginterval.

    Note: An input voltage of greater than 30 volts on the input could causeloss of data or damage to data logger.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    wires

    PULSE110

     

    Battery Warning STATE101

  • 8/18/2019 Quick Start Guide-MadgeTech

    37/57

    MadgeTech Page 73 Quick Start Guide 

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download the

    data from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 74 Quick Start Guide

    To start the STATE101 refer to “Quick Start”. After starting, theSTATE101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The STATE101 has a two-position removable screw terminal connection.The STATE101 accepts 2-wire configurations. To use, connect IN+ topositive signal (typically the red wire) and IN- to the negative signal(typically the black wire). Fig. A

    Fig. A

    The STATE101 records the state of the input of the device at theselected reading rate. To ensure that the STATE101 captures the stateof the input, be sure to have a contact closure that lasts at least thelength of the selected reading rate. The device will only record if thestate has changed when the input is sampled. The date and time isrecorded along with the change of state.

    Note: An input voltage of greater than 30 volts on the input could causeloss of data or damage to data logger.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    wires

    STATE101

     

    Battery Replacemen t STATE110

  • 8/18/2019 Quick Start Guide-MadgeTech

    38/57

    MadgeTech Page 75 Quick Start Guide 

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 76 Quick Start Guide

    To start the STATE110 refer to “Quick Start”. After starting, theSTATE110 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    The STATE110 has a two-position removable screw terminal connection.The STATE110 accepts 2-wire configurations. To use, connect IN+ topositive signal (typically the red wire) and IN- to the negative signal(typically the black wire). Fig. A

    Fig. A

    The STATE110 records the state of the input of the device at theselected reading rate. To ensure that the STATE110 captures the stateof the input, be sure to have a contact closure that lasts at least thelength of the selected reading rate. The device will only record if thestate has changed when the input is sampled. The date and time isrecorded along with the change of state.

    Note: An input voltage of greater than 30 volts on the input could causeloss of data or damage to data logger.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    wires

    STATE110

     

    Battery Warning PRHTEMP101

  • 8/18/2019 Quick Start Guide-MadgeTech

    39/57

    MadgeTech Page 77 Quick Start Guide 

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    MadgeTech Page 78 Quick Start Guide

    To start the PRHTEMP101 refer to “Quick Start”. After starting, thePRHTEMP101 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device is

    stopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or other

    liquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

  • 8/18/2019 Quick Start Guide-MadgeTech

    40/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    41/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    42/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    43/57

  • 8/18/2019 Quick Start Guide-MadgeTech

    44/57

     

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate

    SHOCK101-EB

  • 8/18/2019 Quick Start Guide-MadgeTech

    45/57

    MadgeTech Page 89 Quick Start Guide 

    the data set is accurate.

    MadgeTech Page 90 Quick Start Guide

    To start the SHOCK101-EB, refer to “Quick Start”. After starting, theSHOCK101-EB is armed, data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The SHOCK101-EB records peak acceleration levels for a specifiedreading interval. This means that the SHOCK101-EB will record thelargest acceleration level on each of the axis’s during each readinginterval. For example, if the reading rate is set at 2 seconds, then thelargest acceleration value during the 2 seconds will be recorded for eachof the x, y and z axis’s. Due to the noise of the electronics, there istypically ± 0.2 to 0.3 g of noise on each of the axis’s.

    During real-time operation, the SHOCK101-EB will only take onemeasurement during the reading interval, therefore in real-timeoperation, the device is not measuring the peak acceleration levels.

    When setting up the SHOCK101-EB, notice which way the axis’s are todetermine which direction the shock or acceleration occurs. When theSHOCK101-EB is laying flat with the label facing up, the x-axis is the onethat reads left to right and will read around 0 when still. The y-axis reads

    top to bottom and will read around 0 when still. The z-axis readsperpendicular to the other two axis’s, and will read around 1 when still.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the bottom of the largecase. Disconnect the battery connector to allow the cover to completelyopen. Remove the 6 D-cell batteries from the holder. Replace thebatteries with 6 D-cell batteries. Reconnect the battery connector, closethe case and secure with the screws.

     

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    ULTRASHOCK

  • 8/18/2019 Quick Start Guide-MadgeTech

    46/57

    MadgeTech Page 91 Quick Start Guide 

    t e data set s accu ate

    MadgeTech Page 92 Quick Start Guide

    To start the ULTRASHOCK refer to “Quick Start”. After starting, theULTRASHOCK is armed, data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The ULTRASHOCK records peak acceleration levels for a specifiedreading interval. This means that the ULTRASHOCK will record thelargest acceleration level on each of the axis’s during each readinginterval. For example, if the reading rate is set at 2 seconds, then thelargest acceleration value during the 2 seconds will be recorded for eachof the x, y and z axis’s. Due to the noise of the electronics, there istypically ± 0.2 to 0.3 g of noise on each of the axis’s.

    During real-time operation, the ULTRASHOCK will only take onemeasurement during the reading interval, therefore in real-timeoperation, the device not measuring the peak acceleration levels.

    When setting up the ULTRASHOCK, notice which way the axis’s are todetermine which direction the shock or acceleration occurs. When theULTRASHOCK is laying flat with the label facing up, the x-axis is the onethat reads left to right and will read around 0 when still. The y-axis reads

    top to bottom and will read around 0 when still. The z-axis readsperpendicular to the other two axis’s, and will read around 1 when still.

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched betweenthe cover and the case.

     

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

    PHONEPROCESS

  • 8/18/2019 Quick Start Guide-MadgeTech

    47/57

    MadgeTech Page 93 Quick Start Guide 

    MadgeTech Page 94 Quick Start Guide

    The PHONEPROCESS differs from other MadgeTech data loggers bycommunicating through an onboard modem. In order to communicate tothe PHONEPROCESS, the data logger must be connected to a phoneline. To use the software to connect to the data logger, the correct COMport must be selected in our software. If you are unsure which COM portthe modem is on, you can use Window’s Device Manager to locate theport number. The PHONEPROCESS communicates at 2400 baud;therefore, the correct baud rate in the software must be selected.

    To connect, use the “Modem Connection” feature of the MadgeTechsoftware located under the communication menu. In the “telephonenumber” field, enter the telephone number of the line that thePHONEPROCESS is on. The PHONEPROCESS will then answer the

    call and connect to the MadgeTech software. To start thePHONEPROCESS refer to “Quick Start” and skip ahead to section 4. After starting, the PHONEPROCESS is armed; data will start recording.The device will stop recording data when the end of memory is reachedor the device is stopped. At this point the device cannot be restarteduntil it has been re-armed by the computer.

    Once started, the LED will flash at the selected reading rate to indicatethat the device is running.

    The PHONEPROCESS allows for user defined units to be programmedinto the device. To use the programmable engineering units, refer to the“Engineering Units” section. (page ?)

    The PHONEPROCESS has a two-position removable screw terminalconnection. The PHONEPROCESS accepts 2-wire configurations.Insure that the device that is to be measured can be connected to thePHONEPROCESS input by selecting an output connection with lead

    wires, or by attaching an adapter that will allow you to connect wire leadsto the device to be measured.

    For a two-wire loop, connect IN+ to positive signal (typically the red wire)and IN- to the negative signal (typically the black wire). Fig. A

  • 8/18/2019 Quick Start Guide-MadgeTech

    48/57

     

    IROCTTEMP Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,

  • 8/18/2019 Quick Start Guide-MadgeTech

    49/57

    MadgeTech Page 97 Quick Start Guide 

    To start the IROCTTEMP refer to “Quick Start”. After starting, theIROCTTEMP is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The IROCTTEMP has eight SMP connections which allows the user toinsert subminiature thermocouple plugs into the connectors on thedevice. Fig. A

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. B

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish. 

    Fig. B

     – I nput

    Individual Connector

    for thermocouples

    + Input

    Connector for IFC101

    12

    34

    56

    78

    IROCTTEMP

    MadgeTech Page 98 Quick Start Guide

    g gg yincinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched betweenthe cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

    IRQUADTEMP Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,

  • 8/18/2019 Quick Start Guide-MadgeTech

    50/57

    MadgeTech Page 99 Quick Start Guide 

    To start the IRQUADTEMP refer to “Quick Start”. After starting, theIRQUADTEMP is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-armed by the computer.

    The IRQUADTTEMP has four SMP connections which allows the user toinsert subminiature thermocouple plugs into the connectors on thedevice. Fig. A

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. B

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish.

    Fig. B

     – I nput

    Individual Connector

    for thermocouples

    + Input

    12

    3

    4

    IRQUADTEMP

    Connector for IFC101

    MadgeTech Page 100 Quick Start Guide

    incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    Battery Replacemen t

    Open the case by removing the four screws on the top of the case.Disconnect the battery from the battery connector. Replace the batterywith a Lithium U9VL-J. Close the case and secure with screws, makingsure that the wires for the battery connector do not get pinched betweenthe cover and the case.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

    IRTC110

    T t t th IRTC110 f t “Q i k St t” Aft t ti th IRTC110

    Fig. B

  • 8/18/2019 Quick Start Guide-MadgeTech

    51/57

    MadgeTech Page 101 Quick Start Guide 

    To start the IRTC110 refer to “Quick Start”. After starting, the IRTC110is armed; data will start recording. The device will stop recording datawhen the end of memory is reached or the device is stopped. At thispoint the device cannot be restarted until it has been re-armed by thecomputer.

    This device communicates at 57,600 baud. If you do not have thesoftware set to “Always Auto Configure”, you must select this baud ratefrom the communication menu.

    The IRTC110 has a three position removable screw terminal. Thisallows the IRTC110 to be connected to most 2-wire thermocouples withlead wires, and the ability to shield the thermocouple with the ground

    connection. Be sure to connect the thermocouple with the right polarityas marked on the case of the device, the ground connection is optional.Fig. A

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. B

    2. Click on the change button in the Thermocouple Type window.

    3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish. 

     – Input

    Connector for

    removable screw

    terminal

    + Input

    wires

    GroundIRTC110

    MadgeTech Page 102 Quick Start Guide

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw. For 10+year operation, be sure to set the reading rate to no more than onereading every 15 minutes.

    Note: Before replacing the battery, be sure to stop and download thedata from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

    IRTC4000

    To start the IRTC4000 refer to “Quick Start” After starting the

    Fig. B

  • 8/18/2019 Quick Start Guide-MadgeTech

    52/57

    MadgeTech Page 103 Quick Start Guide 

    To start the IRTC4000 refer to Quick Start . After starting, theIRTC4000 is armed; data will start recording. The device will stoprecording data when the end of memory is reached or the device isstopped. At this point the device cannot be restarted until it has been re-

    armed by the computer.

    Depending on the physical configuration of the IRTC4000, it will eitherhave an input for a subminiature thermocouple connector or a removablescrew terminal.

    The standard connector for the IRTC4000 is the SMP connection whichallows for the user to insert the subminiature thermocouple plug into theconnector on the device.

    The non-standard connector for the IRTC4000 is the removable screwterminal. This allows the IRTC4000 to be connected to most 2-wirethermocouples with lead wires. Be sure to connect the thermocouplewith the right polarity as marked on the case of the device. Fig. A

    Fig. A

    The software accepts eight different thermocouple types: J, K, T, E, R, S,B and N. To select the correct thermocouple type in the MadgeTechsoftware to match the thermocouple that is being used:

    1. Select “Thermocouple Type” from the device menu. Fig. B

    2. Click on the change button in the Thermocouple Type window.3. Select the correct thermocouple type from the drop down list.

    4. Click on the Save button to store the thermocouple type in thedevice.

    5. Click Ok, to finish.

     – I nput

    Connector for

    removable screw

    terminal

    + Input

    wires

    IRTC4000

    MadgeTech Page 104 Quick Start Guide

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF

     ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.Battery Replacemen t

    Open the case by removing the screw in the back. Remove the batteryfrom the battery sockets. If necessary, trim the new battery leads tomatch the length of the existing battery leads. Replace the battery with aLithium LTC-7PN. Close the case and secure with the screw.

    Note: Before replacing the battery, be sure to stop and download the

    data from the logger. This will ensure that the date and time stamp ofthe data set is accurate.

     

    BRIDGE110

    Battery Warning

    ENGINEERING UNITS

    To use the programmable engineering units for applicable devices

  • 8/18/2019 Quick Start Guide-MadgeTech

    53/57

    MadgeTech Page 105 Quick Start Guide 

    Battery Warning

    MadgeTec h data loggers contain a lithium battery. Do not cut open,incinerate, heat above 85 ºC, recharge, submerge under water or otherliquid, or expose the battery to rain or excess humidity. Dispose perlocal regulations. FOR QUESTIONS REGARDING THE SAFETY OF ANY MADGETECH PRODUCTS, PLEASE CALL OUR FACTORY.

    MadgeTech Page 106 Quick Start Guide

    To use the programmable engineering units for applicable devices,select “Engineering Units” from the device menu in the software. Thereare two options for entering the engineering units.

    The first option is to use the “Engineering Units Wizard”. This methodwill automatically calculate the gain and offset values based on theinformation provided. To start the wizard, click on the “Wizard” button inthe “Engineering Units” window and follow these simple instructions:

    1. The Engineering Units Wizard window will appear. The firstscreen is an introduction to the Wizard, click next to continue.

    2. Enter the label of the units that are to be used. Click next tocontinue.

    3. Enter the description of the units to be used. Click next tocontinue.

    4. Enter the range in which the device will be measuring. If thedevice will read 0-100 mA, but only 4-20 mA is going to be used,enter a low limit of 4 and a high limit of 20. Click next tocontinue.

    5. Enter the range in which you want the data to be displayed. Forexample if you want to see 0-14 pH, then enter 0 in the low limit

    box, and enter 14 in the high limit box. Click next to continue.

    6. Click finish to return to the “Engineering Units” screen.

    When finished, click save to store the engineering units in the device. After the software has downloaded all the data, click ok to exit out of theEngineering Units setup.

    The second option is to calculate the gain and offset by hand and enterthose values in the Engineering units gain and offset fields. Engineering

    units consist of two parts, the gain and the offset. In order to calculatethe gain and offset, look at the example below. Here we are trying toconvert mA on a scale of 4 to 20 to pH on a scale of 0 to 14.

    1.14 pH – 0 pH 14 pH

    Gain =20 mA – 4 mA

    =16 mA

    = 0.875

     

    2.

    After calculating the gain and offset enter those values in their

    Offset = 14 pH – Gain * 20 mA = -3.5 pH Calibration 

    MadgeTech’s software allows for our data loggers to be calibrated by the

  • 8/18/2019 Quick Start Guide-MadgeTech

    54/57

    MadgeTech Page 107 Quick Start Guide 

     After calculating the gain and offset, enter those values in theirrespective fields. You will also need to enter a unit description and labelin their respective fields. When finished, click save to store theengineering units in the device. After the software has downloaded all

    the data, click ok to exit out of the Engineering Units setup.

    MadgeTech Page 108 Quick Start Guide

    MadgeTech s software allows for our data loggers to be calibrated by theuser without mechanical calibration. To calibrate MadgeTech devices,simply compare the readings taken by the device to actual knownparameters. To open the calibration window, select “Calibration” under

    the device menu in the software. There are two options in setting thecalibration values into the device.

    The first option is to use the “Calibration Wizard”. This method willautomatically calculate the gain and offset values based on theinformation provided. To start the wizard, click on the “Wizard” button inthe Calibration window and follow these simple instructions:

    1. The Calibration Wizard window will appear. The first screen isan introduction to the Wizard, click next to continue.

    2. Depending on the device to be calibrated, a channel may requirea one or two point calibration.

    a. If the channel requires a one point calibration, then enterthe actual known value in the “Calibration Point” field.Enter the value read by the device in the “Value Read bythe Device at the Calibration Point” field. Click next tocontinue.

    b. If the channel requires a two point calibration, there willbe two sets of two fields. One set will be the “LowMeasurement and Reading (LMR)”, the other set will bethe “High Measurement and Reading (HMR)”. In theLMR section, enter the actual known value in the topbox. Enter the value read by the device in the “Reading”field. In the HMR section, enter the actual known valuein the top box. Enter the value read by the device in the“Reading” field.

    3. Repeat step 2 for each channel the device has. When finished,click finish to return to the “Calibration” screen.

    When finished, click save to store the calibration values in the device. After the software has downloaded all the data, click ok to exit out of thecalibration setup.

     

    The second option is to calculate the calibration values by hand andenter those values in the appropriate fields. Depending on the device tobe calibrated, a channel may require a one or two point calibration.

    NET101

  • 8/18/2019 Quick Start Guide-MadgeTech

    55/57

    MadgeTech Page 109 Quick Start Guide 

    For a one point calibration, there is only the offset to calculate. In orderto calculate the offset, take the value read by the device and subtract it

    from the actual known value. Enter the result in the respective offsetvalue field.

    For two point calibration, there are two parts, the gain and the offset. Inorder to calculate the gain and offset, look at the example below. At thefirst point, the device read 25.5 ºC, but the actual known value was 25ºC. At the second point, the device read 73.0 ºC, but the actual knownvalue was 75 ºC.

    1.

    2.

     After calculating the gain and offset, enter those values in theirrespective fields. You will also need to enter a unit description and labelin their respective fields. When finished, click save to store theengineering units in the device. After the software has downloaded allthe data, click ok to exit out of the Calibration setup.

    73 ºC – 25.5 ºC 47.5 ºC

    Gain = 75 ºC – 25 ºC = 50 ºC = 0.95

    Offset = 73 ºC – Gain * 75 ºC = 1.75 ºC

    MadgeTech Page 110 Quick Start Guide

  • 8/18/2019 Quick Start Guide-MadgeTech

    56/57

     

    MadgeTech Return Policy

    If you have a MadgeTech product that requires replacement, repairs,

    Non-warranty credits will be issued on a case by cases basis with a10% restocking fee.

    The shipping charges for non-warranty repairs will the responsibilityf th t

  • 8/18/2019 Quick Start Guide-MadgeTech

    57/57

    MadgeTech Page 113 Quick Start Guide 

    or upgrading, contact Customer Service at (603) 456-2011, by fax at(603) 456-2012 or by e-mail at [email protected] forMadgeTech RMA #. To process your return, we require the

    following information:

    a) MadgeTech RMA #

    b) Product Description

    c) Product Serial Number

    d) Original MadgeTech Invoice number or your original PO number

    e) Reason for return

    f) If you will be requiring a repair, replacement or a credit for theproduct.

    Please note the above information must be on the paperworkaccompanying the return and that returns will not be accepted without the above information.

     All returned products must be shipped freight prepaid to MadgeTech.For products repaired under warranty, MadgeTech will return theproduct via the same shipping method and will pay the return freightcharges to return the repaired units. If the returned products are notfound to be defective there will be a minimum $25.00 charge as wellas return shipping costs. An additional diagnosis time fee may alsobe charged at the standard rate.

    Products repaired under warranty will retain their original warrantyperiod.

    Products that were originally purchased with a calibration certificatewill be provided with an updated certificate as part of the warrantyrepair.

    Non-warranty repairs are warranted for a period of 90 days from thedate of return shipment.

    MadgeTech Page 114 Quick Start Guide

    of the customer