Purpose Estimate ‘Capsulo Zonular stress’ caused by AcrySof Multipiece IOL when injected without any rotation of the Injector (Shooter / Monarch / Royale).
Recommend a safe technique for Implantation for AcrySof Multipiece IOL.
Analyze the unfolding motions of the AcrySof Single piece IOL.
Recommend easy technique for delivery of AcrySof single piece IOL in bag.
Capsulo Zonular Stress - IndicatorsAvailable space in Ant Chamber
Mean Anterior Chamber Depth (2.93 mm)1 + Mean thickness of Crystalline lens(4.2 mm)2 = 7.13 mm
If the Space Required for Unrestricted Unfolding is > 6mmLeading haptic has to be bent against PC for safe unfolding in AC / Cap bag
CZ stress is directly proportional to:1. Space required for unrestricted unfolding and2. Force required to bend the leading haptic.
1. Tanja M. Rabsilber, MD, Ramin Khoramnia, Gerd U. Auffarth, MD: Anterior chamber measurements using Pentacam rotating Scheimpflug camera: J Cataract Refract Surg 2006; 32:456–459
2. Vargas LG, Peng Q, Escobar-Gomez M, Apple DJ. Overview of modern foldable intraocular lenses and clinically relevant anatomy and histology of the crystalline lens. Int Ophthalmol Clin. 2001;41(3):1-15.
Materials & Methods
AcrySof MA60AC(MP) Mutipiece & SN60WF (IQ) Single piece IOLs
• IOLs injected into transparent plastic chamber filled with OVD.
• Operating microscope used to capture top view. • Detached Slit lamp head used for front view.
Calibration & Measurements
Magnification Correction• Caliper jaws set to 1 mm & Immersed in OVD chamber• Still Image extracted & Scale with smallest marking corresponding to distance between caliper jaws Superimposed •Used this as a ‘mm’ scale for measurements.
Haptic bending force measurement – Set up
• IOL partially delivered till stage requiring maximum vertical space • Haptic pushed on the pan of a precision laboratory weighing scale to
bend on itself till 3 mm • Force required for every mm of bending noted.
Observation - AcrySof Multipiece MA60AC Vertical space required for unrestricted unfolding
Leading haptic points to the right within cartridge barrelTherefore,To ensure that the leading haptic points left and down – ‘Injector’ needs to be turned clockwise by 160° before beginning delivery of IOL
Leading haptic orientation – Prior to Unfolding
Observation -‘Arc of Sweep’ of Leading hapticAcrySof MA60AC - Multipiece
• Relevant only till the final stages of optic delivery• IOL injected without any rotation of delivery system• Superimposition of 1st frame over last frame during unfolding shows ‘arc of sweep’ of leading haptic
Angle between haptic position at onset and completion of unfolding = 170 ̊ clockwise Therefore, Angle by which Injector is to be rotated to maintain constant haptic position throughout Unfolding = 170 ̊ Anticlockwise
170°
Vertical Space required – Direct Shoot – AcrySof IQ
• The ‘Turn, Shoot & Unturn’ technique reduces CZ Stress of AcrySof MP IOL by keeping the leading haptic horizontal and reducing the vertical space required for unfolding.
• The ‘Shoot & Turn’ technique delivers the AcrySof SP IOL in the capsular bag in a single action by countering the rotation of the IOL as it unfolds.