Page 1: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

Quality Control

Page 2: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

HIGH QUALITY FILMS

REPRODUCES ANATOMICAL IMAGE EXACTLY

FILM HAS: CORRECT TECHNICAL FACTORS NO MOTION NO ARTIFACTS PROCESSED CORRECTLY

Page 3: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

RESOLUTION

ABLE TO VISUALLY SEPARATE OBJECTS

BONE VS. TISSUE-- IE: SPATIAL RESOLUTION CR VS CONVENTIONAL FILM

DIFFERENCES BETWEEN DIFFERENT TISSUES-- IE: CONTRAST RESOLUTION CR VS. CONVENTIONAL FILM

Page 4: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

NOISE

FILM GRAININESS (INHERENT) SIZE OF FILM CRYSTALS

STRUCTURE MOTTLE (INHERENT), PHOSPHOR IN SCREEN

QUANTUM MOTTLE

Page 5: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

QUANTUM MOTTLE

LOW NUMBER OF X-RAY PHOTONS INTERACTING WITH SCREEN. INSUFFICIENT AMOUNT OF PHOTONS STRIKING INTENSIFYING SCREEN CAUSING AN INSUFFICIENT LIGHT TO EMIT FROM PHOSPHOR CRYSTALS.

How do we adjust technique for this? CR and Quantum Mottle

Page 6: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

SPEED

FILM’S ABILITY TO RESPOND TO X-RAY EXPOSURE

IF FILM “A” RECORDS MORE DENSITY SOONER THAN FILM “B”, WHICH FILM HAS A HIGHER SPEED?

A

Page 7: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

SENSITOMETRY

MEASUREMENT OF SENSITIVITY TO LIGHT AND THE FILM’S RESPONSE TO CHANGES IN TECHNIQUE AND PROCESSING.

Page 8: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

DENSITOMETRY

MEASUREMENT OF THE DENSITY OF THE FILM BASED ON THE PERCENTAGE OF LIGHT TRANSMITTED THROUGH THE FILM.

Page 9: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

ANATOMY OF A CHARACTERISTIC(H & D CURVE)

Page 10: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.
Page 11: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

OVERHEADS!!!!!!!

Page 12: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.
Page 13: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

RECIPROCITY LAW

OD REMAINS THE SAME NO MATTER THE LENGTH OF EXPOSURE

100 mA X 1 sec = 100 mAs ↓↓

200 mA X .5 sec= 100mAs ↓↓

400 mA X .25 sec =100mAs →→same OD 500 mA X .2 sec = 100mAs ↑↑ Applicable to direct screen exposures

Page 14: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

RECIPROCITY LAW FAILURE

FILM/SCREEN RADIOGRAPHY EXPOSURE TIMES LESS THAT 10

MILLISECONDS EXPOSURE TIMES GREATER THAN

5 SECONDS INTERVENTIONAL STUDIES MAMMOGRAPHY

Page 15: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

RADIOGRAPHIC CONTRAST

IMAGE RECEPTOR CONTRAST

PLUS

SUBJECT CONTRAST

Page 16: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

IMAGE RECEPTOR CONTRAST

Formerly known as film contrast. Why the change in name?

Inherent to film Affected by processing

Page 17: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

SUBJECT CONTRAST

SIZE SHAPE X-RAY ATTENUATION OF THE

ANATOMY AND ENERGY OF BEAM

Page 18: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

REVIEW

L ATITUDE (REMEMBER THIS IN REFERENCE TO THE FILM)

W I D E LATITUDE = LOOOOONNNNG GRAY AKA LOW OR LONG SCALE CONTRAST

NARROW LATITUDE = SHORT SCALE =HIGH CONTRAST

Page 19: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

WIDE LATITUDE FILM, BECAUSE OF LONG GRAY SCALE ALLOWS AS MUCH AT 15% ERROR BY RADIOLGRAPHER

Page 20: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

MAGNIFICATION FACTORS

MOST IMAGES ARE LARGER ON RADIOGRAPH THEN IN ACTUALITY DUE TO SID AND SOD!

FOR EVERY 1” INCREASE IN OID, SID MUST BE INCREASED BY 7” TO COMPENSATE FOR MAGNIFICATION

Page 21: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

SID AND SOD?

SID: SOURCE TO IMAGE DISTANCE

SOD: SOURCE TO OBJECT DISTANCE

MAG. FACTOR IMAGE SIZE OR SID OBJECT SIZE SOD

TO FIND SOD SUBTRACT THE OID FROM THE SID

Page 22: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

IF YOU COULD MEASURE A HEART AND IT MEASURED 12.5 CM BUT ON THE RADIOGRAPH IT MEASURED 14.7 CM, THE MAGNIFICATION FACTOR WOULD BE CALCULATED AS 14.7 DIVIDED BY 12.5 WHICH EQUALS A MAGNIFICATION FACTOR OF 1.17.

Page 23: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

WAYS TO DECREASE MAGNIFICATION

IF OID IS INCREASED? INCREASE SID! CHEST LATERAL AND OBLIQUE C-SPINE

Page 24: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

DISTORTION

Foreshortening Object at an angle and CR is perpendicular

to the part that is angled-trauma elbow Spatial

Tube is angled to separate body parts Clavicle

Elongation object is parallel but tube is angled sigmoid

Page 25: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

FOG LEVEL/CONTRAST AND THE PROCESSOR

As time extends in the developing process the fog level____(toe of the characteristic curve) and the contrast____

increases decreases

Page 26: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

90 seconds is the optimal developer time

A variation of 5 seconds either way can affect quality of film.

Page 27: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

Developer temperature

As the temperature increases the fog level______

increases optimum temperature is usually 90

degrees F.

Page 28: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

SUBJECT CONTRAST

Review: Rad. Contrast=film contrast plus

subject contrast film contrast is controlled latitude of

film, and processing factors

Page 29: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

Subject contrast

Patient (Size) Tissue (KUB) Atomic # ( higher

produces high contrast)

object shape

Page 30: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

MOTION

PATIENT TUBE MOVEMENT

TOMOGRAPHY

Page 31: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

MOTION

LARGE SID = LESS FOCAL SPOT BLUR

SMALL SID = MORE FOCAL SPOT BLUR

Page 32: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

FOCAL SPOT BLUR

Page 33: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

KVP

GIVES ELECTRONS ENERGY TO HIT THE ANODE

HIGH AMOUNT OF KINECTIC ENERGY IN AN ELECTRON PRODUCES A HIGHLY ENERGIZED X-RAY PHOTON

BETTER PENETRATION=LOTS OF GRAYS

Page 34: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

KVP

LOW KVP PRODUCES A BLACK AND WHITE IMAGE.

TO PRODUCE AN ADEQUATE OD MUST USE HIGHER MAS

HIGHER MAS MEANS MORE EXPOSURE TO THE PATIENT

Page 35: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

HIGH KVP HIGH MAS

+Greater margin of error

+Reduction in patient dose

+Wider latitude -increase scatter -increase fog

+Higher OD +Lower Rad noise -increased pt.

Exposure -less margin of

error

Page 36: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

The Radiographer controls

kVp mAs focal spot size distance added filtration screen speed

Page 37: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

REMEMBER

200 mA increased to 300mA is a 50% increase in electrons being produced

As mA increases, time must decrease to maintain same density

Distance Maintenance formula--New Sid over Old Sid Squared

Page 38: Quality Control. HIGH QUALITY FILMS REPRODUCES ANATOMICAL IMAGE EXACTLY FILM HAS:  CORRECT TECHNICAL FACTORS  NO MOTION  NO ARTIFACTS  PROCESSED CORRECTLY.

REMEMBER

Small mA station uses smaller focal spot

smaller focal spot = more detail Inherent = .5 mm additional 2 mm of aluminum between

housing and collimator collimator adds 1 mm