Top Banner

of 20

QC_FB01

Apr 07, 2018

Download

Documents

nhutquatest2
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/6/2019 QC_FB01

    1/20

  • 8/6/2019 QC_FB01

    2/20

    Definitions (1)

    s Quality Control - QC refers to the measures that must be included

    during each assay run to verify that the method is working properly.

    s

    Quality Assurance - QA is defined as the overall program that ensuresthat the final results reported by the laboratory are correct.

    s The aim of quality control is simply to ensure that the results

    generated by the method are reliable. However, quality assurance is

    concerned with much more: that the right method is carried out on the

    right specimen, and that the right result and right interpretation isdelivered to the right personat the right time

  • 8/6/2019 QC_FB01

    3/20

    Definitions (2)

    s Quality Assessment - quality assessment (also known as

    proficiency testing) is a means to determine the quality of

    the results generated by the laboratory. Quality assessmentis a challenge to the effectiveness of the QA and QC

    programs.

    s Quality Assessment may be external or internal, examples

    of external programs include regulatory agencies andreference laboratories.

  • 8/6/2019 QC_FB01

    4/20

    Variables that affect the quality of results

    The educational background and training of the

    laboratory personnel

    The condition of the specimensThe controls used in the test runsReagentsEquipmentThe interpretation of the resultsThe transcription of resultsThe reporting of results

  • 8/6/2019 QC_FB01

    5/20

    Errors in measurement

    s True value - this is an ideal concept which cannot be

    achieved.

    s Accepted true value - the value approximating the true

    value, the difference between the two values is

    negligible (within acceptance criteria).

    s Error - the discrepancy between the result of ameasurement and the true (or accepted true value).

  • 8/6/2019 QC_FB01

    6/20

    Sources of error

    Input data required - such as standards used, calibration values, andvalues of physical constants.

    Inherent characteristics of the quantity being measured

    Instruments used - accuracy, repeatability.

    Observer fallibility - reading errors, blunders, equipment selection,analysis and computation errors.

    Environment - any external influences affecting the measurement.

    Theory assumed - validity of mathematical methods andapproximations.

  • 8/6/2019 QC_FB01

    7/20

    Random Error

    An error which varies in an unpredictable manner, in magnitude

    and sign, when a large number of measurements of the same

    quantity are made under effectively identical conditions.

    Random errors create a characteristic spread of results for any test

    method and cannot be accounted for by applying corrections.Random errors are difficult to eliminate but repetition reduces the

    influences of random errors.

    Examples of random errors include errors in pipetting and changes

    in incubation period. Random errors can be minimized by training,

    supervision and adherence to standard operating procedures.

  • 8/6/2019 QC_FB01

    8/20

    Random Errors

    x

    x x

    x x

    True x x x x

    Value x x x

    x x x

    x

    x

    x

  • 8/6/2019 QC_FB01

    9/20

    Systematic Error

    An error which, in the course of a number of measurements of

    the same value of a given quantity, remains constant when

    measurements are made under the same conditions, or varies

    according to a definite law when conditions change.

    Systematic errors create a characteristic bias in the test results

    and can be accounted for by applying a correction.

    Systematic errors may be induced by factors such as variations in

    incubation temperature, blockage of plate washer, change in the

    reagent batch or modifications in testing method.

  • 8/6/2019 QC_FB01

    10/20

    Systematic Errors

    x

    x x x x x x x

    True x

    Value

    Systematic + BIAS

    Strongly suggest a process error

  • 8/6/2019 QC_FB01

    11/20

    Internal Quality Control Program

    An internal quality control program depend on the use of internal

    quality control (IQC) specimens and the use of statistical methods

    for interpretation.

    Internal Quality Control Specimens

    IQC specimens comprises either (1) in-house sera (single or pooled

    samples), or (2) Reference samples provided by a reference

    laboratory with values within significant analytical ranges (MRL

    minimally but Idealy; 0.5*MRL, MRL and 2*MRL).

  • 8/6/2019 QC_FB01

    12/20

    Quality Control Scheme

    The quality control scheme depend on the use of IQC specimens and isdeveloped in the following manner:

    Use IQC specimen analysed in duplicate for each analytical runs and

    record the individual values.

    Weekly/Monthly/yearly, Calculate the mean and standard deviations(SD) of the individual values obtained

    Make a plot with the assay run number on the x-axis, and measured

    values on the y axis.

    Draw the following lines across the y-axis: mean, -3SD, -2SD, -2SD,

    1SD, 2SD, and 3 SD. (based on normal gaussian distribution) Plot the concentration obtained for the IQC specimen for all analytical

    runs (trend analysis)

  • 8/6/2019 QC_FB01

    13/20

    Statistical Basis for Evaluation

    Mean

  • 8/6/2019 QC_FB01

    14/20

    14

    Objectives of an analytical procedure

    s Able to quantify as accurately as possible each unknownquantity to be determined.

    s After analysis: the difference between returned result xreturned result x and the

    unknown true value true value TT be small or < acceptance limitacceptance limit :

    -- < x - < x - TT

  • 8/6/2019 QC_FB01

    15/20

    15

    Mesured Accuracy profile

    bias (%)

    concentration

    +

    C1 C2 C3 C4

    LLQ ULQRANGE

    a

    ccept a

    ncelim

    its

    dosage interval

    0

    + BIAS

    - BIAS

    Acceptable BIAS

  • 8/6/2019 QC_FB01

    16/20

    Intra Batch Quality Control Estimator

    +15%

    -15%

    0.5*MRL MRL 2*MRL

  • 8/6/2019 QC_FB01

    17/20

    Inter Batch Quality Control Trend Analysis

    1.98 SD

    - 1.98 * SD

    95%CI

    Trend analysis and performance evaluation are vital to assess if the

    method performed with an acceptable accuracy.

  • 8/6/2019 QC_FB01

    18/20

    Warning rules

    Warning 1 : It is violated if 1/3 of IQC bias exceeds 15%

    Warning 2 : It detects systematic. BIAS is systematicaly negative

    or positive.

    Warning 3 : It is violated if consecutive IQC bias lead from apositif to a negative Bias of from a negative to a positive bias.

    This observation may indicate the need to perform instrument

    verification or maintenance.

  • 8/6/2019 QC_FB01

    19/20

  • 8/6/2019 QC_FB01

    20/20

    Follow-up action in the event of a violation

    There are three options as to the action to be taken in the event of aviolation of a pre-established QC rules:

    Accept the test run in its entirety - this usually applies when

    only a warning rule is violated.

    Reject the whole test run - this applies only when a mandatoryrule is violated.

    Accept only results that are within the analytical range that is

    not affected by a violation of either a warning or mandatory

    rule. Re-analyse samples that are affected by rule violation

    (this is always risky)