QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

Mar 31, 2015

Download

Documents

Veronica Pritt
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

QA/QC OF IMAGE DISPLAY EQUIPMENT

Page 2: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

FILM VIEWING EQUIPMENT ("VIEWING BOXES")

The quality and characteristics of the viewing equipment play a decisive role in the recognition of diagnostic details in the radiological image.

The measurable, physical parameter for viewing equipment is the light stream (light quantity) which is emitted from the viewing surface of the equipment.

Page 3: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

If the light stream L (lat. Lumen) is put in relation to the light emitting surface, you produce the measurement parameter

Luminance L: L = Candela / Square meter (cd/m2).

Page 4: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

LUMINANCE

A properly exposed X-Ray film has a mid-optical density of 1.2

To be able to diagnose details in the lower density range of the density curve (D = 0.3 – 0.6), the luminance of the viewing device must be able to be reduced.

Page 5: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

To view high density (Dopt > 2.2) areas of X-Ray films, a strong light source with an iris shutter (“hot light”) is needed. This light source must have a luminance of at least 5000 cd/m2.

Due to their great importance in making diagnostic details of radiographs visible, all viewing devices must undergo an acceptance test and routine constancy tests.

Page 6: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

REPEAT TESTS FOR VIEWING DEVICES, WHICH ARE ROUTINELY USED FOR DIAGNOSIS:

Page 7: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

ACCEPTANCE TEST REQUIREMENTS

Page 8: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.
Page 9: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.
Page 10: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

CONSTANCY TEST REQUIREMENTS

Page 11: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.

IMAGE VIEWING AND DISPLAY DEVICES

Page 12: QA/QC OF IMAGE DISPLAY EQUIPMENT. FILM VIEWING EQUIPMENT ("VIEWING BOXES") The quality and characteristics of the viewing equipment play a decisive role.