Publishing Title: Evaluation of the Systematic Accuracy of Cutting Edge Technology for Stereotactic Radiosurgery Treatment Session title: Physics 1 - Innovative Planning and Delivery Strategies Date/Time: 2014-09-16 13:00 Location: Room D-1 Monitor number: 3 Ning Wen Department of Radiation Oncology Henry Ford Health System 2799 W. Grand Blvd Detroit, MI48202
6
Embed
Publishing Title: Evaluation of the Systematic Accuracy of Cutting Edge Technology for Stereotactic Radiosurgery Treatment Session title: Physics 1 - Innovative.
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Publishing Title: Evaluation of the Systematic Accuracy of Cutting Edge Technology for Stereotactic Radiosurgery
Treatment Session title: Physics 1 - Innovative Planning and Delivery
StrategiesDate/Time: 2014-09-16 13:00
Location: Room D-1Monitor number: 3
Ning WenDepartment of Radiation Oncology
Henry Ford Health System2799 W. Grand Blvd
Detroit, MI48202
The Edge system at Henry Ford
E: 6XFFF, 10XFFF2.5XFFF, HD120MLC
MV DetectoraS1200, 43x43 cm2
Calypso
PerfectPitch® 6D Couch
OSMS x 3
IR camera
Calypso x 3
EDGE E2E Phantom (Prototype) Outer cube dimensions: 15 x 15 x 15 cm3,
Coincidence of isocenters of the imaging systems and Linac is within 1 mm.
The localization accuracy of the Edge platform can achieve submillimeter level.
The targeting accuracy for multiple lesions with shared isocenter is promising. However lesion size and the distance from the isocenter should be evaluated before planning.
Other factors (CT, MR, imaging fusion, contour, tumor motion) should be considered in the SRS/SBRT treatment.