2008-09-08 Pawel Olko 1 Proton beams at IFJ Krakow - present status and future developments Pawel Olko, Institute of Nuclear Physics (IFJ), Kraków
23

Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

Jul 13, 2018

Download

Documents

nguyen_ngoc
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 1

Proton beams at IFJ Krakow -present status and future developments

Pawel Olko, Institute of Nuclear Physics (IFJ), Kraków

Page 2: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 2

Outline

1. Calibration needs for space dosimetry

2. Photons at IFJ - 300 kVp X-rays therapeutic lamp - Cs-137 calibration room- Theratron 780 Co-60- 6-18 MV X-rays (COOK Krakow)

3. Protons at IFJ 2.5 MeV protons from Van de Graaff60 MeV from AIC-144 cyclotron

4. Project of the National Centre for Hadron Radiotherapy(60 -250 MeV protons with scanning beam)

5. Do we need a dedicated calibration protocol (calibration site?) for space dosimetry?

Page 3: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 3

Calibration of space dosimeters

1. Primary and secondary fields complicated- protons + heavy ions- electrons- neutrons- gamma- others

2. There is no single one relevant calibration field for space dosimetry

3. No specific protocol

Page 4: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 4

CERF calibration field for aircrew dosimetry

Page 5: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 5

CERF calibration field for aircrew dosimetry

Page 6: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 6

Calibration laboratory

Cs-137 : 100 nGy/h – 1 Gy/h

Philips 300 kVp –X-rays ; 1 Gy/min

Traceability: Polish Main Office of Measurements

Page 7: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 7

THERATRON 780E

60Co γ-raysDose rate 0.5 -2 Gy/minTraceability: Secondary Standard Laboratory, Centre of Oncology Warszawa (->IAEA)

Page 8: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 8

Van de Graaff accelarator

Beams: 2.5 MeV p2 µm beam spot at 100 pA

Applications: µbeam, single proton irradiation

Page 9: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 9

Van de Graaff accelarator - µbeam

2 µm beam spot at 100 pA

Page 10: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 10

Single Proton Irradiationfaclility at IFJ

2.5 MeV protons

CR-39

Page 11: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 11

AIC-144 cyclotron

- isochronous

- developed at IFJ

- 60 MeV protons

- internal 10 µA

- extracted 2 µA

Page 12: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 12Treatment room(year 2000)

FACILITYFACILITY

Therapy room

Page 13: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 13

Protontherapy room at IFJ

Page 14: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 14

Protontherapy room at IFJ

Page 15: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 15

Dosimetric equipment

Dosimetric diodePTW 60012Farmera chamber

PTW 30010 Unidos dosemeter PTW 10001Markus chmberPTW 23343

Transsmision chamberPTW 7862

Thimble chamberPTW 31002

Unidos dosemeter PTW T-10021

Page 16: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 16

Depth-dose distribution

Range in water - 29 mm Energy at the izocenter 58 MeV

Page 17: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 17

Beam profile

Y - profilefor 40 mm beam diameter

Page 18: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 18

Project on theNational Centre for Hadron Radiotherapy

-NCRH- is on the list of the projects funded by EUStructural Funds with 25 M€

- Pre-agreement signed with Ministry of Science in Oct. 2007

-The decision for locaction of the new facility obtained in May 2008

- Agreement to be signed to the end 2008

- Facility ready 2013

Page 19: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 19

Cyclotrons considered

IBA -235 MeV cyclotron

Accel -250 MeV superconductingcyclotron, power 450 kW

Page 20: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 20

Outline of the NCRH- Phase 1facility with the gantry

EnergyselectorGantry

Cyclotron225-250 MeV

Experimentalroom

Patientpreparation

Radiobiology lab

Eye therapyroom

Page 21: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 21

Outline of the NCRH- Phase 1facility with the gantry

Page 22: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 22

New ion facilities in Europe..

Protons…++ Facilities

- PSI (Eros Pedroni)

- St. Petersburg (Dmitri Seliverstov, Grigory Feofilov)

- IFIMED / Valencia (Jose Barnabeu)

-- Nordic

- CPO, Rinecker, Essen, Trento …

Ions Facilities

- HIT (Thomas Haberer, Udo Weinrich)

- CNAO

- Med-Austron (Erich Griesmaier)

- ETOILE (Jacques Balosso, Marcel Bajard)

- Marburg (Gerhard Kraft)

- MAASTRO / The Ned. (Madelon Pijls-Johannesma)

- BHTC / Belgium (Germaine Heeren, Roger De Croock)

Page 23: Proton beams at IFJ Krakow - present status and future ...wrmiss.org/workshops/thirteenth/Olko.pdfpresent status and future developments ... Calibration needs for space dosimetry 2.

2008-09-08 Pawel Olko 23

Summary

1. Proton and γ-ray calibration available at IFJ for WRIMIS participants

2. In 2013 60 to 250 MeV proton beams will be available at IFJ , including scanning beams.

3. Shall we consider to develop a dedicated calibrationsite? Within the FP7? ESA?