Programming data for Spectroquant ® Test Kits for DR 5000 and DR 6000 photometer from Hach, Hach-Lange Date: January 15, 2013 1 of 49 General information for DR 5000 and DR 6000 when using Spectroquant ® Test kits from Merck. Wavelength specification of the instrument: 190 to 1100 nm Photometric range: ± 3.0 A Standard cells which can be used: 16 mm (only US version), 13 mm (only European version), check your instrument before ordering the cell tests 10, 50 mm, 1-inch round and 1-inch rectangular cells (use the sample cell adapters coming with the instrument) Important: All Merck Spectroquant ® cell tests have a diameter of 16 mm and can only be used in the DR 5000 and DR 6000 US-version. Our Spectroquant ® round tubes 16 mm are not applicable for the 13 mm diameter tubes from the DR 5000 and DR 6000 European version because of the 13 mm shaft. All our Spectroquant ® reagent test kits using the 10, 20 and 50 mm cells can be measured with all DR 5000 instruments. The programming data can be found in the following list. Information about Spectroquant ® test kits: In our Internet homepage www.analytical-test-kits.com users can find all information about our test kits. All package inserts as well as the manuals from e.g. our Pharo spectrophotometers can be found there. This information helps to get a quick overview about the easy handling of our test kits. The package inserts include detailed descriptions about all handling steps, colour reaction, sample preparation if required, interference with other substances, analytical quality assurance data and more. Zeroing the instrument: Before measuring the sample users must enter the “Zero” value for each range. In the below enclosed programming tables it is mentioned for each parameter and measuring range, what “Zero” value is necessary. You can choose between the following options: 1) Own blank (= distilled water and reagents) 2) Cell with blank (white screw cap) 3) Distilled water 4) Clear sample without reagents The advantages of Spectroquant ® Tests kits are, that in some cases users can run as “Zero” distilled water or use the cell with blank (white screw cap) an therefore need not to waste always on determination. This makes live easier and reduce costs. The zero should be entered when the instruments request “Zero”. Message “Under range!” after zeroing the instrument: The instrument must be zeroed for each test or series of tests to establish a zero reference for the measurement. This is done by placing a zero solution (distilled water and reagents) in the cell holder and pressing ZERO. While entering a value of e.g. “25” for the low limit - and not “0” (like Hach) - the message “Under range!” appears in the display. This message doesn’t influence the results. Just go forward with the measurement. If you implement “0” for the low limit this message will not appear (identical as if you use an “original” Hach program). Please contact our local Merck companies or Merck agents for any questions or use the contact form in our internet homepage. No endorsement by Hach Company is implied or intended.
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 1 of 49
General information for DR 5000 and DR 6000 when using Spectroquant® Test kits from Merck. Wavelength specification of the instrument: 190 to 1100 nm Photometric range: ± 3.0 A Standard cells which can be used: 16 mm (only US version), 13 mm (only European version), check your instrument before ordering the cell tests 10, 50 mm, 1-inch round and 1-inch rectangular cells (use the sample cell adapters coming with the instrument) Important: All Merck Spectroquant® cell tests have a diameter of 16 mm and can only be used in the DR 5000 and DR 6000 US-version. Our Spectroquant® round
tubes 16 mm are not applicable for the 13 mm diameter tubes from the DR 5000 and DR 6000 European version because of the 13 mm shaft. All our Spectroquant® reagent test kits using the 10, 20 and 50 mm cells can be measured with all DR 5000 instruments. The programming data can be found in the following list.
Information about Spectroquant® test kits: In our Internet homepage www.analytical-test-kits.com users can find all information about our test kits. All package inserts as well as the manuals from e.g. our Pharo spectrophotometers can be found there. This information helps to get a quick overview about the easy handling of our test kits. The package inserts include detailed descriptions about all handling steps, colour reaction, sample preparation if required, interference with other substances, analytical quality assurance data and more. Zeroing the instrument: Before measuring the sample users must enter the “Zero” value for each range. In the below enclosed programming tables it is mentioned for each parameter and measuring range, what “Zero” value is necessary. You can choose between the following options:
1) Own blank (= distilled water and reagents) 2) Cell with blank (white screw cap) 3) Distilled water 4) Clear sample without reagents
The advantages of Spectroquant® Tests kits are, that in some cases users can run as “Zero” distilled water or use the cell with blank (white screw cap) an therefore need not to waste always on determination. This makes live easier and reduce costs. The zero should be entered when the instruments request “Zero”. Message “Under range!” after zeroing the instrument: The instrument must be zeroed for each test or series of tests to establish a zero reference for the measurement. This is done by placing a zero solution (distilled water and reagents) in the cell holder and pressing ZERO. While entering a value of e.g. “25” for the low limit - and not “0” (like Hach) - the message “Under range!” appears in the display. This message doesn’t influence the results. Just go forward with the measurement. If you implement “0” for the low limit this message will not appear (identical as if you use an “original” Hach program). Please contact our local Merck companies or Merck agents for any questions or use the contact form in our internet homepage. No endorsement by Hach Company is implied or intended.
Cell size used 16 mm round 16 mm round 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 4 of 49
Merck order no. 1.14752 1.14752 1.14752 1.00683.0001 1.00683.0001
Program name Ammonium14752 Ammonium14752 Ammonium14752 Ammonium00683 Ammonium00683
Unit mg/l mg/l mg/l mg/l mg/l
Wavelength 690 nm 690 nm 690 nm 690 nm 690 nm
Resolution 0.01 0.001 0.001 0.1 0.1
Chemical Form 1 NH4-N NH4-N NH4-N NH4-N NH4-N
Calibration “a” 0 0 0 0 0
Calibration “b” 1.23 0.483 0.245 27.8 55.6
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 3.00 1.18 0.500 75.0 150.0
Lower limit 0.05 0.02 0.010 2.0 5.0
Timer 1 (mm:ss) 05:00 05:00 05:00 15:00 15:00
Timer 2 (mm:ss) 05:00 05:00 05:00 ----- -----
Additional Merck information
1) 1) 1) 1) 1)
Cell size used 10 mm 1 inch 50 mm 10 mm 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 5 of 49
Merck order no. 1.14752 1.14752 1.14752 1.00683.0001 1.00683.0001 1.00675.0001*
Program name Ammonium14752 Ammonium14752 Ammonium14752 Ammonium00683 Ammonium00683 AOX00675
Cell size used 10 mm 1 inch 50 mm 10 mm 10 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 10 mm 1 inch 16 mm round 16 mm round 10 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 7 of 49
Merck order no. 1.14834.0001* 1.01745.0001 1.01745.0001 1.01745.0001 1.00858.0001* 1.14815.0001 1.14815.0001
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 8 of 49
Merck order no. 1.00858.0001* 1.14815.0001 1.14815.0001 1.00858.0001* 1.14815.0001 1.14815.0001 1.19251.0001
see appendix page 47 see appendix page 47 180 *)
Program name Calcium00858 Calcium14815 Calcium14815s Calcium00858 Calcium14815 Calcium14815s Carbohyrazide see Oxygen Scavengers
Cell size used 16 mm round 10 mm 10 mm 16 mm round 10 mm 10 mm 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 10 mm 10 mm 50 mm 16 mm round 16 mm round 16 mm round 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 1 inch 50 mm 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 11 of 49
Merck order no. 1.00086/1.00087
1.00088* 1.00086/1.00087
1.00088 1.00608.0001 1.00608.0001 1.00608.0001
Program name Chlorine00086/87/88
Chlorine00086/87/88
Chlorine dioxide00608
Chlorine dioxide00608
Chlorine dioxide00608
Unit mg/L mg/L mg/l mg/l mg/l
Wavelength 550 nm 550 nm 550 nm 550 nm 550 nm
Resolution 0.01 0.001 0.01 0.01 0.001
Chemical Form 1 Cl2 Cl2 ClO2 ClO2 ClO2
Calibration “a” 0 0 0 0 0
Calibration “b” 2.47 0.67 6.33 2.49 1.27
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 6.00 1.000 10.00 3.94 2.000
Lower limit 0.03 0.010 0.10 0.04 0.020
Timer 1 (mm:ss) 01:00 01:00 02:00 02:00 02:00
Timer 2 (mm:ss) ----- ----- 01:00 01:00 01:00
Additional Merck information
3) 3) 3) 3) 3)
Cell size used 16 mm round 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 13 of 49
Merck order no. 1.18750.0001* 1.18751.0001* 1.18752.0001* 1.18753.0001* 1.14560.0001* 1.01796.0001* 1.14540.0001*
Hach Program 431 *) 430 *) 435 *) 435 *)
Program name COD COD COD COD COD14560 COD01796 COD14540
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round
*) The above-mentioned Merck COD cell tests can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach COD tubes. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents). The blank for COD is stabile for e.g. 3 month when storing dark. * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 14 of 49
Merck order no. 1.14895.0001* 1.14690.0001* 1.14541.0001* 1.14691.0001* 1.14555.0001* 1.01797.0001* 1.09772.0001*
Program name COD14895 COD14690 COD14541 COD14691 COD14555 COD01797 COD (Hg free)09772
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents). The blank for COD is stabile for e.g. 3 month when storing dark. * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 15 of 49
Merck order no. 1.09773.0001* 1.17058.0001* 1.17059.0001* 1.14553.0001* 1.14767.0001 1.14767.0001 1.14767.0001
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 16 of 49
Merck order no. 1.14561.0001* 1.09701.0001 1.09701.0001 1.09701.0001 1.14800.0001 1.14800.0001 1.14800.0001
Program name Cyanide14561 Cyanide09701 Cyanide09701 Cyanide09701 Cyanide14800 Cyanide14800 Cyanide14800
Cell size used 16 mm round 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 17 of 49
Merck order no. 1.19250.0001 1.19253.0001 1.19251.0001 1.14557.0001* 1.14557.0001 1.00809.0001* 1.00809.0001
Cell size used 1 inch 1 inch 1 inch 16 mm round 50 mm 16 mm round 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box.As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 18 of 49
Merck order no. 1.14598 1.14598 1.00822.0250 1.14500.0001* 1.14678.0001 1.14678.0001 1.14678.0001
190 *)
Program name Fluoride14598 Fluoride14598 Fluoride Formaldehyde 14500
Cell size used 10 mm 10 mm 1 inch 16 mm round 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 19 of 49
Merck order no. 1.14821.0001 1.09711.0001 1.09711.0001 1.09711.0001 1.14731.0001* 1.14731.0001
see appendix page 47
Program name Gold14821 Hydrazine09711 Hydrazine09711 Hydrazine09711 Hydrogen peroxide14731
Cell size used 10 mm 10 mm 1 inch 50 mm 16 mm round 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 20 of 49
Merck order no. 1.18789.0001 1.18789.0001 1.19251.0001 1.00606.0001 1.00606.0001 1.00606.0001
Cell size used 10 mm 1 inch 1 inch 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 22 of 49
Merck order no. 1.19251.0001 1.14833.0001* 1.09717.0001 1.09717.0001 1.09717.0001 1.00815.0001*
183 *)
Program name Isoascorbic acid see Oxygen Scavengers
Cell size used 1 inch 16 mm round 10 mm 1 inch 50 mm 16 mm round
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to DR 5000 US version. DR 5000 European versions (with 13 mm cell holder) do not allow tests with 16 mm round cells!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 23 of 49
Merck order no. 1.00816.0001* 1.01846.0001 1.01846.0001 1.01846.0001
Program name Manganese 00816
Manganese 01846
Manganese 01846
Manganese 01846
Unit mg/l mg/l mg/l mg/l
Wavelength 445 nm 565 nm 565 nm 565 nm
Resolution 0.01 0.01 0.001 0.001
Chemical Form 1 Mn Mn Mn Mn
Calibration “a” 0 0 0 0
Calibration “b” 4.13 1.67 0.657 0.334
Calibration “c” 0 0 0 0
Calibration “d” 0 0 0 0
Upper limit 5.00 2.00 0.790 0.400
Lower limit 0.10 0.05 0.020 0.005
Timer 1 (mm:ss) 02:00 10:00 10:00 10:00
Timer 2 (mm:ss) 05:00 ----- ----- -----
Additional Merck information
2) 1) 1) 1)
Cell size used 16 mm round 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 24 of 49
Merck order no. 1.01739.0001 1.01739.0001 1.01739.0001 1.14770 1.14770 1.14770 1.19251.0001
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 1 inch 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 26 of 49
Merck order no. 1.01632.0001 1.01632.0001 1.01632.0001 1.14554.0001* 1.14785.0001 1.14785.0001 1.14785.0001
Cell size used 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 27 of 49
Merck order no. 1.14542.0001* 1.14563.0001* 1.14764.0001* 1.00614.0001* 1.14542.0001* 1.14563.0001* 1.14764.0001*
Program name Nitrate14542 Nitrate14563 Nitrate14764 Nitrate00614 Nitrate14542 Nitrate14563 Nitrate14764
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 28 of 49
Merck order no. 1.00614.0001* 1.14773.0001 1.14773.0001 1.14773.0001 1.14773.0001
Program name Nitrate00614 Nitrate14773 Nitrate14773 Nitrate14773 Nitrate14773
Unit mg/l mg/l mg/l mg/l mg/l
Wavelength 340 nm 525 nm 525 nm 525 nm 525 nm
Resolution 1 0.1 0.1 0.1 0.1
Chemical Form 1 NO3- NO3-N NO3-N NO3
- NO3-
Calibration “a” 0 0 0 0 0
Calibration “b” 568 9.62 3.79 42.6 16.8
Calibration “c” 0 0 0 0 0
Calibration “d” 0 0 0 0 0
Upper limit 996 20.0 7.9 88.5 34.8
Lower limit 102 0.5 0.2 2.2 0.9
Timer 1 (mm:ss) 05:00 10:00 10:00 10:00 10:00
Timer 2 (mm:ss) ----- ----- ----- ----- -----
Additional Merck information
2) 1) 1) 1) 1)
Cell size used 16 mm round 10 mm 1 inch 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 29 of 49
Merck order no. 1.09713 1.09713 1.09713 1.09713 1.09713 1.09713
Program name Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713 Nitrate09713
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 30 of 49
Merck order no. 1.14556.0001* 1.14942.0001 1.01842.0001 1.01842.0001 1.14556.0001* 1.14942.0001 1.00609.0001*
Cell size used 16 mm round 10 mm 50 mm 50 mm 16 mm round 10 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 31 of 49
Merck order no. 1.14547.0001* 1.14776 1.14776 1.14776 1.14547.0001* 1.14776 1.14776
Program name Nitrite14547 Nitrite14776 Nitrite14776 Nitrite14776 Nitrite14547 Nitrite14776 Nitrite14776
Cell size used 16 mm round 10 mm 1 inch 50 mm 16 mm round 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 32 of 49
Merck order no. 1.14776 1.00613.0001* 1.14537.0001* 1.14763.0001* 1.19251.0001 1.19251.0001 1.19251.0001
Hach Program 180 *) 181 *) 182 *)
Program name Nitrite14776 Nitrogen00613 Nitrogen14537 Nitrogen14763 Oxygen Scavengers Carbohyrazide
Cell size used 50 mm 16 mm round 16 mm round 16 mm round 1 inch 1 inch 1 inch
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 33 of 49
Merck order no. 1.19251.0001 1.19251.0001 1.14694.0001* 1.00607 1.00607 1.00607 1.01744.0001*
Cell size used 1 inch 1 inch 16 mm round 10 mm 1 inch 50 mm 16 mm round
*) The above-mentioned Merck test can be used with Hach factory-programmed instrument calibrations. No calibration table is necessary. Handling is identical to the corresponding Hach test. Details can be found in the package insert, which is included in the box. As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 34 of 49
Merck order no. 1.14551.0001* 1.00856.0001 1.00856.0001 1.00856.0001 1.00856.0001 1.00474.0001* 1.14543.0001*
Program name Phenol14551 Phenol00856 Phenol00856 Phenol00856 Phenol00856 Phosphate00474 Phosphate14543
Cell size used 16 mm round 1 inch 10 mm 1 inch 50 mm 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to DR 5000 US version. DR 5000 European versions (with 13 mm cell holder) do not allow tests with 16 mm round cells!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 37 of 49
Merck order no. 1.14848 1.14848 1.14848 1.00798.0001 1.00798.0001 1.14848
Program name Phosphate14848 Phosphate14848 Phosphate14848 Phosphate00798 Phosphate00798 Phosphate14848
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 10 mm 1 inch 16 mm round 10 mm 1 inch 16 mm round 10 mm 1 inch As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 10 mm 1 inch 50 mm 10 mm 1 inch 10 mm 1 inch 50 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 43 of 49
Merck order no. 1.14831.0001 1.14831.0001 1.00885.0001* 1.14548.0001* 1.00617.0001* 1.14564.0001* 1.14791.0001
*) *) *)
Program name Silver14831 Silver14831 Sodium00885 Sulfate14548 Sulfate00617 Sulfate14564 Sulfate14791
Cell size used 10 mm 1 inch 16 mm round 16 mm round 16 mm round 16 mm round 10 mm
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 44 of 49
Merck order no. 1.01812.0001 1.01812.0001 1.01812.0001 1.14779.0001 1.14779.0001 1.14394.0001* 1.14394.0001*
*) *) *)
Program name Sulfate01812 Sulfate01812 Sulfate01812 Sulfide14779 Sulfide14779 Sulfite14394 Sulfite14394
Cell size used 10 mm 1 inch 50 mm 10 mm 50 mm 16 mm round 16 mm round
*) Instrument-specific calibration necessary! As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 50 mm 50 mm 10 mm 10 mm 16 mm round 16 mm round 16 mm round 16 mm round As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
Cell size used 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 16 mm round 10 mm As blank can be used: 1) own blank (= distilled water and reagents) 2) cell with blank (white screw cap) 3) distilled water 4) clear sample without reagents * Only applicable to US versions (16 mm round cells) of DR 5000 and DR 6000. The EU version (13 mm cells) doesn’t allow tests!’’
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
from Hach, Hach-Lange
Date: January 15, 2013 48 of 49
Appendix: Measuring range with appendix Each test kit contains a package insert with detailed information regarding the standard procedure of the measurement. In some cases, however, special points have to be considered which are not mentioned in the package inserts. In this appendix you will find specific applications and modified methods for analyses for sensitive measurements indicated with “see appendix page 47”. Calcium: with Spectroquant® Calcium Test, Cat. No. 1.14815.0001 Measuring range: 1.0 – 15.0 mg/L Ca 2.5 – 37.5 mg/L CaCO3 1.4 – 21.0 mg/L CaO Procedure: Add 10 ml of reagent Ca-1A and 8 drops each of reagent Ca-2A and Ca-3A to 1 ml of the sample solution and mix. Leave to stand for 8 min for colour development. After filling into 10 mm cells measure in the photometer at 550 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 1 ml of distilled water instead of the sample solution. Fluoride: with Spectroquant® Fluoride Cell Test, Cat. No. 1.14557.0001 Measuring range: 0.025 – 0.500 mg/l F- Procedure: Pipette 10 ml of sample solution into the reaction cell, mix and add 1 blue microspoon of reagent F-1K. Close the cell tightly and dissolve the solid by shaking. Leave to stand for 5 min for colour development. After filling into 50 mm cells measure in the photometer at 620 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 10 ml of distilled water instead of the sample solution. Fluoride: with Spectroquant® Fluoride Cell Test, Cat. No. 1.00809.0001 Measuring range: 0.025 – 0.500 mg/l F- Procedure: Pipette 10 ml of sample solution into the reaction cell, mix and add 1 blue microspoon of reagent F-1K. Close the cell tightly and dissolve the solid by shaking. Leave to stand for 15 min for colour development. After filling into 50 mm cells measure in the photometer at 620 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 10 ml of distilled water instead of the sample solution Hydrogen peroxide: with Spectroquant® Hydrogen peroxide Cell Test, Cat. No. 1.14731.0001 Measuring range: 0.25 – 5.00 mg/l H2O2 Procedure: The test method of the package insert can be used without any alteration. After filling into 50 mm cells measure in the photometer at 410 nm
Programming data for Spectroquant® Test Kits for DR 5000 and DR 6000 photometer
2- Procedure: Place 1 microspoon of reagent SO3-1K in the cell, close tightly and shake until the complete dissolution of the solid. Then pipette 7 ml of the sample solution into the cell and mix. After the reaction time of 2 min fill into 50 cells and measure in the photometer at 410 nm. Prepare the blank required for the measurement according to the above-mentioned procedure using 7 ml of distilled water instead of the sample solution. Please contact our local Merck companies or Merck agents for any questions. Alternatively contact: [email protected]. No endorsement by Hach Company is implied or intended.