Top Banner

of 72

Prof.arupKRaychaudhuri

Apr 05, 2018

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/2/2019 Prof.arupKRaychaudhuri

    1/72

    Unit for nanoscience and Theme Unit ofExcellence in Nanodevices

    S.N. Bose National Centre for Basic SciencesKolkata-700098

    www.bose.res.in

    Basics of Scanning probe microscopy

    A.K. Raychaudhuri

    SNBNCBS and Bruker School

    December 14-15, 2011

  • 8/2/2019 Prof.arupKRaychaudhuri

    2/72

    Basic concepts

    Simple components of SPM

    Cantilever Statics and Dynamics

    The different modes of SPM

  • 8/2/2019 Prof.arupKRaychaudhuri

    3/72

    I will assume:

    You have used SPM in some form before andhave some acquaintance with it.

    However, the talk is not for experts.

  • 8/2/2019 Prof.arupKRaychaudhuri

    4/72

    SPM

    The Scanning Probe Microscope

    What are the basic components of a SPM

    Localized Probethat has an

    interactionwith

    the substrate tobe imaged

    A nano-positioningmechanism that canposition the probe in

    close proximity

    of the surface

    A system tomeasure the

    interaction of theprobe with the

    substrate

    A mechanism toscan the probe

    relative to thesubstrate andmeasure the

    interaction as

    function of position

  • 8/2/2019 Prof.arupKRaychaudhuri

    5/72

    STM- Quantum mechanical tunneling between atip and the substrate. The contrast comes fromspatial variation of local electronic desnsity of

    states.

    AFM- Localized mechanical (attraction orrepulsion) interaction between tip and surface.

    Physical mechanism and contrast

    Any microscopy will depend on some physicalmechanism to create a contrast spatially.

    It will also need a way to measure the contrastwith spatial resolution.

  • 8/2/2019 Prof.arupKRaychaudhuri

    6/72

    If the process of scanning does not measure thecontrast that has a spatial dependence you will not

    get any image in any scanning microscope.

    Being a computer operated system, any periodic

    noise in the system can create images because thescanning process can add it up to the main signal.These are plain artifacts.

    How to detect artifacts ? A quick thumb rule

  • 8/2/2019 Prof.arupKRaychaudhuri

    7/72

    In contrast to TEM or Optical microscope there isno diffraction and reconstruction of diffracted wave

    front in SPM.

    Advantage:

    Resolution is not diffraction limited.

    Here the limitation comes from the tip size thatinterrogates and of course some fundamentallimitations on detection process and electronics.

  • 8/2/2019 Prof.arupKRaychaudhuri

    8/72

    Different SPMs and different modesThe nature of the tip surface interaction gives

    different types of microscopy.

    The way we detect the response gives us thedifferent modes of SPM.

  • 8/2/2019 Prof.arupKRaychaudhuri

    9/72

    SPM

    The Scanning Probe Microscope (SPM) family

    STM (Tunneling) SFM (Force)

    Scanning ThermalMicroscope (Local

    Temperature)

    STS,STP,ScanningElectrochemical

    Microscope

    Scanning Near FieldOptical Microscope(Optical imaging)

    Atomic ForceMicroscope (AFM)

    Lateral Force (LFM)

    Magnetic Force

    (MFM)

    Electrostatic Force

    (EFM)

    C-AFM

  • 8/2/2019 Prof.arupKRaychaudhuri

    10/72

    Scanning Force Microscope

    It is nothing but a spring balance (the cantilever)

    that is scanned over a surface.

    The cantilever is the precision force detectionelement- we can detect atomic forces

    Type of force of interaction between the tip andsubstrate will determine what we are measuring andthe mechanism that makes the contrast.

    How large are the atomic forces and can we reallydetect them by a cantilever that is much larger?

  • 8/2/2019 Prof.arupKRaychaudhuri

    11/72

    How big is the Atomic Force

    The atomic spring constant

    What is the value of the springconstant of the bond connecting to

    atoms ?

    2 keff/M

    - Is typically in IR range for atomic vibration

    ~ 1013 - 1014cps, M ~ 5 x 10-26 Kg,

    keff= 2

    M ~5 x (1-102

    ) N/m

  • 8/2/2019 Prof.arupKRaychaudhuri

    12/72

    3

    3

    Et wk =

    4L

    1 kf =

    2 m*

    One can make a cantilever as a force measuring elementthat can have the same order of k as that of a molecule.

    w

    L L

    Si elastic modulus (E)[111] Young's modulus= 185GPa

    [110] Young's modulus=170 GPa

    [100] Young's modulus= 130 Gpa

    Si3N4~300 Gpa

    For a Si cantilever :

    t = 5m, w= 20 m, L= 200 mk=10N/m

    It can be softer than atomic

    spring constant

  • 8/2/2019 Prof.arupKRaychaudhuri

    13/72

    L2

    b

    w

    L1

    t: thickness

    m*~0.24(mass of the cantilever)

    3 3 3

    3

    1 2 2

    Et wbk =2b(L -L ) + 6wL

    Engineering cantilevers with different springconstant k- need for different applications

    Advantages:

    1.Less prone to vibrationalnoise.

    2. Can go to lower k orresonance frequency.

  • 8/2/2019 Prof.arupKRaychaudhuri

    14/72

    Estimated radius of curvature of the tip Rt ~ 30 nm

    Kc=0.1 N/m

    Tip

    Engineering cantilevers with different springconstant k-a real triangular cantilever

    Much softer than an

    atomic spring !!!!

    Cantilever

    What ever you do with SFM,the cantilever is the key.

    You need to know it.

  • 8/2/2019 Prof.arupKRaychaudhuri

    15/72

    Some feeling for numbers

    We have a cantilever as a force measuring element.F = k.

    If I want to measure F=1nN, k=1N/m. I should beable to measure a displacement =1 nm.

    Entering the world of nano

  • 8/2/2019 Prof.arupKRaychaudhuri

    16/72

    At the heart of all scanning probe microscope is thecantilever with a tip.

    How we position the tip?How we scan the tip?How we measure deflection of the cantilever?

  • 8/2/2019 Prof.arupKRaychaudhuri

    17/72

    Demystifying AFM-A simple AFM(Home made)

    Laser

    QPD

    Inertial drive piezo

    Scan Piezo

    Electronics

    L. K. Brar, Mandar Pranjape,Ayan Guha and

    A.K.RaychaudhuriDesign and development of

    the scanning forcemicroscope for imaging and

    force measurement with

    sub-nanonewtonresolution

    Current Science , 83,1199 (2002) X-Y micrometer stage

    h i f

  • 8/2/2019 Prof.arupKRaychaudhuri

    18/72

    Schematic of SFM

    DEFLECTION

    SENSOR

    FEEDBACK

    LOOP

    CANTILEVER

    Z-PIEZO

    PROBE

    TIP

    COMPUTER

    XY-PIEZO SCANNER

    Keeps cantileverdeflection or oscillation

    amplitude constant

  • 8/2/2019 Prof.arupKRaychaudhuri

    19/72

    Practical Considerations for AFM/SFM

    1. Cantilever deflection detection system.

    2. Type of cantilevers that can be used.3. Coarse and fine approach mechanism.4. No net relative motion between sample, cantilever and

    detection system.

    5. Scanner range and type of encoder for large sizescanner.6. Data acquisition system ,processing and display

    software.7. Accessibility to all the parts of the SFM and capability of

    using image processing software on stored data.

    Where do the SPM sold by different vendors

    differ?

    B i h ti f SPM

  • 8/2/2019 Prof.arupKRaychaudhuri

    20/72

    Scanner

    Feedback

    A-B

    Pre-Amplifier

    A B

    Quadrant

    Photo Detector

    Tip &

    Cantilever

    Basic schematic for SPM

    To Z-Piezo

    Laser

    ADC

    DAC

    Need for

    calibration

    Keeping something

    constant, need for feed back

    X-Y scanner

    Z-scanner

    Coarse approach vs fine approach

    Pixels

    bits

    PID

    C lib i f i f SFM i i l

  • 8/2/2019 Prof.arupKRaychaudhuri

    21/72

    Calibration of scanning stage of SFM using commercial2-D grating

    The grating has 2160 lines/mm

    1000m/2160=0.46mThe calibration: 40nm/V

    Brar et.al

    (2002)

  • 8/2/2019 Prof.arupKRaychaudhuri

    22/72

    Topography

    Can take care of

    image distortion

    Arranging spheres of PS in an array by self-assembly

    Sub 500nm level calibration, works fine to 20nm

    Can find the size by Electron microscope or DLS

    Soma Das

    (2008)

    C lib ti i t i

  • 8/2/2019 Prof.arupKRaychaudhuri

    23/72

    Mica

    Freshly cleaved

    7 nm x 7 nm

    Calibration in atomic range-A freshly cleaved surface

    Can we assume a linear calibration ?

    The piezo -scanner is non-linear and

    has hysteresis

  • 8/2/2019 Prof.arupKRaychaudhuri

    24/72

    Other calibrations:Z-Calibration- large scale vs small scale

    Force calibration-detection of exact k?

    O ti l h d d D t ti l t i f i

  • 8/2/2019 Prof.arupKRaychaudhuri

    25/72

    Optical head and Detection electronics for scanning

    Scanner

    Feedback

    A-B

    Pre-Amplifier

    A B

    Quadrant

    Photo Detector

    Tip &

    Cantilever

    To Z-Piezo

    Laser

    ADC

    DAC

  • 8/2/2019 Prof.arupKRaychaudhuri

    26/72

    Optical lever =

    = 500 -100(for l=100mm)

    L(Length of the laser path)

    l(Length of the cantilever)

    Main components of the optical

    stage:

    1. Laser diode

    2. Cantilever

    3. Quadrant photo-detector (QPD)

    4. Collimating lenses5. Mirror

    QPD is used as a position

    sensitive detector, its output

    signal is proportional to the

    position of the laser spot.

    Why we need smaller cantilever ?

    C lib ti f th ti l t

  • 8/2/2019 Prof.arupKRaychaudhuri

    27/72

    Calibration of the optical stage.

    0 1 2 3

    -2

    0

    2

    A-B(V)

    Z-displacement(cm)

    Region of Gradient: 1000m

    Detects 4V for 1000m movement

    1mV electrical noise , positional reolution~1/4m

    Using optical lever of 100, we can detect cantilever deflectionof ~ 1/400 m=2.5 nm.

    Source of noise in AFM

  • 8/2/2019 Prof.arupKRaychaudhuri

    28/72

    Atomically resolved steps in Ti terminated SrTiO3substrate-reaching the limits

    Size of step (1/2 unit cell) ~0.38nm

    Courtesy Dr. Barnali Ghosh.

    Taken in CP-II

    Resolution from optical detection

  • 8/2/2019 Prof.arupKRaychaudhuri

    29/72

    Resolution from optical detection

    0 1 2 3

    -2

    0

    2

    A-B(V)

    Z-displacement(cm)

    Region of Gradient: 1000m

    Detects 4V for 1000m movement, 1mV electrical noise~1/4m.

    Reduce noise to 0.1 mV,

    Using optical lever of 100, we can detect cantilever deflection

    of ~ 1/4000 m=0.25 nm.

    Often it is good to have acantilever tip rest on asurface and record the

    output as a function of time

    We have the base responseof the QPD, need to enhance

    optical lever and reduceelectrical noise to get better

    resolution

  • 8/2/2019 Prof.arupKRaychaudhuri

    30/72

    Quadrant photo-detectors

    Why use 4 quadrant detector ?

    Vertical deflection of cantilever-Topography

    Lateral deflection of cantilever-

    Lateral Force Microscopy (LFM)

  • 8/2/2019 Prof.arupKRaychaudhuri

    31/72

    Thermal Noise limited resolution

    If k is reduced the force sensitivity is increased

    Cantilever displacement = Force/k

    K ~ 0.1N/m , displacement of 1nm will come from aforce of 100pN

    Does any thing limit us ?

    Yes it is the thermal noise.

    It can be very high for soft cantilevers (those withvery small k)

    Th l N i li i d l i

  • 8/2/2019 Prof.arupKRaychaudhuri

    32/72

    Thermal Noise limited resolutionFor any oscillatory system we can apply

    Equi-partition theorem

    2/12/1

    2

    2*2

    2*2

    ,,

    2

    1

    k

    Tkz

    Vmzksystemharmonic

    VmzkTk

    B

    B

    For a 0.1N/m cantilever the thermal noise inducedroot mean-square amplitude 0.14 nm.For a deflection of 1nm of the cantilever it is asubstantial amount.

    Force uncertainty~(10014)pN

  • 8/2/2019 Prof.arupKRaychaudhuri

    33/72

    I have discussed some of the basic concepts of theSFM and the main components that go with it and

    their functions as well as limitations.

    Cantilevers and force detection, Scanner

    calibrations, Optical detections and sources ofnoise

    It will be best if your reflect upon your experienceof using SFM and connect to this presentation

  • 8/2/2019 Prof.arupKRaychaudhuri

    34/72

    Cantilever Statics and Dynamics

    The different modes of SPM

  • 8/2/2019 Prof.arupKRaychaudhuri

    35/72

    Source: PhD thesis Soma Das , SNBNCBS

  • 8/2/2019 Prof.arupKRaychaudhuri

    36/72

  • 8/2/2019 Prof.arupKRaychaudhuri

    37/72

    Statics and Dynamics of cantilever

    Interaction between the tip and the substrate willdecide the nature of force and hence the statics and

    dynamics of the cantilever

    Tip sampleinteraction model

    D i f til

  • 8/2/2019 Prof.arupKRaychaudhuri

    38/72

    Dynamics of cantilever

    Any force velocity willadd to damping andreduce amplitude of

    vibration-dissipation

    Any force displacement will change

    the frequency ofvibration

    Different types of force microscopy depends onthe dynamics of cantilever and the mode of

    detection

    tjFekzdtdz

    dtzdm 2

    2

    Simple ball and spring model

    Driving term

    for dynamicmode

  • 8/2/2019 Prof.arupKRaychaudhuri

    39/72

    Static mode (contact mode)

    AFM

    tjFekz

    dt

    dz

    dt

    zdm

    2

    2

    Fkz=0

  • 8/2/2019 Prof.arupKRaychaudhuri

    40/72

    Static mode:

    Mostly for contact-mode the cantilever deflection

    is such that the bending force is balanced by theforce of interaction:

    F(z) =-U/z=-k.z

    U = Total energy that includes the surface as well aselastic deformation energy.

    26)(

    z

    HR

    zf

    tTS

    5.10

    0

    )( )(*3

    4

    62

    zaREa

    HRf t

    tzTS

    a0~Atomic dimension (hard sphere)

    E*~ Effective elastic constant

    Rt- Tip radius of curvature.

    H=Hamakar cosntant

  • 8/2/2019 Prof.arupKRaychaudhuri

    41/72

    26

    )(z

    HRzf tTS

    5.10

    0

    )( )(*3

    4

    62

    zaREa

    HRf t

    tzTS

    Elastic force winsover. The deformation

    of the surface shouldbe larger than thefeatures you would

    like to see

  • 8/2/2019 Prof.arupKRaychaudhuri

    42/72

    Si tip pressingon Si substrate

    One can evaluatethe contact radius

    Herzian contact

    The contact areadepends on Elastic

    modulus

    A th b l t l t til i t t

  • 8/2/2019 Prof.arupKRaychaudhuri

    43/72

    A thumb rule to select cantilever in contactmode imaging

    Cantilever touching a surface is like two springsconnected back to back, The force applied isbalanced by displacement

    The softer spring wins

    substratecantilevereff

    eff

    appl

    substrate

    appl

    cantilever

    appl

    substratecantilevertotal

    kkk

    kF

    kF

    kF

    111

    .

    A th b l t l t til i t t

  • 8/2/2019 Prof.arupKRaychaudhuri

    44/72

    A thumb rule to select cantilever in contactmode imaging

    A surface with mixed k (elastic constants) like acomposite of soft and hard matter will not image thetopography. What you image is actually a mixtureof both

    substratekeffkcantileverksubstratek

    cantileverk

    effk

    cantileverk

    substratek

    ,

    ,

    Correctcondition fortopography incontact mode

    The softer spring wins

    Will image theelasticallydeformed surface

  • 8/2/2019 Prof.arupKRaychaudhuri

    45/72

    Some tips for good contact mode imaging

    Get a soft cantilever that is realistically needed.

    Do a force spectr0scopy (F-d) curveHave some idea about the elastic modulus of the

    surface you image.

    For soft materials when you cannot have verysoft cantilever use LFM

  • 8/2/2019 Prof.arupKRaychaudhuri

    46/72

    ODT self-assembled monolayer on Ag

    Sai and AKR, J.Phys.D Appl. Phys. 40, 3182 (2007)

  • 8/2/2019 Prof.arupKRaychaudhuri

    47/72

    Some useful applications of

    contact mode AFM

    Force spectroscopy

    Piezo-force spectroscopy

    Conducting AFM

    Local charge measurements

  • 8/2/2019 Prof.arupKRaychaudhuri

    48/72

    tj

    Fekxdt

    dx

    dt

    xdm

    2

    2

    Dynamic mode

    Driving force

    Controlled byexperimenter

    Force of interaction oftip with substrate and

    surrounding

  • 8/2/2019 Prof.arupKRaychaudhuri

    49/72

    Dynamic mode (all non-contact modes):

    Cantilever is modulated at resonance frequency and

    the shift in resonance frequency , phase oramplitude measures the force gradient

    -F/z=-k+(2U/z2)

    2)(6

    ))((tz

    HRtzf tTS

    5.10

    0

    ))(( ))((*3

    4

    62

    tzaREa

    HRf t

    ttzTS

    i d h d d

  • 8/2/2019 Prof.arupKRaychaudhuri

    50/72

    Dynamic mode -what do we do ?

    Oscillate the cantilever at close to resonance

    frequency

    Interaction with the substrate will change theresonance frequency and /or amplitude of

    oscillation (through the viscous force on thesurface)

    Detect the departure from resonance or damping

    detected by amplitude, phase or frequency shift asthe cantilever scans the surface

    This leads to contrast and the imaging

  • 8/2/2019 Prof.arupKRaychaudhuri

    51/72

    Dynamics of cantilever

    22222

    0

    0

    22222

    0

    22

    00

    0

    2

    2

    )(

    )/)(()Im(

    )(

    ))(/()Re(

    )(

    mFz

    mFz

    eztz

    Fekzdt

    dz

    dt

    zd

    m

    j

    tj

    In dynamic mode spectroscopy the resonance curve

  • 8/2/2019 Prof.arupKRaychaudhuri

    52/72

    In dynamic mode spectroscopy the resonance curveand its modifications during imaging provides the

    image

    what happens to resonance frequency in dynamic

  • 8/2/2019 Prof.arupKRaychaudhuri

    53/72

    what happens to resonance frequency in dynamicmode when there is additional force

    z

    Uf

    z

    fk

    z

    Ukkeff

    ,)( 02

    2

    0

    eff

    omk0Start with a cantilever that is free

    Shift in resonance frequency when the interaction is turned on

    eff

    effeffeff

    eff

    m

    f

    m

    f

    m

    f

    m

    k

    2

    0

    '

    0

    '

    0

    20

    '

    0

    '

    2

    0

    '

    0

    2

    1

    2

    11

    Force derivative is theimportant parameter in

    dynamic mode

  • 8/2/2019 Prof.arupKRaychaudhuri

    54/72

    NC

    Tapping

    55

  • 8/2/2019 Prof.arupKRaychaudhuri

    55/72

    Force Derivative

    NC

    Tapping

  • 8/2/2019 Prof.arupKRaychaudhuri

    56/72

    Two paradigms of dynamic mode

    Detection by amplitude modulationIf the resonant frequency of a cantilever shifts, then theamplitude of cantilever vibration at a given frequencychanges. Near a cantilevers resonant frequency, this changeis large.

    Non-contact (tip does not touch the substrate,) -This also encompasses the EFM and MFM.

    Tapping or IC mode (the tip touches the surface atsome part of the swing)

  • 8/2/2019 Prof.arupKRaychaudhuri

    57/72

    The set frequency is somewhat larger than the freeresonance frequency.

    Non-contact

    IC/tapping-mode

  • 8/2/2019 Prof.arupKRaychaudhuri

    58/72

    The set frequency is somewhat smaller than the free

    resonance frequency.

    F i l ti f d t h t h t th

  • 8/2/2019 Prof.arupKRaychaudhuri

    59/72

    From simulation of data-what happens to theresonance curve in Tapping mode

    Das, Sreeram,AKR , Nanotechnology 18, 035501 (2007),Nanotechnology21, 045706 (2010),Journal of Nanoscience and Nanotechnology 7, 2167

    (2007)

    Sample: MicaK= 0 68N/m

  • 8/2/2019 Prof.arupKRaychaudhuri

    60/72

    -1.0 -0.5 0.0 0.5 1.0 1.5 2.00.000000000

    10.000000475

    20.000000950

    30.000001425

    40.000001900

    50.000002375

    60.000002850

    70.000003325

    80.000003800

    Sample:Mica

    Amplitude(nm)

    Tip-sample separation (m)

    approach(41nm)

    retract(41nm)

    approach(70nm)

    retract(70nm)

    approach(90nm)

    retract(90nm)

    Amplitude vs. distance curves for mica for three different free vibration amplitude of

    the cantilever.

    K= 0.68N/mResonance Frequency = 86KHz

    61

  • 8/2/2019 Prof.arupKRaychaudhuri

    61/72

    Amplitude vs Height

    (in absence of feedback)

    Application of Non contact mode

  • 8/2/2019 Prof.arupKRaychaudhuri

    62/72

    Application of Non-contact mode

    Magnetic Force Microscopy

    MFM

    Measuring long-range force

    Any force that decays slower than

    inverse square

    26

    )(z

    HRzf tTS

    2,)( nz

    Azf

    nlong

  • 8/2/2019 Prof.arupKRaychaudhuri

    63/72

    26

    )(z

    HRzf tTS

    This mode is realized by employing suitable probes

  • 8/2/2019 Prof.arupKRaychaudhuri

    64/72

    This mode is realized by employing suitable probes(magnetic tip) and utilizing their specific dynamicproperties.

    MFM is an important analytical tool whenever thenear-surface stray-field variation of a magneticsample is of interest.

    MFM can be used to image flux lines in low- andhigh-Tc superconductors . MFM have evenextended local detection of magnetic interactions

    to eddy currents and magnetic dissipationphenomena .

  • 8/2/2019 Prof.arupKRaychaudhuri

    65/72

    The interpretation of images acquired bymagnetic force microscopy requires some basicknowledge about the specific near-fieldmagnetostatic interaction between probe andsample.

    How to take care of the topography ???

    The magnetic stray field produced by a magnetized

  • 8/2/2019 Prof.arupKRaychaudhuri

    66/72

    g y p y gmedium and the contrast mechanism

    effm

    F2

    0

    '

    0

    '

    0

    2

    1

    The shift in frequency the

    MFM detects is the gradient

    of the magnetic force

    Magnetic Force Microscopy of hard disk

  • 8/2/2019 Prof.arupKRaychaudhuri

    67/72

    g py

    (No applied field)

    MFM maps themagnetic domains on

    the sample surface

    Stored data in a

    hard disk

    The stray field is maximumwhen the anisotropy is

    perpendicular

    Magnetic Force Microscopy (with applied field)

  • 8/2/2019 Prof.arupKRaychaudhuri

    68/72

    g py ( pp )

    Requirements for MFM tips

  • 8/2/2019 Prof.arupKRaychaudhuri

    69/72

    Requirements for MFM tipsThese tips can be coated with a thin layer of magneticmaterial for the purpose of MFM observations.

    A lot of effort has been spent on the optimization ofmagnetic tips in order to get quantitative information fromMFM data .

    The problem is that in the coating of conventional tips, apattern of magnetic domains will arrange, which reducesthe effective magnetic moment of the tip. The exact domainstructure is unknown and can even change during MFM

    operation.Best tip is the one that has a single mono-domainmagnetic particle !!!!!

  • 8/2/2019 Prof.arupKRaychaudhuri

    70/72

    Lorentz Microscopy of fieldaround a tip

    Effect of tip sharpness

  • 8/2/2019 Prof.arupKRaychaudhuri

    71/72

    Stray field line scan

    Observed

    Simulated

    Ordinary tip Mono-domain

    tip

    In SFM , what ever you do the most significant

    role is played by the tip and the cantilever

  • 8/2/2019 Prof.arupKRaychaudhuri

    72/72

    I have tried to give a basic introduction to SFM andsome of its different modes and shared my

    experience with you.SFM images are not just picture gallery

    The more knowledge you acquire and morequantitative you become you can get more value

    from your SFM.

    Thank you