Top Banner
Prof. of Clinical Chemistry, Mansoura University
30

Prof. of Clinical Chemistry, Mansoura University.

Dec 15, 2015

Download

Documents

Reyna Skilton
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Prof. of Clinical Chemistry, Mansoura University.

Prof. of Clinical Chemistry, Mansoura University

Page 2: Prof. of Clinical Chemistry, Mansoura University.

Implies a self – regulating process with an element of feed-back which detect any tendency to malfunction

and readjusts the equipment so that it continues to function correctly.

Page 3: Prof. of Clinical Chemistry, Mansoura University.

Definitions and processing concepts:

Continuous flow analyser: The reagent are pumped continuously by

syring pump The samples are introduced at regular

intervals The flow is segmented to separate one

specimen from anothor This type is used as batch analyzer e.g

technicon .Two types :

a) single channel continuous flow: analyse one constituent in each specimen .

b) Multichannel continuous flow: determination of several components in each sample

Page 4: Prof. of Clinical Chemistry, Mansoura University.

Discrete analyser:

Samples are processed in separate

reaction tubes.

The instrument contain a variety of

receptacle in which the sample and

reagents are mixed.

The reaction occurs: cuvettes, dry

film slides.

Some containers contain the

reagents and only the specimen

need to be added.

Others add both reagents and

specimen.

Page 5: Prof. of Clinical Chemistry, Mansoura University.

Reaction vessels and cuvets in discretes: Reused or discarded

Used as cuvet or just reaction tube,

aspirated in tube then to another tube

for measurement For wash reused reaction vesseles or cuvets

aspiration of the reaction mixture

a detergent ,alkaline or acid wash soln . is

then dispensed into and aspirated out of

the cuvets

Page 6: Prof. of Clinical Chemistry, Mansoura University.

Parallel analysis: All specimens are subjected to a series of

analytical processes at the same time in a

parallel fashion:

Discretionary multiple – channel

analysis. The specimen can be analyzed by any

one or by more than one of the

available processes (methods,

channels).

Page 7: Prof. of Clinical Chemistry, Mansoura University.

- A number of specimens are processed in the same analytical session , or run:

Batch analyzer.

Page 8: Prof. of Clinical Chemistry, Mansoura University.
Page 9: Prof. of Clinical Chemistry, Mansoura University.

The components steps in automated system

1. Specimen identification.2. Specimen preparation.3. Specimen handling, transport and delivery.4. Specimen processing.5. Sample transport and delivery. 6. Reagent handling and storage. 7. Reagent delivery.8. Chemical reaction phase.9. Measurement approach.10. Signal processing data handling and process

control.

These operations are usually performed sequentially but in some instrument, they may be combined and occur in parallel.

Page 10: Prof. of Clinical Chemistry, Mansoura University.

Specimen identification:

Page 11: Prof. of Clinical Chemistry, Mansoura University.

Specimen Preparation:

Page 12: Prof. of Clinical Chemistry, Mansoura University.

Specimen Handling, Transport and Delivery:

Page 13: Prof. of Clinical Chemistry, Mansoura University.

Specimen processing:

Page 14: Prof. of Clinical Chemistry, Mansoura University.

Sample transport and delivery:

In continuous – flow system: the sample is aspirated through the

sample probe into continuous reagent stream

In Discrete analyzers: the sample is aspirated into sample probe and

delivered, with reagent into reaction tube or cup .

Page 15: Prof. of Clinical Chemistry, Mansoura University.

Carry over:

Carry over occurs when a previous sample or reagent contaminates

successive tests in a run causing the next sample to have an aberrantly

higher or lower results . Carry over occurs when anlytes occur in extremely high levels e.g. enzyme

or h CG in pregnancyCarry over also occurs in systems

that reuse cuvettes that are insufficiently washed after each

testing cycle.In discrete systems with disposable reaction vessels and measuring cuvettes carry over is caused by the pipeting system

Page 16: Prof. of Clinical Chemistry, Mansoura University.
Page 17: Prof. of Clinical Chemistry, Mansoura University.
Page 18: Prof. of Clinical Chemistry, Mansoura University.
Page 19: Prof. of Clinical Chemistry, Mansoura University.
Page 20: Prof. of Clinical Chemistry, Mansoura University.
Page 21: Prof. of Clinical Chemistry, Mansoura University.

Disadv: one analyte per rotor but now recent models allow multiple analysis 24 rotors can be loaded at one time.

In Continuous flow: by glass coils passing through the samples and

reagents.

Page 22: Prof. of Clinical Chemistry, Mansoura University.

Measurement approaches:

Spectrophotometry Fluorimetry e.g. IMX

Chemiluminescenc e.g. IMMULITE

Electrochemical for electrolytes

Page 23: Prof. of Clinical Chemistry, Mansoura University.

Examples:

Page 24: Prof. of Clinical Chemistry, Mansoura University.
Page 25: Prof. of Clinical Chemistry, Mansoura University.

Automated immunoassay systems differ from chemical analyzers in that they require the use of:

1.solid – phase reagents to separate bound and unbound

2.More sensitive detection methods e.g fluorescence and chemiluminescence

3.Special handling of the reagents:

e.g. 1. Thermal conditioning to keep chemilum substrate stable. 2. Constant agitation to keep microparticles suspended.

Page 26: Prof. of Clinical Chemistry, Mansoura University.

4-Multiple Calibrators: (upto 6)

• Calibration usually nonlinear

Example : ACS : 180

• benchtop • R-A • Imm . AS.S ay

IMMULITE :

• bench – top • R-A • Imm. Assay

Automated immunoassay

Page 27: Prof. of Clinical Chemistry, Mansoura University.

Selection of instruments

Role of an automated system in the workflow of

the laboratory.

Identification of candidate system (RA or

batch).

Cost effectiveness of the system.

User-friendiness (interface between the

machine and the human operator).

Analytical performance.

Page 28: Prof. of Clinical Chemistry, Mansoura University.

Advantages of automation:

Large number of samples may be processed with minimal technician.

Two or more methods may be performed simultaneously.

precision is superior to that of manual performance.

calculations may not be required.

Page 29: Prof. of Clinical Chemistry, Mansoura University.

Disadvantages of automation: limitation of the methodology "closed

system".Expensive.

Many systems, are impractical to use for small numbers of samples.

Therefore, back up manual methods may be required for individual

emergency analysis. Back up methods must also be available in the event of

instrumental failures.

Page 30: Prof. of Clinical Chemistry, Mansoura University.