PRO OX ® -100 Kit PROGRAMMABLE DIGITAL OXYGEN MONITOR US & FOREIGN PATENTS ISSUED & PENDING STATE-OF-THE-ART TECHNOLOGY FOR PRECISE OXYGEN MONITORING MADE IN THE USA
Jul 25, 2016
PRO OX®-100 Kit
PROGRAMMABLE DIGITALOXYGEN MONITOR
US & FOREIGN PATENTS ISSUED & PENDING
STATE-OF-THE-ARTTECHNOLOGY FORPRECISE OXYGENMONITORING MADE IN THE USA
About
• The PRO OX-100 programmable handheld digital oxygen monitor is designed to give
operators the most precise oxygen level readings possible.
• The PRO OX-100 comes equipped with many advanced features such as its data logging
capabilities, which allow operators to create permanent records of real time data (at 15
second intervals) and export up to 50 data points to Microsoft® Excel and plain text format.
• The PRO OX-100 also offers a built-in, programmable multi-language feature with on-screen
instructions that are available in English, Spanish, German, and Portuguese. Together with
different voltage and frequency specifications and compatibilities, the PRO OX-100 is the
most universal oxygen weld monitor available on the global market today.
Features and Benefits
1) Data Logging Capabilities
Download Driver Software by completing this form.
2) Archive Records with Software Reporting
•Create permanent records of real time data (at 15 second intervals for continuous mode or as
needed in spot mode) of oxygen levels for critical welding operations
3) Easy-to-Use Conversion Software
•Capture and export 50 data points in just clicks to Microsoft® Excel and plain text format
4) Convenient and Quick Data Offload
•The PRO OX-100 offloads data to a PC via a convenient USB interface at a high-speed, ensuring
data integrity
5) SPOT:
•Use this mode if the purging process requires more than 20 minutes to extend battery life
•Use this mode for simultaneous weld applications
6) CONTINUOUS :
•Use this mode when permanent records are required (in conjunction with data logging)
•Use this mode when PPM is near required level
7) Self-Calibrating and Certified
•Calibrates using ambient air. No special calibration gas required.
•The PRO OX-100 Oxygen Monitor is factory calibrated and certified, accompanied by a dated
certificate of authentication
•Meets requirements specified by the National Institute of Standards and Technology
8) Self-Calibrating and Certified
•Calibrates using ambient air. No special calibration gas required.
•The PRO OX-100 Oxygen Monitor is factory calibrated and certified, accompanied by a dated
certificate of authentication
•Meets requirements specified by the National Institute of Standards and Technology
9) Audiovisual Alarm
•Eliminates guesswork as audiovisual alarm can be set to desired O2 PPM value
•Unit will produce an intermittent beep sound simultaneously as a green light flashes
•Operator is alerted even in noisy or poorly lit environments and can continue evacuating oxygen,
reset the alarm or commence welding
10) Rechargeable 9V Battery & Charger Designed For Universal Voltage
•Compatible with different voltage and frequency specifications from around
the world
•Can operate on 100VAC/60Hz, 120VAC/60Hz and 220VAC/50Hz
configurations
•Rechargeable battery is the environmentally friendly solution
•Can be used over and over, saving on purchases
•Any non-rechargeable 9V battery can be substituted
•Charger also serves as a standalone power supply when operation will exceed
one hour in length
11) CE Designation
The PRO OX-100 is in compliance with EC Directives
12) Languages
Programmable in four popular languages to satisfy international customer needs
English
German
Portuguese
Spanish
How PRO OX-100 Works
Installing the Neoprene Extension Tubing:
Attach neoprene extension tubing with female fitting to the Quick Connect “male” fitting on the top
right side of the unit
Pull gently to ensure the connection is securely in place
Insert the stainless steel probe to the end of the neoprene tube without a fitting
Before taking an oxygen measurement, ascertain both connections are tightly secured
To Set Language:
Press and hold the “MENU” button
Next press the “RIGHT” → button until you see “SETUP: LANGUAGE” displayed
Press the “ENTER” button to select language setup
Choose the preferred language (German, Spanish, Portuguese, English)
Press “ENTER” button to Save and Exit
Press “MENU” button to return to main screen
To Set Date and Time:
Press the “Menu” button to display the “SETUP” menu
Select the “SET DT/TM” option by pressing “ENTER”
Use the “RIGHT” → and “LEFT” ← arrow buttons to increase or decrease “HOUR”, “MINUTE”,
“SECOND”, “MONTH”, and “YEAR” value, using the “ENTER” button to save settings
To Set Alarm ON/OFF:
Press the “Menu” button and use the “RIGHT” →Arrow button until “SETUP: SET ALARM” is
displaying
Press the “ENTER” button to select “ALARM SETUP”
Use the “RIGHT” → and “LEFT” ← Arrow buttons to increase or decrease “ALARM” set value
When Alarm is disabled (factory setting), use the “RIGHT” and “LEFT” ← arrow buttons to display PPM
value
5 Attempting to decrease the set point below 100 PPM, will disable the alarm, by pressing the “LEFT” Arrow
button
6. Press “Enter” to save settings and exit
To View Log:
Press the “Menu” button and use the “RIGHT” → arrow button until “SETUP: VIEW LOG” is
displaying
Press the “ENTER” button to select the “LOG” viewer
Use the “RIGHT” → and “LEFT” ← arrow buttons to navigate through the log. The screen will
display up to a total of 50 data points
Press “Enter” to save settings and exit
To Clear Log:
Press the “MENU” button and use the “RIGHT” → arrow button until “SETUP: ERASE LOG” is
displaying
Press the “ENTER” button to begin erasing all data points in the log
3. Once the screen displays “DONE”, all the data points have been erased from the log
4. Press “Enter” to save settings and exit
Calibrating the PRO OX-100:
When the PRO OX-100 unit is at the main screen, the operator can press the “AUTO CAL” button to
begin calibration
The unit will automatically begin to sample ambient air and automatically calibrate
Once the calibration is complete, the unit will display “CALIBRATION DONE”, and immediately return
to the main screen
The unit is now ready to use
Spot Reading Mode:
When the PRO OX-100 unit is at the main screen, press the “SPOT” button, and the unit will then begin
running
Once the spot reading is complete, the PRO OX-100 will display the oxygen percentage value
Press the “LEFT” ← arrow button to return to main screen
Continuous Reading Mode:
When the PRO OX-100 unit is at the main screen, press the “CONT” button
During operation, the unit will constantly update the latest oxygen level
To exit out of continuous mode, press the “CONT” button, after the screen displays a reading
NOTE: Audiovisual Alarm is factory set to “Alarm Disabled”
Download PRO OX 100 Brochure here: http://www.aquasolwelding.com/wp-content/uploads/2014/09/pro-ox-100-brochure.pdf
Types & Sizes: PRO OX-100 Kit Programmable Digital Oxygen Monitor