PRINCIPLES OF BEAM DIRECTION AND USE OF SIMULATORS Moderator- Dr Bhavana Rai
PRINCIPLES OF BEAM DIRECTION AND USE OF SIMULATORS
Moderator- Dr Bhavana Rai
What is beam direction
During treatment planning whole plan of the treatment is worked out in advance of the actual treatment and certain devices are used to direct the beam towards the tumor , so that better tumor control is achieved with minimum complications
Co-60Tumor
Beam direction
Why it is required?
• In order to reduce unnecessary irradiation to patient
- beam size should be minimum required to cover desired . treatment zone
-beam must be accurately aimed towards the zone
Better tumor control Low normal tissue complication
Better Therapeutic ratio
How is Beam Direction achieved?
Using Beam direction device in radiotherapy machine
Planning in advance
Accurate calculations
Beam Direction Device
Collimators
Front and Back pointer
Pin and arc
Isocentric Mounting
Laser
Light Field
Distance Meter
Breast Cone
What is Collimation?
• Some radiation is emitted out in every direction,
• Even the useful beam is far greater than is needed for most treatments
• Methods, to produce beam of the shape and sizes, so that desired zone is irradiated and normal tissues are spared is k/a COLLIMATION
Collimators
• Radiation source is mounted, surrounded with metal box, except for a thin ‘window’- ‘Fixed’ or ‘Master’ collimator
Fixed Collimators
• Radiation allowed to escape is usually accepted for which primary exposure rate at the edge is 50% of that at centre
20°
• Protects the patient from bulk of the radiation.
• Dictates the maximum field size for the machine.
Maximum field sizeCo-60 -> 35 35 cm Linac -> 40 40 cm
Treatment or movable collimator
• Placed below master collimator required for
- cutting down the penumbra
- defining the required field size
• Types - Applicators - Movable jaws
APPLICATORS
• As electron scatter is so significant, applicators are necessary to delineate the field sharply close to the patient
Applicators: Design• Applicators are constructed
of low Z materials (e.g. aluminium, plastic, etc.) to minimise x-ray contamination.
• A frame is generally provided near to or at the final section, into which irregular or non-standard size cut-outs can be positioned for tertiary field shaping.
Metal Plate with hole
Metal Sheet
Box
Plastic Cap
• They indicate immediately the shape and size of beam
• Give useful indication of its direction
• Position the patient at desired distance
• Acts as compressor- by pressing the applicator end into the tissues
Jaws/Movable Diaphragms
• Are blocks of lead which by a suitable mechanical linkage can be made to move in and out at the turn of a handle
• With the gantry vertical and the head at 0 degree-
• Y2 - at the gantry end X2 - on right side (i.e. supine patient’s left) as viewed facing the gantry.
• A strip, a centimetre or so wide, of copper-tungsten alloy is placed on the inner faces of the jaws
• Also called heavy alloy• It has more attenuation
than lead
Front and Back Pointer method
• Based on the principle that any line can be defined by two points
• Entry point- A Tumor centre- T Exit point- B
Pros and cons
Pros• Easily available• Easy to use• Cheap• Accurate
Cons• Accurate localization of
tumor required• Requires skin marks-
unreliable• Front and back points must
be acessible
Pin And Arc
• Consists of - Ruler(R) - Slider(S) - Arc(T) - Pin(V) - Carriage(U)• Based on principle of
Parellogram
Isocentric Mounting
• The axis of rotation of the three structures:– Gantry– Collimator– Couch
coincide at a point known as the Isocenter.
Why Isocentric Mounting
• Better accuracy• Faster setup• Easier transfer from
simulator to machine
LASERS
• Lasers used to identify isocentre• Laser lines pass through isocentre
perpendicular to each other• A typical set of lasers comprises -Cross lasers (coronal and lateral
lines) placed on both walls lateral to isocentre
- Line laser (sagittal) placed at foot of bed passing through isocentre
-Cross laser (lateral and sagittal) placed on ceiling above isocentre
LIGHT FIELD• In order to enable the treatment
staff to see where the radiation beam will be located, a light beam is generated that coincides with the useful limits of the beam.
• by using a high intensity light bulb containing a small filament that must be optically coincident with the source position
• Cross on field indicates central axis (and wedge direction)
DISTANCE METER• Uses the light field and a
second light source to indicate distance from radiation source to patient surface
• Second light source marked with lines at set distances
• Intersection of the two light beams varies with distance to surface – indicating distance
Steps of advance treatment planning
Positioning Immobilisation Localisation Field
selectionDose
distribution Calculations Execution & verification
PATIENT POSITIONING
• Defn-• It is one of the weakest link in treatment
planning process• Good patient position is ALWAYS:
– Stable.– Comfortable.– Minimizes movements.– Reproducible.
Standard Positions
• MC used body position.• Also most comfortable.• Best and quickest for setup.• Minimizes errors due to
miscommunication.
• Best for treating posterior structures like spine
• In some obese patients setup improved as the back is flat and less mobile.
Supine
Prone
Positioning aids/ device• Help to maintain patients in non standard
positions.• These positions necessary to maximize
therapeutic ratio.• Accessories allow manipulation of the non rigid
human body to allow a comfortable, reproducible and stable position.
Breast Board
Other Positioning Devices
IMMOBILIZATION
Mould
Thermocast
• Thermoplastics are long polymers with few cross links.
• They also possess a “plastic memory” - tendency to revert to normal flat shape when reheated
Other Immobilization devices
Localization
• The target volume and critical normal tissues are delineated with respect to patient’s external surface contour.
• What to localize?– Tumor– Organ
• Methods?– Clinical examination– Imaging
Clinical localization
Advantages;
– Available everywhere. – Cheapest and quickest.– Needs little additional
equipment.
Disadvantages;– Error prone in the wrong
hands.– Accessible areas
required.– Volumetric data not
easily obtained.
Clinical localization is mandatory despite advanced imaging – need to know what to image!
Estimation of depth
• From data gained by localization studies:– CT / MRI – Accurate data– Lateral height method– Tube shift method
• Depth estimation necessary for:– Calculations– Selection of beam energy
Lateral height method
d
d
H1 + H2
2d =
H1H2
Tube shift method
• Image shift and tube shift are interrelated WHEN the tube to target distance remains constant.
• Goal: To obtain a graph of different object heights against the tube shift.
• Serial measurements of image shift measured (for same tube to film distance) while varying the height of the markers above the table.
Tube shift principles
Marker
d2
yf
S
Tumor
x1x2
d1
Calculation
d1
fy
d2
x1x2
x1
S= d1
f – d1
S
x2
S= d2
f – d2
y = d2 – d1
= fx2 + S
x2 -x1 + S
x1
Tumor
Marker
Imaging Localization
• Imaging:– X-rays– USG scans– CT scans– MRI scans – PET scan– Fusion imaging
• Type of study selected depends on:– Precision desired.– Cost considerations– Time considerations– Labour considerations
X-Ray localization • X-rays made to pass through
concerned site, which is captured by photographic film or digital detector
• Pros - cheap - takes less time - easily available• Cons - radiation exposure - 2D view
Follow up
USG scan• Uses ultrasound waves• USG generated by transducer
focussed on concerned part echoes received transformed in digital image
• Pros - cheap - no raditation exposure
• Cons - needs expertise to interpret
CT Localization• An X-ray procedure that
combines many X-ray images with the aid of a computer to generate three-dimensional images of the internal organs and structures of the body
MRI Localization
• Uses a magnetic field and pulses of radio wave energy to make pictures of organs and structures inside the body
• Soft tissue structures are better visualised
PET Localization• Entails imaging of biodistirbution of a
radiolabeled compound selected based on its biochemical behaviour
• FDG-PET-> demonstrate metabolically active disease
• 18F-FLT(flouro deoxy thymidine)-a marker for cellular proliferation can be used to assess response to radiation therapy( FDG is inferior as radiation causes inflammation which is picked up by FDG)
• Cu-ATSM demonstrateshypoxia ---> radiation resistance
• 18F-FMISO• Choline PET- recurrent prostate cancer• 11C-metomdate-adrenocortical tumors
Fusion Imaging
• Includes MRI-CT fusion and MRI-PET fusion, PET-CT fusion
• Bony landmarks visualized better in CT
• Soft tissues better visualized in MRI,
• Metabolically active disease better in PET
Field Selection
•Single field•2 Field•Multiple field
Tumor can be
irradiated by one or
more radiation
beam
Single Field
• Criteria for acceptability:1. Dose distribution to
be uniform (±5%)2. Maximum dose to
tissues in beam ≤ 107%.
3. Critical structures don’t receive dose exceeding their normal tolerance.
• Situations used:– Skin tumors– CSI– Supraclavicular
region– Palliative treatments
2 Field techniques• Can be :
– Parallel opposed– Angled
• Perpendicular• Oblique
– Wedged pair
• Advantages:– Simplicity– Reproducibility– Less chance of
geometrical miss– Homogenous dose• Dose homogeneity depends on:
– Patient thickness– Beam energy– Beam “flatness”
Multiple fields• Used in 3DCRT & IMRT• Used to obtain a “conformal” dose
distribution in the modern radiotherapy techniques.
• Disadvantages:– Integral dose increases– Certain beam angles are prohibited due to
proximity of critical structures.– Setup accuracy better with parallel opposed
arrangement.
Field Shaping
Custom blocking
Applicators
Independent jaws
Multi leaf collimators
Breast cone
Custom blocking
Multi leaf collimators• Developed in Japan in 1980s• In Europe first used by Scanditronix f/b Philips(Elekta) and
Varian and rest• Instead of a single block of metal, these have up to 80 pairs of
leaves that can move independently, allowing any beam shape to be produced subject to the width of the leaves.
Features • Used to shape radiation field• Can interdigitate• Fields set automatically• 52 – 160 leaves• Moves independently• Typically 60mm tall• 2.5% leaf transmission; 4% interleaf leakage• Tongue and groove offset• Focused in 1 direction• 7mm penumbra
Types of MLCsThere are three main types of MLC:a) Type A (e.g. Scanditronix and Siemens): The MLC provides
all the collimation (except for the primary collimator) and completely replaces the standard collimation system
b) Type B (e.g. Elekta): The MLC provides field shaping but additional shielding is provided by backup collimators. The MLC (including the backup collimators) replaces the standard collimation system
c) Type C (e.g. Varian): The MLC provides field shaping in addition to the standard collimation system. Although part of the head assembly, it is externally mounted and complements the standard collimation system
Dose distribution analysis
• Done manually or in the TPS.• Manual distribution gives a hands on idea of
what to expect with dose distributions.• Inefficient and time consuming.• Pros:
– Cheap– Universally available– Adequate for most clinical situations.
Calculations
SSD technique (PDD method)
SAD technique
Clarkson’s technique
Computerized
Verification Of Treatment
• The primary purpose of treatment verification is to verify the treatment volume under actual conditions
• Done when the patient is positioned on couch in treatment room just before giving radiation
Various Methods
Port films
• Films are of poor quality• Viewing is delayed as time required for processing is long
Electronic portal imaging
• Possible to view images instantaneously• Although images are of poor quality
Cone beam CT
• Can be kilovoltage or megavoltage type• Acquires images in therapy room coordinates• kV CBCT has far superior quality
Execution
• When everything is verified and treating doctors feel that beam direction is adequately achieved then execution of plan is done by giving prescribed radiation dose.
RADIOTHERAPY SIMULATORS
RADIOTHERAPY SIMULATORS
• Is an apparatus that uses a diagnostic x-ray tube but duplicates a radiation treatment unit in terms of its geometrical, mechanical, and optical properties
• Called simulation because the treatment is being "simulated," or not really given.
Need Of Simulators(a) Manual marking not accurate due to anatomical
uncertainities.(b) Geometrical relationship between the radiation beam and
the external and internal anatomy of the patient cannot be duplicated by an ordinary diagnostic x-ray unit.
(c) Field localization directly with a therapy machine by port film quality is poor.
(d) Field localization is time-consuming, if carried out in the treatment room.
(e) Unforeseen problems with a patient set-up or treatment technique can be solved.
Function• It displays the treatment fields so that the target volume may
be accurately encompassed without delivering excessive irradiation to surrounding normal tissues
• By radiographic visualization of internal organs, correct positioning of fields and shielding blocks can be obtained in relation to external landmarks.
Types
X-ray simulators
CT simulators
4D CT
MRI Simulator
X-Ray(conventional) Simulators• Major sub-systems are-gantry,
collimator, x-ray tube, imaging unit, patient support system(couch), remote control console
• The variable focus to axis distance makes it suitable for therapy simulation for number of teletherapy machines
• X-ray passes through one side of the patient and attenuated beam is converted to digital image by image intensifier
Pros and Cons of X Ray simulator
Pros • Cheapest among all
simulators• Less radiation exposure
than CT• Fast and efficient
Cons • Only 2 D planning possible• Inability to accurately
distinguish the different densities of areas- making it difficult to visualize anatomy
Best for checking radiotherapy plans and planning palliative treatments very quickly and efficiently
CT Simulator• A dedicated radiation therapy
CT scanner with simulation accessories such as flat table, lasers, immobilization and image registration device, and software for virtual simulation
• Uses CT scanner to localize the treatment fields
• Positioning, levelling snd imobilization is done in treatment position
• Image are sent electronically to preserve electron density data
What is Virtual Simulation• A software(exclusively written
for simulation) provides outlining of contours, target volumes and critical structures, interactive portal displays and placement, review of multiple plans, and a display of isodose distribution- k/a virtual simulation
• It is so called as patient is represented by CT images and radiation machine by beam geometry and expected dose distirbution
Pros and Cons of CT Sim
Pros• Gives volumetric
distribution of dose• DRR, 3DCRT, IMRT, VMAT
techniques can be done• Treatment planning
algorithms are developed specifically for CT-attenuation coefficient can be identified precisely
Cons • More radiation to Patient• Suboptimal tissue contrast• Lack of functional
information• Difficulty in visualize small
group of cancer cells that are separated from gross tumor
4D-CT Breathing/organ motion -problems with accurate target definition (moving targets may appear with distorted shapes and in wrong locations on CT) -increased irradiation of normal tissues (larger fields are often used to ensure that the tumor is not missed). 4D CT• Takes respiratory motion into account that
can be used for planning, delivery and verification
• Takes phase images of deep inspiration-> mid inspiration-> mid expiration->deep expiration
• Other images such as MIP, minMIP can be generated
How does it worksInfrared marker-
moves with breathing motion
Infrared camera -records movement
Converted to digital data
Encorporated to CT imaging
Pros and Cons of 4D CT Simulator
Pros • Takes respiratory/organ
motion into account• Highly precise techniques
such as gating can be done
Cons • Respiratory motion is not
constant
MRI Simulators• No MRI only simulator at present
• Due to major drawback - Inherent geometric distortion in the images - Low signal intensity of bone causes difficulty in obtaining DRRs - lack of inherent electron density information
• But it’s demand is rising due to - Better soft tissue contrast - No radiation exposure - By using MR gel dosimetry that utilises radiosensitive material that can be irradiated and then ‘read out’ by virtue of the changes in magnetic properties thus,a true 3D dosimetric analysis may be performed with good accuracy, if properly callibrated
Treatment Plan
Two volumes overlayed
MRI dosimetry data
Current Practice
Is still to fuse to CT for this purpose
Proper fusion is not possible until and unless MRI is taken in treatment position
Treatment position not possible every time in radio diagnosis MRI machine due to narrow bore of gantry and curved couch
Newer MRI machines have come up with 70 cm wide bore and flat top and also lasers installed to facilitate CT fusion
Future Aspects of MRI simulators
• Work is going to develop planning algorithms for MRI based planning
• Pseudo DRRs are being researched where CT data is used as a surrogate for MR signal intensity
• MR gel dosimetry is being further researched and calibration mapping is going on, which can provide better dosimetric calculations
IT IS THE FUTURE OF SIMULATION
CONCLUSION
• Accuracy is vital for treatment • Better to use kilovoltage beam with careful beam
direction than megavoltage beams without it• To neglect extra accuracy that can be gained by beam
direction is to throw away much of the value of powerful and expensive apparatus now in use in radiotherapy
• Simulators is an important part of advance planning where treatment planning can be made and reviewed
thankyou