PRE-KLENZ™ Point of Use Processing Gel Dispensing and Cling Properties

Technical Data Monograph

POINT OF USE PROCESSING GELC L E A N S T A R T S H E R E

Page 2: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

2

Table of Contents

1. Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2. Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Dispensing and Coverage properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Cling properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

3. Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Dispensing and Coverage properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Cling properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

4. Conclusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Page 3: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

3

1Purpose

Pretreatment of soiled surgical instrumentation at point of use is recommended by multi-industry guidelines1,2,3 in order to begin and facilitate the decontamination process. A variety of pretreatment products are available in the market today, although many have drawbacks when it comes to ease of use and efficacy. PRE-KLENZ Point of Use Processing Gel is a ready-to-use, neutral pH gel designed to keep soils moist and initiate the cleaning process on reusable surgical instruments at the point of use. The gel technology allows PRE-KLENZ Point of Use Processing Gel to cling to instrument surfaces longer than products dispensed as liquids or foams, and the easy-dispense package enables instrument coverage with the simple push of a button, at any angle.

2Methods

Dispensing and Coverage propertiesTo compare the dispensing and coverage properties of PRE-KLENZ Point of Use Processing Gel versus other commercially available pretreatment products, the products were dispensed onto approximately 18.5 cm diameter paper circles. Product dispensing and cov-erage patterns were documented photographically.

Cling propertiesTo compare the cling properties of PRE-KLENZ Point of Use Processing Gel versus other commercially available pretreatment products, the products were dispensed onto pre-weighed stainless steel panels. After application, the panels were weighed, suspended vertically for 30 seconds, and then weighed again. The percentage of the product remaining on the panels was then calculated.

1 Association for the Advancement of Medical Instrumentation (AAMI). ANSI/AAMI ST79:2010/ A4:2013: Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities, Section 6.4. AAMI, Arlington, VA; 2013.

2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials, Test Methods, and Acceptance Crite-ria for Cleaning Reusable Medical Devices, Section 5.2.2. AAMI, Arlington, VA; 2011.

3 Association of periOperative Registered Nurses (AORN). Guideline for cleaning and care of surgical instruments, Recommendation III G. In: Guidelines for Periop-erative Practice. AORN, Inc, Denver, CO; 2015.

Page 4: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

4

Dispensing and Coverage propertiesThe dispensing and coverage patterns of a pretreatment product are indicative of the efficacy and ease of use of the product. In order to adequately pretreat soiled instrumentation, a product needs to provide thorough coverage in an efficient manner. The product also needs to have controlled application that does not splash off the surface to which it is applied. Such splashing can carry soil off instru-ments onto surrounding surfaces or into the air.

PRE-KLENZ Point of Use Processing Gel dispensed easily with a one-touch button actuator and quickly provided broad, thorough, uniform coverage to the area in which it was applied. This can be seen in Figure 1, where PRE-KLENZ Point of Use Processing Gel covered and wetted the circle to the point of wrinkling the surface. Comparatively, other commercially available pretreatment products utilize traditional trigger-sprayers that dispense product in ways that produce splattering or that make it difficult to ensure soiled instruments are properly covered with product in an efficient manner. Although the coverage area of OptiPro™ Gel4 Instru-ment Pre-Cleaner was broad, the coverage itself was uneven, leaving pockets of only light to essentially no treatment throughout the sprayed area. Both Prepzyme® Multi-Tiered™ Enzymatic Foam Spray5 and Blu62™ Pretreatment Foam6 provided a limited, narrow area of coverage with each spray, which would make covering an entire tray of instruments quite inefficient and ergonomically more difficult. The EmPower Foam™ 7 sprayer propelled product out at erratic angles such that the majority of the product dispensed did not land where intended making adequate tray coverage extremely challenging, and creating a product exposure hazard. Renuzyme Foam Spray8 dispensed in a strong stream, leading to significant amounts of splashing that could propel soil out of the instrument tray being treated, thus creating a biohazard.

3Results

4 OptiPro™ is a trademark of Ecolab USA, Inc.5 Prepzyme® and Multi-Tiered™ are trademarks of The Ruhof Corporation6 Blu62™ is a trademark of Stryker Instruments7 EmPower Foam™ is a trademark of Metrex Research8 Renuzyme Foam Spray is distributed by Getinge USA, Inc.

Page 5: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

5

Figure 1: Dispensing pattern of pretreatment products

Cling propertiesIn order for a pretreatment product to initiate the cleaning process by loosening soil, it must remain on instruments for extended periods of time without running off. PRE-KLENZ Point of Use Processing Gel clings to soiled instruments, remaining on the sur-faces longer than many other commercially available pretreatment products. Figure 2 demonstrates the amount of product remaining on the surface of a stainless steel panel after vertical suspension for 30 seconds. Approximately 80% of PRE-KLENZ Point of Use Processing Gel remained on the panel as compared to the competitive products, some of which maintained only around 10% of the initial application amount. Figure 3 depicts the superior cling of PRE-KLENZ Point of Use Processing Gel when compared to other pretreatment products.

PRE-KLENZ Processing Gel EmPower Foam OptiPro Gel

Prepzyme Multi-Tiered Enzymatic Renuzyme Foam Blu62

Page 6: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

6

Figure 2: Percentage of product remaining on a surface after 30 second vertical suspension

0

10

20

30

40

50

60

70

80

90

100

PRE-KLENZ™ Point of Use

Processing Gel

OptiPro™ Gel Instrument Pre-Cleaner

Bio-ZolveInstrument

Presoak SprayGEL

EmPower Foam™

Prepzyme® XFExtreme Foam

RenuzymeFoam Spray

Prepzyme® Multi-Tiered™

Enzymatic Foam Spray

Blu62™ Pretreatment

Foam

% R

emai

ning

on

Pane

l

Cling Evaluation on Stainless Steel PanelsVertical suspension for 30 seconds

Page 7: PRE-KLENZ Point of Use Processing Gel Dispensing and Cling ... · 2 Association for the Advancement of Medical Instrumentation (AAMI). AAMI TIR30:2011: Compendium of Processes, Materials,

7

Figure 3: Cling property

PRE-KLENZ OptiPro GelPrepzyme Multi-

Tiered Enzy-matic

Bio-Zolve Spray Gel EmPower Foam

Blu62 Pretreatment

Foam

Initial

After 30 sec in vertical position

The controlled application of PRE-KLENZ Point of Use Processing Gel provided broad, uniform coverage versus other commercially available pretreatment products. Additionally, PRE-KLENZ Point of Use Processing Gel exhibited superior surface clinging ability, allowing for longer retention on soiled instrument surfaces.