200544 D Pre-Installation Instructions & Check List Note: You must complete and return the check list on the last page of these instructions in order to schedule set-up and training for your TD-100. Manufactured By: 6810 Kitimat Rd, unit 1 Mississauga, ON L5N-5M2 Phone: (800) 701-7450 Phone: (905) 826-6330 Fax: (905) 826-3288
8
Embed
Pre-Installation Instructions Check List · Pre-Installation Instructions & Check List ... Clamps to secure the plumbing installation As Req ... Fax completed checklist to Phoenix
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
200544 D
Pre-Installation Instructions
&
Check List
Note: You must complete and return the check list on the last page of these instructions in order to schedule set-up and training for your TD-100.
Manufactured By:
6810 Kitimat Rd, unit 1
Mississauga, ON
L5N-5M2
Phone: (800) 701-7450
Phone: (905) 826-6330
Fax: (905) 826-3288
2 of 8 200544 D
Introduction
Before beginning this installation, make certain that you have read and understand
these instructions. If you are uncertain of any of these requirements please call
Customer Service at Phoenix Airmid Biomedical (800) 701-7450. This installation
should conform to your local building code requirements and be performed by
qualified personnel. Upon completing the Pre-Installation, send photos and
completed check list on the last page to Phoenix Airmid Biomedical.
CAUTION! Failure to follow the instructions could result in voiding
the warranty.
TD-100 Pre-Installation Kit is intended for use by building engineering to
prepare the room for the TD-100.
Space Planning
Your TD-100 is a floor standing unit that is secured to the wall using brackets that
are provided. Make sure to allow space beside and above the TD-100 as shown in
figures 1 and 2.
Figure 1. Typical Setup for the TD-100.
3 of 8 200544 D
The on/off switch, electrical and fluid connections are on the left side of the TD-
100 and will require at least 6” of clear space. Where possible the right side should
allow at least 6” for the probe connector bracket as shown in figure 1.
Figure 2. Mounting and space requirements.
If you plan to have additional equipment installed with your TD-100 please plan
for the additional space at this time.
Electrical Planning
A 20 Amp hospital grade GFCI receptacle is required near the TD-100. The TD-
100 comes with a 6’ cord that attaches near the top left side. The receptacle must
be within reach of the cord.
4 of 8 200544 D
Plumbing Planning – Water Supply
A high quality COLD potable water source that meets Federal Clean Water
Standards at the point of use is necessary. The water source must be regulated to
20-25 psi. The source must be within 8’ of the left side or 6’ of the right side of
the TD-100.
Plumbing Planning – Drain
The drain is one of the most important parts to the TD-100s fume management
system. The drain has two jobs to do with the TD-100, first is to carry off the
waste fluids and second is to help contain the smell of the disinfectant. The TD-
100, a trap, drain fitting with the drain filter having a direct connection to the riser
is the ideal setup.
Given practical limitations, the TD-100 may have to share a drain or connect to a
lateral. If TD-100 shares a drain connection the TD-100 should be after any shared
traps and the lowest connection on the shared drain line to prevent fumes from
escaping from the other drains. See figure 3.
“S” traps and good pitch may allow other drain configurations to work. Remember
that the key is not to allow the liquids from the TD-100 to accidentally run into
another trap that is open to the room.
Items included in Pre-Installation Kit
Match all the items in the kit with those on the list below. If any items are missing
or damaged, immediately contact CS Medical.
Item Part # Quantity Pre-Installation Instructions 200544 1
Drain Fitting 200498 1
Drain filter 200499 1
Quick Disconnect ¼” MNPT 200500 1
5 of 8 200544 D
Materials Required but not Supplied
Description Quantity P-Trap 1-1/2” (that will accept 1-1/2” PVC) 1
Water shut-off valve 1
Pressure Regulator (Watts 263A) ¼” NPT outlet 1
Pressure gauge for regulator (0-60 psi) 1
Assorted pipe As Req
Clamps to secure the plumbing installation As Req
Installation Step 1 – Install the GFCI Receptacle
A hospital grade GFCI receptacle is required where the TD-100
™ will be plugged
in. The receptacle should be located as close as possible to the TD-100™
(Six foot
power cord supplied with TD-100).
The following diagram is a typical installation. Please place items in Pre-
Installation Kit as needed to meet the installation requirements shown in the
diagram.
Figure 3. Electrical and Plumbing Locations.
6 of 8 200544 D
Installation Step 2 – Assemble the Drain Fitting
Assemble the 1½” drain and P-trap as shown below; making sure that the top of
the pipe is no higher than 18”. See figure 3. Cement (or attach with a Fernco
coupling or similar) the drain fitting PN 200498, supplied in Pre-Installation kit to
the P-trap as shown below.
Installation Step 3 – Place the Drain Filter on the Drain Fitting
Place the drain filter PN 200499 on top of the drain fitting as shown in below. DO
NOT CEMENT. This filter is intended to be replaced.
CAUTION! DO NOT BOND THE DRAIN FILTER IN PLACE. IT
IS INTENDED TO BE REPLACED.
Drain Fitting
P/N 200498
Drain Filter
P/N 200499
7 of 8 200544 D
Installation Step 4 – Regulated Water Supply
Install a water pressure regulator (Watts 263A recommended) to a high quality
potable COLD water supply (high quality potable water will meet the federal clean
water standards at the point of use) with a shut-off valve. Adjust regulator outlet
pressure to 20-25 psi static. Install the quick disconnect fitting (PN 200500). The
flow rate should be a minimum of 1gpm @ 20 psi.
Installation Step 5 – Schedule Set-up and Training
Email photos showing the completed area including the plumbing preparation
along with the completed check list for the TD-100 to [email protected] to
schedule your set-up and training.
You may also fax the completed Pre-Installation Check list to Phoenix Airmid
Biomedical to schedule your TD-100 set-up and training. FAX# (905) 826-3288.