POINT OF CARE ANEMIA DEVICE BME 400 Final Report December 11, 2015 Team Members: Carly Hildebrandt – Team Leader Katie Barlow – Communicator Jeff Wu – BPAG Lizzie Krasteva – BSAC Michal Adamski - BWIG Advisor: Dr. Thomas Yen Client: Dr. Philip Bain
30

POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

Nov 12, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

POINTOFCAREANEMIADEVICE

BME400FinalReportDecember11,2015

TeamMembers:CarlyHildebrandt–TeamLeader

KatieBarlow–CommunicatorJeffWu–BPAG

LizzieKrasteva–BSACMichalAdamski-BWIG

Advisor:Dr.ThomasYenClient:Dr.PhilipBain

Page 2: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

2

AbstractAnemia affects twenty-five percent of the global population, being most prevalent in the

underdeveloped countries of Africa. Most types of anemia are preventable, but lack of funding andresources forcompleteblood testspreventsclinicians frommakingaproperdiagnosisandsuggestingappropriatetreatment.Thegoalistodevelopacost-efficient,portable,andaccuratedevicetocountredbloodcellsandmeasuretheirmeancorpuscularvolume.Apreviousteamdevelopedaproof-ofconceptusingamicrofluidicschannelandacomputerinterfacetomeasuretheresistancechangeofmicroparticlesmoving through themicrochannel.Therearemany improvements tobemadeon thedevice,and thefocuswillbeonpumpingandfilteringmethods.Eachdesignwasevaluatedusingadesignmatrix,andtheteam decided to move forward with the syringe pump and cell filters. Cell filters were testing andstatistical analysis showed significant differences before and after filteringmeaningmore thanwhitebloodcellswerelostintheprocess.Activeandpassivepumpingtechniqueswereevaluatedusingdilutedporcine blood samples to determine themicrochannel’s ability to count cells. The fluctuations of thechannelresistanceduringactivepumpingledtoinconclusiveresultsandwhilepassivepumpingshowedafewpeakscorrelatingtocells,notenoughwerecountedtomakeanyclaimsatthistime.

Page 3: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

3

TableofContentsIntroduction.........................................................................................................................................4

Motivation................................................................................................................................................4ProblemStatement..................................................................................................................................4PastWork.................................................................................................................................................4CompetingDesigns...................................................................................................................................6

DesignAlternatives..............................................................................................................................7PumpingTechniques................................................................................................................................7FilteringTechniques.................................................................................................................................9

DesignEvaluation...............................................................................................................................12DesignMatrices......................................................................................................................................12ProposedFinalDesign............................................................................................................................14

Fabrication/DevelopmentProcess......................................................................................................15Materials................................................................................................................................................15Methods.................................................................................................................................................15FinalPrototype.......................................................................................................................................15Testing....................................................................................................................................................16

Results...............................................................................................................................................18Filtration.................................................................................................................................................18PumpingDataCollection........................................................................................................................19

Discussion..........................................................................................................................................21Ethics......................................................................................................................................................21NecessaryChanges.................................................................................................................................21SourcesofError......................................................................................................................................22

Conclusions........................................................................................................................................22Futurework............................................................................................................................................23

References.........................................................................................................................................25

Appendix...............................................................................................................................................iDataAnalysis...........................................................................................................................................iiiMaterials..................................................................................................................................................v

Page 4: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

4

Introduction

MotivationAnemiaisadeficiencyofthehemoglobinintheblood,usuallycharacterizedbyabnormalsize,

shape, and reduced number of red blood cells and quantified by measuring hemoglobin and meancorpuscularvolume.Commonsymptomsofanemiaarefatigue,dizziness,rapidheartbeat,andshortnessof breath. [1] There are many types of anemia, but the team’s focus is microcytic, normocytic, andmacrocyticanemia.Microcyticischaracterizedbyabnormallysmallredbloodcells,usuallyintherangeof 4 to6µm indiameter and less than80 fL volume, and is causedby irondeficiency.Normocytic ischaracterized by having normal-sized red blood cells (6-8 µm, 80-100 fL) but is caused by decreasedproduction or increased destruction of red blood cells, such as hemolysis. Finally, macrocytic ischaracterizedbyabnormallylargeredbloodcells,usuallyintherangeof8-10µmindiameterandgreaterthan100fLvolume,anditiscausedbyvitaminB12deficiencyorhypothyroidism.[2]

Anemia is very prevalent in underdeveloped countries, due most commonly to malnutrition.Anemia affects roughly twenty-five percent of the global population, but the highest prevalence is inAfrica,makingupaboutsixtypercentofthosegloballyaffectedbyanemia.Mosttypesofanemiaarenotonlytreatablebutcanbeprevented.However,lackoffundingandresourcesforcompletebloodtestsinthese developing countries block the ability for clinicians to properly diagnose anemia and suggesttreatment. [3] The client, Dr. Philip Bain, volunteerswithGlobal Brigades, an international non-profitorganizationthathelpscommunitiesaroundtheworldmeettheirhealthcareandeconomicneeds.[4]After volunteering inGhana,Africa, hewanted to initiate thedevelopmentof an inexpensive anemiadetectiondevicetoallowclinicianstomakeaproperdiagnosisandgiveappropriatetreatmentoptions.

ProblemStatementAnemia affects many people worldwide and disproportionately affects those in developing

countriesduetoalackinmedicalinfrastructuretoproperlydiagnosetheblooddisorder.Aportable,easytouse,andcost-effectivedevice isneededtodiagnosethecondition inthesecountriesatthetimeofinitialmedicalcare.Anemiacanbediagnosedbyevaluatingredbloodcellsizeusingthemeancorpuscularvolume(MCV).Thegoal is to fabricateamicrofluidicdevice thateffectivelymeasures theMCVof redblood cells to determine if a patient has normocytic, macrocytic, or microcytic anemia with resultscomparabletocurrentcellcountingtechniques.

PastWorkLastyear,ateamdevelopedaproof-of-conceptdeviceforthisproject,consistingofa

microfluidicschannelandcomputerinterfaceusingLabVIEW.Theyusedapassivepumpingmethodinoneinlet-outletpairtotransport10µmmicroparticlesthroughamicrochannel,andthentheymeasuredtheresistancechangeacrosstheotherinlet-outletpair(Seefigures1,2).Eachpeakinfigure3representedamicroparticlemovingfromtheleftsideofthechanneltotheright,whiletheheightofthepeakdeterminedtheparticle’srelativesize.

Page 5: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

5

Figure1:Channeldesign.Theinletandoutletrepresenttheflowofthemicroparticlesthroughthechannel,andE1andE2representtheelectrodes.[5]

Figure2:Circuitschematicoftheelectrodesthatmeasuretheresistancechangeacrossthemicrochannel.[5]

Figure3:Voltagevstimewithcoincidenteventofamicroparticlemovingthroughthemicrochannel.[5]

Page 6: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

6

Thereisalotofworklefttobedoneforthedevice,however.Ameetingwiththepreviousteamwasarrangedwherepotentialimprovementsontheirprototypewerediscussed.Theymentionedthedifficultyofpassivepumping,asaddingtheproperamountofsolutiontoboththeinletandtheoutletforthepassivepumpingtoworkistedious.Inaddition,theyneverwereabletotestwholebloodforthedevice,andafilterisneededforthedevicetoremoveunwantedbloodcells,suchaswhitebloodcellsandplatelets.

CompetingDesignsInadditiontoresearchingthepastworkofthepreviousteams,wealsolookedintoother

competingdesignsalreadyavailable.ThetwodesignsclosesttowhatwearetryingtoaccomplisharetheCoultercounterandtheLabonaChipBloodCountingDevice.Bothdeviceswouldworkverywellintheoryforsolvingthecurrentproblem,butalsooffersignificantdisadvantagestothecurrentworkthathasbeendone.

CoulterCounterTheCoultercounteroffersmanyhospitalsandclinicstheabilitytocountandsizevariouscells

andparticlesinanelectrolytesolution.[6]Theadvantagesofthisdevicearethatwholebloodcanbeused,anditisversatileinthatitcanbeusedformorecountingstrategiesthansimplydetermininganemia.Despitebeingadvantageousinallowingmanycountingstrategies,thisisalsosomewhatofadisadvantagewhencomparedtoouridealdevicebecauseitisabitmorecomplexthanwhatweneed.Thedevicedoesalsocomeatasignificantcost-mostCoultercountersareontheorderof$2,000ormore,andwearelookingforadevicethatcanbeboughtinanunderdevelopedcountryforaround$50.Inaddition,theCoultercounterisabitlarge,thereforemakingitmoredifficulttotransport.

Figure4:Coultercounter[7]

LabonaChip-On-chipsamplepreparationforcompletebloodcountfromrawbloodTheseconddevicewefoundwascreatedattheUniversityofTorontoinToronto,ON,CA.The

devicecanalsousewholebloodandworksbyfilteringwhitebloodcellsfromredbloodcells,thendiluting,lysing,andmeasuringthecellsincountandsize.[8]ThedeviceismuchsmallerthantheCoultercounter,asitisamicrofluidicdevicemadeofPDMSandisthesizeofamicroscopeslide,makingthedeviceveryportable(seeFigure5).Thedevicealsousesapneumaticpressuredevicetodrivethe

Page 7: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

7

fluidsthroughthevariouschannels,whichiseasiertousethanthepassivepumpingtechniquethepreviousteamusedforourdevice.However,thedeviceisstillmorecomplicatedthanwhatweneed,sinceweareonlyinterestedinredbloodcellcountandsize,andthepneumaticpressuredevicetodrivethefluidthroughiscostlyinitofitself,makingtheoverallcostofthedeviceontheorderof$500.Therefore,wedecidedtocontinueonwherethepreviousdesignteamleftoffonourdeviceandmakethenecessarychangesneeded.

Figure5:On-chipsamplepreparationforcompletebloodcountfromrawblood[8]

DesignAlternativesInordertocontinuewiththeprototypeanddesignofthepreviousteam,twocomponentswere

evaluatedforintegrationintothedesign.Whilethepreviousteamprovedtheeffectivenessofamicrofluidiccountingchannel,bloodwasneveranalyzedwiththedevice.Duetoblood’sheterogeneouscompositionandhighcellconcentration,itmustgothroughapreparationprocesspriortobeingsentthroughthechannel.Thisinvolvesfilteringthesample,soonlyredbloodcellscrosstheaperture,anddilutingtoalowconcentrationsothatonlyonecellpassesthroughthechannelatatime.Thetwodesigncomponentsthatwereevaluatedwerepumpingandfilteringtechniques.

PumpingTechniquesThepumpingmethodisnecessarytoevaluatethebesttechniquetopassthebloodsample

throughthedevicequicklyandeffectively.

PassivePumpThepassivepumpingtechniqueworksbyallowingthecellstotravelthroughthechannelsdown

theirconcentrationgradient.ThecountingchannelwouldbefilledwithPBSorsomesimilarconductingsolution.Whenadropletofadilutedbloodsolutionisplacedononeendofthechannelthesolutionwill

Page 8: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

8

spreadandmovetotheoppositesideofthechannelsimplybecauseitwantstotravelfromhighconcentrationtolow.Anoverviewofthisprocesscanbeseeninfigure6.

Figure6:Overviewofpassivepumpingtechnique.Dropletofbloodsolutionisaddedtoinletendofchannelandbloodcellswillflowtowardstheoutletend,downtheirconcentrationgradient.

Thistechniquerequiresverylittleadditionalmaterials.Apipettewouldbeneededtoaddadropletofknownvolumetothechannelandthatisall.Thisallowsthisdesigntobeverylowcostperuse.Therearequiteafewdisadvantagesthataccompanythisdesign.Itisverydifficulttoplaceadropletonsuchasmallsurfaceandrequiresaskilleduser.Thisincreaseinuserdifficultycanhinderthetimeittakestorunthediagnostictestandincreasestheriskinwastingmaterialsiftheclinicianismorepronetomistakes.Thereisalsoverylittlecontrolovertherateofflowofthecellsthroughthechannel.Additionally,thiscanincreaseusagetimesanddecreasetheamountofteststhatcanbeperformedinsomeallottedamountoftime.Alsosincethereisnotanyactivepumpingtakingplace,thedirectionofflowmightnotbeunidirectionalwhichwouldgreatlyskewresults.

SyringePumpUsingasyringetomanuallypassthebloodsamplethroughthedeviceisanotherpumping

option.Thisdesignincludesasmallsyringefilledwiththedilutedbloodsolutionandatubingchannelthatconnectstothemicrofluidiccountingchannel.Theuserwouldmanuallycompressthesyringetosendthesampleslowlythroughthesample.Therateofflowcouldbemeasuredbasedonthespeedatwhichtheusercompressedthesyringeandthediameterofthesyringeopeningwherecellswillbeleavingandenteringthechannel.Theset-uptousingasyringeasthemodeofpumpingbloodthroughthechannelcanbeseeninfigure7.

Page 9: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

9

Figure7:Syringeattachedtoamicrofluidicchannel.Userwouldmanuallycompresssyringetosendbloodsamplethroughthechannel.Thesyringewouldbeintegratedintothechannelthroughanadditionaltubeas

shown.

Someadvantagesofthisdesignistheoverallsimplicityandeaseoffabrication.Allcomponentsofthispumpcanbeboughtcommerciallyandeasilyintegratedintotheexistingdevice.Sincemanycliniciansarefamiliarwithoperatingasyringe,itshouldalsobesimpletouse.Thedisadvantagesincludeitstendencyforvariationbetweeneachtest.Becausetheflowratedependsmuchontheuserandthespeedatwhichtheusercompressesthesyringe,thismaycausedifferencesinflowbetweensamples.Theuserwouldneedtoapplyacontinuouspressureataspecifiedspeedtoperformthediagnostictestcorrectly.

PeristalticPumpAperistalticpumptosendthebloodsamplethroughthechannelisacommonmicrofluidic

techniqueandanotherdesignalternativethatwasconsidered.Thispumpincorporatesflexibletubingandarotortosendthebloodsolutionthrough.Therotorcompressesthetubingasitrotatespushingfluidthroughthetubingunevenly.Aperistalticpumpsimilartowhatwouldbeintegratedintoourdesignisshowninfigure8.

Figure8:Theperistalticpumptubingandrotorshownhere.Therotorrotates,compressingthetubingandsendingthefluid(whichcanbeseeninblue)throughthechannel.[9]

Theadvantagesofusingthistypeofmechanicalpumpisitsautomationandeaseofuse.Thepumponlyneedstheusertoloadthebloodsampleandthepumpwillautomaticallysendthefluidthroughthechannel.Becauseofthisautomation,theflowratewillalsobeconsistentthroughouteachtestandhavenodependencyontheuser.Becauseofthepumpsmechanicalpropertiesandvarietyofmovingparts,thisdesignwillrequiremoremaintenancethantheotheroptions.Foruseindevelopingcountrieswhereresourcesarelimited,thiscanbeveryproblematic.

FilteringTechniques Anemiacanbediagnosedbyanalyzingredbloodcellvolumeandsize.Bloodismadeupofmorecomponentsthanjustthedesiredredbloodcellsandinordertogetthemostaccuratemeasurement,afilteringtechniqueisnecessarytopreparethesamplesothatonlyredbloodcellsremain.Witha

Page 10: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

10

filteringmechanism,thegoalistohaveonlyRBCssentthroughtheapertureofthemicrofluidiccountingchannel.

CellFilterAfterthecellsaredeliveredintothedeviceviaapumpingmechanism,theymustbefilteredto

isolatetheredbloodcellsfromtheothercomponentsofblood.Oneoptionisasyringefilter,showninFigure9,whichcontainsmixedcelluloseesterfiltermembranesinplastichousings.[10]Thisfilterwouldbeattachedtothepump,andcellsolutionwouldbepumpedthroughthefilterbeforeitissentthroughthemicrofluidicportionofthedeviceforanalysis.Thefilterporesizewouldbeabout11μminordertoexcludewhitebloodcells,whichhaveadiameterof12-15μm,andenableentryofredbloodcells,whichhaveadiameterbetween4-10μm.Thewiderangeofredbloodcellsizesreflectsthetypeofanemia,wheremicrocyticanemiaisdisplayedbyanRBCdiametercloserto4μm,andmacrocyticanemiaismanifestedbyadiametercloserto10μm.Thediameterofplateletsisontheorderof1-3μm,sotheywillpassthroughthefilter,buttheanalysisisnotpresumedtobeaffectedbytheirpresence.

Figure9:Asyringefilterisconnectedtothepumpandallowsforcellsolutiontobepumpedthroughthefilter,removingwhitebloodcells.[10]

Thissinglesyringefilterwouldbeeasytomanufacturesincethefilterunitsareavailableonthemarket,andtheywouldonlyneedtobeadaptedtothepumpandmicrofluidicchipportionsofthedesignusingasimplemethodsuchasattachingthepumpoutflowtothefilterusingatubeandthefilteroutflowtothechipusinganothertube.Operationtimewouldbelowwiththisdesignsincethecellsolutionwouldonlybepassingthroughasinglefilter.However,thisfilterwouldhaveshortlifespansinceitisnotreusable.Indevelopingcountrieswithfewresources,itmaybeunfeasibletoreplenishstocksoffiltersonaregularbasis.

CascadingFilter Anotherfilteroptionisamultiple-layeredcascadingfilterthatwouldbemadeupofthreefiltersofdecreasingporesize.Thefilterwouldbethesizeofa5mLsyringeandcontainthreedisksmadeofpolyethylenesecuredtothewallsofthecylindricalsyringe,asshowninfigure10.Thesethreedisksthatcontainperforationswillallowforsuccessivemovementofredbloodcellsthroughbasedondiameter.Atthetoplayer,allbutthelargestcells(WBCsof~15μmdiameter)willbeallowedtoflowthroughtothelayerunderneath,andthelayerunderneathwillcontainsmallerporesthatexcludecellsofa13μm

Page 11: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

11

diameter.Thefinallayerwouldblockpassageofcellsunder11μmwide,effectivelyisolatingredbloodcellsfromwhite.

Figure10:Acascadingfilterisolatesredbloodcellsbyremovinglargerwhitebloodcellsfromthecellsolutionusingasuccessionofthreelayers,eachcontainingasmallerporesizethantheoneabove.

Thisdesignwoulddecreasecellaccumulationwithinthefilterbyhavingmultiplelayerswhichallowcellstoflowgraduallythroughthedeviceinsteadofbuildingupandclusteringtogetherabovethefilter.However,thedesignwouldbedifficulttofabricatesincethistypeoffilterisnotcurrentlyavailableonthemarketandwouldhavetobebuiltbythedesignteam.Inaddition,alongeroperationtimewouldbenecessarywiththisfiltersincethecellsolutionwouldhavetopassthroughmultiplefiltersinsteadofjustonefilterasinthecaseoftheaforementionedcellfilter.

Built-inMicrofluidicFilter Afinaloptionforthefilterisabuilt-inmicrofluidicfilterthatwouldbeincorporatedintothemicrofluidicchipapparatususedforcellanalysisbuiltbythepreviousdesignteam.Asshowninfigure11,arowofpillarswith11μmopeningswouldbeconstructedonthechiptoallowredbloodcellstopassthroughtheopeningswhileblockingpassageofthewhitebloodcells,whicharelarger.

Page 12: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

12

Figure11:Abuilt-inmicrofluidicfilterconsistsofchannelsthatallowforpassageofredbloodcellsintothemicrofluidicdeviceandexclusionofwhitebloodcells.[11]

Thisbuilt-infilterwouldbeeasytousesinceitisintegratedintotheexistingdeviceandwouldallowfiltrationtobeaccomplishedwithoutusingaseparatephysicalcomponent.Ontheotherhand,thefilteritselfwillhavetobemadeonthemicroscale,suggestingthatfabricationwillbedifficultsincemuchsmallerpartsareinvolved.Additionally,whitebloodcellsanddebrismayclogthefilterquicklysincethereareonlyacertainnumberofopeningstoallowforcellstopassthrough.

DesignEvaluation Pumpingandfilteringtechniqueswereanalyzedandevaluatedbasedonavarietyofcategoriesinordertoproposethemosteffectivefinaldesign.

DesignMatrices

Page 13: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

13

Figure12:Designmatrixforalternativepumpingtechniques.

Adesignmatrixwascreatedforthethreeaforementionedpumpingtechniques(Figure12).Thedesignswererankedonthesixfollowingcategoriesinorderofdecreasingweight:easeofuse,timeofuse,cost,easeofmanufacturing,size,andsafety.

Easeofuseisofhighprioritytoourclient,Dr.Bain.Hesaidthattheexperiencelevelofcliniciansatpointofcareclinicsindevelopingcountriesvaries.Hewantedourteamtodesignadevicethatwouldallowforcliniciansofanyskillleveltouse.Thepassivepumpingtechniqueisadifficultproceduretosetup,andclinicianswhohaveneverusedapassivepumpingtechniquebeforewouldneedtobetrainedtoperformthistechnique.Althoughaperistalticpumpisconvenientbecauseitcanregulatetheflowrate,thispieceofequipmentisexpensiveandbulky.Therefore,thesyringescoredhighestforeaseofuse.Cliniciansofvaryingskilllevelshavemostlikelyusedasyringebefore,eitherfordrawingbloodoradministeringshots.Theyarealreadyfamiliarwiththispieceofapparatus,sotheywouldnotneedanyadditionaltrainingtoutilizeourdevice.

Timeofuseisalsoofhighprioritytoourclientandwasweightedthesameaseaseofuse.Dr.Bainsaidpointofcareclinicsseeabout300-500patientsperday,soourdeviceneedstobequicktouse.Thesyringeandperistalticpumptiedinthiscategory,andthepassivepumpscoredthelowest.Thepreviousdesignteamsaidthatittookthemafewtriestogetthepassivepumptowork,sothatiswhyitscoredthelowestinthiscategory.

Thecostcategorywasdeterminedbyadditionalcoststothecurrentdevice.Thepassivepumpinginletsandoutletsarealreadybuiltintothecurrentdevice,sonochangesneedtobemadetothecurrentdesign.Thesyringeandperistalticpumpingtechniqueswouldrequireadditionalequipmenttobepurchased.Thiscausedthesetwotechniquestoscorelowerinthiscategory,withtheperistalticpumpscoringlowestbecauseit’ssignificantlymoreexpensivethantheothertwodesignalternatives.

Easeofmanufacturingwasbasedonhowmuchadditionalworkwouldbeneededtoadaptthepumpingtechniquestothecurrentdesign.Again,sincepassivepumpingisalreadypartofthecurrentdesign,itscoredhighestinthiscategory.Thesyringescoredsecondhighestbecausetheinletsofthecurrentdesignwouldhavetobemodifiedtofitthesyringe.Theperistalticpumpscoredthelowestbecauseinadditiontomodifyingtheinletsofthecurrentdesign,theotherendoftheperistalticpumpwouldhavetobefittedtothebloodsamplecontainer.

Sizewasbasedoffofthecurrentdevice.Therefore,thepassivepumpscoredhighest.Thesyringescoredsecondhighestbecauseit’sanadditionalpieceofequipmentthatneedstobeintegratedintothecurrentdesign.Theperistalticpumpscoredlowestbecauseitisabulkypieceofequipmentthatismuchlargerthanthemicrofluidicdevice.

Safetywasnotaconcernforthethreedifferentdesigns.Clinicianswhoworkinclinicsaretrainedinhandlingbodilyfluids.Thethreeproposeddesignsdonotexposetheclinicianstoanythingtheyarenotalreadyexposedto.

Page 14: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

14

Figure13.Designmatrixforalternativefilteringtechniques.

Aseconddesignmatrixwascreatedforthethreefilteringtechniques(Figure13).Thedesignswererankedonthesevenfollowingcategoriesinorderofdecreasingweight:easeofuse,timeofuse,cost,easeofmanufacturing,size,lifespan,andsafety.

Easeofuseandtimeofusefollowthesamecriteriaasmentionedinthepumpingtechniquessection.Thebuilt-infilterscoredhighestinthesetwosectionsbecausethefilteringcomponentisincorporatedintotheexistingdesignandnoadditionalequipmentisneeded.Thecellfilterscoredhighestforcostandeaseofmanufacturing.Cellfiltersarereadilyavailableandcanbepurchasedinbulkatalowcost.Althoughmicrofluidicdevicesarefairlyinexpensivetomanufacture,theresearchanddesignneededtogettoafunctionalprototypecanbequiteexpensive.Masksneededtofabricatethemicrofluidicsdevicescancostover$100permask,andit’snotguaranteedthatourmicrofluidicfilterdesignwouldwork.Thebuilt-infilterwillbeintegratedintotheexistingmicrofluidic,soitscoredhighestinthesizecategory.Thecascadingfilterandbuilt-infiltertiedforlifespanbecausethesetwodesignsarereusable.Safetywasalsonotaconcernforthesetechniquesasthecliniciansshouldbetrainedtoproperlyhandlebodilyfluids.

ProposedFinalDesign Asseenfromthedesignmatrices,thesyringeandcellfilterscoredthehighestandwillbeincorporatedintothefinaldesign.Thefamiliaritythatclinicianshavewithsyringeswasabigfactorinchoosingthisdesign.Inaddition,syringesarereadilyavailableandoflowcost.Similarly,cellfiltersarealsoreadilyavailableandoflowcost.Althoughcellfiltersarenotreusablelikethecascadingandbuilt-infilter,ourteamfeltthatdevelopingcountrieslackthenecessaryresourcesandtimetokeepthereusablefilterscleanedandmaintained.Thus,thecheaperandeasieralternativeistogowiththedisposablecellfilter.

Page 15: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

15

Fabrication/DevelopmentProcess

MaterialsForthissemester’swork,filteringwasdeterminedtobeamajorareaoffocus.Forthispartof

thedesign,asamplepackofvariouslysized,roundfilterswaspurchased.Thesefiltershaveaporesizeofabout10microns;thiswasdeterminedtobeanadequatesizetoletredbloodcellsthroughwhileatthesametimetrappingthewhitebloodcellsinthecollectedsamples.

PDMSwasthematerialofchoiceforthefabricationofthemicrochannelsusedinthisproject.ThepreviousgroupthatworkedonthisprojectusedPDMSwithgeneralsuccessandasubstantialamountremainedfromtheirwork.Therefore,itwasusedagainforthecontinuationofthisproject.

SyringetochannelattachmentrequiredaseriesofLuerlocksandtubing.TheseweresecuredtothePDMSchannelinletportwithastrongclearadhesivetopreventleakageandbackflow.

MethodsThechannelitselfwasmoldedfromPDMS,vacuumedandallowedtosetoverthemasterdisc

designedbythepreviousteam.Itwasfabricatedsothatthethicknesswasapproximately1cmhigh.AfterthePDMShadcured,thechannelwasremovedfromthemasterdiscandplasmatreatedtoaglassslide.Thechanneldesigncontainedfourports,twoforelectrodeplacement,aninletandanoutlet.AllfourwerepunchedfromthePDMSusingabiopsypunch.ALuerlockadapterwasplacedintotheinletport,whereitwassecurelysealedwithastrongadhesive.Themicro-scalesizeoftheaperturerequiresastrongsealattheinletporttominimizeleakageandbackflow.Becauseofthis,thethicknessofthePDMSwasvitalinassistingthesecurementoftheLuerlockadaptersincethreadedendhasalargesurfaceareatoholdontowithoutblockingtheopeningofthechannel.ThetubingconnectedfromtheLuerlockadaptertoanotheradapterwherethesyringeisattached.

FinalPrototypeInfigure14,thechannelisattachedatitsinletporttothetubingandaseriesofLuerlock

adapters.Theadaptersallowthesyringe,tubingandchanneltobemanagedseparatelywhenneedbe.Eachisabletobedisconnectedforreplacementofindividualpartswhichisnecessaryifthedeviceistobeusedinpractice.

Figure14:PDMSchannelwithsyringeconnectedthroughtubingandLuerlockingadapters

Page 16: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

16

Theoveralldevicesizeandportabilityrequirementsledtothefabricationoftheboxencompasseddeviceasseeninfigure15.Allcomponentswerecombinedintoanelectricalboxinordertopresentthedeviceinaclean,easytousemanner.Thisincludedthecircuitry,theNI-DAQwhichwasusedfordataacquisition,andthechannel.Asyringeconnectstoaninletportatthetopoftheboxwhichattachestothechannelthroughtubing.TheboxallowstheNI-DAQtobeconnectedtoacomputerwiththeappropriatesoftwareforcollectingdata.

Figure15:Finalboxenclosure.Syringeattachesattopofboxandsendsadilutedbloodsamplethroughtubingandthemicrochannel

Testing Thecellfilteringdesignaswellastheactivepumpingsyringedesignwerebothevaluated.Todeterminetheeffectivenessoftheproposedfiltrationmethod,cellcountsweremeasuredbeforeandafterbeingfiltered.Experimentsweresetuptoevaluatetheabilityofthemicrochanneltocountcellsastheycrosstheapertureafterbeingmanuallypassedthroughbysyringe.

FiltrationToobtainanaccurateRBCcountusingthechannel,afilterwasdevisedtoseparateWBCsfromRBCs.The10umporesizewaschosentoexcludeWBCs,whichhaveadiameterof12-15um,fromRBCs,whichare6-8umwide.[13]Theporesize,whichisintermediatebetweenthatofredandwhitebloodcells,allowsmacrocyticRBCsupto10umindiametertopassthroughthefilter.Thefiltrationdevicewasfabricatedbyplacingacircularpolyestermembranefilterwithadiameterof47mmandaporesizeof10umontotheopenendofa50mLcentrifugetube.Aroundcutwasmadeontheplasticcaptoallowfilteredbloodtoflowdirectlyintoatube,andthecapwasscrewedon,securingandstretchingoutthepolyestermembrane(Figure16).AsolutionofbloodinPBSdiluted1:5000(1uLofbloodin5mLPBS)waspouredintothetubecontainingthefilter,andcellcountsoftheunfilteredandfilteredsolutionswereobtainedviauseofahemocytometer.

Page 17: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

17

Figure16:Thefabricatedfilteringmethodusedtogatherfiltrationdata

Itisexpectedthatcellcountbeforeandafterfilteringwillgiveanindicationoftheamountofcellslostthroughtheoverallfilteringprocess.Forthepurposeofthisdevice,ideallyonlylargewhitebloodcellswillberemovedthroughouttheprocess.Whitebloodcellsmakeuponly1%ofwholeblood,soinordertosupportthehypothesisthatonlyasmallportionofcells(whitebloodcells)arebeingfiltered,thecellcountsbeforeandafterfilteringneedtoshownosignificantdifference.[14]

PumpingandDataAcquisitionTrialswereperformedusingbothactiveandpassivepumpingtechniquestosendadiluted

bloodsamplethroughthedevice.Electrodeswerefittedeachtimetomeasurechangesinvoltageascellspassedthroughtheaperture.

ActiveSyringePumpingThechannelwasfittedwiththeLuerlockadaptersandtubingtotesthowwellthechannel

couldmeasurechangesinvoltageascellspassedthroughitsaperture.Thechannelwasplacedunderamicroscopesothemovementofcellscouldbeviewedandmatchedwithcollecteddata.Avoltagedividercircuitwascreatedwiththechannelrepresentingoneresistorofthatcircuit.(FigureFROMPASTWORK)Ascellspassthroughtheaperture,theresistanceofthechannelisexpectedtoincrease,thereforeincreasingtheoutputvoltagewhichismeasuredthroughanNI-DAQdataacquisitiondevice.Theoverallset-upcanbeseeninfigure17.

Page 18: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

18

Figure17:Theexperimentalset-upusedforactivefiltering.Themicrochannelisplacedunderthemicroscopewhileelectrodesfromavoltagedividerareattachedatportsinthechannel.AnNI-DAQ

measurestheoutputvoltage.

Aporcinebloodsamplewasdiluted1:5000inPBS.Thesyringewasfilledwiththedilutedbloodsolutionandattachedtothetubingandchannel.Labviewsoftwarewasrunasthebloodsamplewasmanuallypassedthroughthemicro-channel.Videowasalsorecordedascellsflowedthroughtheaperture.

PassivePumpingAnidenticalset-upasabovewasperformedwithoutthesyringeandtubingattachment.Afresh

PDMSchannelwasplasmacoatedtoaglassslideandplacedunderthemicroscope.Electrodeswereplacedintheportsofthechannelastheywerewithactivepumpingtrialsanddatawasrecordedascellspassedthroughtheaperture.

Withpassivepumping,adropofdilutedbloodsolutionwasplacedontheinletportofthechannelandcellsflowedfreelydownthechanneltowardstheoutletport.

Results

FiltrationStatisticalanalysisrevealedthattherewasasignificantdifference(pvalue<0.0001)between

cellconcentrationintheunfilteredandfilteredsamples(seeTable1).Thelargedisparitybetweencellcountsindicatesthatthefilterwasnotworkingasexpectedsinceitwashypothesizedthatfiltrationwouldonlyremoveasmallportionofcells.Thisanalysisleadstothebeliefthatredbloodcellsarealsobeingremovedinthefilteringprocess,whichwouldaffecttheaccuracyoftheoveralldevice.Fromthisexperimentaldataandanalysis,itwasdeterminedthatafilterwouldnotbenecessarysincewhitebloodcellsonlymakeupasmallpartoftotalcellcount,whichshouldnotdrasticallyaffectcountingwithinthe

Page 19: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

19

channel.Additionally,byaddingafiltrationcomponent,designcomplexitywouldbeincreased,whichisnotidealforasystemusedinlowresourcesettings.

Table1:CellConcentrationinUnfilteredvs.FilteredBloodSamples

PumpingDataCollection DatawascollectedinLabViewandanalyzedwithapythonscriptwrittenbypreviousteammemberRussellLittle.Theanalysissoftwarefiltersandfitsthedatatoacurvelevelingvoltagemeasurementsandremovinginstrumentaldrift.

Active Themotionofmanuallypumpingabloodsolutionthroughthechannelshowedtogivegreatfluctuationsinvoltage.(Figure18)Ideallytherewouldbeaconstantbasevoltagewhichrepresentstheresistanceofthechannelwithnocellspresentandascellspassthroughtheaperture,distinctspikesinvoltagearerecorded.Overmanytrials,noneoftheseexpectedspikeswereabletobedistinguished;evenmicroscopingvideoofthechannelverifiedthatcellswerepassingthroughtheaperture.Theconstantfluctuatingchangeinvoltageovertimeisbelievedtobecausedbythevaryingpressureappliedtothesyringe.Asdifferentpressuresareappliedtothechannel,changesinresistanceandnoisecausethedatatofluctuate.Therearemanymorecontributingfactorsthatpreventedtherecordingofvoltagechangesandarediscussedfurtherinthesourcesoferror.

Page 20: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

20

Figure18:Voltageovertimeasbloodcellswereactivelypumpedthroughmicro-channel.(a)Therawoutput

data(b)Filtereddata(c)Curvefitdatatolevelvoltage

PassiveUtilizingthepassivepumpingtechniqueresultedinafewtrialswheredistinctdipsinvoltage

wererecorded.(Figure19)Withthemicroscopeitwasverifiedthatthesedipsdidcorrelatewithcellspassingthroughtheaperturebutnotallcellscausedacorrespondingchangeinvoltage.Overtimeresultsbecamemoreinconclusivewithminimaldistinctchangesinvoltage.Manytrialswereperformedbutduetothelackofvoltagechanges,resultswereinconclusive.Potentialexplanationsforthelackinsubstantialdataisdiscussedfurtherinsourcesoferrorandfuturework.

Figure19:Voltageovertimeasbloodcellswerepassivelypumpedthroughmicro-channel(a)Therawoutputdata(b)Filtereddata(c)Curvefitdatatolevelvoltage

Page 21: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

21

Discussion Forthefilteringtestingresults,itisapparentthatthereisasignificantdifferenceincellnumberbeforeandafterfiltering.Thisisconcerningbecauseitisobviousthatourfilterwasremovingredbloodcellsandwhitebloodcellsfromthesolution,whentheoriginalideaofthefilterwastoremoveonlythewhitebloodcells.Becausewhitebloodcellsonlymakeupapproximately1percentoftheblood,thereshouldnotbeasignificantdifferenceinthecellcounts,butthewhitebloodcellswouldstillhavebeenremovedfromthesolution.[14]Thoughwedeemedthefilterunnecessaryfortesting,thereisstillapotentialforwhitebloodcellstoenterthedevice.Duetotheirsize,theycouldblockthechannelandpreventanaccuratereading.Therefore,furtherresearchandbrainstormingwillneedtobedonefordetermininganoptimalmethodforremovingthewhitebloodcellspriortothebloodenteringthechannel.

Forthepumpingresults,itwasfoundthatthesyringedidnotdeliveraconsistentflowrate,thereforecausingerraticfluctuationsinthedatacomparedtothepassivepumpingtechnique.Itisthoughtthatanautomatedpumpingsystemcouldsolvethisproblemtodeliveraconsistentflowrate.Inthe“LabonaChip”paperfound,theteamofresearchersfromTorontousesapneumaticpressuredevicetoconsistentlymovethecellsthroughthechannels.[8]Thistechniquemaybesomethingtoconsider,andisdiscussedfurtherinfuturework.

EthicsTheassessmentofthisdevicereliesontheavailabilityofappropriatebloodsamples.This

projecttakesasmallbloodsampleandrunsitthroughamicrochanneltotestforthepresenceofanemiainthepatientandwhattypeofanemiaitis.Itisimportantthatallbloodsamplesusedinthetestingofthisdevicecomefromconsentingsourcesandallapprovalneededpriortousinghumansampleshasbeenmade.

Theporcinebloodthatwasusedforthetestingofthisprojectwashumanelygatheredanddonatedtobeusedbyalaboratoryoncampus.

NecessaryChangesTheresultsgatheredabovedemonstratetheneedtoredesignsomedevicecomponents.The

fluctuatingvoltageobservedthroughmanuallypumpingsolutionthroughthechannelshowstheneedforanautomatedpumpingmechanism.It’spredictedthatapplyingaconstantpressuretothesyringewouldeliminatethemajorityofthesefluctuationsandyieldbaselineresultssimilartothoseofpassivepumping.

Duetobudgetconstraints,thechanneldesignfromthepreviousteamwasusedfortesting.Thepreviousteamdesignedthechanneltoworkwellwhilepassivepumpingbutsomeminortweakstothechannelcouldmakeitmorereliablewithactivepumping.Asuggestedchanneldesignisshowninfigure20.Therewasatendencyforsolutiontotraveldowntheshortestandnearestpathsothenewdesignincorporatestwoequallengthoutletports(anelectrodewillbeplacedineither).Thiswilleliminatethefavorabilityofoneoutletovertheotherandallowcomparableflowtoeach.Inadditiontoanew

Page 22: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

22

channel,buildingcupsaroundallchanneloutlets,allowingsolutiontopool,wouldpreventtheoverflowofsolution.

Figure20:Apotentialfuturechanneldesignwheretheoutletportsareofequallengththereforeflow

doesnotfavoroneovertheother.

SourcesofErrorTherearesomeknownvariablesthatmayhaveaffectedtheresultsofthissemester’swork.In

ordertobuildthePDMSchannel,themastermoldfromthepreviousteamwasused.Overtime,thephotoresistofthemasterstartedtowearandflakeoff.Thismadethefabricationofeachchannelmoredifficultandalteredtheoveralldimensionsandshapewhichmayhavehadaneffectonthetestingresults.

Avarietyofionsolutionswereusedtodilutebloodsamplesincluding1XPhosphateBufferSolution(PBS),anisotonicNaClsolutionandahypertonicNaClsolution.Theconductivityofeachsolutionmayhavehadamoresignificanteffectontheresultsthanexpected.Thiswasatradeoffbetweenusingamoreconductivesolutionthatwouldaffectcellsovertimeandalessconductivesolutionthatwouldhaveminimaleffectsonthecells.

Channelblockingwasarecurringconcernwiththechannelbeingthattheaperturewasonly15uminsize.Anysortofdebrisorlargersizedcellcouldclogtheapertureandpreventfutureparticleflow.Theclogwouldinturnaffecttheresistanceofthechannel.WithpassivepumpingaclogcouldeasilybereversedbylightlytappingthetopofthePDMSbutactivepumpingcausedmoresevereeffects.Thepressureofthesyringelodgesthedebrisfurtherintotheaperturemakingthenearlyimpossibletoreleaseandrenderingthechanneluseless.

Adifferentpotentialsourceoferrorintheresultswastheageoftheporcinebloodthatwasused.Thebloodwasacquiredfromalocallaboncampusanditwasstoredinafridgewithoutanyothermethodofpreservingit.Itislikelythatovertimethebloodcouldhavedeterioratedandformedclumps.Thiswouldthrowoffthefilteringdataasitwouldbehardtostudyaconsistentsampleoftheblood.

Conclusions Aportable,easytousedeviceisneededtodiagnoseanemiaindevelopingcountrieswheretypicaldiagnosticmethodsarenotavailable.ThisprojectintendedtoimproveuponaparticlecountingmicrofluidicdevicewiththeendgoalofdevelopingadevicethataccuratelymeasuresMCVofredbloodcells.Thefocusofthisprojectwasontwomodificationstothepreviousmicrofluidicdevice.The

Page 23: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

23

microfluidicchannelwasfittedwithasyringepumpingmechanisminordertoquicklymeasureredbloodcellsizeandbloodfiltrationwasanalyzedwiththeintentionofremovingwhitebloodcellsfromadilutedbloodsolution.

Theresultsfromfilteringadilutedbloodsampleshowedasignificantdifferenceincellcountbeforeandafterfiltration.Thisledtoaconclusionthatfilteringthebloodsampleremovesmorethanjustlargewhitebloodcellssincethesemakesuchasmallportionofwholeblood.Theremainderofthedesignforwentthefilteringprocess.Activeandpassivepumpingofadilutedbloodsamplethroughthemicrochannelwasanalyzedtofindthatactivepumpingcauseslargefluctuationsinchannelresistancereducingtheabilitytodistinguishsmallvoltagechangescausedbycells.Passivepumpingshowedevidenceofvoltagechangesascellspassthroughthechannelaperturebuttrialsovertimeshowedinconsistentresultswhichpreventedanysignificantconclusions.

FutureworkThereisamultitudeofworkthatcanbedoneonthisproject,themostpressingbeingto

redesignthedevicetofunctionaswasoriginallyintended.Thismeanstakingabloodsamplefromapatient,directlyinsertingitintothedeviceandgettingausablereadingthatwoulddiagnoseanemiaaswellasthetypeofanemiapresent,withinaminute,thatwouldbeaccuratetoabout95percent.Atthiscurrenttime,itwasshownthatsyringepumpingbyhandisnotpracticalbecauseitcreatesunsteadyflowwhichcausesavaryingsignalthatisnotreadable.Somesortofautomationwouldneedtobeintroducedintothedevicetocreateasmoothandcontinuousflow,however,thisisdifficultconsideringthecostandsizeconstraintoftheproject.

Moreworkwouldneedtobedoneinordertoimprovethefunctionalityofthechannels,thisincludesmakingtheflowmorepredictablethroughthechannel.Anotherkeyimprovementthatneedstobemadeistheoverallproductionofthemicrochannels,currentlyittakesoveranhourandahalftocreateasinglechannel.Thisprocessneedstobemadequickerandmoreefficienttohelpmaketestingnewdesignsfaster,aswellasmakingiteasiertofabricatelargerquantitiesatatime.Alongwiththis,abettermethodtoattachthetubingtothechannelsisneeded.Currently,thesealbetweenthechannelandthetubingleaksandthisleakagecouldharmthecircuitryovertime.

Usingthemicrochannels,acurvewouldbemadethatrelatesvoltagestodifferentsizedmicroparticles.Microparticlesofvariousknownsizeswouldbeused,andbecausetheyarespherical,theirvolumecouldbesimplycalculatedusingtheirdiameters.Thiswouldthenbeusedtocorrelatevolumestovoltages.Usingthiscurve,itwouldbepossibletodeducethesizeofthecellsthatpassthroughthechannelbyrelatingthevoltagetotherelatedvolume.

Someotherworkthatcouldbedoneinthefutureincludescleaninguptheprototypetobeapresentable,marketableproduct,alsomakingitmorecompact.ThiswouldalsoincludeswitchingfromtheNI-DAQunittoamicrocontrollertosimplifythecircuitryaswellasremovingtheneedforalaptoporcomputer.Atthispointascreenwouldbeattachedtothedevicetogivethereadoutsoftheresultsofthetest.Itwouldalsobebeneficialtotheendconsumertomakethevariouspartsofthedeviceeasiertoinstallandreplace,includingthenecessarymicrochannelsandthevariouswiringandelectrodesassociatedwiththecircuitry.

Page 24: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

24

EventuallyitwouldbecomenecessarytogainIRBapprovaltotestwithhumanblood.Thiswouldbeanecessarysteptobeabletoprovethatthedevicefunctionswithhumanblood.

Anecessarystepthatneedstobedoneinthefutureiscontinuingtestingofthedeviceandoptimizingthedesignbasedonnetheobservationsandresultsoftheseteststomakesurethatitfunctionsasintendedandthattheresultsoftheanemiatestsareaccurateandreproducible.Thedevicewillbeupdatedasnecessaryuntilthisisachieved.

Anothernecessarystepinthefuturewillbetotestthefunctionofthedevicewithanaveragepersonwhohashadnoexperiencewithmicrofluidicchannelsordevicessimilartothisone.Thistestwillhelpmakeitclearwhetherornotthedeviceneedstobesimplifiedinordertobeusedbyageneraltechnicianatahealthclinicinathirdworldcountry.Inconjunctionwiththistest,ageneralmanualorprocedureofoperationforthedevicewillneedtobewrittenup.Thismanualwillbegivenatthetimeoftheabovetest,andwillberevisedasneededtobeunderstandableandexecutablewithoutanyotherpriortraining.Basedontheseresults,thedesignofthedevicewillbeadjustedasneededuntilthisgoalismet

Page 25: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

25

References[1]Mayoclinic.org,'AnemiaSymptoms-MayoClinic',2015.[Online].Available:http://www.mayoclinic.org/diseasesconditions/anemia/basics/symptoms/con-20026209.

[2]V.Planche,S.Georgin-Lavialle.“Etiologiesanddiagnosticwork-upofextrememacrocytosisdefinedbyanerythrocytemeancorpuscularvolumeover130fL:Astudyof109patients,”AmJ.Hemotol.,vol.89,no.6,pp.665-666,2014.

[3]Who.int,'WHO|Worldwideprevalenceonanemia1993-2005',2015.[Online].Available:http://www.who.int/vmnis/anaemia/prevalence/summary/anaemia_status_summary/en/.

[4]Globalbrigades.org,'GlobalBrigades|GlobalBrigades',2015.[Online].Available:https://www.globalbrigades.org/.

[5]Chiang,M.,Choi,U.,Enge,J.,Little,R.BME402.“DevelopmentofAnemiaDetectionDeviceforLow-ResourceSetting,”2015.

[6]W.R.Hogg,W.Coulter;Apparatusandmethodformeasuringadividingparticlesizeofaparticulatesystem;UnitedStatesPatent3557352

[7]BeckmanCoulter.Multisizer™3COULTERCOUNTER®,ItemNo:6605697.Beckmancoulter.com

[8]J.Nguyen,Y.Wei,Y.Zheng,C.WangandY.Sun,'On-chipsamplepreparationforcompletebloodcountfromrawblood',LabChip,vol.15,no.6,pp.1533-1544,2015.

[9]Micro-milling.Digitalimage.Z-microsystems-Mouldingof'Lab-On-A-Chip'UnderCleanRoomConditions.Z-micosystems,n.d.29Sept.2015.[Online].Available:http://www.medicaldevice-network.com/contractors/nanotechnology/z-microsystems/z-microsystems5.html.

[10]Sigma-AldrichCo.LLC."Millex®SyringeFilterUnits,Disposable,MixedEsters."PoreSize0.22μm,Diam.33mm.N.p.,n.d.[Online].Available:http://www.sigmaaldrich.com/catalog/product/sigma/f8148?lang=en&region=US

[11]Shields,C.,Reyes,C.,Lopez,G.“Microfluidiccellsorting:areviewoftheadvancesinseparationofcellsfromdebulkingtorarecellisolation,”LabonaChip,2015.

[12]Guckenberger,DeGroot,Wan,Beebe,&Young.“Micromilling:Amethodforultra-rapidprototypingofplasticmicrofluidicdevice,”LabonaChip,15(11),2364-2378,2015.

[13]ThroughtheMicroscope.WadsworthCenter.NewYorkStateDepartmentofHealth.http://www.wadsworth.org/chemheme/heme/microscope/rbc.htm

[14]BloodBasics.AmericanSocietyofHematology.http://www.hematology.org/Patients/Basics/

Page 26: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

Appendix

ProductDesignSpecificationsClient:Dr.PhilipA.BainAdvisor:ProfessorThomasYenTeam:

CarlyHildebrandt [email protected] [email protected] [email protected] [email protected] [email protected]

LastUpdated:Decmeber9,2015FunctionAnemiaaffectsmanypeopleworldwideanddisproportionatelyaffectsthoseindevelopingcountriesduetoalackinmedicalinfrastructuretoproperlydiagnosetheblooddisorder.Aportable,easytouse,andcost-effectivedeviceisneededtodiagnosetheconditioninthesecountriesatthetimeofinitialmedicalcare.Anemiacanbediagnosedbyevaluatingredbloodcellsizeusingthemeancorpuscularvolume(MCV).ThegoalistofabricateamicrofluidicdevicethateffectivelymeasurestheMCVofredbloodcellstodetermineifapatienthasnormocytic,macrocytic,ormicrocyticanemiawithresultscomparabletocurrentcellcountingtechniques.Clientrequirements

● Deviceshouldprovideanaccuratediagnosisofanemiaanddifferentiatebetweenmicrocytic,normocytic,andmacrocyticanemia

● Deviceshouldbelow-costandadaptabletoresource-limitedenvironments● Aclinicianshouldbeabletousethedeviceeasilyandreliablyafterpropertrainingwithan

intuitiveuserinterface● Thedeviceshouldbeabletodiagnoseanemiaatthepointofcare

DesignRequirements1. PhysicalandOperationalCharacteristics

a.Performancerequirements:Thedeviceshouldbeabletodiagnoseanemia,byidentifyingpatients’RedBloodCell(RBC)count,MeanCorpuscularVolume(MCV),andhemoglobin(Hb)levels.Testinganddiagnosistimeshouldtakelessthan30minutes.

b.Safety:Bloodsamplesmustbeintroducedtothedeviceinacontainedenvironmentsothatnocontaminationoccursbetweendeviceusesandbetweentheuser.Thedevicemustbeusedwithproperbloodcollectiontechniques.c.AccuracyandReliability:RBC,MCV,and/orHblevelsshouldbemeasuredwithatleast95%accuracywhencomparedtostandardcountingtechniques(i.e.CoulterCounter)toallowfordiagnosisofanemia.Diagnosiswith95%accuracyshouldtakenolongerthan30minutes.

Page 27: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

ii

d.LifeinService:Devicehardwareshouldfunctionforatleast5years,andthebloodcollectionplatformwillbereusable.Softwareshouldbeabletobeupgradedwhennecessary.e.ShelfLife:Deviceandallattachmentsshouldhavetheabilitytobestoredfor5yearsfromthetimeofmanufacture.f.OperatingEnvironment:Thedevicewillbeprimarilyusedindevelopingcountries(i.e.countriesinAfrica,SoutheastAsiaandSouthAmerica),soavailableresourcesshouldbetakenintoconsideration.Generallyaccesstomostresourcesislimitedsothedeviceshouldbeabletostandaloneorrunwithminimaloutsidehelp(i.e.batteriesorasmallgenerator).g.Ergonomics:Thedeviceshouldbeeasilytransportableandusablebycliniciansofvaryingexperienceandeducationalbackgrounds.h.Size:Thedeviceshouldbesmallenoughtofitonabenchtopinaclinicalsettingwhilemaintainingenoughportabilitytotransporttoareasinneed.Thecircuitryhousingshouldbenolargerthan4”x6”x10”.Themicrofluidicdeviceshouldbenolargerthan4”x4”.i.Weight:Thedeviceshouldweighnomorethanfivepounds(2.3kg).j.Materials:Materialsshouldbelow-costanddurable.Theyshouldalsobebiocompatibleastonotdisruptcellstructureasthebloodismovingthroughthedevice(i.e.PDMSforthemicrochannels).k.Aesthetics,Appearance,andFinish:DeviceshouldhaveanoutputscreentoviewRBC,MCV,Hblevels,andclassificationofanemia.Datashouldbeeasilyinterpretedbytheuser.Circuitryshouldbehiddenfromview.

2.ProductionCharacteristics

a.Quantity:Oneprototypeisneededforproofofconcept.b.TargetProductCost:Productshouldnotcostmorethan$200.

3.Miscellaneous

a.StandardsandSpecifications:Tobedeterminedatalatertime.Devicewillbeusedindevelopingcountries,notcoveredbytheFDA.FDAguidelineswillbefollowed,butapprovalisnotneeded.b.Customer:Cliniciansofvaryingskilllevelsindevelopingcountries.

c.Patient-relatedconcerns:Shouldbeabletogivepatientadiagnosisatpointofcare.d.Competition:Existingdevicesincludethefollowing:

● CoulterCounter:AcoultercountermeasurestheMCV,hemoglobinandtheredbloodcelldistributionwidth.Thecoulterprinciple,whichusesthedirectcurrentimpedancemethod,governstheuseofthedevice.

● HemoGlobe:HemoGlobeisanon-invasivedevicedesignedbystudentsfromJohnsHopkinsUniversitythatmeasuresthehemoglobinlevelofabloodsample.Thedevicemeasurestheparameterandthensendsthedatatoacenterforfurtherdataanalysis.

Page 28: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

iii

● MicrofluidiccardforRBCanalysis:Patent-US8034296B2:Thecartridgeenablesacompletebloodcountincludingredbloodcellanalysisofvariousparameters.Thecartridgechannelmayalsobesubjecttoanelectricormagneticfieldduringoperation.

DataAnalysisCreatedbyRussellLittleimportutilsfromparse_fileimportparsefromplotimportplotfromcalcimportpolyfit,find_peaks,mean,moving_averagedefprocess_raw_data(filename=None,maxima=None,binsize=20,deg=3,delta=.004,xlabel='Time(s)',title=None,**kwargs):'''Args:=======->maximaT/Fdeteckpeaksasmaximaorminima->labelThisfunctionisthemainentrypoint.givenafileitwilldothewholeprocessing...thing...thisismainlyjusttomakeiteasier.thisisverymuchdependantonthestructuerofthedatafile!Ifyouchangeacquisition!!!changethis!!!!'''#ASSUMPTIONS!!!!!!changethese.theremaybemoresubtleassumptions#furtherin.inputV='Voltage_0'#inputsignal,totrackphasechangesoutputV='Voltage_1'#outputsignal,totrackparticlesx='X_Value'#timearray,r=parse(filename=filename,**kwargs)#parsethedatafileassert(inputVinr)#__ASSUMPTIONS__weknowthistobetheinputsignalassert(outputVinr)#__ASSUMPTIONS__weknowthistobetheoutputsignal

Page 29: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

iv

#addsomeusefulinfor['inputave']=[mean(r[inputV])]*len(r[inputV])r['outputave']=[mean(r[outputV])]*len(r[outputV])#============================='''Smooththeoutputtoremovelargepeaksresultingfromrarenoiseevents'''r['smoothedoutput']=moving_average(r[outputV],n=binsize)r['binsize']=binsize#============================='''removeinstrumentaldrift.Byfittingapolynomialtotheoutputsignal,thensubtractingit.'''r['polyfittooutput']=polyfit(r[x],r[outputV],d=deg)r['deg']=degr['outputadjusted']=r['smoothedoutput']-r['polyfittooutput']#=============================='''Runthedetectpeakalgorithmtocountthepeaksinthedatafile.'''mx,mn=find_peaks(r[x],r['outputadjusted'],delta=delta)ifmaxima:r['particles']=mxr['numberofparticles']=len(mx)else:r['particles']=mnr['numberofparticles']=len(mn)r['max']=maximar['delta']=delta#============================='''Thehardpartisdone.justplottheresults.Iwanttoproduceonesummarygraph.thatdisplaystheprocessingpipelineandonethatisjusttheresults.'''plots=[('RawOutput',r[outputV]),('SmoothedOutput',r['smoothedoutput']),('Result',r['outputadjusted'])]plot(r[x],plots,x_title=xlabel,plot_title=title,measurements=[])#=============================#andwe'redone.returntheresultsjustforgoodnesssake.returnr

Page 30: POINT OF CARE ANEMIA DEVICE€¦ · Pumping Data Collection ... There are many types of anemia, but the team’s focus is microcytic, normocytic, and macrocytic anemia. Microcytic

v

process_raw_data()

Materials

Product PlacePurchased PricePolyester10umFilterSamplePack Steriltech 30.98ElectricalBox Amazon 7.45NI-DAQ BMEDepartment donated1,3mLSyringes BMEDepartment donatedLuerLocksandtubing Dr.ThomasYen donated