Page 1: PHYSICS OF PROTON THERAPY

PHYSICS OF PROTON THERAPY

Basit Athar, PhD DABRHampton University Proton Therapy Institute

1

Page 2: PHYSICS OF PROTON THERAPY

PROTONS

•Protons are charged particles and directly interact with electrons.

•Their mass(Mp) is 938MeV/c2 in comparison Me = 0.511MeV/c2

•Knock-out electrons easily.

•Impart their energy to electrons thus to the medium.

•This energy imparted gives rise to the idea of DOSE.

2

Page 3: PHYSICS OF PROTON THERAPY

CONCEPT OF DOSE Dose is a measure of the amount of energy deposited in a small volume as a result of the radiation - be that energy deposited locally, or brought to the point of interest by secondary radiation generated at some distance from the primary interaction.

Dose is defined as the energy deposited in a volume divided by its mass.

Dose is expressed in units of Gray (Gy) ~ 1Gy = Joule/kg

To understand this energy deposition, need to understand charge particle interaction with matter.

3

Page 4: PHYSICS OF PROTON THERAPY

BASIC INTERACTIONS

•Energy Loss - Coulomb interactions with electrons. Defines the range and shape of dose profile along the beam line.

•Scattering - Coulomb interactions with target nuclei. Defines the shape of lateral profile.

•Nuclear Interactions - Inelastic nuclear interaction with target nuclei. Modifies the depth dose and lateral dose distribution.

4

Page 5: PHYSICS OF PROTON THERAPY

WHY HEAVY CHARGED PARTICLES?

Heavy charged particle therapy can reduce the dose load “intergral dose” to normal tissues surrounding the tumor target volume by a factor of 2-3 (reduced “dose bath”)

5

80

100

120

140

160

180

200

0 5 10 15 20 25 30Depth (cm)

Dos

e (%

)

60

40

20

0

10 MV Photon

Proton SOBP (18cm)

Page 6: PHYSICS OF PROTON THERAPY

WHY HEAVY CHARGED PARTICLES?

Heavy charged particle therapy can reduce the dose load “intergral dose” to normal tissues surrounding the tumor target volume by a factor of 2-3 (reduced “dose bath”)

5

80

100

120

140

160

180

200

0 5 10 15 20 25 30Depth (cm)

Dos

e (%

)

60

40

20

0

10 MV Photon

Proton SOBP (18cm)

Excessive dose

Tumor

Page 7: PHYSICS OF PROTON THERAPY

WHY HEAVY CHARGED PARTICLES?

Increased dose conformality - dose gradient between tumor target volume and surrounding normal tissue.

Physics: ionization of atoms and possible direct breakage of one or both DNA strands.

Chemistry: creation of free radicals and other chemical changes.

Biology: damage to DNA and other intracellular targets - cell kill

6

Page 8: PHYSICS OF PROTON THERAPY

ENERGY LOSS•Protons predominantly lose energy via coulomb interactions with the outer shell electrons of the target.

excitation and ionization of atoms

•loss per interaction is small - “continuously slowing down approximation” (CSDA) is valid.

•Range of secondary e+ is < 1mm - locally absorbed dose.

•Protons are ~2000 times heavier than electrons - no significant deflection.

•Energy loss is given by Bethe-Bloch equation.

7

Page 10: PHYSICS OF PROTON THERAPY

ENERGY LOSS: BETHE-BLOCH EQUATION•δ is density correction - Screening of outer electrons from

inner shell electrons. Effect is larger in dense medium.•C is shell correction - important for low energies where the particle velocity is similar to velocity of atomic electrons.•Z, A - atomic number, atomic mass of absorber.• Tmax - max energy transfer to free electrons•I - mean excitation energy

•To first order -dE/dx ~ 1/β2 ⋍ 1/v2

•Tmax ⋍ 4Tmec2/mpc2 •T⋍ 200MeV → Tmax ⋍ 0.4MeV corresponds to electron range of 1.4mm. •For most electrons KE is much lower energy

http://physics.nist.gov/PhysRefData/Star/Text/PSTAR.html

0"

10"

20"

30"

40"

50"

60"

70"

80"

90"

0" 50" 100" 150" 200"

Stop

ping(pow

er([M

eV/(g/cm

2)](

Kine7c(Energy([MeV](

Stopping(Power(of(Protons(in(Water(

high E (large velocity) small

losses

low E (small velocity) large

losses

8

Stopping power = -dE/dx

Page 11: PHYSICS OF PROTON THERAPY

ENERGY LOSS: BRAGG PEAK

•Protons go almost straight in water - no significant deflection.•High energy protons - small energy losses •Low energy protons - large energy losses•Not all protons lose same energy - give rise to range straggling.

Depth in material

9

Averangerange R

Fraction of particlessurviving

Braggcurve

100 %

Most of the energydeposited at end of track

Width of the peak dependent on range straggling and initial energy spectrum

dE/dx

Energetic protons with minimal energy loss

dE/dx

Page 12: PHYSICS OF PROTON THERAPY

ENERGY LOSS: RANGE STRAGGLING

Convolution for range straggling

•Range Straggling:Protons lose their energy in individual collisions with electrons.Some protons lose more energy earlier in their journey they do not get to travel that far.Some protons suffer less collisions and get to travel farther.Protons with the same energy E0 have slightly different ranges - Range straggling is Gaussian and is 1% of R0.

10

Page 13: PHYSICS OF PROTON THERAPY

ENERGY LOSS: RANGE STRAGGLING

Convolution for range straggling

•Range Straggling:Protons lose their energy in individual collisions with electrons.Some protons lose more energy earlier in their journey they do not get to travel that far.Some protons suffer less collisions and get to travel farther.Protons with the same energy E0 have slightly different ranges - Range straggling is Gaussian and is 1% of R0.

10

Page 14: PHYSICS OF PROTON THERAPY

ENERGY LOSS: RANGE•Energy-range relationship, protons in water

11

Page 15: PHYSICS OF PROTON THERAPY

ENERGY LOSS: RANGE•Energy-range relationship, protons in water

11

Page 16: PHYSICS OF PROTON THERAPY

SPREAD OUT BRAGG PEAK RANGE & SOBP (MODULATION)

12

3 cm

50% 8 cm

75%

range 90%

depth!

Dose!

distal end

Page 17: PHYSICS OF PROTON THERAPY

SPREAD OUT BRAGG PEAK RANGE & SOBP (MODULATION)

12

3 cm

50% 8 cm

75%

range 90%

depth!

Dose!

distal end

3 cm

50% 8 cm

75%

modulation 90%

depth!

Dose!

distal end

proximal end

Page 18: PHYSICS OF PROTON THERAPY

COULOMB SCATTERING•Protons predominantly scatter due to elastic coulomb interaction with target nuclei•Many small angle deflections.•For radiotherapy a pure Gaussian approximation gives excellent result.•Full description is given by Moliere scattering and later by Highland approximation.

•Ɵ0 ∝ 1/pv ⋍ 1/(2T) ⋍ 1/K.E - for T<< 938MeV; T=1/2pv•Ɵ0 ∝ 1/LR0.5 •1.0 g/cm2 of water (LR=36.1 g/cm2) Ɵ0 ~5 mrad for 200MeV protons•1.0 g/cm2 of lead (LR=6.37 g/cm2) Ɵ0 ~14mrad for 200MeV protons•Large angle scattering in high Z materials.

13

Page 19: PHYSICS OF PROTON THERAPY

COULOMB SCATTERING

14

Page 20: PHYSICS OF PROTON THERAPY

NUCLEAR INTERACTION

•About 20% of incident protons have inelastic nuclear interactions with the target nuclei.•Reduction of primary proton fluence with depth.•Secondaries charged (p,d,α,recoil target nuclie) ~ 60% of energy - absorbed locally.

neutral (n,γ) ~ 40% of energy - absorbed in surrounding tissues.

•Production of unstable recoil particles (activation) radiation safety dose verification using PET/CT

15

Page 21: PHYSICS OF PROTON THERAPY

NUCLEAR INTERACTION

16

secondary protons generated in nuclear

interaction

Nuclear interactions take away the dose from the peak

Harald Paganetti

Page 22: PHYSICS OF PROTON THERAPY

NUCLEAR INTERACTION

•A certain fraction of protons have nuclear interactions with the absorbing matter (tissue), mainly with 16O.

•These protons are lost from the beam.•Janni data suggests 1% loss per cm

17

Page 23: PHYSICS OF PROTON THERAPY

NUCLEAR INTERACTION

•A certain fraction of protons have nuclear interactions with the absorbing matter (tissue), mainly with 16O.

•These protons are lost from the beam.•Janni data suggests 1% loss per cm

Depth (mm)

Arb.

Units

~70%

17

Page 24: PHYSICS OF PROTON THERAPY

ANATOMY OF PENCIL BEAM

18

overall shape from increase of dE/dx as

proton slows

1/r2 and transverse size set peak to entrance ratio

nuclear reactions take away from the peak and

add to this region

width from range straggling and beam

energy spread

Along the beam direction Transverse to the beam direction

Page 25: PHYSICS OF PROTON THERAPY

PENCIL BEAM INTO BROAD BEAM

Dividing the problem into two parts:

The Lateral Problem: How to spread the beam laterally to reach a uniform field up to 24cm or more in diameter?

The Longitudinal Problem:How to modulate the beam energy to reach a uniform field from a given range down to skin?

19

Page 26: PHYSICS OF PROTON THERAPY

BROAD MODULATED BEAM•Lateral scattering use thick scatterer or beam scanning (scanning magnets) makes the beam laterally uniform

20

Page 27: PHYSICS OF PROTON THERAPY

BROAD MODULATED BEAM•Lateral scattering use thick scatterer or beam scanning (scanning magnets) makes the beam laterally uniform

ScatteringDS

ScanningUS

20

Page 28: PHYSICS OF PROTON THERAPY

BROAD MODULATED BEAM•Lateral scattering use thick scatterer or beam scanning (scanning magnets) makes the beam laterally uniform

ScatteringDS

ScanningUS

20

•Pristine Bragg peaks with shifted ranges are superimposed with correct weights and is achieved by pulling back the range - stepped wheel spinning at high speed.

Page 29: PHYSICS OF PROTON THERAPY

DOUBLE SCATTERING (DS)

!  Block and Range compensator"!  Second Scatterer"

All layers in a SOBPs are delivered at the same time.

Range Modulator revolves at 600 RPMs

21

Pristine Bragg peak

lateral profile

depth profile

Nozzle

Page 30: PHYSICS OF PROTON THERAPY

UNIFORM SCANNING (US)

The

!  Fixed Scatterer"

!  Range Modulator

!  Scanning Magnet

The

Beam is delivered layer by layer. At any given time beam is a pristine Bragg peak of certain range

22

Block and compensator

Pristine Bragg peak

lateral profile

depth profile

Nozzle

Page 31: PHYSICS OF PROTON THERAPY

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-100 -80 -60 -40 -20 0 20 40 60 80 100

Nozzle EntranceAfter FS and RMAfter Scanning MagnetsAfter block and RC

Nozzle Entrance

After FS & RM

After Scanning Magnets

After block and RC

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-250 -200 -150 -100 -50 0 50 100 150 200 250

Nozzle EntranceAfter FS and RMAfter SSAfter block and RC

Nozzle Entrance

After FS & RM

After SS

After block & RC

LATERAL PROFILE

23

DS US

Page 32: PHYSICS OF PROTON THERAPY

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-100 -80 -60 -40 -20 0 20 40 60 80 100

Nozzle EntranceAfter FS and RMAfter Scanning MagnetsAfter block and RC

Nozzle Entrance

After FS & RM

After Scanning Magnets

After block and RC

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-250 -200 -150 -100 -50 0 50 100 150 200 250

Nozzle EntranceAfter FS and RMAfter SSAfter block and RC

Nozzle Entrance

After FS & RM

After SS

After block & RC

LATERAL PROFILE

23

DS US

These protons are stopped in the nozzle, snout, and aperture and generate neutrons.

Page 33: PHYSICS OF PROTON THERAPY

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-100 -80 -60 -40 -20 0 20 40 60 80 100

Nozzle EntranceAfter FS and RMAfter Scanning MagnetsAfter block and RC

Nozzle Entrance

After FS & RM

After Scanning Magnets

After block and RC

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-250 -200 -150 -100 -50 0 50 100 150 200 250

Nozzle EntranceAfter FS and RMAfter SSAfter block and RC

Nozzle Entrance

After FS & RM

After SS

After block & RC

LATERAL PROFILE

23

DS US

These protons are stopped in the nozzle, snout, and aperture and generate neutrons.

Lateral profile becomes important in clinical beam - is a function of depth

Page 34: PHYSICS OF PROTON THERAPY

NEUTRON YIELD

Brass aperture Brass aperture

Opening (target) Opening (target)

Small fields~20% of the beam treats~80% of the beam produces neutrons

Large fields~60% of the beam treats~40% of the beam produces neutrons

24

Page 35: PHYSICS OF PROTON THERAPY

LATERAL PENUMBRA

25

•What contributes to lateral penumbra

Page 36: PHYSICS OF PROTON THERAPY

LATERAL PENUMBRA

25

80 %

20 %

Penumbrawidth

width80-20 ~ σtot

•What contributes to lateral penumbra

Page 37: PHYSICS OF PROTON THERAPY

LATERAL PENUMBRA

25

80 %

20 %

Penumbrawidth

width80-20 ~ σtot

•What contributes to lateral penumbra

zp

zap

σ tot = σ 2source

zp − zapzap

$

% & &

'

( ) )

2

+σ rc2 +σ pat

2

(Virtual) Source

Aperture

geometric contribution

Page 38: PHYSICS OF PROTON THERAPY

LATERAL PENUMBRA

26

σ2 (

)

σtot

σsource

σpat + σrc

22

2

2patrc

ap

appsourcetot z

zzσσσσ ++"

"#

$%%&

' −=

Source: S. Safai

At shallow depth lateral penumbra is dominated by the source size.

As depth in the patient increases patient scatter starts to increase.

Beyond ~16cm patient scatter becomes dominant.

Page 39: PHYSICS OF PROTON THERAPY

RELATIVE BIOLOGICAL EFFECTIVENESS (RBE)

What is the difference in biological effectiveness between particles and photons?

Dosis

0 1 2 3 4 5 6 7 8

Survival

0.001

0.01

0.1

1

Dose [Gy]

Sur

vivi

ng F

ract

ion

RBE=Dx/Dp

X-rays

Particles

27

Page 40: PHYSICS OF PROTON THERAPY

RELATIVE BIOLOGICAL EFFECTIVENESS (RBE)

Dose per fraction [Gy]

1 10

RB

E

0.5

1.0

1.5

2.0

2.5

1.07±0.12

RBE = 1.1 for protons (general consensus)

RBE for heavier particles (C12) can be as high as 3.5

28

Page 41: PHYSICS OF PROTON THERAPY

RELATIVE BIOLOGICAL EFFECTIVENESS (RBE)

RBE increases with depth because linear energy transfer (LET) increases with depth.

For protons we assume 1.1 flat for the entire SOBP.

Biological effective range increases by 1-2mm. 29

Carbon 12

Protons

Page 42: PHYSICS OF PROTON THERAPY

LINEAR ENERGY TRANSFER (LET)LET of charged particle in a medium is the ratio of dE/dl, where dE is the average energy locally imparted to the medium by a charged particle in traversing a distance of dl.

LET < 10 KeV/μm low LETLET > 10 KeVμm high LET

250 kVp X-Rays: 2 keV/μm3 MeV X-Rays: 0.3 keV/μm14 MeV neutrons: 12 keV/μmheavy charged particles: 100-200 keV/μm10 keV electrons: 2.3 keV/μm

30

Page 43: PHYSICS OF PROTON THERAPY

DOSE CALCULATIONPencil beam formalism

Dose from a single beam at a point (x’,y’,z’) relative to the source position of the pencil beam

31

Page 44: PHYSICS OF PROTON THERAPY

DOSE CALCULATIONPencil beam formalism

Dose from a single beam at a point (x’,y’,z’) relative to the source position of the pencil beam

31

)',','()'()',','( zyxOzCzyxd p ⋅=

Central Axis term

C(z / ) = DD(deff )ssd0 + deff

z /"

# $

%

& '

2

Broad beam depth dose distribution. Measured in the

water phantom and modified by an inverse square law €

O(x',y ',z') =1

2π σ tot (z')[ ]2exp x'2 +y'2

2 σ tot (z')[ ]2$

% & &

'

( ) )

Off-Axis term

Is the lateral flux distribution from the radial emittance suffered by protons

directed along the axis of the pencil beam

Page 45: PHYSICS OF PROTON THERAPY

DOSE CALCULATIONPencil beam formalism for broad beam - sum of multiple pencil beams

32

Page 46: PHYSICS OF PROTON THERAPY

DOSE CALCULATIONPencil beam formalism for broad beam - sum of multiple pencil beams

32

Page 47: PHYSICS OF PROTON THERAPY

DOSE CALCULATIONPencil beam formalism for broad beam - sum of multiple pencil beams

32

Dose at a point in a beam is a integral of over all pencil beams:

Dp (x,y,z) = dx 'dy' C(x ',y',z)2π σ tot (x ',y',z)[ ]2

exp − (x '−x)2 + (y '−y)2

2 σ tot (x',y ',z)[ ]2%

& ' '

(

) * * ∫∫

Page 48: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

Page 49: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

High-DensityStructure

BodySurface

CriticalStructure

TargetVolume

Beam

Page 50: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

High-DensityStructure

BodySurface

CriticalStructure

TargetVolume

Beam

water equivalent thicknesssets the range for treatment

Page 51: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

Aperture

High-DensityStructure

BodySurface

CriticalStructure

TargetVolume

Beam

water equivalent thicknesssets the range for treatment

Page 52: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

Aperture

High-DensityStructure

BodySurface

CriticalStructure

TargetVolume

Beam

water equivalent thicknesssets the range for treatment

Page 53: PHYSICS OF PROTON THERAPY

FIELD DOSE SHAPING

33

Aperture

High-DensityStructure

BodySurface

CriticalStructure

TargetVolume

Beam

water equivalent thicknesssets the range for treatment

Page 54: PHYSICS OF PROTON THERAPY

DOSE PLANNING

34

Typically 120 keV

CT-scan in Hounsfield units

It measures linear attenuation

10002

2 ⋅−

−=

airOH

OHXHUµµ

µµ

Substance HU

Air -1000

Fat -120

Water 0

Muscle +40

Bone +400 or more

Page 55: PHYSICS OF PROTON THERAPY

DOSE PLANNING

35

0

0.25

0.5

0.75

1

1.25

1.5

-1000 -750 -500 -250 0 250 500 750 1000 1250 1500

CT Value

Rel

ativ

e st

oppi

ng p

ower TPS

Measured

Water

Lung

Air

Bone

Metal

CT vs relative electron density for photons

CT vs relative stopping power for protons

Page 56: PHYSICS OF PROTON THERAPY

DOSE PLANNING

35

0

0.25

0.5

0.75

1

1.25

1.5

-1000 -750 -500 -250 0 250 500 750 1000 1250 1500

CT Value

Rel

ativ

e st

oppi

ng p

ower TPS

Measured

Water

Lung

Air

Bone

Metal

Photons ProtonsElectron density Proton stopping power

80

100

120

140

160

180

200

0 5 10 15 20 25 30Depth (cm)

Dos

e (%

)

60

40

20

0

10 MV Photon

Proton SOBP (18cm)

Any error in CT will result in uncertainty in

proton range calculations

CT vs relative electron density for photons

CT vs relative stopping power for protons

Page 57: PHYSICS OF PROTON THERAPY

DOSE PLANNING

%5 error in CT calibrationcan have couple of % error in photon depth dose. Tumor is still covered

Page 58: PHYSICS OF PROTON THERAPY

DOSE PLANNING

%5 error in CT calibrationcan have couple of % error in photon depth dose. Tumor is still covered

Photons ProtonsElectron density Proton stopping power

0

20

40

60

80

100

120

140

160

180

200

0 5 10 15 20 25 30 35

Photon (10MV)

Photon (10MV)

Proton SOBP (18cm)

Proton SOBP (18cm)

Unrealistic5% errorin CT conversion

5% error in CT can change the proton range.- miss part of tumor- overshoot tumor into critical structure like cord, brainstem- RBE is also large at the end.

Page 59: PHYSICS OF PROTON THERAPY

SUMMARY

37

Page 60: PHYSICS OF PROTON THERAPY

SUMMARY•Physics: Energy loss, scattering, nuclear interactions lead to Bragg peak shaped dose distribution.•Pencil beam is converted to a broad beam by lateral and longitudinal spreading.•Range straggling modifies the beam at distal end along the direction of beam ~1% of the Range.•Lateral penumbra modifies the beam in a transverse direction to the beam ~2% of the Range.•RBE modifies the does depths and increases range by ~2mm.•Dose calculation based on pencil beam algorithm. •Sculpting raw proton beam into clinical deliverable beam.

37

Page 61: PHYSICS OF PROTON THERAPY

THANK YOU

38