Top Banner
PGA309 EVM Multi-Cal-Tools- V1.6a User’s Guide Iven Xu April. 25 th , 2013 Thanks to Ian Williams 1
41

PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

Mar 26, 2015

Download

Documents

Luke Wentworth
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

PGA309 EVM Multi-Cal-Tools-V1.6a User’s Guide

Iven Xu

April. 25th, 2013

Thanks to Ian Williams

1

Page 2: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

Agenda

• USB_DAQ and Multi-Cal-Tools overview

• Hardware configuration

• Software configuration

2

Page 3: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

3

a) USB_DAQ and Multi-Cal-Tools overview

Page 4: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview

• USB_DAQ Board Default Configuration:

4

Page 5: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multi-Cal-Tools Board:

5

Page 6: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multi-Cal-Slave Board(optional board):

6

Page 7: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multi-Cal-Interface PCA board:

7

Page 8: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multi-Cal-Interface Cable:

8

Page 9: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multi-Cal-Power Cable:

9

Page 10: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Final setup overview:

10

This is our test board, you can connect your PGA309 board to J* or T* of interface board.

Page 11: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

a) USB_DAQ and Multi-Cal-Tools overview• Multimeter:

– Agilent 34401A (GPIB, Real RS232)

– TH1961 (USB-to-RS232)

• Our software “***.v1.6a” can support those three different interface– Download link:

– ftp://ftp.ti.com/pub/linear_apps/Multi-Cal-PGA309/

11

Page 12: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

12

b) Hardware configuration

Page 13: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:

• Step 1:

13

Page 14: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:• Step 2:

14

Page 15: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:• Step 3:

15

This is our test board, you can connect your PGA309 board to J* or T* of interface board.

Page 16: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:

• Step 4:

(4-wire voltage output)

Notes:

(Same GND for the

below two power supply)

Power supply #1:

+Vdut =5V

-Vdut=GND

Power supply #2:

+-15V

GND16

Page 17: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:

• Step 4:

(4-wire voltage output)

Notes:

(Same GND for the

below two power supply)

Power supply #1:

+Vdut =5V

-Vdut=GND

Power supply #2:

+-15V

GND17

Page 18: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:

• Step 5:

(4-wire current output)

Notes:

(Float GND for

power supply #1)

Power supply #1:

+Vdut =24V

-Vdut≠GND

Power supply #2:

+-15V

GND18

Page 19: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

b) Hardware configuration:

• Step 5:

(4-wire current output)

Notes:

(Float GND for

power supply #1)

Power supply #1:

+Vdut =24V

-Vdut≠GND

Power supply #2:

+-15V

GND19

You can use your own Multi-meter with current measurement!

Page 20: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

Extending the system with Slave Board:

20

Page 21: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

21

c) Software configuration

Page 22: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:

22

Windows 7 Attention

Page 23: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Press “Select Model File” to select correct model file.

• Also, you can use “”Create Model File” to create a new model file.

23

Page 24: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:

• Default Model Files for you:

24

Page 25: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Press “Start Cal” and then press the four Yes and Continue with Calibration

25

Page 26: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Press “Continue” to do the Fast Scan.

• For the below picture, channel 0 and 1 is ok because I only connected two sensor modules to channel 0 and 1.

26

Page 27: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Press “Continue” to do the calibration

• Below result was tested with TH1961 (with USB-232 rather than GPIB).

27

Page 28: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• The other thing you need to pay more attention is “Address of TH1961”,

• The current address of TH1961 we used is “COM 6”.

• You need to change it in DMM script file.

• Detailed step is in the below slides:

28

Page 29: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• How to create a new model file to match the address of TH1961.

29

Page 30: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Take voltage output as an example:

• Select “4wire_4p096_diode_USB.txt”

30

Page 31: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Take voltage output as an example:

• Press “Edit DMM Script” and select “Voltage_TH1961_USB.txt”

31

Page 32: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Take voltage output as an example:

• Change the COM port in the open script file. Make sure it is the same as your Multimeter(TH1961) address.

32

Page 33: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Take voltage output as an example:

• After you change the script file, please save the file and re-select the script file again. Then press “save” to save the Model File.

33

Page 34: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

c) Software configuration:• Take voltage output as an example:

• Save the Model File:

34

Page 35: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

• Select the new Model File, then you can press “Start Cal” to do the calibration.

35

c) Software configuration:

Page 36: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

4-wire voltage output result:

• After the calibration:

36

Page 37: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

For current output:

• Model File:– Iout_4p096_diode_USB.txt

• DMM script file:– Current_TH1961_USB.txt

Also need to pay more attention on the address of DDM!

37

Page 38: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

Thanks!

38

Page 39: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

Back Up!If you are using Agilent 34401, pls see

the below two slides of configuration

files.39

Page 40: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

If you use Agilent 34401A with GPIB:

• Voltage output:– Model File:

• 4wire_4p096_diode.txt

– DMM script file:• Volt_34401_GPIB.txt

• Current output:– Model File:

• Iout_4p096_diode.txt

– DMM script file:• Current_34401_GPIB.txt

Also need to pay more attention on the address of DDM!

40

Page 41: PGA309 EVM Multi-Cal-Tools-V1.6a Users Guide Iven Xu April. 25 th, 2013 Thanks to Ian Williams 1.

If you use Agilent 34401A with RS232:

• Voltage output:– Model File:

• 4wire_4p096_diode.txt

– DMM script file:• Voltage_34401_rs232.txt

• Current output:– Model File:

• Iout_4p096_diode.txt

– DMM script file:• Current_34401_rs232.txt

Also need to pay more attention on the address of DDM!

41