Patient Setup for Precision Proton Therapy Hsiao-Ming Lu, Ph.D.
48

Patient Setup for Precision Proton Therapy

Jan 06, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Patient Setup for Precision Proton Therapy

Patient Setup for Precision Proton Therapy

Hsiao-Ming Lu, Ph.D.

Page 2: Patient Setup for Precision Proton Therapy

• What’s so special about setup for PT?• Anything different in immobilization?• Imaging systems• Organ motion issues• Future tools

Content

Page 3: Patient Setup for Precision Proton Therapy

• Clinical data show significant decrease in tumor control (≥ 5%) when tumor dose decreased by 4 - 5 %

• Therefore, “significant”portion of CTV cannot fall outside high dose region more than once in fractionated course of radiation L. Vehey UCSF

What Precision Is Needed?

Page 4: Patient Setup for Precision Proton Therapy

What Precision Is Needed?

• The uncertainty of the target position is more important than the uncertainty of the dosimetry– Daily dose errors of ≥ 20 % possible with

positioning error– Dose errors due to errors in dose

calculation probably not more than 5 - 10%

L. Vehey UCSF

Page 5: Patient Setup for Precision Proton Therapy

Proton Dose Distribution Extra SensitiveVery Sensitive to WED Variations

Over or Undershoot!

Page 6: Patient Setup for Precision Proton Therapy

• Body posture variationsIntra or inter-fractionalCauses different tissue build up in beam path

Requires solid immobilization/registration• Lateral displacement

Causes misalignment of compensatorSetup tolerance must be consistent

not only with planning marginsbut also with compensator smearing

WED Variations due to Set Up

Page 7: Patient Setup for Precision Proton Therapy

Inter-fraction Set Up Variation

Head and Neck

Pelvis

Rosenthal 2004

Page 8: Patient Setup for Precision Proton Therapy

Hair

• Style & Placement• Clips (even plastic)• Gel & Water

Potential Range Perturbation

Rosenthal 2004

Page 9: Patient Setup for Precision Proton Therapy

• Generic requirementsSmall inter- and intra-fractional variationUncertainty well-understoodPatient comfortableFixed to treatment table, if possiblePractical

Immobilization

• Proton specificNo potential range perturbations

e.g., passing part of the frameHardware clearance

Page 10: Patient Setup for Precision Proton Therapy

• Dedicated group (3)• Dedicated equipment• Dedicated session

before CT

No Small Matter

Page 11: Patient Setup for Precision Proton Therapy

Immobilization devices

• Alpha Cradle• Bean Bag• Duncan Head Rest• Leg Abductor• Wing Board

• Base of Skull *• IC-Mask (reinforced)• Modified GTC• Neck Cast *• Para-Nasal Sinus *

Non-rigid DevicesRigid Devices

Engelsman 2005

Page 12: Patient Setup for Precision Proton Therapy

Immobilization For Proton Therapy

Allow treatment snout proximity to cervical/thoracic spine

Prevent Articulation of head/neck

Base of Skull (BOS)Contoured Carbon Base (0.3 - 0.5 cm)Mask w. Low Profile Pin Arms

Rosenthal 2004

Page 13: Patient Setup for Precision Proton Therapy

Immobilization For Proton TherapyIntra-Cranial (IC)

Curved Wafer Filled Carbon BaseDense Foam PadReinforced Mask

Rosenthal 2004

Page 14: Patient Setup for Precision Proton Therapy

Immobilization For Proton TherapyParanasal Sinus (PNS) Frame

Sturdy Lucite Base and ArchDense Foam Head PadRigid Bite Mold - Prosthedontal

Table 1A: Measured motion between set-up and post treatment film pairs using the PNS immobilization system (N=20).

(Mean time between films = 16 ± 4 minutes (SD)) Average Standard

Deviation A-P 0.3 mm 0.4 mm S-I 0.0 mm 0.5 mm Lat. -0.4 mm 0.5 mm 3-D Displacement 0.9 mm 0.4 mm 2-D Displacement 0.7 mm 0.4 mm Roll 0.07 degrees 0.45 degreesPitch 0.03 degrees 0.42 degreesTurntable -0.14 degrees 0.67 degrees

Rosenthal 2004

Page 15: Patient Setup for Precision Proton Therapy

In-House DevicesRadionics Gill-Thomas-Cosman (GTC) Frame

Occipital Cup with Heavy-Body Reprosil ®Velcro Head RestraintBite Mold

Bussiere 2006

Page 16: Patient Setup for Precision Proton Therapy

GTC ModificationsReplace High Density Occipital Cup with MoldcareTM Cushion

Bussiere 2006

Page 17: Patient Setup for Precision Proton Therapy

GTC ModificationsCushion

Bussiere 2006

Page 18: Patient Setup for Precision Proton Therapy

Immobilization For Proton Therapy

Medulloblastoma FrameFull Body Prone MoldMask, Head/Body Articulation Minimal

Rosenthal 2004

Page 19: Patient Setup for Precision Proton Therapy

CSI (Cranio-Spinal Immobilization)

Rosenthal 2004

Page 20: Patient Setup for Precision Proton Therapy

Medulloblastoma – CSI + posterior fossa boost

Match lines

Rosenthal 2004

Page 21: Patient Setup for Precision Proton Therapy

2 Spine fields matching at anterior cord surface

Dose %11010510210090805010

Protons Photons

Non-aligned divergence

Rosenthal 2004

Page 22: Patient Setup for Precision Proton Therapy

Current• Orthogonal X-ray images (DIPS)

Digital image panelsIntegrated with TPSLandmark matching to DRR(bony anatomy or implants) 6D correction

• Ultrasound (prostate only)

Imaging System

Page 23: Patient Setup for Precision Proton Therapy

Radiographic Set Up w.Diagnostic X-rays DRR From CT

1993 Implanted Fiducials Digitized from Film 3D - High Accuracy - 1/2 mm (Gall, Verhey ‘93)

1996 Digital Radiography – Filmless CCD Camera

Flat Panel Imager Receptors at NPTCSmaller Lighter Device, Accurate Couch, High Resolution

1999

1980

History proton X-ray positioning

Digirad – Application Integrates Planning and Set Up2001

2003/4DIPSTM2 – Dicom Database, TCS Integration Ready, Feature Matching Automation Rosenthal

2004

Page 24: Patient Setup for Precision Proton Therapy

Repositioning For Proton Therapy

X-rays ( anterior-posterior )

Orthogonal X-ray Tubes

Simultaneous Lateral and AP Radiographs

Rosenthal 2004

Page 25: Patient Setup for Precision Proton Therapy

Repositioning For Proton Therapy

X-Ray Tubes

Nozzle Snout (withaperture & compensator)

6-axis patient positioner

Rosenthal 2004

Page 26: Patient Setup for Precision Proton Therapy

Cage Source

Cage Detector

Nozzle Detector

Nozzle Source

Hitachi gantry with orthogonal X-ray system

Zhu, 2007

Page 27: Patient Setup for Precision Proton Therapy

Repositioning For Proton Therapy

Amorphous Silicon Flat Panel Imagers

Rosenthal 2004

Page 28: Patient Setup for Precision Proton Therapy

Repositioning For Proton Therapy

DIPS Radiographic Set Up Software

Gold Seeds in Prostate

Rosenthal 2004

Page 29: Patient Setup for Precision Proton Therapy

Diagnostic x-ray positioningDIPSTM Relies on Identifying Landmarks

Rosenthal 2004

Page 30: Patient Setup for Precision Proton Therapy

1) Patient immobilization2) Initial positioning (lasers and skin marks)3) Take AP/LAT orthogonal image pair

General position adjustment(translation and rotation)

4) Take beam-line image (BEV)Small adjustment (translation only)(Tolerance must agree with smearing)

Setup Process

Page 31: Patient Setup for Precision Proton Therapy

Set Up Uncertainty - Summary

Intercranial tissues movement

• 1 mm invasive skull fixation

• 2-3 mm high quality head mask

• 3 mm neck articulation

Associated with breathing and cardiac motion:

• Lung tumors• Diaphragm• Liver• Kidneys• Pancreas

3.9 mm12-17 mm10-25 mm11-19 mm20 mm

K. Langen NAC Capetown SA

Digestive and Muscular Variation:

• Prostate• Bladder• Rectum• Abdomen

Highly Unpredictable

Daily Verification or Tracking Required

Rosenthal 2004

Page 32: Patient Setup for Precision Proton Therapy

Sonarray System3D ultrasound to CT contours

Ultrasound for Prostate

Page 33: Patient Setup for Precision Proton Therapy

Workstation

CT-Sim Room

U/S-SIM

Ultrasound to Ultrasound The RESTITU Platform

Treatment Room

U/S-GUIDE

Page 34: Patient Setup for Precision Proton Therapy

STAR (Stereotactic Alignment radiosurgery)Fixed Beam SRS/SRT

Bussiere 2006

Page 35: Patient Setup for Precision Proton Therapy

Imaging/Alignment for STARFilm (3D) → Digital (2+D) → Digital 3D

Varian PaxScan 4030: 127 μm pixel

(Gadolinium Oxysulfide Scintillator)

Bussiere 2006

Page 36: Patient Setup for Precision Proton Therapy

Ocular Treatment

Bussiere 2006

Page 37: Patient Setup for Precision Proton Therapy

Imaging uncertainty due to respiration

Organ Motion

• Instant X-ray image

• Averaged DRR from 3D CT

Page 38: Patient Setup for Precision Proton Therapy

Example: LiverRL fieldUse same range, mod, compensator for 50% and 0% phases 50% 0% (50% field)

Proton Range Variation

In many cases, gating is the only option

Page 39: Patient Setup for Precision Proton Therapy

• Gated imaging• Use bony landmarks

for pitch, roll, and rotation

• Use seed implants for translation

Patient Set Up for Gating

Page 40: Patient Setup for Precision Proton Therapy

• Verify patient posture• Verify treatment

position correction• Monitor intra-

fractional motion• Respiratory gating by

tracking multiple points

Future – Surface Imaging

Bert, C. 2004

Page 41: Patient Setup for Precision Proton Therapy

• Fast digital fluoroscopy panels• Organ motion management• Direct target volume tracking• Respiratory gating by target position

Future – Fluoroscopy

Page 42: Patient Setup for Precision Proton Therapy

Fluoroscopic Tracking - Liver

Visualize Clips Computer Tracks Clips

Rosenthal 2004

Page 43: Patient Setup for Precision Proton Therapy

• 3D target verification

• Adaptive therapy

• On-line range verification

Future – Cone Beam CT/CT

Page 44: Patient Setup for Precision Proton Therapy

The ultimate proof!But, post-analysis only, non-interactive

In Vivo Range VerificationActivated PET imaging, in-beam, or offline

Page 45: Patient Setup for Precision Proton Therapy

Detectors: MOSFET, TLD, DiodesSurface, cavity or implant

What not Point Dose Measurement

Widely practiced in photon/electron therapy

Page 46: Patient Setup for Precision Proton Therapy

Dose at point A does not verify distal coverage at point B

Not So Easy for Protons

Also need residual proton range at point A

Page 47: Patient Setup for Precision Proton Therapy

Stanley RosenthalMarc BussiereMartijn EngelsmanKatia ParodiAlejandro MazalRonald Zhu……

Acknowledgment