PATIENT LIFT DIGITAL 700 SCALE INSTALLATION GUIDE FOR MODELS MDS400EL/450EL/600EL/700EL Product does not contain Dry Natural Rubber Latex. MDSMELSCALE
6

PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

Jun 24, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

PATIENT LIFT DIGITAL 700 SCALEINSTALLATION GUIDEFOR MODELS MDS400EL/450EL/600EL/700EL

Product does not contain Dry Natural Rubber Latex.

MDSMELSCALE

Page 2: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

2 MEDLINE MDSMELSCALE | PIC-00364 (Rev: 05-03-17)

1. Load cell scale ----------------1

2. Bushing -----------------------1

3. Universal bracket ------------1

4. Pin ----------------------------2

5. Key ring -----------------------2

The digital MDSMELSCALE is designed to use with the lifts as shown above. The scale may be set to read 0-700 lbs or 0-317 Kgs. The scale will not increase the weight capacity of the lift and therefore never attempt to lift anyone whose weight is above the lifts’ safe working load. For accurate reading a secure hardware fit is required therefore do not attempt to use other scale brands on these lifts.

Each scale is calibrated at the factory prior to shipping. The calibration certificate is in thebottom of the carton specific to the unit serial number printed on the right side of the scale. The scale is powered by one replaceable 9-volt battery, which is stored inside the scale. Removing the black rubber plug on the right underside of the scale will access the battery. Replace batteries as required. Properly dispose used batteries.

1

5

43

2! Please make sure all parts shown are included in the

package.

Page 3: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

1-800-MEDLINE (633-5463) | medline.com 3

!• Use of the MDSMELSCALE will not increase the lift capacity of the lift. Always

refer to the lift safe working load limit and do not attempt to lift more than the lifts’ load limit.

• Perform all lift procedures as described in your owner’s manual. Always check that all parts are safely attached and secured prior to attempting a lift.

• Torn, frayed or damaged slings could fail and result in serious injury or death.

Check slings carefully prior to each lift and discard and replace any that show signs of wear and tear. Never use a sling with a lower safe working load limit than the weight of the resident/patient/user.

Prior to assembly please ensure all parts as noted above are present. Then follow these step-by-step instructions:

MDSMELSCALE ASSEMBLY

Step 1: Remove the spreader bar from the lift boomStep 2: Place the metal bushing (Part #2) in the hole in the upper scale unit as shown above Step 3: Attach the upper part of the scale to the boom using the clevis pin and ring clip

previously on the spreader bar.Step 4: Attach the universal scale bracket (Part #3) to the bottom of the scale using one pin

(Part #4) and one ring clip (Part #5)Step 5: Re-attach the spreader bar to the bottom of the scale unit using the remaining pin

(Part # 4) and ring clip (Part #5)

Page 4: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

4 MEDLINE MDSMELSCALE | PIC-00364 (Rev: 05-03-17)

HOW TO USE THE MDSMELSCALEStep 1: Place the sling to be used for weighing on the spreader bar without patient/resident/

user.Step 2: Press the On/Zero button on the scale faceplate to “zero-out” the reading (the weight

of the sling will remain zeroed out until the Off button is pressed)Step 3: Press the “Kg/Lb” button to select the weight option you require.Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

the weight is supported by the lift. Stabilize the movement of the sling after lifting.Step 5: Press the “Operate” button to display the weight of the patient/ resident/user in the

previously selected Kg or Lb unitStep 6: Lower the patient safely and either continue to next patient with same sling zeroed

out or change sling and repeat steps 1-5.Step 7: Press the “Off” button to turn off the scale when weighing procedure has been

completed.Step 8: Repeat Steps 1-7 for each subsequent weighing procedure.

MAINTENANCE

1. Always keep the scale in horizontal position.2. Keep the scale in a cool and dry place.3. Clean the scale with a damp cloth and avoid getting water inside the scale. DO NOT use

chemical/abrasive cleaning agents.

WARRANTYThere is a one year limited warranty against manufacturers defects, excluding normal batery replacement.

Page 5: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

1-800-MEDLINE (633-5463) | medline.com 5

PREPARE 400LB CERTIFIED WEIGHT.

1. Press the button to turn on the scale. Press the button to set weight unit

to “Lb”. Then press the button to turn off the scale.

2. With the scale turned off, press and hold the button and then press the button

once. Release the button when “CAL?” appeared on the LCD display.

3. Press the button one more time, LCD display will show “YES”. Then it will show

“LoAd”, “200.0_”, “or”, “400.0_” in sequence. Then it will alternatively display “200.0”, “or”,

“400.0”.

4. Attach 400lb weight to the scale, and the LCD display will show “400.0_”. For accurate

results, it’s recommend applying the calibration load directly onto the scale

5. Press the button to confirm. The LCD display will show “DonE” then “400.0_”. Then

the LCD display will begin to show count down “99999”, “88888”, “77777”, “66666”, “55555”,

“44444”, “33333”, “22222”, “11111”, “00000”.

6. After the count down, the LCD display will show “400.0”. Now turn off the scale by pressing

the button.

7. Unload the 400lb weight.

8. Calibration completed.

(Note: Tolerance +/- 2lbs.)

MDSMELSCALE CALIBRATION INSTRUCTIONS

Operate/Hold

lb

kgZero

Hold

Page 6: PATIENT LIFT DIGITAL 700 SCALE › media › catalog › Docs › MKT › MAN... · Step 4: Carefully place the patient/resident/user in the sling and perform a lift so that all of

Medline México01-800-831-0898www.medlinemexico.com | [email protected]

Medline Canada1-800-396-6996www.medline.ca | [email protected]

Medline United States1-800-MEDLINE (633-5463)www.medline.com | [email protected]

Some products may not be available for sale in Mexico or Canada. Please contact your Account Representative for additional information. We reserve the right to correct any errors that may occur within this instruction sheet. ©2017 Medline Industries, Inc. Medline is a registered trademark of Medline Industries, Inc.

MDSMELSCALE | PIC-00364 (Rev: 05-03-17)

Medline Industries, Inc. Northfield, IL 60093 USA1-800-MEDLINE (633-5463)medline.com | [email protected]

FOLLOW US BLOG