Top Banner
Parameters & Observation s
20

Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Dec 19, 2015

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Parameters &

Observations

Page 2: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Level 1: Patients Level 2:

Physicians

User Level

Level 3: Developer

Page 3: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

User Level

Level 1: Patients Level 2:

Physicians

Level 3: Developer

Page 4: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Use MobileEyes

Change parameters using MobileEyes.

MobileEyes screens the parameter, depending on the type of user.

All values are recorded in millimeter and second

Page 5: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

To access the parameters that have any significant change in the behavior of the chair, the Parameter

Priority has to be set to Expert.

The Higher Parameters

Page 6: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Quite fast. Work with speed/velocity parameters

Starts and ends with a jerk.

Turning (especially through a small angle) is not smooth.

Does not go through doorways.

Almost runs into obstacles, before coming to a stop.

Perhaps parameters for rotation?

Padding parameters

Padding parameters

Main Concerns

Perhaps parameters for acceleration?

Page 7: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Each Drive Mode has a different set of parameters e.g.

Changing parameter in one list, has no effect on the parameters of the other list, even though the parameters may have the same name.

Parameter Lists

Page 8: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Parameter list – Express Drive

Robot Config - a short list of speed related parameters.

Path Planning Setting - full list of parameter for Smart Drive.

Page 9: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Parameter list - Smart Drive

• TeleOp Settings - full list of parameter for Smart Drive

Page 10: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

List of Modified Parameters

MaxVelTransAccel TransDecel

SideClearanceAtSlowSpeedFrontPaddingAtSlowSpeedFrontPaddingAtFastSpeed

ForwardSideClearanceAtSlowSpeed

Page 11: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Procedural Standards

Did test trials: Number of test trials ranged from 6 to 15.

All test trials were done over a distance of 20ft.

When testing each parameter:

simulate a hospital setting.

judged from a patient’s point of view.

Page 12: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

MaxVel Controls the speed of the chair directly.

• In Express Drive: MaxVel = 600

On average 8.3sec to cover a distance of 20ft.

• In Smart Drive: MaxVel = 900

On average 8.6sec to cover a distance of 20ft.

Page 13: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

TransAccelIncreasing the value, decreases the time taken by

the wheelchair to reach maximum speed.

• In ExpressMode:TransAccel = 425

Lower value results in jerkier start.

Page 14: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Slows down the wheelchair.Braking distance is 1.5m.

• In Express Mode:TransDecel = 175

If TransDecel < 175, stops smoothly but overshoots goal.

TransDecel

Page 15: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

TransDecelOvershooting the Goal

Page 16: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Side Padding Any object at this specified distance from its side,

is considered an obstacle.

• In Express Drive:SideClearanceAtSlowSpeed = 50

• In Smart Drive:ForwardSideClearanceAtSlowSpeed = 25

Both values set to allow the wheelchair to go through door ways.

Page 17: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Front PaddingAny object at this specified distance from its

front, is considered an obstacle.

• In Express Drive:FrontPaddingAtSlowSpeed = 300Applies when speed < than MaxVel.

FrontPaddingAtFastSpeed = =500Applies only if speed of wheelchair is at

maximum.

Page 18: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Results

Starting and ending are smoother.

Speed is comfortable.

Can go through door ways which are at least 33.5 inches (851mm) wide.

Stops at a safe distance from the obstacle.

Turning is smoother than before but still a bit jerky.

Page 19: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.

Challenges

Overcoming the fear of making changes.

Could not figure out a way to measure the effect of RotVelMax.

Page 20: Parameters & Observations. Level 1: Patients Level 2: Physicians User Level Level 3: Developer.