Top Banner
Pacemaker troubleshooting- single chamber pacemakers
63
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Pacemaker troubleshooting- single chamber pacemakers.

Pacemaker troubleshooting-single chamber pacemakers

Page 2: Pacemaker troubleshooting- single chamber pacemakers.

Reasons for evaluation

• Patient symptomatic– Palpitation– Syncope – presyncope

• Pacemaker malfunction suspected– ECG– Telemetry– Ambulatory ECG

• Routine pacemaker follow up

Page 3: Pacemaker troubleshooting- single chamber pacemakers.

Patient details

• Indication for pacing• Implant operative note• Diagnosis• Medication –can alter pacing tresholds• DC ,MRI,electrocautery• Trauma,electrical current exposure• Concurrent medical problems-CRF,hyperkalemia• Prior ECG,CXR

Page 4: Pacemaker troubleshooting- single chamber pacemakers.

Pacemaker system– manufacturer– Model– current programming– date of implant– special features– Sensing and pacing tresholds– impedance

Page 5: Pacemaker troubleshooting- single chamber pacemakers.

Lead system– manufacturer– model– polarity– Fixation– insulation and date

Page 6: Pacemaker troubleshooting- single chamber pacemakers.

Causes of pacemaker malfunction

• Pacing stimuli present with failure to capture• Pacing stimuli present with failure to sense• Pacing stimuli absent– Oversensing– No output

• Pseudomalfunction

Page 7: Pacemaker troubleshooting- single chamber pacemakers.

Failure to capture

• No evidence of depolarization after pacing artifact

Loss of capture

Page 8: Pacemaker troubleshooting- single chamber pacemakers.
Page 9: Pacemaker troubleshooting- single chamber pacemakers.
Page 10: Pacemaker troubleshooting- single chamber pacemakers.
Page 11: Pacemaker troubleshooting- single chamber pacemakers.

Early post device implantation

• CXR abnormal– Lead dislodgement

• Downward migration of PG• Failure to secure anchoring sleeves properly• Too little or too much slack

– Lead perforation

• Elevated impedance– Loose set screw– failure to seat lead pin properly in header

• Normal CXR,normal impedance– Micro lead dislodgement– Inflammatory response

Page 12: Pacemaker troubleshooting- single chamber pacemakers.

Late post device implantation

• Battery depletion• Twiddler s syndrome• Abnormal myocardium• Insulation failure• Conductor failure– Mechanical stress on lead

• Anchoring sleeve• Interaction with generator in pocket• b/w clavicle and first rib

Page 13: Pacemaker troubleshooting- single chamber pacemakers.

• Increase the energy in the output pulse– Run a capture threshold test– Adjust the output parameters, if necessary• Pulse amplitude (V)• Pulse width or duration (ms)

– It is generally more efficient to increase the pulse amplitude

• Investigate possible lead problems• Reprogram device polarity

Page 14: Pacemaker troubleshooting- single chamber pacemakers.

Undersensing

• An intrinsic depolarization that is present, yet not seen or sensed by the pacemaker

P-wavenot sensed

Atrial Undersensing

Page 15: Pacemaker troubleshooting- single chamber pacemakers.

• Undersensing occurs when the pacemaker does not detect intrinsic activity that really is there

• Undersensing causes the pacemaker to pace more than it should

Page 16: Pacemaker troubleshooting- single chamber pacemakers.
Page 17: Pacemaker troubleshooting- single chamber pacemakers.
Page 18: Pacemaker troubleshooting- single chamber pacemakers.
Page 19: Pacemaker troubleshooting- single chamber pacemakers.

Undersensing May Be Caused By:

• Inappropriately programmed sensitivity• Lead dislodgment• Lead failure:– Insulation break; conductor fracture

• Lead maturation• Change in the native signal • Functional – Magnet– Noise reversion

Page 20: Pacemaker troubleshooting- single chamber pacemakers.

• Adjust the sensitivity setting– Run a sensing threshold test– Measure the intrinsic signals– Adjust the sensitivity appropriately• To increase sensitivity, decrease the mV setting

– Make all changes to sensitivity settings in small steps since large changes may only introduce new sensing problems

Page 21: Pacemaker troubleshooting- single chamber pacemakers.
Page 22: Pacemaker troubleshooting- single chamber pacemakers.

Oversensing

Ventricular Oversensing

Marker channel shows intrinsic activity...

...Though no activity is present

• The sensing of an inappropriate signal – Can be physiologic or nonphysiologic

Page 23: Pacemaker troubleshooting- single chamber pacemakers.

• Oversensing occurs when the pacemaker inappropriately “thinks” that it sees intrinsic activity that is not there

• Oversensing causes the pacemaker to inhibit the pacing output pulse, even though the device should be pacing

Page 24: Pacemaker troubleshooting- single chamber pacemakers.
Page 25: Pacemaker troubleshooting- single chamber pacemakers.
Page 26: Pacemaker troubleshooting- single chamber pacemakers.

• Myopotentials– Pectoralis,rectus abdominis,diaphragm– Bipolar leads are less susceptible

• Oversensing of normal intracardiac signals– Reduce sensitivity of affected lead to eliminate

oversensing

• EMI• Provocative maneuvers may be needed if

intermittent symptoms• Magnet eliminates pauses-oversensing

Page 27: Pacemaker troubleshooting- single chamber pacemakers.

• Reprogram the sensitivity– Conduct a sensing threshold test– Adjust the sensitivity by making the device less

sensitive (increase the mV setting)– Make only small changes

• Extend the refractory period

Page 28: Pacemaker troubleshooting- single chamber pacemakers.

No Output

• Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate

Pacing output delivered; no evidence of pacing spike is seen

Page 29: Pacemaker troubleshooting- single chamber pacemakers.
Page 30: Pacemaker troubleshooting- single chamber pacemakers.

No Output May Be Caused By:

• Poor connection at connector block• Lead failure• Battery depletion• Circuit failure

Page 31: Pacemaker troubleshooting- single chamber pacemakers.

• Steps to take for possible loss of output– Verify all lead connections– Check lead integrity– Evaluate battery status– Contact the device manufacturer

• Loss of output may require the replacement of all or part of the pacing system

Page 32: Pacemaker troubleshooting- single chamber pacemakers.

Pseudomalfunctions

Pseudomalfunctions are defined as: Unusual,Unexpected ECG findings that appear

to result from pacemaker malfunction but that represent normal pacemaker function

Page 33: Pacemaker troubleshooting- single chamber pacemakers.

• Hysteresis• Magnet rate• rate responsive pacing• Noise reversion• Rate drop response• Sleep rate algorithm

Page 34: Pacemaker troubleshooting- single chamber pacemakers.

Hysteresis

• Allows a lower rate between sensed events to occur; paced rate is higher

Lower Rate 70 ppm Hysteresis Rate 50 ppm

Page 35: Pacemaker troubleshooting- single chamber pacemakers.

Magnet Operation

• Magnet application causes asynchronous pacing at a designated “magnet” rate

Page 36: Pacemaker troubleshooting- single chamber pacemakers.

• Threshold Margin Test (TMT)• Three beats at 100 bpm, followed by a magnet rate of 85• Third beat has an automatic pulse width decrement of 25% • Elective replacement indicators-change the rate from 85 to

65

• Extended TMT.– TMT is performed at 100 ppm– Pulse width reduced by 25% on 3rd , 50% on 5th , and 75%

on 7th

Page 37: Pacemaker troubleshooting- single chamber pacemakers.

Rate Responsive Pacing

• An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior

VVIR / 60 / 120

Page 38: Pacemaker troubleshooting- single chamber pacemakers.

Electrical Reset and Battery Depletion• Reset may occur due to exposure to EMI electrocautery, defibrillation, causing reversion to a “back-

up” mode– Rate and mode changes will occur– Device can usually be reprogrammed to

former parameters• Elective replacement indicators (ERI) can resemble back-up

mode– Interrogating device will indicate ERI

(“Replace Pacer”)

Page 39: Pacemaker troubleshooting- single chamber pacemakers.

A Change in Pacing Modes May Be Caused By:

• Battery depletion indicators (ERI/EOL)• Electrical reset• Mode switching• Noise reversion

Page 40: Pacemaker troubleshooting- single chamber pacemakers.

Noise Reversion

• Sensing occurring during atrial or ventricular refractory periods will restart the refractory period. Continuous refractory sensing is called noise reversion and will:– Cause pacing to occur at the sensor-indicated rate

for rate-responsive modes– Cause pacing to occur at the lower rate for non-

rate-responsive modes

Page 41: Pacemaker troubleshooting- single chamber pacemakers.

Noise reversion

VT in a patient with VVI-pacing occurs at lower rate due to noise reversion

Page 42: Pacemaker troubleshooting- single chamber pacemakers.

Rate drop response

Delivers pacing at high rate when episodic drop in rate occurs

Page 43: Pacemaker troubleshooting- single chamber pacemakers.

Muscle Stimulation May Be Caused By:

• Inappropriate electrode placement near diaphragm or nerve plexus

• Break in lead insulation • Unipolar pacing

Page 44: Pacemaker troubleshooting- single chamber pacemakers.

Flouro diagnosis

Page 45: Pacemaker troubleshooting- single chamber pacemakers.
Page 46: Pacemaker troubleshooting- single chamber pacemakers.
Page 47: Pacemaker troubleshooting- single chamber pacemakers.
Page 48: Pacemaker troubleshooting- single chamber pacemakers.
Page 49: Pacemaker troubleshooting- single chamber pacemakers.

1.ECG showsA.Failure to senseB.Failure to captureC.HysteresisD.Oversensing

Page 50: Pacemaker troubleshooting- single chamber pacemakers.

2.ECG showsA.Failure to senseB.Failure to captureC.HysteresisD.Oversensing

Page 51: Pacemaker troubleshooting- single chamber pacemakers.

3.ECG showsA.Failure to senseB.Failure to captureC.HysteresisD.Oversensing

Page 52: Pacemaker troubleshooting- single chamber pacemakers.

• 4.ECG demonstratesa.Failure to senseb.Failure to capturec.Functional non captured.fusion

Page 53: Pacemaker troubleshooting- single chamber pacemakers.

5

ECG shows A.Failure to sense B.Failure to capture C.Hysteresis D.Functional non capture

Page 54: Pacemaker troubleshooting- single chamber pacemakers.

6

ECG showsA.Functional non captureB.Failure to captureC.HysteresisD.Oversensing

Page 55: Pacemaker troubleshooting- single chamber pacemakers.

7.possibilities are• A.VVI in VOO mode• B.magnet kept• C.persistent ventricular undersensing• D.Noise reversion

Page 56: Pacemaker troubleshooting- single chamber pacemakers.

• 8.make and break potentials usually cause• A.undersensing• B.oversensing• C.functional non capture• D.failure to capture

Page 57: Pacemaker troubleshooting- single chamber pacemakers.

• 9.elevated pacing threshold and elevated impedance can be caused by

• A.lead fracture• B.loose set screw• C.insulation failure• D.battery depletion

Page 58: Pacemaker troubleshooting- single chamber pacemakers.

• 10.elevated threshold with decreased impedance caused by

• A.lead fracture• B.loose set screw• C.insulation failure• D.battery depletion

Page 59: Pacemaker troubleshooting- single chamber pacemakers.

11.What are appropriate in this patientA.sensitivity setting to be adjustedB.lower the programmed rate C.introduce hysteresisD.check capture treshold

Page 60: Pacemaker troubleshooting- single chamber pacemakers.

12.What is the next step in this patient– A.adjust sensitivity-increase mV setting– B.adjust sensitivity-decrease mV setting– C.pacing treshold test-adjust output parameters– D.activate hysteresis

Page 61: Pacemaker troubleshooting- single chamber pacemakers.

13.What is appropriate in this patient– A.adjust sensitivity-increase mV setting– B.adjust sensitivity-decrease mV setting– C.pacing treshold test-adjust output parameters– D.activate hysteresis

Page 62: Pacemaker troubleshooting- single chamber pacemakers.

14.What is appropriate in this patient– A.adjust sensitivity-increase mV setting– B.adjust sensitivity-decrease mV setting– C.pacing treshold test-adjust output parameters– D.activate hysteresis

Page 63: Pacemaker troubleshooting- single chamber pacemakers.

15.What is appropriate in this patient– A.adjust sensitivity-increase mV setting– B.adjust sensitivity-decrease mV setting– C.pacing treshold test-adjust output parameters– D.activate hysteresis