Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006 PACS Overall Analysis 1 Overall PACS EQM IMT Analysis Contributions by PACS ICC
18

Overall PACS EQM IMT Analysis Contributions by PACS ICC

Jan 12, 2016

Download

Documents

osric

Overall PACS EQM IMT Analysis Contributions by PACS ICC. PACS EQM IMT Contents. Short Functional Test (SFT) warm (3 x)~ 1h Short Functional Test (SFT) cold HeII(not done)(~ 1h) PACS Integrated Module Test: Full Functional Test (FFT) 7.7h - PowerPoint PPT Presentation
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 1

Overall PACS EQM IMT Analysis

Contributions by PACS ICC

Page 2: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 2

PACS EQM IMT Contents

21-22 Nov 2005

H-Field Tests incl.

threshold measurements

23-25 Nov 2005

E-Field Tests

• Short Functional Test (SFT) warm (3 x) ~ 1h• Short Functional Test (SFT) cold HeII (not done)(~ 1h)

PACS Integrated Module Test:• Full Functional Test (FFT) 7.7h• Short Performance Test (SPT) 12.8h• AOT Test 9.7h• EMC Test (Spec+Phot) ~ 5d

Separate, joint SPIRE/PACS procedure:• PACS/SPIRE Parallel Mode Test (PACS part)~4.7h

• Extra straylight tests (~3d)• Short Functional Test (SFT) cold HeI ~ 1h

Page 3: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 3

Available Reports/Notes • 3 reports on short functional tests warm• 1 report on short functional test HeI

• Performance test of PACS chopper and duty cycle• Internal calibration sources performance tests• Cryostat background adjustment• Emissivity/RSRF on internal calibration sources• Function and performance of PACS grating• Performance test PACS Photometer filter wheel• Ge:Ga detector dark current test on cold internal calibration sources• Ge:Ga bias loop tests and detector performance• Dead/suspicious pixel maps from RSRF, bias loop and dark current tests• Ge:Ga time constants for flux changes • Low frequency noise measurements of the bolometer• Photometer FOV maps

• Summary note on AOT related tests

Page 4: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 4

Warm Electronics and OBSW

• SPU-L failure: loose FPGA removed, repaired and re-installed• DPU AVM1 failure: broken inductor removed and repaired, exchange with CFM2 but that one generated corrupted output only re-installed AVM1 • SPU OBSW: upgraded from V11.1 to 11.7 to prevent IMT data losses due to a known bug, detected during ILT• SPU OBSW: Data rate / CPU workload too high (as expected from ILT) • DPU HW/OBSW: Few known and new bugs detected, but no major show-stopper (Corrupted packet checksums, duplicate HK packets, SSC discontinuities)• BOLC: 20Hz command had unexpected impact on supply groups

Page 5: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 5

Test Results (1)• Chopper

- duty cycle: comparable to ILT (but too small)- noise: slightly worse than in ILT, but still in spec.- Slight angular offset detected: contact with housing has been removed after the ILT

Page 6: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 6

Test Results (2)• Calibration sources

- Nominal performance: total heating time and stability in spec

Fluctuations = 50 unitsSpec.= 7mK = 70 units

Page 7: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 7

Test Results (3)• Grating: Suffered from a DMC FPGA problem (introduced after ILT)

Spontaneous grating controller self-disable

Excessive noise on position readback

Page 8: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 8

Test Results (4)• Photometry Filter Wheel

- nominal performance, transition time even slightly shorter than in ILT

Position A

Position B

Page 9: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 9

Test Results (5)

• Temperature sensors: - nominal performance for all but the two blue detector sensors- 1 sensor was recovered, apparently a contact problem on a DMC connector

• Blue detector heater: - o.k. (but thermal shimming to be improved)

• Optical performances:- no particular tests done, since no telescope/OGSE

available- straylight- FoV maps do not indicate any other problem

Page 10: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 10

Test Results (6)• Ge:Ga Detector performance:

- IMT performance similar to ILT- Differences due to other background conditions: absolute level and stability - Lost a number of pixels

Rint=64, Cint=0.3pF

Rint=64, Cint=0.3pF

Page 11: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 11

Test Results (6)• Ge:Ga Detector performance:

- Lost a number of pixels

Blue array :   - 2 new high noise pixels

Red array :  - 7 new dead pixels   - 5 new high noise pixels   - 1 dead pixel in ILT seems to react better to flux again, but very noisy.

Page 12: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 12

Test Results (7)

• Bolometer performance: - IMT performance similar to ILT- Differences due to shorter test time and other background conditions: absolute level and stability

ILT IMT

Page 13: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 13

Test Results (8)

• Bolometer performance: - IMT performance similar to ILT- Differences due to shorter test time and other background conditions: absolute level and stability

ILT IMT

Page 14: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 14

Further Tests

• Not reported in this presentation:- Optimum Ge:Ga detector Bias

- S/N as a function of reset interval- Emissivity of calibration sources- Quick wavelength check (several executions

failed, one was successful)

- Memory Management- SAFE mode transition- Non-Prime PACS Switch-on- Bolometers Saturation check- Detector selection table test- Compression/Reduction Modes Test (Phot & Spec)- Time constants for flux changes in spectroscopy

• Failed Tests: (due to grating problem)- Grating relative spectral response function- Emissivity of calibration sources (failed partly)- Medium sampling grating test

Page 15: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 15

AOT Tests (1)

• Overview:- Spectroscopy “TuttiFrutti” AOT ( o.k.)- Photometry “TuttiFrutti” AOT ( o.k.)[Test of all AOT relevant OBCPs, verification of PACS sub-unit synchronisation and building blocks for real AOTs]

• Spectrometer Line Scan AOT - Calibration blocks concepts - Line scan concepts - Frequency switching concepts• Spectrometer Range Scan AOT• Photometer AOT - Calibration block concepts - Nominal photometry concepts - Photometry in PACS/SPIRE parallel mode

Page 16: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 16

AOT Tests (2)

• Consolidated set of AOT performance tests established for future test campaigns• Although the AOT tests are highly valuable, not everything could be tested: pointing modes, space environment effects, no external sources

• Report on “AOT related test during the EQM IMT” is in preparation, results to be presented at next PACS ICC Feb 23/24

Page 17: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 17

PACS/SPIRE Parallel Mode Test (1)

Concept:- “Shifted” (by ~25-30min.) cooler recycling- SPIRE activities- Thermal behaviour test (with SPIRE on in parallel) - Two test cases: Single and Dual band photometry

- Only Line Scan AOT option tested - Dual band: blue LW, double compression

red, nominal compression - Single band: only red channel, nominal compression - 15 TM packets per sec available for PACS

data rates: PACS ~72 kbit/s SPIRE ~57 kbit/s

Page 18: Overall PACS EQM IMT Analysis Contributions by PACS ICC

Herschel EQM – Results and Lessons Learnt 6/7 Feb 2006

PACS Overall Analysis 18

PACS/SPIRE Parallel Mode Test (2)

• Results:

- Thermal behaviour: o.k.- Single band photometry: o.k.- Dual band photometry: Bug in CUS script, double reduction in blue channel has been set unfortunately to nominal again

Data loss due to violation of bus bandwidth, however correctly handled by DPU.

- Overall: Successful as seen by PACS next steps to be carried out elsewhere (CUS, HSPOT…)