Top Banner
OPTIONS to AIRPORT OPTIONS to AIRPORT ALI’S TOWNCAR: 206 ALI’S TOWNCAR: 206 - - 459 459 - - 7903 7903 $28 for 1 $28 for 1 - - 2 Riders; $5 ea 2 Riders; $5 ea add’l add’l up to 4 total up to 4 total Call 2 hours ahead Call 2 hours ahead Direct to Airport Direct to Airport SHUTTLE EXPRESS VAN 800 SHUTTLE EXPRESS VAN 800 - - 487 487 - - 7433 7433 $23.50 for 1 $23.50 for 1 st st ; $8 for 2 ; $8 for 2 nd nd ; $5 ea ; $5 ea add’l add’l up to 6 up to 6 May stop elsewhere if not full May stop elsewhere if not full Call 2 hours ahead Call 2 hours ahead SHUTTLE EXPRESS TOWNCAR (same #) SHUTTLE EXPRESS TOWNCAR (same #) $50 for up to 4 riders (Direct to Airport) $50 for up to 4 riders (Direct to Airport) Call 2 days ahead Call 2 days ahead
66

OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Apr 24, 2018

Download

Documents

ngoxuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

OPTIONS to AIRPORTOPTIONS to AIRPORT•• ALI’S TOWNCAR: 206ALI’S TOWNCAR: 206--459459--79037903

–– $28 for 1$28 for 1--2 Riders; $5 ea 2 Riders; $5 ea add’l add’l up to 4 totalup to 4 total–– Call 2 hours aheadCall 2 hours ahead–– Direct to AirportDirect to Airport

•• SHUTTLE EXPRESS VAN 800SHUTTLE EXPRESS VAN 800--487487--74337433–– $23.50 for 1$23.50 for 1stst; $8 for 2; $8 for 2ndnd; $5 ea ; $5 ea add’l add’l up to 6 up to 6 –– May stop elsewhere if not fullMay stop elsewhere if not full–– Call 2 hours aheadCall 2 hours ahead

•• SHUTTLE EXPRESS TOWNCAR (same #)SHUTTLE EXPRESS TOWNCAR (same #)–– $50 for up to 4 riders (Direct to Airport)$50 for up to 4 riders (Direct to Airport)–– Call 2 days aheadCall 2 days ahead

Page 2: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Use of the R/F Accreditation Use of the R/F Accreditation Phantom for Phantom for

Fluoroscopic System EvaluationFluoroscopic System Evaluation

Beth A. Schueler, Ph.D.Mayo Clinic

Robert L. Dixon, Ph.D.Wake Forest University

Charles R. Wilson, Ph.D.Medical College of Wisconsin

Page 3: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

OverviewOverview•• ACR R/F PhantomACR R/F Phantom•• Fluoroscopic System Evaluation InstructionsFluoroscopic System Evaluation Instructions•• Sample ResultsSample Results•• Corrective Actions for System OptimizationCorrective Actions for System Optimization•• Additional Phantom CharacterizationAdditional Phantom Characterization

–– BackscatterBackscatter–– Equivalent water attenuationEquivalent water attenuation

•• ACR VascularACR Vascular--Interventional Accreditation Interventional Accreditation ProgramProgram

Page 4: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

IntroductionIntroduction•• ACR R/F Accreditation Program includes ACR R/F Accreditation Program includes

evaluation of:evaluation of:–– equipment performanceequipment performance–– personnel qualificationspersonnel qualifications–– quality controlquality control–– clinical image qualityclinical image quality

Page 5: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

IntroductionIntroduction•• Equipment performance is assessed with a Equipment performance is assessed with a

standard phantom and image quality test toolstandard phantom and image quality test tool•• Sites submit the following:Sites submit the following:

–– radiographic images from each xradiographic images from each x--ray system ray system • table • upright bucky

–– spot films from each fluoroscopic systemspot films from each fluoroscopic system–– fluoroscopic system evaluation performed by a fluoroscopic system evaluation performed by a

medical physicistmedical physicist

Page 6: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR R/F PhantomACR R/F Phantom

–– PatientPatient--equivalent equivalent acrylic and aluminum acrylic and aluminum phantomphantom

–– Image quality test toolImage quality test tool

•• Design is based on the Design is based on the CDRH fluoroscopic CDRH fluoroscopic phantomphantom

Page 7: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

CDRH Fluoroscopic PhantomCDRH Fluoroscopic Phantom•• Dimensions:Dimensions:

–– 7.6” (19.3 cm) acrylic 7.6” (19.3 cm) acrylic + 0.18” (4.6 mm) Al+ 0.18” (4.6 mm) Al

–– Leg supports to raise Leg supports to raise phantom 3.9” (10 cm)phantom 3.9” (10 cm)

Page 8: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

CDRH Abdomen/Spine PhantomCDRH Abdomen/Spine Phantom•• Fluoroscopic phantom design is derived from the Fluoroscopic phantom design is derived from the

LucAl phantomLucAl phantom–– Simulates a 21.5 cm AP abdomen Simulates a 21.5 cm AP abdomen

thickness in absolute and spectral thickness in absolute and spectral transmissiontransmission

–– Aluminum spine strip is includedAluminum spine strip is included

•• For the CDRH fluoroscopy For the CDRH fluoroscopy phantom, the Al spine strip is phantom, the Al spine strip is extended over entire thicknessextended over entire thickness

Page 9: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

CDRH Fluoroscopic PhantomCDRH Fluoroscopic Phantom

Mesh Patterns

Page 10: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR R/F PhantomACR R/F Phantom•• Same acrylic and aluminum Same acrylic and aluminum

thickness as the CDRH phantom thickness as the CDRH phantom

3”

0.18” Al

3”

1.6”•• Cut into slabs Cut into slabs •• Increase size to 10” by 10” Increase size to 10” by 10”

(25.4 by 25.4 cm)(25.4 by 25.4 cm)•• Lead markers included as a Lead markers included as a

reference for repeatable reference for repeatable collimationcollimation

Page 11: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR R/F PhantomACR R/F Phantom•• Overtable xOvertable x--ray tube configuration ray tube configuration

(radiography or fluoroscopy)(radiography or fluoroscopy)

0.18” Al

1.6” Lucite

3” Lucite

3” Lucite

–– Al plate is placed toward Al plate is placed toward the xthe x--ray tuberay tube

Page 12: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR R/F PhantomACR R/F Phantom•• Undertable xUndertable x--ray tube ray tube

configurationconfiguration

3” Lucite

3” Lucite

1.6” Lucite

3.9” air gap

0.18” Al

–– Legs raise phantom 3.9” Legs raise phantom 3.9” (10 cm) above the (10 cm) above the tabletoptabletop

–– Exposure measurements Exposure measurements made at the tabletop are made at the tabletop are approximately freeapproximately free--inin--airair

Page 13: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Image Quality Test ToolImage Quality Test Tool

91

12

3

45

6

7

8

9 2

3

45

6

7

8

Lucite Base: 10” length and width,

0.375” thickness

Mesh Patterns:0.75” diameter

Al Disk: 2” diameter,

0.080” thickness

Low Contrast Holes in Al Disk:

0.375” diameter

Contrast Detail Holes

Al Plug

Page 14: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Image Quality Test Tool Image Quality Test Tool

•• For fluoroscopic and spot imaging, only the For fluoroscopic and spot imaging, only the central portion of the test tool is usedcentral portion of the test tool is used

91

12

34

56

7

8

9 2

3

45

6

7

8Mesh Patterns:

Mesh # Lines/in1 122 163 204 245 306 407 508 609 80

Low Contrast Holes in Al Disk:

Hole # _Depth1 0.068”2 0.049”3 0.035”4 0.025”5 0.018”6 0.0126”7 0.0091”8 0.0063”9 0.0040”

Page 15: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Low Contrast HolesLow Contrast Holes•• Percentage contrast of the low contrast holes has been Percentage contrast of the low contrast holes has been

calculated using numerically generated spectra*calculated using numerically generated spectra*

0

5

10

15

20

25

1 2 3 4 5 6 7 8 9Hole Number

% C

ontr

ast 60 kVp

80 kVp100 kVp120 kVp

* * Suleiman OH, et al (1997). "Nationwide survey of fluoroscopy: Radiation dose and image quality." Radiology 203:471-476.

Page 16: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Fluoroscopic System EvaluationFluoroscopic System Evaluation•• Equipment SetEquipment Set--upup

–– Undertable xUndertable x--ray tube systemray tube system–– Overtable xOvertable x--ray tube systemray tube system

Page 17: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Equipment SetEquipment Set--up: Undertable Tubeup: Undertable Tube

3” Lucite

3” Lucite

1.6” Lucite

0.18” Al

1. Assemble the phantom 1. Assemble the phantom with the leg supports so with the leg supports so that the Al plate is toward that the Al plate is toward the xthe x--ray tube and the lead ray tube and the lead markers are toward the markers are toward the image receptor.image receptor.

Page 18: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Equipment SetEquipment Set--up: Undertable Tubeup: Undertable Tube2. Center the image quality test tool on the table top 2. Center the image quality test tool on the table top

underneath the phantom.underneath the phantom.3. Center the 3. Center the

phantom in the phantom in the FOV under FOV under fluoroscopy.fluoroscopy.

4. Position the II 4. Position the II tower so that it tower so that it rests on top of rests on top of the phantom.the phantom.

Page 19: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Equipment SetEquipment Set--up: Undertable Tubeup: Undertable Tube5. Set the II FOV to mode closest to 9” (23 cm).5. Set the II FOV to mode closest to 9” (23 cm).6. Collimate to the lead markers.6. Collimate to the lead markers.

Page 20: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Equipment SetEquipment Set--up: Overtable Tubeup: Overtable Tube

0.18” Al

1.6” Lucite

3” Lucite

3” Lucite

1. Assemble the phantom so 1. Assemble the phantom so that the lead markers are that the lead markers are toward the image receptor on toward the image receptor on the table top and the Al plate the table top and the Al plate is toward the xis toward the x--ray tube.ray tube.

Page 21: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Equipment SetEquipment Set--up: Overtable Tubeup: Overtable Tube

2. Center the image quality 2. Center the image quality test tool on top of the test tool on top of the phantom.phantom.

3. Center the phantom in 3. Center the phantom in the FOV under the FOV under fluoroscopy.fluoroscopy.

Page 22: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Image Quality EvaluationImage Quality Evaluation1. Adjust the fluoroscopic imaging settings to those 1. Adjust the fluoroscopic imaging settings to those

routinely used for a barium enema exam. routinely used for a barium enema exam. Dose mode, continuous or pulsed exposure, selectable Dose mode, continuous or pulsed exposure, selectable

filtration, grid position, selectable kVp setting, …filtration, grid position, selectable kVp setting, …For digital systems, window/level or other image For digital systems, window/level or other image

processing settings.processing settings.2. Dim the room lights.2. Dim the room lights.3. View the live fluoroscopic image.3. View the live fluoroscopic image.

Page 23: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Image Quality EvaluationImage Quality Evaluation4. Count mesh patterns 4. Count mesh patterns

visible. visible. 5. Count low contrast 5. Count low contrast

holes visible. Adjust holes visible. Adjust monitor brightness and monitor brightness and contrast to optimize contrast to optimize image.image.

6. Note actual kVp and 6. Note actual kVp and calculate % contrast.calculate % contrast.

Page 24: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Entrance Exposure Rate Entrance Exposure Rate

2. For undertable 2. For undertable tube systems, tube systems, position an ion position an ion chamber under chamber under the phantom on the phantom on the table top.the table top.

1. Remove the image quality test tool.1. Remove the image quality test tool.

Page 25: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Entrance Exposure RateEntrance Exposure Rate

4. Center the ion chamber in 4. Center the ion chamber in the FOV.the FOV.

5. Using same fluoroscopic 5. Using same fluoroscopic imaging settings as used to imaging settings as used to evaluate image quality, evaluate image quality, measure exposure rate.measure exposure rate.

3. For overtable tube 3. For overtable tube systems, position an ion systems, position an ion chamber 30 cm above the chamber 30 cm above the table top.table top.

Page 26: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Entrance Exposure RateEntrance Exposure Rate6. Place a lead attenuator 6. Place a lead attenuator

(at least 0.25” thick) on (at least 0.25” thick) on phantom and measure phantom and measure the maximum exposure the maximum exposure rate.rate.

Page 27: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

SpotSpot--film Device Image Acquisitionfilm Device Image Acquisition•• For fluoroscopic systems capable of multiple For fluoroscopic systems capable of multiple

types of spot imaging, the site selects the most types of spot imaging, the site selects the most commonly used mode for image submissioncommonly used mode for image submission–– filmfilm--screen cassettescreen cassette–– photospot cameraphotospot camera–– digital photospotdigital photospot

•• Entrance exposure is measured with a Luxel Entrance exposure is measured with a Luxel dosimeterdosimeter

•• Same equipment setSame equipment set--up as for fluoroscopyup as for fluoroscopy

Page 28: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

FilmFilm--screen Spot Image Acquisitionscreen Spot Image Acquisition

2. Select the 2. Select the 1 on 1 1 on 1 format.format.

1. Load a 10x12” (24x30 cm) or 9.5x9.5” (24x24 1. Load a 10x12” (24x30 cm) or 9.5x9.5” (24x24 cm) cassette into the spotcm) cassette into the spot--film device.film device.

Page 29: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

FilmFilm--screen Spot Image Acquisitionscreen Spot Image Acquisition3. Expose and process the 3. Expose and process the

film. film. 4. Measure the OD in the 4. Measure the OD in the

center of the test tool center of the test tool image. OD should be image. OD should be between 1.0 and 1.8.between 1.0 and 1.8.

ODOD

5. Repeat the exposure 5. Repeat the exposure with the dosimeter with the dosimeter covering a mesh pattern.covering a mesh pattern.

Page 30: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Digital Photospot Image AcquisitionDigital Photospot Image Acquisition1. Select the image processing 1. Select the image processing

options routinely used for options routinely used for clinical imaging and expose.clinical imaging and expose.

2. Print a hardcopy in the 2. Print a hardcopy in the most commonly used film most commonly used film format and size. (No OD format and size. (No OD criteria.)criteria.)

3. Repeat the exposure with 3. Repeat the exposure with the dosimeter covering a the dosimeter covering a mesh pattern.mesh pattern.

Page 31: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Sample ResultsSample Results•• FluoroscopyFluoroscopy

–– Example ACR R/F Phantom imagesExample ACR R/F Phantom images–– 1991 Upper GI Fluoroscopy NEXT (Nationwide 1991 Upper GI Fluoroscopy NEXT (Nationwide

Evaluation of XEvaluation of X--ray Trends) survey results obtained ray Trends) survey results obtained with the CDRH fluoroscopic phantom*with the CDRH fluoroscopic phantom*

* * Suleiman OH, et al (1997). "Nationwide survey of fluoroscopy: Radiation dose and image quality." Radiology 203:471-476.

•• Spot imagesSpot images–– Example ACR R/F Phantom imagesExample ACR R/F Phantom images

Page 32: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Fluoroscopy Example 1Fluoroscopy Example 1

Continuous fluoroContinuous fluoroGrid inGrid in

110 kVp / 2.4 mA110 kVp / 2.4 mA5 mesh patterns (30 lines/in)5 mesh patterns (30 lines/in)

7 holes (1.6% contrast)7 holes (1.6% contrast)8 R/min8 R/min

Page 33: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Fluoroscopy Example 2Fluoroscopy Example 2

Pulsed fluoro (15 pps)Pulsed fluoro (15 pps)Grid outGrid out

90 kVp / 4 mA90 kVp / 4 mA5 mesh patterns (30 lines/in)5 mesh patterns (30 lines/in)

6 holes (2.6% contrast) 6 holes (2.6% contrast) 1 R/min1 R/min

Page 34: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Fluoroscopy Example 3Fluoroscopy Example 3

Continuous fluoroContinuous fluoroGrid outGrid out

80 kVp / 2.6 mA80 kVp / 2.6 mA5 mesh patterns (30 lines/in)5 mesh patterns (30 lines/in)

7 holes (2.1% contrast)7 holes (2.1% contrast)3 R/min3 R/min

Page 35: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

NEXT Survey Results NEXT Survey Results Fluoroscopy: Mesh Visibility

0%

10%

20%

30%

40%

< 12 12 16 20 24 30 40 50 60

Mesh Observed (lines/in)

Faci

litie

s (%

)

n=352 sites

Mean: 24 lines/in(4 mesh patterns)

Page 36: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

NEXT Survey ResultsNEXT Survey ResultsFluoroscopy: Low Contrast Holes

0%

10%

20%

30%

0 1 2 3 4 5 6 7 8

Holes Observed

Faci

litie

s (%

)

n=352 sites

Page 37: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

NEXT Survey ResultsNEXT Survey ResultsFluoroscopy: Low Contrast Visibility

0%

10%

20%

30%

1-1.9 2-2.9 3.3-9 4-4.9 5-5.9 6-6.9 7-7.9 >8

Percentage Contrast Observed

Faci

litie

s (%

)

n=352 sites

Mean: 4.9%

Page 38: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

NEXT Survey ResultsNEXT Survey ResultsFluoroscopy Exposure Rate

0%

10%

20%

< 2 2-2.9 3-3.9 4-4.9 5-5.9 6-6.9 7-7.9 >8

Entrance Exposure Rate (R/min)

Faci

litie

s (%

)

n=340 sitesMean:

5.0 R/min

Mean kVp: 102

Page 39: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Spot Image Example 1Spot Image Example 1

FilmFilm--screenscreen85 kVp / 17 mAs85 kVp / 17 mAsSmall focal spotSmall focal spot

8 mesh patterns (60 lines/in)8 mesh patterns (60 lines/in)9 holes9 holes300 mR300 mR

Page 40: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Spot Image Example 2Spot Image Example 2

FilmFilm--screenscreen120 kVp / 4 mAs120 kVp / 4 mAsLarge focal spotLarge focal spot

6 mesh patterns (40 lines/in)6 mesh patterns (40 lines/in)7 holes7 holes200 mR200 mR

Page 41: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Spot Image Example 3Spot Image Example 3

Digital PhotospotDigital Photospot60 kVp / 50 mAs60 kVp / 50 mAsSmall focal spotSmall focal spot

6 mesh patterns (40 lines/in)6 mesh patterns (40 lines/in)9 holes9 holes400 mR400 mR

Page 42: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Spot Image Example 4Spot Image Example 4

Digital PhotospotDigital Photospot80 kVp / 3 mAs80 kVp / 3 mAsSmall focal spotSmall focal spot

6 mesh patterns (40 lines/in)6 mesh patterns (40 lines/in)8 holes8 holes55 mR55 mR

Page 43: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Spot Image Example 5Spot Image Example 5

Digital PhotospotDigital Photospot120 kVp / 1 mAs120 kVp / 1 mAsSmall focal spotSmall focal spot

6 mesh patterns (40 lines/in)6 mesh patterns (40 lines/in)7 holes7 holes60 mR60 mR

Page 44: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Suggested Corrective ActionsSuggested Corrective Actions•• If phantom image quality If phantom image quality

or dose does not meet or dose does not meet minimum pass/fail minimum pass/fail criteria, the medical criteria, the medical physicist may be called physicist may be called on to suggest corrective on to suggest corrective actionaction

Page 45: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• Phantom entrance exposure rate too high Phantom entrance exposure rate too high

–– Image intensifier exposure rate is set too high Image intensifier exposure rate is set too high • 1-3 µR/frame in the 9” (23 cm) FOV recommended

–– Increase filtrationIncrease filtration• 3-3.5 mm Al at 80 kVp recommended

–– Remove grid during fluoroscopyRemove grid during fluoroscopy

Page 46: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• Number of mesh patterns observed is too lowNumber of mesh patterns observed is too low

–– Image intensifier focus is poor Image intensifier focus is poor

Page 47: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• High contrast spatial resolution High contrast spatial resolution

measurementmeasurement–– Measure at center and edge of imageMeasure at center and edge of image–– 1.25 lp/mm (line bar pattern) or 30 1.25 lp/mm (line bar pattern) or 30

wires/in (mesh pattern) in the 9” FOV wires/in (mesh pattern) in the 9” FOV recommended recommended

Page 48: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• Number of mesh patterns observed is too lowNumber of mesh patterns observed is too low

–– Image intensifier focus is poorImage intensifier focus is poor–– Focal spot size is too large Focal spot size is too large

Page 49: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective ActionsFocal Spot Blur

0

20

40

60

80

100

0 1 2 3 4

Actual Focal Spot Size (mm)

Lim

iting

Res

olut

ion

(mes

h)

UndertableTube (m=1.8)

OvertableTube (m=1.4)

Page 50: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective ActionsTV Chain Resolution

0

20

40

60

80

100

525 1024

Number of TV Scan Lines

Lim

iting

Res

olut

ion

(mes

h)

UndertableTube (m=1.8)

OvertableTube (m=1.4)

Page 51: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• Number of mesh patterns observed is too lowNumber of mesh patterns observed is too low

–– Image intensifier focus is poorImage intensifier focus is poor–– Focal spot size is too large Focal spot size is too large –– Poor monitor performancePoor monitor performance–– Excessive image noiseExcessive image noise–– Poor contrastPoor contrast

Page 52: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Corrective ActionsCorrective Actions•• Percentage contrast detectability is too highPercentage contrast detectability is too high

–– Image intensifier exposure rate is set too low Image intensifier exposure rate is set too low • 1-3 µR/frame in the 9” (23 cm) FOV recommended

–– Consider a replacement II or replacement monitorConsider a replacement II or replacement monitor

Page 53: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Additional Phantom CharacterizationAdditional Phantom Characterization1. Backscatter:1. Backscatter:

Measurement of the amount of backscatter produced by Measurement of the amount of backscatter produced by the ACR R/F phantom at the exposure measurement the ACR R/F phantom at the exposure measurement locationlocation

2. Equivalent Water Attenuation:2. Equivalent Water Attenuation:Comparison of the ACR R/F phantom transmission Comparison of the ACR R/F phantom transmission

with other common phantomswith other common phantoms

Page 54: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR R/F Phantom: BackscatterACR R/F Phantom: Backscatter•• The CDRH fluoroscopy The CDRH fluoroscopy

phantom design allows for phantom design allows for measurement of the entrance measurement of the entrance exposure with minimal exposure with minimal contribution from backscattercontribution from backscatter

–– The Al plate is positioned on the The Al plate is positioned on the entrance side of the phantomentrance side of the phantom

3” Lucite

3” Lucite

1.6” Lucite

3.9” air gap

0.18” Al

–– Leg supports raise the phantom Leg supports raise the phantom 3.9” (10 cm) from the tabletop3.9” (10 cm) from the tabletop

Page 55: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Backscatter: MethodsBackscatter: Methods•• For validation, the backscatter factor (BSF) was For validation, the backscatter factor (BSF) was

measured as a function of distance from the measured as a function of distance from the phantomphantom–– BSF = Exposure with phantom in place BSF = Exposure with phantom in place ÷ Exposure Exposure

without phantomwithout phantom–– FOV was collimated to the lead markersFOV was collimated to the lead markers–– Measurements made at 60, 80, 100 and 120 kVpMeasurements made at 60, 80, 100 and 120 kVp

Page 56: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Backscatter: ResultsBackscatter: Results

1

1.05

1.1

1.15

0 5 10 15Distance from Phantom (cm)

BSF 60 kVp

80 kVp100 kVp120 kVp

Page 57: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Attenuation: MethodsAttenuation: Methods•• Measure the equivalent water thickness of the Measure the equivalent water thickness of the

ACR R/F PhantomACR R/F Phantom–– Matching exposures determined by matching Matching exposures determined by matching

radiographic screenradiographic screen--film optical densities (to within film optical densities (to within 0.02 OD)0.02 OD)

–– FOV was collimated to the lead markers FOV was collimated to the lead markers –– Measurements made at 60 and 100 kVpMeasurements made at 60 and 100 kVp–– Repeat for an ANSI abdomen phantomRepeat for an ANSI abdomen phantom

Page 58: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ANSI Abdomen PhantomANSI Abdomen Phantom•• Acrylic and aluminum (type 1100 alloy) phantom Acrylic and aluminum (type 1100 alloy) phantom

–– PatientPatient--equivalent in absolute and spectral equivalent in absolute and spectral transmissiontransmission

1 mm Al

6” acrylic(15 cm)

2 mm Al

Page 59: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Attenuation: ResultsAttenuation: Results

17.4

23

17.4

23

0

5

10

15

20

25

ACR R/F AbdomenPhantom

ANSI AbdomenPhantom

Equi

vale

nt W

ater

Thi

ckne

ss (c

m)

60 kVp100 kVp

7.6” acrylic + 4.6 mm Al

6” acrylic + 3 mm Al

Page 60: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR VACR V--I Accreditation ProgramI Accreditation Program•• Vascular, interventional, neurovascular Vascular, interventional, neurovascular

procedures procedures •• Accreditation Program includes evaluation ofAccreditation Program includes evaluation of

–– equipment performanceequipment performance–– personnel qualifications (physician, RT, MP, nurse)personnel qualifications (physician, RT, MP, nurse)–– quality controlquality control–– clinical image qualityclinical image quality

Page 61: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR VACR V--I Accreditation ProgramI Accreditation Program•• Quality controlQuality control

–– Initial and annual medical physics testsInitial and annual medical physics tests–– More frequent technologist testsMore frequent technologist tests

•• Patient dose monitoring recommendedPatient dose monitoring recommended–– dosedose--area product meterarea product meter–– skin dose monitorskin dose monitor–– entrance skin dose calculationentrance skin dose calculation–– estimation from fluoroscopy time and number of estimation from fluoroscopy time and number of

imagesimages

Page 62: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR VACR V--I PhantomI Phantom•• Same as the R/F phantom, with the addition of an Same as the R/F phantom, with the addition of an

artery blockartery block•• Artery block includes iodineArtery block includes iodine--filled simulated filled simulated

arteries of various diameters with stenoses and arteries of various diameters with stenoses and aneurysmsaneurysms

Page 63: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Artery BlockArtery Block

Page 64: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR VACR V--I PhantomI Phantom•• Subtraction imaging of the artery blockSubtraction imaging of the artery block

–– One 3” section of One 3” section of the phantom is the phantom is replaced with a slot replaced with a slot blockblock

–– Artery block is Artery block is passed through the passed through the slot: blank side for slot: blank side for mask, artery side mask, artery side for live imagefor live image

Page 65: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

ACR VACR V--I PhantomI Phantom•• DSA image DSA image

of the artery of the artery blockblock

Page 66: OPTIONS to AIRPORT a 21.5 cm AP abdomen ... 8 0.0063 ” 9 0.0040” Low ... TV Chain Resolution 0 20 40 60 80 100 525 1024 Number of TV Scan Lines Limiting Resolution

Use of the R/F Accreditation Use of the R/F Accreditation Phantom for Phantom for

Fluoroscopic System EvaluationFluoroscopic System Evaluation

Beth A. Schueler, Ph.D.Mayo Clinic

Robert L. Dixon, Ph.D.Wake Forest University

Charles R. Wilson, Ph.D.Medical College of Wisconsin