Top Banner
OPTICAL FIBERS FOR CHEMICAL SENSING AND BIOSENSING Vlastimil Matejec , Miroslav Chomat, Ivan Kasik, Ondrej Podrazky, Marie Pospisilova Institute of Photonics and Electronics AS CR, v.v.i., Chaberska 57, 182 51 Prague 8, Czech Republic Project BIO-OPT-XUV, Workshop W1, Kladno, October 3, 2011
55

Optical fibers for chemical sensing and biosensing (V. Matejec)

Feb 10, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Optical fibers for chemical sensing and biosensing (V. Matejec)

OPTICAL FIBERS FOR CHEMICAL SENSING AND

BIOSENSINGVlastimil Matejec,

Miroslav Chomat, Ivan Kasik, Ondrej Podrazky, Marie Pospisilova

Institute of Photonics and Electronics AS CR, v.v.i., Chaberska 57, 182 51 Prague 8, Czech Republic

Project BIO-OPT-XUV, Workshop W1, Kladno, October 3, 2011

Page 2: Optical fibers for chemical sensing and biosensing (V. Matejec)

OUTLINE• INTRODUCTION – Performance of IPE• EVANESCENT-WAVE FIBER-OPTIC

SENSORS- Ways for increasing the sensitivity

• FIBER-OPTIC TIPS- Detection in cells and small volumes

• SPECIAL SENSING STRUCTURES- Microstructure fibers, Whispering Gallery Mode (WGM) microresonators

• CONCLUSIONS

Page 3: Optical fibers for chemical sensing and biosensing (V. Matejec)

INTRODUCTIONInstrumental possibilities of IPE

Technology of optical fibersModified Chemical Vapour Deposition – MCVD

Dip-coating devicesFiber drawing towers

Measurement techniquesAbsorption spectrometers, fluorescence

spectrometer Horiba (Jobin Yvon), AFM, spectral elipsometry available in the

Institute, fiber-optic measurements

Page 4: Optical fibers for chemical sensing and biosensing (V. Matejec)

TechnologyMCVD

Dip coating

Sol-gelPreparation of preforms for drawing of optical fibers

Page 5: Optical fibers for chemical sensing and biosensing (V. Matejec)

Fiber DrawingPreform

Graphite furnace

Fiber

Fiber diameterIndicator

Feedback

Spool

Polymerc oating die

Curing furnace

Pulley

Page 6: Optical fibers for chemical sensing and biosensing (V. Matejec)

Measurement devices

FluorologSpectrometers

Page 7: Optical fibers for chemical sensing and biosensing (V. Matejec)

Know-howNearly 30 years of expertise at the preparation of

optical fibers 1981-1989: Optical fibers for telecommunications

Polymer-clad silica (PCS) fibersMultimode graded-index (MM GI) fibers

Single mode (SM) fibers1987 - : Special optical fibers for fiber lasers and

amplifiers and sensors

Page 8: Optical fibers for chemical sensing and biosensing (V. Matejec)

Fiber waveguides

100 μm

PCS

100 μm

Multimode

100 μm

SM

Page 9: Optical fibers for chemical sensing and biosensing (V. Matejec)

Attenuation of telecommunication fibers

800 1000 1200 1400 1600

1

10

100 10-10

0,01

0,1

0,316

0,794

PCS GI SM

Atte

nuat

ion

[dB

/km

]

Wavelength [nm]

Fully comparable with world producers

Page 10: Optical fibers for chemical sensing and biosensing (V. Matejec)

Fiber-Optic SensorsFiber-optic hardware for sensors employingEvanescent-wave detection principle

Fiber tipsMicrostructure fibers

Whispering Gallery Mode ResonatorsLong-period gratings in special fibers

Issue:How to control and increase the sensitivity of optical fibers to model chemicals or temperature?

Page 11: Optical fibers for chemical sensing and biosensing (V. Matejec)

Detected variables change optical properties in the cladding which is detected at the fiber output from changes of the transmitted light.

Evanescent-wave sensors

d

ncl

nco

Guidedwave

Evanescentwave

Radius (r)

Intensity Ez

Radius ( r )

Refractive index n( r )

Core

CladdingDetection site

Page 12: Optical fibers for chemical sensing and biosensing (V. Matejec)

Advantages- Possibility to control the detection length and access of chemicals into the cladding .

- Detection principles based on refractometry, absorption and fluorescence spectroscopy could be employed

Limitations- Less than 1% of the optical power is transmitted in the cladding in standard fibers → low detection sensitivity.

- Evanescent waves penetrate into the cladding on micrometer distances.

? Ways for improving the detection sensitivity of standard optical fibers

Evanescent-wave sensors

Page 13: Optical fibers for chemical sensing and biosensing (V. Matejec)

Attenuated Total Reflection – ATR

2

2

2.

2

1 ⎟⎟⎠

⎞⎜⎜⎝

⎛−

⎟⎟⎠

⎞⎜⎜⎝

−≈

c

c

clcocl

nnd

θθ

θθ

πλεγ

L) exp(- = 0 γii PP

θ= π/2 − Ψ − complementary angle to the angle of reflection Ψ

Higher detection sensitivity ↔ Pi ↓

Page 14: Optical fibers for chemical sensing and biosensing (V. Matejec)

Ways for increasing the sensitivity standard fibers

- Increase of the detection length L - Increase of bulk absorption coefficient εcl

optochemical transducers pH indicators detection of pH (CO2, NH3 )Ru complexes with the fluorescence quenched by oxygen – detection of O2

Enzymes such as glucoseoxidase – detection of glucose

Page 15: Optical fibers for chemical sensing and biosensing (V. Matejec)

Ways for increasing the sensitivity- Materials of the core and cladding ncl →nco - PCG

fibers, detection membranes- Decrease of d - sectorial fibers,D-fibers

V. Matejec et al., Sens. Actuators B39 , 334 (1997)- Control of the reflection angle θ → θc

excitation by an inclined collimated beam, beveled fibersA. Abdelghani et al., Sens. Actuators B 44 , 495 (1997)inverted-graded fibersV. Matejec et al. Sens. Actuators B 51, 340 (1998)bent fibers (U fibers), coiled fibers V. Matejec et al., Sensor Lett. B 80, 132 (2001)

Page 16: Optical fibers for chemical sensing and biosensing (V. Matejec)

Suitable core material

1.36 1.40 1.44 1.48 1.52 1.56 1.60 1.64-4

0

4

8

12

16

20

24

28

PCS PCG-F2 POF

10*lo

g(P/

P 0) [dB

]

Cladding refractive index

Ormocers

Polymer Clad Silica (PCS), Polymer Clad Glass (PCG), Plastic Optical Fibers (POF) – sensitive for different refractive index values

Page 17: Optical fibers for chemical sensing and biosensing (V. Matejec)

Decrease of the core radius

D-fibers (Culshaw-UK-1985)Sectorial (s-) and D-fibers prepared by pulling from a properly ground and polished circular preform

Used mainly for gas and physical sensorsG. Stewart, W. Jin, B. Culshaw, Sens. Act. B 38, 42-47, 1997

Page 18: Optical fibers for chemical sensing and biosensing (V. Matejec)

SPR Platforms for Biosensors• D-polished SM fiber

R. Slavik et al., Novel spectral fiber optic sensor based on surface plasmon resonance, Sens. Actuators B74, 106-111 (2001)

Page 19: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sectorial fibers

Enable to decrease the core dimension a and improve access to evanescent field

Fibers sufficiently robust (due to sectorialpart -0.3 mm)

Core diameter - 30μm (good sensitivity)

Page 20: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sectorial fibers – Spectral response

~5 times higher response than comparable PCS fibers

Page 21: Optical fibers for chemical sensing and biosensing (V. Matejec)

Si Detector

Laser diode - 630 nm

α

Motor

Gas in

Detection membrane

Solution in (out)

Cell

Gas out

Fiber

Excitation by an inclined collimated beam

Kinetics measurements, interaction parameters

-24 -18 -12 -6 0 6 12 18 240.0

0.2

0.4

0.6

0.8

1.0

Fixed angle

N2 Toluene (1.9 vol.%) Hexane (12 vol.%)R

elat

ive

outp

ut p

owe

P/P

max

Angle of inclination - α [degree]

MTES membrane

Page 22: Optical fibers for chemical sensing and biosensing (V. Matejec)

Beveled fibersInstead of excitation by an inclined collimated beam –use of beveled fibers and excitation without the inclination

0 50 100 150 200 250 3000.0

0.3

0.6

0.9

1.2

1.5

1.8

Polymer#2

BEVELED FIBER- sensitivity curves

Beveling 0 o,α50 used

Beveling 36 o, α= 0o

Beweling 18 o, α=0o

10*lo

g[P

(t=0)

/P(e

q.)]

[dB

]

Concentration of toluene [mg/l]

The same sensitivity achieved for beveled fiber and fiber without beveling excited by the inclined beam

Page 23: Optical fibers for chemical sensing and biosensing (V. Matejec)

Multimode IGI fibersSensitivity control through control of the angle of reflection on core/cladding boundary

GI fiber

IGI fiber

Page 24: Optical fibers for chemical sensing and biosensing (V. Matejec)

IGI fibers - ray trajectories

Cladding

Cladding

Axial PointLight Source

Core

Ψ2

Ψ1

h

Proper h → the same values of Ψ

Page 25: Optical fibers for chemical sensing and biosensing (V. Matejec)

Calibration curves of IGI fibers

-0.004 0.000 0.004 0.008 0.012

0

10

20

30

40

IGI fiber

PCS fiber

h=2.0 mm

h=1.2 mm

h =2.0 -1.2 mmFiber endface distance

Out

put s

igna

l [ar

b. u

nit]

Core/cladding refractive-index difference

IGI: tailoring the sensitivity in a RI range 1.41-1.49

Page 26: Optical fibers for chemical sensing and biosensing (V. Matejec)

Bent fibers - U fibers

Page 27: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensitivity to refractive index

1,32 1,34 1,36 1,38 1,40 1,42 1,44 1,460

4

8

12

16

20

24

10 lo

g[P

(dry

)/P]

[dB

]

Refractive index of immersion

R → ∞ R=2 mm R=1 mm

Active length 1 cmλ=670 nm

Page 28: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensitivity to absorption coefficientFiber bending - 2 or 1 mm

0 1 2 3 4 5 6 7 80,0

0,1

0,2

0,3

0,4

0,5

0,6

R=2 mm R=1 mm

Straight fiber

10*lo

g[P

(wat

er)/P

] [d

B]

Concentration of methylene blue [mg/l]

pH=5λ=665 nm

Page 29: Optical fibers for chemical sensing and biosensing (V. Matejec)

Biosensing with U-shaped fibercoated with gold nanoparticles (GNP)

V.V.R. Sai et al., Biosens.Bioelectr. 24, 2804–2809, 2009

Time seconds

a-IgG;b-PBS;c-BSA/PBS;d-PBS;

e-0.06; f-0.6; g-6; h-30 μg/ml

Page 30: Optical fibers for chemical sensing and biosensing (V. Matejec)

IPE Sensor of oxygen and glucose

LED light

PMT Detector

Main unit Communication A/D,

P

RS

Silica or PMMA

Sensing film is placed

Bent plastic fiber, detection membrane of ORMOCER®(n~1.5) with Ru complex and glucoseoxidase.

Detection based on monitoring oxygen consumption at the enzymatic reaction of glucose

Page 31: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensor of oxygen and glucose

0 100 200 300 400 500 600 700 800 900 1000 1100

0.66

0.68

0.70

0.72

0.74

0.76

0.78

0.80buffersolution removed

Air bubbled+1ml 0.1M glucose

+1ml 0.1M glucose

Rel

ativ

e flu

ores

cenc

e in

tens

ity

Time [s]

Time response Calibration

Detection limit 0.2 mM

0.0 0.5 1.0 1.5 2.00.0

0.1

0.2

0.3

0.4

0.5

10*lo

g[P

] [dB

]

Glucose concentration [mM]

Page 32: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensors on Fiber-Optic TipsFiber-optic tip= Optical fiber conically narrowed into a sharp tip (micrometer - nanometer scale).

Coupling of higher modes into lower modes –adiabatic tapersSmall tips modified with optochemical transducers suitable for intracelular measurementsE. J. Park et al., J. Mater. Chem. 15 (2005) 2913 – detection of oxygen

50 nm

Page 33: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensors on Fiber-Optic Tips

Intracellular measurements.

K-selective taper probes in mouse oocyte (100μm)S.M. Buck et al. , Talanta 63 (2004) 41

Page 34: Optical fibers for chemical sensing and biosensing (V. Matejec)

Sensors on Fiber-Tapers

- R. Kopelman et al. (USA)E. J. Park et al., J. Mater. Chem. 15 (2005) 2913 – detection of oxygen M.R. Shortreed et al., Sens. Actuators B 38-39 (1997) 8 - detection of potassium

-T. Vo-Dingh et al., (USA)T. Vo-Dingh et al., J. Nanoparticle Research 2 (2000) 17-B.M. Cullum et al., Tibtech September 18 (2000) 388-reviewT. Vo-Dinh et al., Anal Bioanal Chem 382 (2005) 918 -reviewB. M. Cullum et al., Analytical Biochemistry 277 (2000) 25– benzo[a]pyrenetetrol in mammary carcinoma cells

Optical sensorsO.S. Wolfbeis, J. Mater. Chem. 15 (2005) 2657– fluorescence sensing

Near field transmission measurements are used

Page 35: Optical fibers for chemical sensing and biosensing (V. Matejec)

Approaches of IPE - Reflection

Page 36: Optical fibers for chemical sensing and biosensing (V. Matejec)

Preparation of TapersInput silica fiber

Elongation of the fiber heated by a burner or a CO2 laser

Heating

Coating with ITO or Al protection coating

Precise cutting

Page 37: Optical fibers for chemical sensing and biosensing (V. Matejec)

Fiber tapers (IPE)

Tip with a final diameter of 2 μm, coated with a protective ITO layer (mechanical strength)

Page 38: Optical fibers for chemical sensing and biosensing (V. Matejec)

Detection Layer at Taper TipSilica xerogels embedding BCECF or HPTSHPTS= 8-hydroxypyrene-1,3,6-trisulfonic acid trisodiumsalt (λexc =430 nm, λem = 480 nm) BCECF=2’,7’-Bis(2-carbonylethyl)-5(6)-carboxyfluorescein(Aldrich 14560) (λexc =473 nm, λem = 540 nm)

pH fluorescence indicators used in biochemical measurements

Prepared by contact of the tip end with a sol drop

Page 39: Optical fibers for chemical sensing and biosensing (V. Matejec)

V Taper

Strong fluorescence signals, easy instrumentation

Page 40: Optical fibers for chemical sensing and biosensing (V. Matejec)

V-Taper

4 5 6 7 8 9 10

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

Fluorescence signals 620 nm Fit of the data

Rel

ativ

e ou

tput

pow

er

pH

I=0.15 mol⋅l-1 BCECF

Calibration

Excitation 473 nm

Page 41: Optical fibers for chemical sensing and biosensing (V. Matejec)

Detection of pH in leaf exudates

5.50.16.0

5.60.35.4

5.00.35.4

MeanSt. deviationElectrochem.

Xylem exudateextretedafter cutting off the leaf at its morphological base

Xylem exudateexcreted after cutting off leaf tip

Guttationsolution excreted at the top tip of leaf

pH gradient in leafs has not been detected

Page 42: Optical fibers for chemical sensing and biosensing (V. Matejec)

Cell Measurements in collaboration with the Institute of Experimental Botany AS CR

Difficult penetration of the cell membrane

Page 43: Optical fibers for chemical sensing and biosensing (V. Matejec)

Microstructure fiberssensitive to gaseous oxygen

500 μm

Page 44: Optical fibers for chemical sensing and biosensing (V. Matejec)

Microstructure fibers sensitive to gaseous toluene

1350 1400 1450 1500 1550 1600 1650 1700 1750 18000,0

0,2

0,4

0,6

0,8

1,0

1,2

Atte

nuat

ion

[dB

]

Wavelength [nm]

0,20 mol% 0,27 mol% 0,39 mol% 0,83 mol% N2

On photonics crystal fibers for sensing see e.g.

R.V. Nair, Progress in Quantum Electronics 34, 89–134, 2010M. Skorobogatiy, J. Sensors Volume 2009, Article ID 524237, 20 pages

Page 45: Optical fibers for chemical sensing and biosensing (V. Matejec)

WGM microresonators

Stems of microspheres fixedin silica capillaries for easily handling

K1-365 μmK2-330 μm

Heating a fiber tip with a burner

K3-67 μm

Heating a fiber with a CO2 laser

Page 46: Optical fibers for chemical sensing and biosensing (V. Matejec)

Excitation of WGMs - Taper

Top view Side view

Excitation by a red laser -visualisation

Waist diameter < 5μm

μ-Rezonator

Page 47: Optical fibers for chemical sensing and biosensing (V. Matejec)

WGM resonances-taperMicrosphere K1

Out

putV

olta

ge[m

V]

Time [ms]

---- Microresonator out contact

---- Microresonator in contact

Out of contact

In contact

Q ~ 106 detection length meters

Page 48: Optical fibers for chemical sensing and biosensing (V. Matejec)

Long Period Gratings Preparation

A high- power CO2 laser

crossing the fiber, induces changes of the mechanical tensions in the fiber – period 0.1-0.5 mm

Page 49: Optical fibers for chemical sensing and biosensing (V. Matejec)

IPE - LPGs in IGI Fibers

-6 -4 -2 0 2 4 6

1.450

1.451

1.452

1.453

1.454

1.455

1.456

1.457

1.458

1.459

Ref

ract

ive

inde

x

Radius [mm]

9 azimuthal sections

SM fiber core and IGI (GI) cladding enable us to control the LPGs sensitivity to strain and to refractive-index changes

-30 -20 -10 0 10 20 301,455

1,460

1,465

1,470

1,475

1,480

1,485

1,490

Uniformcladding

Graded-indexcladding

Ref

ract

ive

inde

x

Distance from fiber center [μm]

Core

Page 50: Optical fibers for chemical sensing and biosensing (V. Matejec)

Temperature sensitivity of LPGsSpectral effects

1480 1500 1520 1540 1560 1580

-62

-60

-58

-56

-54

23.9 26.7 30.7 34.5 38.5 43.2

Out

put p

ower

den

sity

(dBm

/nm

)

Wavelength (nm)

Temperature (°C)

Page 51: Optical fibers for chemical sensing and biosensing (V. Matejec)

Conclusions1) IPE performance for the projectPreparation fiber-optic hardware for chemical sensors and biosensors applicable in medicine.Experience with the application of functional layers onto optical fibersExpertise with characterization of sensing modules 2) IPE offersFiber-optic elements for the application of optochemicaland biological transducers and testing their sensing performance- Bent fibers, beveled fibers, fiber tips, microresonatorsEducation of students

Page 52: Optical fibers for chemical sensing and biosensing (V. Matejec)

Optical fibers - Group of IPE

Page 53: Optical fibers for chemical sensing and biosensing (V. Matejec)

THANK YOU

Page 54: Optical fibers for chemical sensing and biosensing (V. Matejec)

8-hydroxypyrene-1,3,6-trisulfonic acid trisodium salt-HPTS

Page 55: Optical fibers for chemical sensing and biosensing (V. Matejec)

2’,7’-Bis(2-carbonylethyl)-5(6)-carboxyfluoresceinBCECF (Aldrich 14560)

Excitation at 473 nm, reference 430 nm